Automated Clinical Analyzer

S e t u p M a n ua l

TOKYO BOEKI MEDISYS INC.


Original Instructions
Manufactured by Tokyo Boeki Medisys Inc.
1-14-21, Higashitoyoda, Hino-shi, Tokyo 191-0052, Japan
Tel: (81)-42-587-7777
Contents

1. Introduction ---------------------------------------------------------------------------------------------- 1
2.Contents -------------------------------------------------------------------------------------------------------------- 1
3. Installation requirements ------------------------------------------------------------------------------------- 2
3.1 Power supply requirements ------------------------------------------------------------------------------ 2
3.2 Recommended conditions --------------------------------------------------------------------------------- 2
3.3 Reservoirs requirements-------------------------------------------------------------------------------- 3
3.4 Layout (Water purification system is available only in Japan.) ------------------ 3
4. Installation procedure flow diagram ---------------------------------------------------------------------- 4
5. Unpacking ---------------------------------------------------------------------------------------------------------- 5
6. Confirmation of the contents in the box ----------------------------------------------------------------- 9
7. Removal of the fixing material --------------------------------------------------------------------------- 10
8. Preparation of reservoir ------------------------------------------------------------------------------------- 14
8.1 Separate collection reservoir -------------------------------------------------------------------------- 14
9. Preparation of detergents ----------------------------------------------------------------------------------- 15
9.1 Acid detergent solution ---------------------------------------------------------------------------------- 15
9.2 Alkaline detergent solution ---------------------------------------------------------------------------- 18
10. Connection with water purification system -------------------------------------------------------- 20
11. Connection of drainage line ------------------------------------------------------------------------------ 21
12. Connection with PC------------------------------------------------------------------------------------------ 24
13. Connection of the power supply cable ---------------------------------------------------------------- 25
13-1. Shipment state of the AC cable-------------------------------------------------------------------- 25
13-2. Connection ------------------------------------------------------------------------------------------------ 26
14.Software install ---------------------------------------------------------------------------------------------- 27
14.1 Disk1 installation ------------------------------------------------------------------------------------- 27
14.2 Disk2 installation ------------------------------------------------------------------------------------- 28
15. Application of electric power ----------------------------------------------------------------------------- 33
16. Movement check ---------------------------------------------------------------------------------------------- 35
17. Input temp parameter -------------------------------------------------------------------------------------- 36
18. Function tests of each unit. ------------------------------------------------------------------------------- 37
18.1 Temperature control ---------------------------------------------------------------------------------- 37
18.2 Sample probe position (STM) confirmation -------------------------------------------------- 39
18.3 Reagent probe position (RTM1,2) confirmation -------------------------------------------- 43
18.4 Washing nozzle position confirmation --------------------------------------------------------- 46
18.5 Mixing unit-1,2 (MU-1,2) position confirmation ------------------------------------------- 47
18.6 Barcode reader adjustment ------------------------------------------------------------------------ 48
19. Priming ----------------------------------------------------------------------------------------------------------- 49
20. Auto-gain and cell check ----------------------------------------------------------------------------------- 51
21. Test run ---------------------------------------------------------------------------------------------------------- 53
21.1 Item parameter setting. ---------------------------------------------------------------------------- 53
21.2 Reagent position setting ---------------------------------------------------------------------------- 55
21.3 Calibration setting ------------------------------------------------------------------------------------ 57
21.4 Order ------------------------------------------------------------------------------------------------------- 57
21.5 Sample and reagent setting ----------------------------------------------------------------------- 58
21.6 Test run start ------------------------------------------------------------------------------------------- 58
22. Shutdown -------------------------------------------------------------------------------------------------------- 59
23. ISE unit setting --------------------------------------------------------------------------------------------- 60
23.1 The structure of the ISE unit is shown below. ----------------------------------------------- 60
23.2 ISE setup preparation ------------------------------------------------------------------------------- 72
23.3 ISE unit setting ---------------------------------------------------------------------------------------- 72

1
Setup Manual

23.4 Measurement preparation ----------------------------------------------------------------------- 74


24. Motor parameter setting ----------------------------------------------------------------------------------- 78
24.1 Motor parameter setting screen ------------------------------------------------------------------ 78

Appendix 1 (Confirmation procedure)---------------------------------------------------------------------- 82


Appendix 2 (Sensor connector panel)----------------------------------------------------------------------- 88
Appendix 3 (Rear panel) ---------------------------------------------------------------------------------------- 89
Appendix 4 (Sample Barcode reader adjustment) ----------------------------------------------------- 90
Appendix 5 (Reagent Barcode reader adjustment) ---------------------------------------------------- 92
Appendix 6 ----------------------------------------------------------------------------------------------------------- 94
(Setting of STM down stop position for using a micro cup on the tube.) ------------- 94

Maintenance procedure for detergent line

2
1. Introduction

This manual describes installation procedure and information on our automated


clinical analyzer
Caution : A designated person of Tokyo Boeki Machinery can only install and transport
this analyzer.

2.Contents

1) Analyzer --------------------------- 1 set

2) Accessories ------------------------ 1 set

(1) Standard accessories

(2) Parts for installation

(3) Parts to be exchanged periodically

(4) Maintenance parts

(5) Tools

1
Setup Manual

3. Installation requirements

To install correctly, the following installation requirements are necessary.


Please refer to Page 2, System setup example.

3.1 Power supply requirements


(1) Power supply capacity should be over 1500VA.
(2) Voltage: AC 100V, 115V, 230V ±10% 50/60HZ ±1HZ
(3) The electric cable of 50i must be connected to distribution board, and
NEVER use 3P or 2P plug. In case you do not have distribution board,
please ask certified person to set up the board. Only certified person
can connect the cable to distribution board.
(4) Grounding: Special 3rd grade Resistive less than 10 ohm
(5) Power must be supplied for 24 hours a day.
(6) Install away from high frequency electromagnetic wave emitting equipment
(such as centrifuge, electric discharge), preferably not in the same room.

3.2 Recommended conditions


(1) Less dust and good ventilation
(2) No direct sun light
(3) Horizontal floor (less than 5mm/m)
(4) Endure weight is over 300kgf/㎡.
(5) Room temperature range is 15~30℃, and deviation during test is ±2℃.
(6) Humidity range is 40~80%RH.
(7) No vibration
(8) Electric voltage variation is less than ±10%.
(9) There are no high frequency electro-magnetic emission apparatus near by.
(10) Power supply switchboard with 3rd grade grounding terminal
Resistance is less than 10Ω.
(11) Power supply capacity should be over 1500VA.
(12) Do not install at the position higher than 2000m.
(13) Minimum working space is AC INPUT

50 cm
the followings.

ANALYZER
50 cm 50cm

FRONT

100 cm

2
3.3 Reservoirs requirements

(1) Separate Collection Reservoir, Water Reservoir must be set below the
instrument level.
(2) Both Acidic and Alkaline cleaning solution reservoirs must be set the
designated position.

3.4 Layout (Water purification system is available only in Japan.)

Water tank

Separate
Collection
Reservoir

<Front view> Detergent bottles


・Front ;ALK
・Back ;ACID

Separate
Collection
Reservoir
Water tank

<Back view>
Layout

3
Setup Manual

4. Installation procedure flow diagram

5. Unpacking ※Refer content numbers from 5 to 22.

6. Confirmation of the contents in the box

7. Removal of the fixing material

8. Preparation of reservoir

9. Preparation of detergents

10 Connection with water purification system

11.Connection of drainage line

12.Connection with PC

13.Connection of the power supply cable

14.Software install

15.Application of electric power

16.Movement check

17.Input temp parameter

18.Function tests of each unit

19.Priming

20.Auto-gain and cell check

21.Test run

22.Shutdown

4
5. Unpacking

For unpacking, follow the procedure below.


※The actual analyzer and container is different from the picture.

5.1. Confirm the production number on the box.


5.2. Cut packing bands (2 bands) by a cutter knife.

Cut these bands (2 bands)

Check the condition

5.3. Open the lid of the outer box and take out two supporting stands.

Open this lid Remove the parts holding the analyzer.

Take out two supporting stands.

5
Setup Manual

5.4. Pull out the outer box.

You may be injured. Please do by


more than 2 persons.

Pulling out the outer box, 4 pillars


which hold the analyzer will fall down.
Please be careful.

5.5. Remove the 4 pillars and take out the 2 slopes.

Take out these slopes.

5.6. Remove the antistatic bag.

Check the condition

6
5.7. Remove 6 cushions at the bottom of the analyzer.
Lift the analyzer a little. Remove the cushions.

The analyzer is very heavy. Please


do by more than 2 persons. Please
be careful.

5.8. Set the supporting stands and the slopes.

The slopes fit this position.

7
Setup Manual

5.9. Push the analyzer and get the analyzer down from the container.

Check whether casters fit in the


slopes.

Holding the analyzer, push the


analyzer. Please be careful.

5.10. Check the movement of front doors. Confirm the working of fixing magnets
on the upper and down parts after take the instrument out.

Upper part

magnets

Down part

magnets

8
5.11. Check the movement of six casters at the bottom. Confirm the instrument fixing
by two adjusters.

6. Confirmation of the contents in the box

Remove the contents in the box.


Check the accessory box and quantity.

6.1. Open the left of front door and take out the accessory box.

6.2. Open the box and check the contents.

Open the box and you can see


a box for reaction cells.

*Please check the contents seeing a packing list.

9
Setup Manual

7. Removal of the fixing material

Open the top cover of the analyzer and remove 7 fixing materials and fixing tapes.

Reagent probe1 CWS MU1 MU2 Sample probe

Reagent probe2
ISE probe

7.1. Remove the fixing parts of sample probe and ISE probe.

Fixing part of the Fixing part of the


sample probe sample probe

10
7.2. Remove the fixing parts of reagent probes (R1, R2).

Fixing part of reagent


Fixing part of reagent
probe(R2).
probe(R1).

7.3. Remove the fixing parts of mixing unit 1, 2.

Fixing part of Fixing part of


mixing unit 1 mixing unit 2

7.4. Remove the fixing part of Cell Washing Station.

Fixing unit of
CWS

11
Setup Manual

7.5 Check the lines of sample probe, R1 R2 probes, ISE probe and stirrers.

Confirm the straight line for probes and stirrers.


If it bends, adjust by fingers lightly. Do not press hard.

7.6 Check the sample tray holder.

Confirm all the positions for upper and down parts.

12
7.7 Check the section plate on the RGT.

Confirm the bending for all the


section plate. The reagent
aspiration will not stable if the
reagent bottle is set incorrect
position by bending plate.

7.8 Check the two fixing screws on the CWS.

The nozzle positions on the CWS are


fixed by two fixing screws.

If the screws are loose, fix it by


fingers.

7.9 Check the CWS tubing.

Confirm all the tubing and


connections on the CWS .

CWS

13
Setup Manual

8. Preparation of reservoir

Prepare a separate collection reservoir.

8.1 Separate collection reservoir

Float switch

1. Loosen the cap and take out the float switch.


2. Confirm that the float switch moves up/down smoothly.
3. Assemble back the float switch and screw the cap tightly.
4. Connect the float switch to a connector of analyzer and set a tank in designated
position. (See Page 3 “Layout”)

Connect the float switch to this


connecter. (TANK・WASTE1)

14
9. Preparation of detergents

Prepare an Acid & an Alkaline detergent solution bottles.

Put “Hanger rack container” into the following place.

Please put the container like this.

9.1 Acid detergent solution


1). Set the Acid detergent solution bottle to the place designated position. (Back side)
2). Take out a cap for detergent bottle (Acid) with level sensor assay from packing
box and set the cap to the detergent bottle.

This item is included in


the accessory box.

Install to the bottle.

15
Setup Manual

3) Aspirate the Acid detergent into the air trap.


Please fill the detergent into the cup.

① Connect the tube ○


A of level sensor ASSY to the air trap ○
A .
② Connect the tube ○
B to the detergent aspiration part on WSP.
③ Remove the cap of tube○
C , aspirate the detergent using syringe.

Remove the cap when aspirate the detergent, replace the cap
after aspiration.

Teflon tube
PVC tube

Fill the detergent


till cover top ○
B

Air trap

16
4) Remove the air from WSP.

① Fix the cap of air trap○


C .
② Remove the joint between ○
X and ○
Y.
③ Connect the Teflon tube to syringe (5mL), insert the joint ○
X to the tube tip.
④ Connect the joint○
X to the tip of detergent dispensation on WSP, aspirate the

air until the air goes out.


⑤ Replace the joint, fix the air trap which keeps horizontal surface.

Upper Connector

X (l arge di ameter)

Lower Connector
(s ma l l di ameter)


Y Connect pos i tion
for s yri nge

The plug is
attached back

5). Connect a float switch to the connecter ACDIC of Analyzer.

Connect the float switch to this connecter


(ACDIC)

17
Setup Manual

9.2 Alkaline detergent solution


1) Set the Alkaline detergent solution bottle to the place designated position.
2) Take out a cap for washing solution (Alkaline) with level sensor assay from packing
box and set the cap to the detergent solution bottle.
3) Remove the air by referring procedure for Acid detergent. (Refer 9.1 Acid detergent)
4) Connect a float switch to the connecter of Analyzer.

Connect the float switch to this connecter


(ALKALINE)

18
Acid side Alkaline side

Acid Alkaline
detergent detergent
bottle bottle

19
Setup Manual

10. Connection with water purification system

Please abide by the following condition.


* Rating pressure from water purification system: 0MPa
* Please install the water purification system in the place lesser than 90cm
from the ground which Analyzer is installed. If the water purification system
is installed in the place higher than CWS, the water will spill out from CWS.

Connection between water purification system and PC.


(1) Insert a blue tube to a buffer tank of your water purification system.
* Be sure to insert to the bottom of the tank.
* If there is no buffer tank in water purification system, please prepare a buffer
tank.

(2) Connect the blue tube to a “WATER2” connecter of the rear of the analyzer.

Blue tube
4×6

(3) Connect a blue tube to “WATER2” connecter of the inside of the rear panel and
connect the other side of the tube to a bottom of a degasser.

Connect the tube to this connecter

20
11.Connection of drainage line

There are 3 variations for connection of drainage line.


1. When you collect the separate collection liquid and other waste liquid in a separate
collection reservoir in a mass.
2. When you collect only the separate collection liquid in a separate collection reservoir
and discard other waste liquid in the outside of the analyzer.
3. When you discard the separate collection liquid and other waste liquid in the outside
of the analyzer.

Please discard the separate collection liquid and other waste liquid according
with environment regulations.

1. When you collect the separate collection liquid and other waste liquid in a separate
collection reservoir in a mass.

1) Set the separate collection reservoir in the left side, insert the separate collection
tube “WASTE2A”, “WASTE2B” and waste line tubes “WASTE1A”、”WASTE1B”.
2) Insert the cooling waste line tube to the separate collection reservoir.
Set the separate collection reservoir in the lower left side, insert the separate
collection tube and waste line tube into the separate collection reservoir.
*In this case, do not soak the tubing line tip of cooling unit in liquid.
When it is soaked, the water leaking happens by the remaining water which is
not drained from cooling unit.
*When the tank become full, an error message is shown.
Please discard the contents of the tank.
Separate
Cooling unit waste collection
line tube. reservoir

*The tubing of cooling


unit line and waste line
are combined after the
VE analyzer.

*Water purification system is not included.

21
Setup Manual

Remove the tubes と


[WASTE2A “WASTE1A”,
WASTE2B]のチ
“WASTE1B” , “WASTE2A” and
ューブを外し、分離回収タンクに
“WASTE2B”
挿入する , insert the separate
collection reservoir.

2. When you collect only the separate collection liquid in a separate collection reservoir
and discard other waste liquid in the outside of the analyzer.

1) Set the separate collection reservoir in the lower left part of the inside of the
analyzer and insert a separate collection tube.
2) Connect a drainage tube to “WASTE2A” “WASTE2B” connecter of rear panel.
* In this case, do not soak the tubing line tip of cooling unit in liquid.
When it is soaked, the water leaking happens by the remaining water which is
not drained from cooling unit.
*When the tank become full, an error message is shown.
Please discard the contents of the tank.

Separate
collection
reservoir
Cooling unit waste
line tube.

*The tubing of cooling


unit line and waste line
are combined after the
VE analyzer.

Remove the tubes of “WASTE2A” and


“WASTE2B” , insert the separate
collection reservoir.

22
WASTE1A and WASTE1B connecter

3) Insert the waste line tubes into the drain.

3. When you discard the separate collection liquid and other waste liquid in the outside
of the analyzer.

“WASTE1A” “WASTE1B”
and
“WASTE2A” “WASTE2B”

(1) Connect a drainage tube to “WASTE1A” “WASTE1B” and “WASTE2A” “WASTE2B”


connecters of the rear panel.
(2) Insert the waste line tubes into the drain.

23
Setup Manual

12.Connection with PC

Connection between PC and the analyzer is as follows.

Don’t forget to turn OFF the analyzer before connection between PC and the
analyzer.

Please refer to the instruction manual which is attached to a PC.

RS232C straight RS232C straight


cable (9 pins) cable Printer
(option)

*Please use RS232C straight cable for connection between the analyzer and the PC
and the printer.

The PC specification for analyzer.

OS Windows 7 Professional / Ultimate (32bit)


CPU Core 2 Duo 3.0 GHz or above
Memory 2 GB or above(Recommend 4GB)
HDD 60 GB or above

24
13.Connection of the power supply cable

The power supply cable of this device is connected as follows.

The construction of power supply cable is as follows.

Do not touch the power supply cable to avoiding the electrical accident.

13-1. Shipment state of the AC cable

Enlarged view
拡大

BIOLIS 50i Superior 本体


Analyzer

AC power supply cable is directly connected


ACケーブルは直出しとする Length 5 meters
長さ5m

AC power supply cable is directly connected with Analyzer and the length of the cable
is 5 meter. The tip of the cable is peeled like the above figure.
The electric cable of 50i must be connected to distribution board, and NEVER use 3P
or 2P plug.

25
Setup Manual

13-2. Connection
This machine is connected directly with a switchboard because consumption electricity
is 1200VA (1.2kVA) .

Voltage Connection
AC100V Breaker terminal is more than 15A
AC115V Breaker terminal is more than 13A
AC230V Breaker terminal is more than 7A

*1 When you connect other device along with Analyzer Superior with switchboard,
please consider consumption electricity.
*2 Please attach Fuse-free breaker and C class ground terminal to the switchboard
box inside. (less than 10Ω)
*3 Please connect a ground wire of Analyzer with a ground terminal of the
switchboard box.
*4 The distance between the machine and the switchboard box is less than 5 meter.
*5 Please confirm the voltage of breaker terminal which is within ±10% for each power supply
before connect the Analyzer Superior.

Switchboard

* Please use Crimp terminals and


tighten screws.

1. L (LIVE)

2. N (Neutral)

Ground

26
14.Software install

Software installation is as follows.


Please be sure to use Windows7 Professional or ultimate (32bt).
You cannot use 64 bit version.
There are two CDs (Disk1 & Disk2). Please be sure to install from Disk1.

14.1 Disk1 installation


1). Insert Disk1 to the PC and installation starts automatically.
2). Select language.
3). Installation starts.

4). Click “Next” button.

5). Click “install” button.

6). Click “Finish” button.

27
Setup Manual

Disk1 installation is finished. Then install Disk2.

14.2 Disk2 installation


1). Insert Disk2 to the PC and installation starts automatically.
2). Select language.
3). This screen is shown. Click “Next” button.

4). Please read “Software license agreement” and click “I accept the terms of
the license agreement”.

5). Please input serial number.

28
*If the number is “5000290212”, please input like the following photo.

6). Please input “User Name” and “Company Name”.

7. Please click “Install” button.

29
Setup Manual

8). Finally please click “Finish” button.

The icon of Analyzer is shown on your desktop.


Please click the icon and confirm whether the installation is completed correctly.

30
14.3 OS setting

1). Date format setting


(1) Click “Start” button in your computer and open “Control Panel” button.
(2) Click “Regional Options” icon in the screen.

(3) Click “Customize…” button.

(4) Click “Date” button.

Short date format;


Select “yyyy-MM-dd”, “MM-dd-yyyy”
or “dd-MM-yyyy”

Date separator;
Select one from “-”(hyphen) , “/” (slash)
or “.”(period)

31
Setup Manual

2). Number setting


(1) Click “Numbers” button in the screen.
At the first line, select “. ” (period) at Decimal symbol.
At the third line, select “, ” (comma) at Digit grouping symbol.

(2) Click “Apply” button and “OK” button.

3). Start-up and Log in screen


*To start-up, input “Service” to [User Name] and click Login button.

Note1 : The “S” of Service is a big letter.


Note2 : Default user name and password are followings.
Note3 : Refer to “Service manual” for the setup.

Priority level User Name Password


1. Service engineer level Service
2. Manager level Manager
3. Operator level User1
4. Simple operation level Easy

32
15.Application of electric power

1). Confirm the system power switch is OFF.


2). Turn ON the main breaker at the rear panel.
- Fans start running at the rear panel.
- Reagent cooling controller (reagent tray) starts operation.
- Start the ISE calibration measurement when the ISE module is
working on the instrument.

System power switch Main breaker

3). Turn ON PC main switch.


- After a while, the main screen appears.

Main screen
4). Turn ON the system power switch at the right side panel.
When measuring the ISE calibration :
Please wait turn ON the system switch until finish the ISE calibration
measurement (10 min.).

33
Setup Manual

The ISE calibration results will disappear if turn ON the system switch
during ISE calibration.

System power switch

- A message of [Program Down-Loading…] is shown in the center of


main control screen and loading (%) is also shown.
When loading is finished and displays 100%, the message disappears.

[Program Down-Loading…] screen


- Then a message [Parameter Down-Loading…] is shown and will be
cleared when loading finished. The system goes back to the Main
Control screen.
- The connection between PC and the analyzer is completed.
- The temperature monitor and the status monitor on the upper right
of the screen show [IDLE.] and [WARM-UP]

Main screen

34
16. Movement check

Initialize the system by following procedure. If the system is initialized


without trouble, the basic movements are considered all right.

1). Click Ready button at lower part of the screen.

Click

Main screen

- Each mechanical unit moves to the home position and then goes back to
the start position.
- During initialization the status monitor shows [INIT].

2). When initialization finishes without trouble, status monitor changes


from [INIT] to [Ready].

Note If any trouble occurs during initialization, an error


message will be shown.
In this case, please refer to “Chapter 6 Alarm function”
of Operation manual to check.

Initialization is possible even if the temperature monitor


shows [WARM-UP]

35
Setup Manual

17. Input temp parameter

Temp parameter sheet is attached in the back of the front right door.

Temp parameter is attached in this area.

Temp parameter sheet

1). Click Maintenance → Temp button.


Open the “Temperature Monitor screen”.
* If the analyzer turned OFF, you can not click this button.
* Be sure to turn ON the analyzer.
2). Click Temp Para button in Temperature Monitor screen.
“Temperature control parameter screen” appears.
3). Input “Temp” values to the following red circle.

Input “Temp” values to


this area.

Temperature Monitor screen Temperature control parameter screen

4). Click “Set” button and “Return” button.


36
18. Function tests of each unit.

This section is for checking the temperature control function and confirming
the relative position between probes and cuvettes.

18.1 Temperature control


- Check that the temperature controllers of reaction tray, reagent tray
and washing water pre-heater are working properly.
- This function check should be done around 30 minutes after the
system switch was turned ON.

1). Click Maintenance button at upper right of the screen.

Click

Main screen

2). Click Temp. button in the maintenance menu.

Click “Temp”

Maintenance screen

37
Setup Manual

3). The temperature monitor screen appears.

Temperature monitor screen

4). The normal temperature of the reaction tray (RCT) and the
washing water pre-heater (WSH) are the following.
- Reaction tray (RCT) : around 37 ±0.3℃
- Washing water pre-heater (WSH) : around 38 ±4.0℃
If the values are out of normal range, turn the system power off
and restart once again to operate, following the above mentioned
1 - 3.

5). Confirm that the reagent tray temperature (RGT) is around 8


Celsius degrees in the digital thermometer temperature display.
Confirm that the sampler temperature (STF) is around 15 Celsius
degrees in the digital thermometer temperature display.
(Open the right front door to read the display of the digital ther
mometer temperature.)

Check this temperature

Before VE After VE
Front view

38
18.2 Sample probe position (STM) confirmation
Check the positions of the sample probe, the sample cup and the
reaction cuvette by the following procedure.

- Confirm that the probe is straight, before starting position


confirmation.
If it is bent, correct it to be vertical.
- Set the sample tray and remove the reaction tray cell cover,
before starting confirmation.

1). Open the top cover.

CWS
MU1 MU2
STM

RTM1

RTM2

2). Click Ready button at lower part of the screen to initialize the
system.

Click

Main screen
3). After initialization, check the position of the sample probe by the manual
operation.

39
Setup Manual

4). The position confirmation on the manual operation screen

(1) Clicking Maintenance at upper right of screen, the following screen


appears.

Click

Maintenance screen
(2). Click M.Function button in the Maintenance screen.
Then the Manual operation screen appears.

Click

Manual operation screen

(3). Check that the tip of the sample probe, when it is initialized, is
in the center of the washing trough position. (Move the probe
down by pressing the arm lightly.)
- If the tip of the sample probe is not in the trough center, bend
the probe by fingers.

Sample trough

40
(4). Click “STM” in the manual function screen.

6. STM ROTATE SAMPLE OUTSIDE + RUN


19. STM UP/DOWN DOWN CUP + RUN

Check that the tip of the sample probe is in the center of the outside
sample cup.

11. STM UP/DOWN INIT + RUN

(5). Next, check the position of the sample probe by the following
procedure.
3. STM Rotate Reaction cell + RUN
16. STM Up/Down DOWN Reaction cell PUSH + RUN

Check that the sample probe tip is in the center of the reaction cuvette.

11. STM UP/DOWN INIT + RUN

41
Setup Manual

(6). Next, check the height of sample probe on the RCT.

- Confirm the height of sample probe which


keeps 9mm(±0.5mm) from the top of
reaction cell to the sample probe tip.

(7). Next, to check the position of the inside the sample cup of sample
probe, move the STF. Select STF from unit list.

10. STF CUP POSITION (IN) + RUN

(8). Then select STM from unit list and back to the STM function
screen.
7. STM Rotate Sample Middle + RUN
19. STM UP/DOWN DOWN CUP + RUN
Check that the tip of the sample probe is in the center of the inside
sample cup.

11. STM Up/Down INIT + RUN


3. STM Rotate Reaction cell + RUN
5. STM Rotate Washing Trough + RUN

Sample probe goes back to the home position.

42
18.3 Reagent probe position (RTM1,2) confirmation
Check the positions of the reagent probe (RTM1,2).

- Confirm that the probe is straight, before starting position confirmati


on.
If it is bent, correct it to be vertical.
- Set the R1 bottle and R” bottle before starting confirmation.
(R1 position number is “1” and R2 position number is “61”.)

The explanation of the procedure followed by “17.2 Sample probe position


confirmation” is as follows.

Manual operation screen

1) Check that R1 probe tip, when it is initialized, is in the center of the


washing trough position. (Move the probe down by pressing the arm
lightly.)
- If the R1 probe tip is not in the trough center, bend the probe
by fingers.

R1 trough

2) Using the manual operation function, move R1 and R2 probe to the


reagent bottle to check that the probe tip is in the center of the
bottle, and move it to the reaction cuvette to check that the probe
is in the center of the cuvette.

43
Setup Manual

For the manual operation, refer to “The position confirmation on the


manual operation screen” in “17.2 Sample probe position (STM)
confirmation”.

3) The procedure after the above mentioned “1. R1 probe position checking”.

10. RTM1 Up/Down INIT + RUN

6. RTM1 Rotate bottle (IN) + RUN


16. RTM1 Up/Down Down Reagent Min Stroke + RUN

4) Check that the R1 probe tip is in the center of the reagent bottle.

10. RTM1 Up/Down INIT + RUN


3. RTM1 Rotate Reaction cell + RUN
14. RTM1 Up/Down Down Reaction cell + RUN

5) Check that the RTM1 tip is in the center of the reaction cell.

10. RTM1 Up/Down INIT + RUN


4. RTM1 Rotate Washing trough + RUN
R1 probe goes back to the washing trough position.

44
6) Check the height of R1 probe on the RCT.

- Confirm the height of R1 probe which


keeps 9mm(±0.5mm) from the top of
reaction cell to the R1 probe tip

7) Next, to check the position of R2 probe, move the RGT. Select RGT
from unit list.
5. RGT2 Position + RUN

8) Then select RTM2 from unit list and goes to the RTM2 function screen.
5. RTM2 Rotate Bottle (OUT) + RUN
16. RTM2 Up/Down Down Reagent Min Stroke + RUN
Check that the tip of the RTM2 is in the center of the reagent bottle.

10. RTM2 Up/Down INIT + RUN


3. RTM2 Rotate Reaction cell + RUN
14. RTM2 Up/Down Down Reaction cell + RUN
Check that the tip of the RTM2 is in the center of the reaction cell.

10. RTM2 Up/Down INIT + RUN


4. RTM2 Rotate Washing trough + RUN
R2 probe goes back to the washing trough position.
Check the height of R2 probe on the RCT.

- Confirm the height of R2 probe which


keeps 9mm(±0.5mm) from the top of
reaction cell to the R2 probe tip

45
Setup Manual

18.4 Washing nozzle position confirmation


Washing nozzle position confirmation is done by the following
procedure.

The explanation of the procedure after “14.3 Reagent probe position


confirmation” is as follows.

CWS

Washing
nozzle
Position confirmation of washing nozzle

Select and click CWS from manual operation screen.


1) Confirmation of the descent movement.

2) CWS Up/Down, down Bottom Position + RUN


At this time, all of the probe tips not scratch inside cuvettes.

2. CWS Up/Down INIT + RUN


Check that all of probes move up smoothly.

46
18.5 Mixing unit-1,2 (MU-1,2) position confirmation
Mixing unit-1,2 position confirmation is done by the following
procedure.
The following explanation is given, assuming that the system
Initialization is completed.
For the manual operation, refer to “The position confirmation on the m
anual operation screen” in 17.2.

MU1
MU2

Stirrer

1) Select and click MU1 from unit list.

4. MU1 Rotate Cell (M1) + RUN


9. MU1 Up/Down Down Reaction Cell + RUN
5. MU1 Up/Down INIT + RUN
3. MU1 Rotate Washing Trough + RUN
Check that MU1 is in the center of the reaction cell and washing trough.

47
Setup Manual

2) Next, select and click MU2 from unit list.

4. MU2 Rotate Cell (M2) + RUN


9. MU2 Up/Down Down Reaction Cell + RUN
5. MU2 Up/Down INIT + RUN
3. MU2 Rotate Washing Trough + RUN
Check that MU2 is in the center of the reaction cell and washing trough.

3)Next, check the height of stirrer tip on the RCT.


(1)Using the manual operation function, move MU1 to the reaction cell
position.
4. MU2 Rotate Cell (M1) + RUN
(2) Put the check tool which is in parts box on the reaction cell.
- The tool slides and sets from the side of stirrer on the cell brock.
-Confirm the height of stirrer. The stirrer position is low if it up by
tool.
(3) Pull the tool. The stirrer position is high if the tool is able to pull
out.
- The tool can’t pull out if the stirrer position is correct position.

Insert from
the side
Confirm the stirrer
can’t pull out
Stirrer tuning
tool
Insert until the back

Please use the tool which is in parts box


for the height tuning. Loose the fixing
screw of stirrer and tune up the height.

Height tuning
tool
18.6 Barcode reader adjustment
Please refer to Appendix 4.

48
19.Priming

Prime operation is to fill the water supply line with pure water.

1. After initialization, click Maintenance button.

Click

Maintenance screen

2. Clicking User Maint. button in Maintenance, the following


Manual maintenance screen appears.

Click “Prim”

User maintenance screen

3. Click PRAIM button. The system starts priming automatically.


When Stop button is clicked, an emergency stop is given.

4. Check that every water supply line is filled with water and there is no
Leaking of water from the sample probe, the reagent probe and the
washing nozzle.

49
Setup Manual

Check that every water


supply lines are filled with
water.

50
20. Auto-gain and cell check

The Auto gain (adjustment of the intensity of light) requires after priming.
Need to check the CWS tubing which has no air bubbles.
Measure the Auto gain after 15 min from the lamp On.

①Clicking Maintenance button at upper right of the screen, the

following is shown.

Click “Auto-gain”

Maintenance screen
② Clicking Auto-gain button , the following screen is shown.

Click “Start”

Auto-gain screen
③Clicking Start button, Auto-gain starts.
④ Confirm the range of each values 28000~54000 after Auto-gain measurement.
Please try again if the values are out of ranges after checking the air bubble in
CWS, reaction cell and lamp conditions.

51
Setup Manual

⑤ Click Save .
⑥ Return to Maintenance screen by clicking Close button.
⑦ Click Ready .
⑧ Click Cell check. The cell check screen is shown.

Click “Start”

Cell check screen


⑨ Click Start . The cell check starts.
⑩ Confirm the cell check results.
Please try Auto-gain first if the cell check values are not correct. Next, try
Cell check again.
⑪ Click Save .
⑫ Click Close .

52
21. Test run

Test run is for confirmation of automatic operation of dispensing, mixing,


measurement and so on.
Please refer to Operator’s manual for the operation in each screen.

- If you make a TEST item, it is easy to use a saline solution


(0.9% NaCL) or de-ionized water instead of sample and reagent.

Main screen
21.1 Item parameter setting.
Click Item button and set the item parameter.
1) Select number “1” of item number to make “test” parameterin.
2) Input the first page of the item parameter screen with the
followings.

DATA INFORMATION CALIBRATION


UNITS ABS TYPE Factor
DECIMALS 0 FACTOR 1000
ANAYLYSIS
TYPE END method
MainW.length 1 340nm
Sub W.length 2 405nm
CORR
SLOPE 1.0
INTER 0.0

53
Setup Manual

Item parameter screen (Page 1)

After checking, click Next Page button.


3) Check that the second page of the item parameter screen is filled
with the followings.

ASPIRATION DATA PROCESS


KIND - Double READ START END
SAMPLE 7 MAIN 74 76
REAGENT1 VOL 140 SUB 33 35
REAGENT2 VOL 35 ABS. LIMIT
R1 Blank - Water-Blank LOW -3
HIGH 3
REACTION MONITOR ABSORBANCE LIMIT
0 LEVEL POINT 1 BLANCK CORREC 1
SPAN 3 ENDPOINT LIMIT 2

Item parameter screen (Page 2)

54
And click Save button and then Return button to go back to
the Main Control screen.

21.2 Reagent position setting


1) Click Reagent button in Main screen to check the bottle position in
the bottle parameter screen.

2) Select one round for current round at the reagent screen.

Reagent setting screen

3) After selecting round, clicking Round Name, the following screen is


shown. Name for round name.

4) Name for a current round and click Save button.

5) Click Return button to go back to the Reagent setting screen.

55
Setup Manual

The registered screen is shown as following.

6) Double-clicking TEST item form the Item List, the test count of reagent
is filled. Click Save button.

7) Click Set Current Round button after save the data.

8) The following message appears. Click OK button to save.

56
21.3 Calibration setting
Open Calibration screen at the upper right of screen. The current round
is displayed. Input “0” in the blank absormance manually and hit Enter
key.

Click

Input 0 and hit Enter key

Calibration parameter screen

After click Save button before change the screen.

21.4 Order
Click Order button in Main screen, and perform the order entry.
1) Input “1” in Tray number. Input “1” in Sample number.
2) Input “10” in # of Asp.
3) Click “TEST”, so the check mark appears on the left of it.

Input “1”. Input “1”. Input “10”.

Click

57
Setup Manual

4) Click G.Order button and then G.Order dialog box comes on.
5) Input “6” for End Sample No. and click OK button.

Input “9” and


click OK.

Click.

Order entry screen (Dialog box)

21.5 Sample and reagent setting


1) Put 20ml of saline solution (0.9% NaCL) or de-ionized water into R1
and R2 bottles and set it to No.1 position on the reagent tray.

2) Put 500 micro litter of saline solution (0.9% NaCL) or de-ionized water
into six sample cups respectively, and set them to No.1 to No.6
positions on No.1 sample tray.

Perform the initialization of the system by clicking Ready button in the Main
screen.

21.6 Test run start


1) Set the sample tray to the fixed position of the system.

2) Click Start button in the Main screen.


- The automatic movement will start.
- Run monitor is displayed on the Main screen.

3) If all the test results printed out are within the range of ±20
and the system stops automatically, the test run is considered to be
completed successfully.
- If the values are out of the range, repeat the priming.

58
22. Shutdown

End of operation and power OFF will be done, following the procedure
below.

1. Click Exit button at the upper right of the screen.

Click

- Program ending confirmation dialog box appears in the center of the


screen

Click

Confirmation dialog box

2. Click OK button in the dialog box.


[System shutdown] message appears in the center of the screen.

3. Turn OFF the system switch at the right side of the system.
- Even if you turn off the above system switch, the reagent cooling
system is still running.

4. Turn OFF the PC power.

59
Setup Manual

23. ISE unit setting

23.1 The structure of the ISE unit is shown below.

60
No. Name Code Contents

① Waste pump RP3 Waste Calibrator liquid from the pot


② Power on LED LED Green light shows power ON.
③ Cable cramp ━ Fixing cable.

Aspirate Calibrator and patient sample. Waste the


④ Aspiration pump RP2
liquid from electrodes.

⑤ Measurement unit ━ Composes electrodes and PCB.


⑥ Pinch valve 2 PV2 Change the line between Cal.1 and Cal 2.
⑦ Pinch valve 1 PV1 Stop the liquid flow through electrode.
⑧ Calibrator cap ━ Connect Calibrator pack.
⑨ Liquid ground ━ Connect the liquid to ground line.
⑩ Nozzle arm ━ Includes liquid level sensor PCB.
⑪ T3 housing ━ Includes driver PCB. Fix each units.
⑫ Calibrator pump RP1 Supply Calibrator liquid to the pot.
⑬ Pinch valve 5 PV5 Change the line between pot and overflow.
⑭ Pinch valve 4 PV4 Change the line between Calibrator 1 and air.
⑮ Pinch valve 3 PV3 Change the line between Calibrator 2 and air.
⑯ XZ component cover ━ Cover the belt and sensor.
⑰ Z axis motor ━ Move the sample nozzle up and down.
⑱ X axis motor ━ Move the sample nozzle horizontally.
⑲ Sample nozzle ━ Aspirate Calibrator and patient sample.
⑳ Calibrator pot ━ Supply Calibrator liquid and wash sample nozzle.

61
Setup Manual

1) Connector panel

62
2) Pump and valve panel

No. Name Code Contents

Connect the tubes between pump tube and Vinyl


① Mini-fitting ━
chloride tube

② Pump tube hook ━ Hook the mini-fitting, fix the pump tube.
③ Pinch valve PV Open and close the flow line.
④ Pinch valve head ━ Parts for open and close the pinch valve tube.
⑤ Spring housing ━ Control the pressure by spring.
⑥ Tube pressure ━ Press pump tube to roller housing.

Pour the liquid by drawing the pump tube based


⑦ Roller housing ━
on turning.
⑧ Y joint ━ Joint to change the flow by pinch valve.

Stainless steel pipe for connecting between


⑨ Tube joint ━
Pharmed tube and Vinyl chloride tube.

63
Setup Manual

3) Electrode

Before electrodes set After electrodes set

64
4) PCBs

Control PCB

Driver PCB

65
Setup Manual

66
5) Nozzle arm

67
Setup Manual

6) Nozzle unit

68
69
Setup Manual

7) Flow

Liquid ground
line

Electrode

Calib.
Waste pot
li

70
8) Tubes
Tube kind and length [mm]

71
Setup Manual

23.2 ISE setup preparation


Setup in program

① Click “System” in menu screen.

Click “Parameter” in system screen.

Check ISE レ

System parameter screen

② Check レ ISE

③ Click Save .

④ Click “Exit” and turn OFF and ON the instrument power switch.

23.3 ISE unit setting


① Set the ISE unit to the instrument from the top.
Please be careful with pinch and crush for the tubing and wiring
② Fix the ISE unit by 4 screws. Refer Fig.1.
③ Fix the nozzle unit by 4 screws. Refer Fig.2
④ Connect the nozzle unit cable to the ISE unit.
⑤ Set the calibrator pot on the nozzle unit.
⑥ Set the tube on the calibrator pot to hook. Refer Fig.2.
⑦ Connect the power supply cable (DC24V)
⑧ Connect PIO and SIO cables
⑨ Set the waste line tube to waste trough
⑩ Hook the calibrator 1 and Calibrator 2 on the instrument

72
Pan screw with cross hole
M3x8L(4)

Fig. 1

Confirm the nozzle and pot position.

Pan screw with cross hole


M3x8L

Hook

Fig. 2

73
Setup Manual

23.4 Measurement preparation


1. Pump tube setting.
Set the roller pump tube for three positions.
The spring housings set off before installation to avoid the tube pinch.

Spring housing
Spring housing “Set”
“Set”

Spring housing
“Set”

2. Setup the electrode.


Set the Reference, Na, K and Cl electrode.

① Remove the electrode from the box.

② Remove the seal tape of both side gently using tweezers.

Seal tape
Seal tape(both side)

Electrode(from left Na, K, Cl, Ref)

Notice :
1) Do not touch the metal part.
2) Remove the seals before setup. Remove the seal paste if it covers the hole.

74
Metal part

Take out the electrode

③ Set all the electrodes on a connecting plate..

Connecting plate

④ Keep holding the plate down, and insert the electrode .

Holding plate Push down

75
Setup Manual

④ Release the lock and up the holding plate gently.

Up the holding plate gently

3. Calibrator setting
Set the Calibrator 1 and 2.
Connect the caps to Cal1 and Cal2., hang the calibrator packs.

4. Waste line
Confirm the waste line tube connect to the waste bottle.

5. ISE probe position


① Power on the main switch at the back. Turn on the LED lamp(Green).
Start the liquid flow and Calibration measurement.
※Do not turn on the system switch on the instrument(10 min) when
measuring the Calibration. The Calibration results will disappear if
turn on the system switch.
② Turn On the PC and power On the system switch at the light side on
the instrument.
③ Initialize the instrument by clicking “Ready” in menu
④ Confirm the ISE probe position by manual operation.
⑤ Click Maint ⇒ ISE⇒ Maintenance. Open ISE maintenance screen.

Position setting

ISE maintenance screen

76
⑥Click Position setting button. The password input dialog will appear.
Click OK. The Nozzle position setting screen appears.

ISE nozzle position screen


- Confirm the probe tip position which is center of each sampling position
on the STF. (Inside, Middle and outside)
- Correct the position by loose the probe fixing screw.
⑦ Set a sample cup Sample position 1.
⑧ Click Out and Position 1 in nozzle position setting screen.
The position 1 on the STF turns to ISE sample aspiration position.
⑨ Click Cup and Adjust Init. The ISE probe down to the Position 1.
⑩ Confirm the nozzle tip position which down to the center of the
cup.
⑪ Click INIT . The ISE probe move to the Reagent pot .
⑫ Set a sample cup Sample position 37 .
⑬ Click Mid and Position 37 in nozzle position setting screen.
The position 37 on the STF turns to ISE sample aspiration position.
⑭ Click Cup and Adjust Init. The ISE probe down to the Position37.
⑮ Confirm the nozzle tip position which down to the center of the
cup.
⑯ Click INIT . The ISE probe move to the Reagent pot .
⑰ Set a sample cup Sample position C,E 1 .
⑱ Click In and Position C1 in nozzle position setting screen.
The position C1 on the STF turns to ISE sample aspiration position.
⑲ Click Cup and Adjust Init. The ISE probe down to the
Position C1 .
⑳ Confirm the nozzle tip position which down to the center of the
cup.
21 After position confirmation for outside, middle and inside, click

Close ⇒ Exit .

77
Setup Manual

24. Motor parameter setting

The pulse value of stop position and the sensing level for liquid level
sensor are able to input in this screen

Note: Please make a memo to keeping the original value before changing.

24.1 Motor parameter setting screen


① Click System button in Main screen.
② Click Configuration button and click OK in password dialogue
box.

Click

Click

③ Click Motor parameter setting button at right bottom site.

Click

④ Motor parameter setting dialogue box appears in the center of the screen

78
The following menu is shown.
Motor Para : The offset pulse value for INIT position is able
to change.
Pump Para : Do not change any values.
Stop Position : Stop position is able to change for each unit.
Action Para : The “level para” input for level sensor PCB is able
to use.
Dilution Para : Do not change any values.

⑤ Setting screen for motor control parameter. The offset pulse value for
INIT position is able to change. Click “Save” after changing the value.

⑥Setting screen for each pump parameter. Do not change any values.

79
Setup Manual

⑦ Setting screen for unit stop position. The units RCT, STF, STM, RTM,
RGT, MU1, MU2, CWS and ISE are able to change the stop position.
Click “SAVE” after changing the value.

⑧ Setting screen for action parameter. Use input the parameter for level
sensor PCB. Do not change other values.
Input procedure for level sensor PCB.
1. Open “Action Para” setting screen.
1) Click “System” in menu screen.
2) Click “Configuration” button.
3) Click “OK”.
4) Click “Motor parameter setting”.
5) Click “Action Para” button.
2. Click “Save”. The sensitivity and delay values input the level
sensor PCB.

80
⑨Setting screen for dilution volume. Do not change any values.

81
Setup Manual

Appendix1 (Confirmation procedure)


If you find out some problems in the analyzer during installation, please check the
following points.

1. The tubes of CWS, STM, RTM and ISE shouldn’t be disconnected.

Please check all the tubes connected firmly to CWS, RTM, STM and ISE. If those
tubes aren’t connected properly, air will be permeated and the result of measurement
will not be correct.

STM

Level sensor cable

RTM*

82
ISE*

*Please make sure to disconnect the level sensor cable from the PCB.

2. The tubes connected to the electronic valves shouldn’t be disconnected.

Please check all tubes connected to the electronic valves. If the tubes are not
connected deeply, the waste liquid will be leaked by the connection part.

83
Setup Manual

3. The tubes of WSH shouldn’t be disconnected and the screws shouldn’t be loose.

The tubes connected to WSH should be inserted deeply because if the tubes are not
inserted correctly, air will be permeated and the results of measurement will not be
correct.

4. The tubes of WSP shouldn’t be disconnected and the screw shouldn’t be loose.

The tubes connected to WSP are fixed by a screw. If the screw is loosened, the
solution and water will be leaked. Please tighten the screw carefully.

84
5. The PCBs are inserted correctly.1

Please check all the cables and connectors are firmly adjusted.

6. The cables connected to PCBs are inserted correctly.

There are a lot of cables connected to the PCBs. If those connecters aren’t connected
correctly, you cannot measure properly. Some error message will be appeared.

× ○

1 If you don’t know how to open the cover of PCB, please refer to Service manual.

85
Setup Manual

7. The tubes connected to SP, RP and PWP shouldn’t be loose.2

Please check all tubes connected to SP, RP and PWP. If the tubes are loosened, air
will be permeated.

8. The screws for fixing each unit shouldn’t be loose.3

Each unit is fixed by three or four black screws. If the screws are loosened, please
tighten by Phillips head screwdriver.

2 If you don’t know how to open the front cover, please refer to Service manual.
3 If you don’t know how to open the upper plane, please refer to Service manual.

86
9. The pillars for fixing all screws shouldn’t be loose.

All black screws are fixed to the gold metal pillar. Please tighten the pillar and the
black screws altogether screws. Even if you tighten only the black screws, the pillar
will not be fixed.

87
Setup Manual

Appendix 2 (Sensor connector panel)

Sensor connector panel

Separate collection reservoir Waste liquid reservoir

Acid detergent bottle

Waste tank reservoir


Alkaline detergent bottle

88
Appendix 3 (Rear panel)

Water purification system 6×8

Waste tank or Drainage 4×6

Separate collection reservoir 4×6

Waste tank or Drainage 4×6

89
Setup Manual

Appendix 4 (Sample Barcode reader adjustment)

1. Go to “Manual Function”.
2. Initializes STF
3. Input 37 at the STF barcode position number ( 13. STF BARCODE POSITION).

1. Initialization

13. Barcode position


Input No.37

4. Set tube with sample barcode label at the position number 37 of STF.
The barcode label should be placed in front of the barcode reader.

Barcode reader

5. Unscrew the two indicated screws and take out the upper panel.

Screws

Upper panel

6. Loosen the two indicated screws and make the barcode reader movable.
90
7. Keep pushing “TEST” button until LED turns on the light and change its mode.
(stop pushing it as quickly as the LED turns on the light)

Test button LED

Barcode reader

Screws

8. When the mode changes, you can check the conditions of the laser light of barcode.
The photo below indicates the bad conditions.

It turns green when


The barcode label
the laser light is on
detection LED turns
green when the good
detection. It turns red
when the bad detection.

9. Adjust the position of barcode reader. When it will be best position, all the LED
light turn on.
Tighten the screws when it is in the best position.
After adjustment, push the “TEST” button once and it will be normal mode.

The conditions becomes


Screws
best when the maximum
number of LED turn on.

10. Check the STF barcode position number 1, 55 and 19 after the adjustment by
using number 37. The method of check is same as you did.
91
Setup Manual

Appendix 5 (Reagent Barcode reader adjustment)

1. Press “Ready”
2. Manual Function - RGT1 and select No.6

3. Manual Function - RGT2 and select No. 6

92
4. Barcode Reader: Press [TEST] button for 5 seconds.

Loosen the 2 screws to move the bar-code reader.


Please adjust the light path touching to the position in the picture below.

93
Setup Manual

Appendix 6
(Setting of STM down stop position for using a micro cup on the tube.)

When you use the micro cup on the tube, please set the down stop position for
STM to adapt the using tube.

STM stop position (STM Up/Down) Pulse setting for a micro cup on the tube.

The default pulse value is set for 75 mm tube.


(Before VE analyzer = 2196, after VE analyzer = 2553)

Please set the pulse for using tube.


(1mm= 38 pulses)
The height becomes high according to decreasing value.

<STM stop position screen>

94
20140611

Maintenance procedure for the detergent lines.


If you don’t use the analyzer for a long time, there is possibility that the flow pass is blocked by
alkaline detergent crystallization.
To solve the problem, please follow the below procedure if necessary.

1. In case that you don’t use the analyzer for one week.
※ We recommend to do “Special wash 1” and “Special wash 2” to clean up the
detergent lines and cells.

1) Set 30% acid in detergent position No.1 and 15% alkaline detergent in
detergent position No.2 for sample, R1 and R2 probes.

< detergent position >

NOTICE Please use an optimal detergent depending on washing type.


Electrolytic substance contamination : 30% acid detergent
Protein contamination : 15% alkaline detergent
*Be careful not to misplace the detergent.

2) Choose “Maintenance” tab and click “User Maint”.

User Maintenance screen

3) Click “Special Wash1” and “Special Wash2” button. Special Wash will be started.
(About 26 minutes.) 90 cuvettes are washed in the same procedure as measurement,
The “Special Wash 1” is used for acid detergent line.
The “Special Wash 2” is used for alkaline detergent line.
Please select proper acid or alkaline detergent wash for contamination type if you use to
cancel the contamination.

Tokyo Boeki Medisys Inc. No.A-OG00018-01


20140611

2. In case that you don’t use the analyzer for a long time

(more than two weeks)

~~~ Before stop of the analyzer~~~

*Please do "Special Wash1 and Wash 2 " by de-ionized water.


(Exchange the detergent bottle into de-ionized water bottle.)

1) Remove the alkaline and acid detergent bottles and set the de-ionized
water to each bottles.
*Set the tube cap with level sensors into the de-ionized water bottles.
*In the inside of the left side front door, there are the detergent bottles.

Front view

2) Set de-ionized water bottles in detergent position No.1 instead of 30%


acid detergent and set in detergent position 2 instead of 15% alkaline
detergent for sample, R1 and R2 probes.

< detergent position >

3) Choose “Maintenance” tab and click “User Maint”.

User Maintenance screen

Tokyo Boeki Medisys Inc. No.A-OG00018-01


20140611

4) Click “Special Wash 1” button. Special wash will be started (About 26 minutes).

5) Click "Special Wash 2" button. Special wash will be started (About 26 minutes).
* All the detergent lines are filled by de-ionized water after “Special Wash 1 and
Wash 2”.

7) Keep on putting the tube cap with level sensor into the de-ionized water until the
analyzer will be started.

NOTICE Drying of the detergent line causes a clot.


Be careful not to be dry.

3. Incase before restart of the analyzer, exchange the de-ionized

water bottle into the detergent bottle.

~~~ Before restart of the analyzer ~~~

1) Remove the de-ionized water bottles and set the alkaline and acid detergent
bottles.
* Set the tube cap with level sensors into the detergent bottles.

NOTICE Be careful not to misplace the alkaline detergent and the acid
detergents.

Front View

2) Set 30% acid detergent in detergent position No.1 and 15% alkaline
detergent in detergent position No.2 for sample, R1 and R2 probes .

< detergent position >

Tokyo Boeki Medisys Inc. No.A-OG00018-01


20140611

NOTICE Please use an optimal detergent depending on washing type.


Metallic contamination : 30% acid detergent
Protein contamination : 15% alkaline detergent)
※ Be careful not to misplace the detergent.

3) Choose “Maintenance” tab and click “User Maint”.

4) Click "Special Wash 1" button for acid detergent line. Special wash will be started
(About 26 minutes). 90 cuvettes are washed in the same procedure as
measurement,

5) Click "Special Wash 2" button for alkaline detergent line. Special wash will be
started (About 26 minutes). 90 cuvettes are washed in the same procedure as
measurement,

* Each detergent lines are filled by acid and alkaline detergent after “Special
Wash 1 and Wash 2”

Tokyo Boeki Medisys Inc. No.A-OG00018-01


Setup Manual Version 1.11
Feb. 10, 2016
Printed in Japan