Professional Documents
Culture Documents
MULTIPLE CHOICE
REF: p. 321
REF: p. 321
REF: p. 321
4. In comparing PET myocardial imaging to SPECT myocardial imaging for myocardial disease,
PET demonstrates:
a. increased resolution, decreased sensitivity, and increased specificity
b. decreased resolution, increased sensitivity, and increased specificity
c. increased resolution, increased sensitivity, and increased specificity
d. decreased resolution, increased sensitivity, and decreased specificity
ANS: C
Since PET scanners have higher sensitivity and resolution than SPECT, PET myocardial
images provide substantially better quality with increased sensitivity and specificity for
detecting coronary artery disease.
REF: p. 321
5. The production of a single PET radiopharmaceutical requires all of the following EXCEPT:
a. a cyclotron
b. a target
c. a hot cell
d. collimator
ANS: D
The production of a single PET radiopharmaceutical requires very expensive equipment,
including the cyclotron, target, automated synthesis module, hot cell (to contain the synthesis
module), and any additional radiopharmaceutical quality control or precursor preparation
equipment.
REF: p. 322
6. Characteristics of the bismuth germinate crystal include that fact that it:
a. is a fast scintillator
b. yields more light than an NaI (Tl) crystal
c. is very good at stopping 511-keV photons
d. has excellent energy resolution
ANS: C
Bismuth germinate (Bi4Ge3O12) is commonly known as BGO. Bismuth has a high atomic
number and is therefore very good at stopping the 511-keV photons. BGO is unfortunately a
relatively slow scintillator and also yields a relatively small amount of light compared with
NaI (Tl). In addition, BGO has poor energy resolution that requires energy windows that span
from about 350 to 650 keV. The stopping power of BGO compensates for its other
deficiencies and it is still a good scintillator for 2D-dedicated PET scanners.
REF: p. 322
7. All of the following have been used as crystals for PET imaging EXCEPT:
a. BGO
b. LSO
c. LYSO
d. YBGO
ANS: D
Some dedicated PET scanners use other scintillation crystal materials. Lutetium orthosilicate
(Lu2SiO5 [Ce]; LSO), lutetium yttrium orthosilicate (LYSO), and gadolinium orthosilicate
(Gd2SiO5 [Ce]; GSO) are now available in some commercial PET systems. These crystals
have slightly lower stopping power than bismuth germinate (BGO).
REF: p. 322
8. In a PET scanner, the determination of the annihilation event is made possible by:
a. electronic collimation
b. parallel hole collimation
c. back-scatter projection
d. filtered back-scatter projection
ANS: A
PET scanners have electronic collimation to determine the location of the annihilation event.
The detectors used for coincidence detection to identify an annihilation gamma pair and
determine a line of response (LOR) path representing the path of the photon pair.
REF: p. 322
REF: p. 322
10. In PET scanners with a “reflective block of crystals” design, reflective material surrounds:
a. each crystal
b. every other crystal
c. every block of crystals
d. every ring of crystals
ANS: A
The alternative reflective block approach is to create a block from a group of small individual
crystals. Each crystal is surrounded by a carefully designed amount of reflective material that
spreads the light in a designated pattern onto the photomultiplier tubes (PMTs); therefore both
the cut block and reflective block designs do the same thing—they use reflective surfaces on
the sides of individual crystals to distribute the scintillation photons to the PMTs.
REF: p. 323
REF: p. 324
REF: p. 326
REF: p. 327
REF: p. 327
REF: p. 328
18. Regarding septa on a PET system, all of the following are true EXCEPT:
a. they limit the field of view of events to those within the same detector ring, or
events with contiguous planes
b. they reduce the number of scatter and random events from outside the plane
c. they allow for the system to operate in 2D mode
d. they allow for the system to operate in 3D mode
ANS: D
Septa limit the field of view of events to those within the same detector ring, or events with
contiguous planes; they also reduce the number of scatter and random events from outside the
plane. With septa in place, the scanner operates in 2D mode. Retracting the septa creates the
ability to place the scanner in 3D detection mode.
REF: p. 337
22. In a PET system, energy window calibration is typically performed after repair as well as:
a. daily
b. weekly
c. quarterly
d. annually
ANS: C
Energy window calibration is typically only performed after repair, during quarterly
preventive maintenance (PM) service, or just before a normalization calibration is done.
REF: p. 342
REF: p. 343
REF: p. 343
REF: p. 343
26. 2D normalization is done to:
a. subtract background from all LORs
b. measure the efficiency for all LORs
c. to average the data from all LORs
d. to determine the dead time across the LORs
ANS: B
2D normalization calibrations are done to measure the efficiency for all LORs. Normalization
calibration is performed by rotating radioactive rod sources, which contain a low activity
source of radioactivity, and acquiring data that will be used to balance the efficiency of all
detectors in the scanner.
REF: p. 344
27. Absolute activity calibrations should be performed after a major service as well as:
a. daily
b. weekly
c. quarterly
d. annually
ANS: C
Well counter calibration is an historical term for a calibration more appropriately referred to
as absolute activity calibration. Absolute activity calibration factors are used to convert pixel
values into a measure of absolute activity per voxel. Absolute activity calibrations are
performed only after all other calibrations have been performed and should be done quarterly
or after major service.
REF: p. 344
REF: p. 345
The distinct advantage of ToF is that there is not a full LOR across the FOV of the scanner but
rather only a short line in which the annihilation event took place. The effect on the
signal-to-noise ratio of a small lesion is the same effect as acquiring at least 2.5 times the
number of events to improve lesion detection. ToF scanners thus can produce high-quality
images using a slightly shorter acquisition time. In ToF imaging, there is image improvement
as the likelihood of detecting random events is reduced.