DATA SHEET

Water Meter Test Bench Model Q5


For residential water meters sizes DN15, DN20 & DN25

Your complete small meters shop in one single equipment...

Volumetric testing method according to ISO


4064-2 and OIML R49-2 international standards
for the determination of the measuring error of
residential water meters sizes DN15, DN20 and
DN25.

3 volumetric reference tanks. Includes high


capacity 350 litre tank for reliable, low uncertain-
ty high flow testing.

Testing flowrate measurement with a choice of


standard rotameters or advance digital display
magnetic flowneters.

Built-in water reservoir and automatic pumping


system eliminates the need for further infrastruc-
ture in your testing lab.

Automatic pumping system based on sub-


mersible pump, constant pressure automatic
VFD control and elevated Constant Head
Tank.

Compact design, minimum space, easy to


install and operate.

Standard Manual Operation, upgradable to


PLC controlled, semiautomatic operation.

www.ibpcal.com “I-B-P-Cal” is the Spanish acronym for


Puebla, Mexico “Test Bench Engineering and Calibrations, Inc”
Model Q5 Technical Data
Testing Capacity: Additional Instrumentation:
5 water meters size DN15 x 110 mm length Thermo-hygrometer for measuring ambient reference conditions:
5 water meters size DN15 x 165 mm length
Water Temperature. Ambient Temperature.
5 water meters size DN15 x 190 mm length
5 water meters size DN20 x 190 mm length Athmospheric Pressure and Relative Humidity.
3 water meters size DN25 x 260 mm length
Water Inlet and Outlet pressure measurement:
Estimated average productivity: 90 meters tested daily
(estimate based on testing 3 flows with Q3=2,5-R80 or N=1,5-Class B water meters) 2-1/2” Dial Analog Pressure Gauge. Range: 0-7 Kg/cm2 .

Minimum division: 0.1 Kg/cm2 .


Volumetric Reference Tanks:

Reading Resolution: 200 millilitres.


Tank 1: Volume 350 litres Integrated Pumping System:
(Optional: Reading Res. 100 millilitres.)

Reading Resolution: 100 millilitres. Integrated Water Reservoir - 1,200 litre capacity
Tank 2: Volume 100 litres
(Optional: Reading Res. 50 millilitres.)
Submersible Pump with Automatic VFD control
Reading Resolution: 10 millilitres.
Tank 3: Volume 10 litres Requires 220 VAC - 25 Amp Power Supply
(Optional: Reading Res. 5 millilitres.)

Quiet and automatic operation


Testing Flowrates:
Elevated Constant Pressure Tank included for low flow testing
With Standard Rotameters: With Optional Mag Flowmeters:
6.0 l/h to 7,000 l/h 2.0 l/h to 7,500 l/h
Testing branch 1: 6.0 a 60 l/h Testing branch 1: 2.0 to 40 l/h General Information
Testing branch 2: 24 a 240 l/h Testing branch 2: 20 to 300 l/h Test-Start and Test-End valves are manually operated
Testing branch 3: 240 a 2,280 l/h Testing branch 3: 300 to 1,500 l/h Test Flow Regulation valves are manually operated
Testing branch 4: 1,800 a 7,000 l/h Testing branch 4: 500 to 7,500 l/h Tank Draining valves are manually operated

Flowmeters data: Mechanical Water Meter Clamping System on work bench

Standard Rotameters: Optional Mag Flowmeters: All valves can be upgraded to semi-automatic
PLC control operation on request.
Acrylic Body 24 VDC Power Supply

Stainless Steel float & guide Programmable Units

l/h, l/m and GPM scales Full color Display

Accuracy ±5% of f.s. Accuracy ±0.5% of flow

INSTALLATION AND OPERATION REQUIREMENTS

A minimum available space is required for the installation and operation of


the Model Q5 Water Meter Test Bench. The figure on the left shows the
general dimensions of the equipment.

An available 220VAC, 25 Amp Electric Supply is required.

The integrated water reservoir and pumping system makes possible to install
the Model Q5 virtually in any place where there is enough available area. No
need for expensive infrastructure or special building construction.

For the low and very low flowrate testing, it is required to install a small
elevated Constant Head Tank provided by IBPCal. See details of this
installation in the Model Q5 Operation & Maintenance Manual, or contact us
for installation details information.

GENERAL DIMENSIONS

5A Old St. Joseph Road, Laventille


EXCLUSIVE REPRESENTATION Trinidad - West Indies
Phone: (868) 624-5617 - (868) 624-5618
FOR ALL CARICOM COUNTRIES email: sales@landwengineering.com
website: www.landwengineering.com

You might also like