Professional Documents
Culture Documents
G-Scan Installation Guide Rev8
G-Scan Installation Guide Rev8
G-Scan Installation Guide Rev8
Service Manual
Installation Guide
© ESAOTE S.p.A.
AG 2011
The reproduction, transmission or
use of this document or its contents
is not permitted without express
written authority. Offenders will be
liable for damages. All rights,
including rights created by patent
grant or registration of a utility model
or design, are reserved.
Replaces: VER.G English
Doc. Gen. Oct/2011
8300096031 VER.8
Document revision level
The document corresponds to the version/revision level effective at the time
of system delivery. Revisions to hardcopy documentation are not
automatically distributed.
Please contact your local ESAOTE office to order current revision levels.
Disclaimer
The installation and service of equipment described herein is to be performed
by qualified personnel who are employed by ESAOTE or one of its affiliates
or who are otherwise authorized by ESAOTE or one of its affiliates to provide
such services.
Assemblers and other persons who are not employed by or otherwise directly
affiliated with or authorized by ESAOTE or one of its affiliates are directed to
contact one of the local offices of ESAOTE or one of its affiliates before
attempting installation or service procedures.
Part 0 Table of Contents
3
8300096031 VER.8 Table of Contents
0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0
4
Welcome to the MR System Tune-Up .................................. 119 Visual Inspection Checks _____________________ 132
Overview ___________________________________119
Outside Marking Checks ...................................................... 132
Scout Acquisition ___________________________120 Cover and Mechanical Component Checks ......................... 132
5
8300096031 VER.8 Table of Contents
0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0
6
7
8300096031 VER.8 Table of Contents
0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0
8
Disassembly ________________________________277
9
8300096031 VER.8 Introduction - Read this first
1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1
10
Introduction
WARNING Make sure you use only non magnetic tools
The following general safety notes are supplemented by specific near the magnet. Do not put any other
descriptions in the documents supplied by us, e.g. product magnetic object near the magnet (e.g. the
documents, ARTD (general guidelines) as well as by written metal support for transporting the magnet).
instructions in specific cases. Be careful that screws, etc. or other metallic
All safety instructions must be observed in the performance of work objects do not fall into the equipment, because
and tests; in addition, the prescribed country-specific requirements they could cause a short circuit.
(e.g. occupational safety and accident prevention regulations) must
be complied with. Changes made in products delivered by us
Changes made in products/systems delivered by us must not be
General Safety Notes implemented without our written release. This applies in particular to
changes which may affect the mechanical and/or electrical safety or
radiation-protection properties of a product (e.g. changing of safety
distances, removal of locks/instructions, enlargement of the
WARNING Magnetic objects may be attracted by the radiation-exit window on the lead cone, etc.).
magnetic force field, and drawn in the
direction of the gantry causing possible
injuries to the patient being examined. Electrical installations
Large moving magnetic objects, such as a Electrical installations of medically used rooms in the Federal
stretcher, for example, could disturb the Republic of Germany must be made in accordance with the
uniformity of the magnetic field inside the requirements of VDE Regulations 0107.
magnet, causing artifacts to appear on the For all other countries, the regulations to be complied with (national
diagnostic image. and VDE regulations) must be taken from the installation sheets for
the project concerned.
11
8300096031 VER.8 Safety - Introduction
2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2
12
13
8300096031 VER.8 Safety - General Safety Notes
2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2
14
NOTICE All screws must be secured in accordance with Fig. 2: Electrostatically sensitive devices
the corresponding data. If "Loctite" has to be used
to secure screws, this is stated in the text.
Batteries
For replacement of batteries, special preventive measures apply.
In the case of lithium batteries, there is as risk of explosion if they are
incorrectly installed.
The notes concerning handling and disposal must be observed in
each case.
Never
● Use defective or damaged tools
● Block emergency exits
● Stretch ropes by twisting them
● Knot ropes or chains
● Stand under heavy loads
● Exceed the load capacity of devices
● Route or stretch ropes, chains, or bands over sharp edges
● Allow personnel to stay in areas where heavy loads are being
moved
15
8300096031 VER.8 Safety - General Safety Notes
2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2
16
If ferromagnetic materials or tools are introduced into the vicinity of Fig. 4: System label
the magnet these objects will become projectiles that could be lethal
to personnel working near the magnet.
Place the following labels on the outside of the pavilion door near the
handle.
17
8300096031 VER.8 Safety - General Safety Notes
2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2
18
Precautions/restrictions implemented
● The project manager has informed employees working in the
vicinity of the magnet about its pending installation
● The prescribed non-magnetized tools are available on-site
● Access to the magnet and examination room must be restricted as
soon as a magnet is installed
Comply with the following signs
19
8300096031 VER.8 Safety - General Safety Notes
2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2
20
Phantom fluids fluid must be disposed of in a container. The same container is used
for the water used to wipe down the contaminated area, the
The functions of MR measurement phantoms require that they be disposable gloves or the water used to rinse contaminated parts of
filled with saline hydrous solutions. When the MR phantoms are used e.g. lab coats or clothing.
properly, operating personnel will not come into contact with the MR
phantom fluid. First aid
Solution ingredients If personnel come into contact with MR phantom fluids despite
following the safety regulations, the following first aid measures apply:
The phantom fluids contain the following ingredients:
In the event of contact with skin, the skin affected should be rinsed
NiCl + NaCl + distilled water thoroughly with soap and water. The contaminated water must be
poured into a bucket, the bucket labeled accordingly, and the water
Handling and storage of phantoms disposed of. Contaminated clothing must be taken off immediately
To prevent damage, the phantoms should be transported, stored or and disposed of.
used at temperature levels between -5 °C to 50 °C. In the event of contact with the eyes, they should be rinsed for
approximately 10 - 15 minutes under running water keeping the
Disposal and use of leaking phantom containers eyelids open. If aerosols are inhaled, the person should be moved
immediately into the fresh air and a physician should be consulted.
The use of leaking phantoms is prohibited. Only authorized
If fluid is swallowed, the person should drink plenty of water and
companies or customer service are permitted to dispose of the
vomiting should be induced. A physician must be consulted
phantom fluid in compliance with the guidelines and safety
immediately.
instructions of the respective country (ARTD-002.731.20.01.02).
21
8300096031 VER.8 System Installation -
3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3
22
Specifications The aim of the Radio frequency Shielding Pavilion is to protect the
inner volume against external electromagnetic fields.
Main characteristics The Shielding Pavilion has been designed to shield a Nuclear
Magnetic Resonance System with an open permanent magnet.
The external field's attenuation performance of the shielding Pavilion
Tab. 1: Technical specifications is greatly influenced by the type and state of the metallic contacts
(copper fingers).
The geometric dimensions and the installation procedure for a
Feature Data
standard Pavilion are given in detail below.
Attenuation 70 dB (f0 = 10 MHz) The main characteristics of the Shielding Pavilion are:
Size 4802.5 x 3850 x 3240 (±10 mm) (outer – Total weight: 1200 kg
dimensions excluding the filter, the ramp at the – RF attenuation: 70 db at 10 MHz (E field)
pavilion entrance and the external door handles
– Transparency (ratio empty/full): 32%
and rail)
Electric continuity is ensured by finger contacts partly placed on the
Door Two sliding doors. Necessary forces to move frame of the sliding door and partly on the jambs of the fixed part. The
doors in the range of usual forces required to contact between the sliding shutters is achieved through a finger-knife
move, open and close shielding room doors less System on their doorstop surface.
than 20 kg
Self-supporting wooden panels placed on steel panels covered with
vinyl compose the floor of the Pavilion.
Assembling
Refer to the draws provided with the Pavilion parts.
Painting
A painting kit is provided with the Pavilion. Use it to complete Pavilion
installation, complying with the Pavilion color. If necessary use the
potassium nitrate solvent.
Grounding
Refer to Console installation paragraph.
Warning/prohibitive signs
Refer to Location of warning/prohibitive signs paragraph.
23
8300096031 VER.8 System Installation - Pavilion installation
3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3
24
System Installation
PLEASE COMPLETE THE FOLLOWING FORM MARKING OFF ALL THE PERFORMED ITEMS AND E-MAIL/FAX BACK TO ESAOTE.
25
8300096031 VER.8 System Installation - System Installation
3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3
26
27
8300096031 VER.8 System Installation - System Installation
3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3
28
Fig. 11: Gravity Point (front view) Fig. 12: Gravity Point (lateral view)
29
8300096031 VER.8 System Installation - System Installation
3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3
30
Magnet Lifting with Crane Fig. 15: Lifting the Magnet and the Crate
Follow this procedure only if you have received a Magnet packed with
crate otherwise skip it and go to the next paragraph.
With Crate
● If the crate has to be lifted by a crane, you must proceed very
carefully and as shown in the next figure. The plastic stripes
(metallic chains are allowed only in the upper part of the crate)
must be set according to the signs present on the Magnet crate,
where reinforcements have been placed. The Gravity Point of the
Magnet is indicated by cross signs on the crate.
31
8300096031 VER.8 System Installation - System Installation
3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3
32
Fig. 17: Lifting the Crate Upper Parts Fig. 18: Lifting the Crate Upper Parts
Fig. 19: Lifting the Crate Upper Parts Fig. 20: Magnet on Pallet
● The only available side for forklifts is the Magnet rear side
● The Magnet is joined to its pallet by four brackets placed on the
corners (as shown in the next figure)
33
8300096031 VER.8 System Installation - System Installation
3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3
34
● It’s possible to lift the System using the kit code 910 3014 000 that Fig. 22: Magnet and Eye Bolts Positions
contains four Eye Bolts and four Shackles. These items are
certified and Stainless Steel made. In the next figure one Eye Bolt
is shown and the certification numbers are visible. This kit is
provided by ESAOTE only if purchased
WARNING This procedure must be carried out only using ● Mount the four Eye Bolts and join the four Shackle to them, the
ESAOTE certified parts! four plastic stripes have to be inserted in advanced, as shown in
the next figures
● The positions useful to screw the Eye Bolts onto the Magnet Iron
Yoke are shown in the next figure
Fig. 23: Eye Bolt with Shackle and Plastic Stripe Fig. 24: Eye Bolts with Shackles
● Now a crane can easily lift up the system hooking the four stripes
with the Crane
35
8300096031 VER.8 System Installation - System Installation
3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3
36
● Put down the Magnet on the dedicated carts and guide on a flat
floor and move it to its final destination NOTICE THE MOVING KIT DESCRIBED IN THESE
PAGES HAS TO BE CONSIDERED AS AN
Fig. 25: Guide (to be placed under the rear Magnet baseside)
EXAMPLE. A SIMILAR KIT MUST BE
PROVIDED AND USED ONLY BY THE
PROFESSIONAL MOVERS THAT PERFORM
THE SYSTEM MECHANICAL INSTALLATION
Fig. 26: One of the three carts (two of them have to be placed under the
front left and right magnet base sides)
● Place wooden thickness between the guide, the carts and the
Magnet base and remember that the Cart thicknesses must meet
the base front feet frames
● When the Magnet is in the final destination lift it using the jacks. To
better move the Magnet using the Carts and the Guide could be
useful to remove the plastic and paper protection because they
hide cables and fragile parts and they increase the shape of the
Magnet. If you remove them leave the blue plastic protection
visible in the next figure till to reach the system final destination
and remove it only when cabling the System
37
8300096031 VER.8 System Installation - System Installation
3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3
38
Fig. 29: Correct Magnet Position inside the ESAOTE Pavilion Fig. 30: One of the four Jacks
Place them in the Magnet front and rear sides and act on them at the
same time in order to keep the Magnet levelled.
WARNING Keep the Magnet levelled rising the four jacks
at the same time.
Then remove the carts and the guide and place the Magnet aluminum
feet in the locations shown in the next figure in order to fit in the
Magnet base frames present in the bottom side.
The feet have different dimensions: one big, two medium and one
small. Be aware that the smallest foot must be placed with the
orientation shown in the following figure.
39
8300096031 VER.8 System Installation - System Installation
3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3
40
● Open the Electronics Unit crate, remove the locking bolts from its
base and lift up the Electronics Unit using a crane and the
provided eyehooks (as shown in the next figure) then remove the
crane base and place the Electronics Unit on a plane floor. When
done remove the eye hooks (store them in a safe place for future
use) and place, in the same positions, the provided bolts
● Move the Electronics Unit (using its wheels) in the pre defined
position (outside the System Pavilion)
Fig. 34: Electronics Unit ● Adjust the Electronics Unit feet in order to level it. When the
Electronics unit is correctly levelled lock the foot pivot nuts
● Open the rear bottom panel to have access to the Electronics Unit
connection panel
● Some cables are already connected and stored into the
Connection Panel cavity: don’t forget to put them through the
bottom Electronics Unit opening (visible in the next figure)
41
8300096031 VER.8 System Installation - System Installation
3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3
42
Fig. 36: Electronics Unit Connection Panel Fig. 37: Electronics Unit Connection Panel
43
8300096031 VER.8 System Installation - System Installation
3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3
44
Computer Unit installation Fig. 41: Signal Box and SB15 Optical Fiber Connector of the ACM card
● Open the Monitor box and place it on the installed table and
connect it to the power plug and PC Unit SVGA digital output
using the provided cable then connect mouse and keyboard
● Connect the provided optical fiber data cable (orange) from the
PC unit DPSM card (PC1 connector) to the ACM card of the
Signal Box present in the Electronics Unit (SB15 connector) as
shown in the following two figures
Fig. 42: PC Unit ● The PC has no ground connection with the Magnet because is
directly powered using the Hospital line while the data connection
is provided by the optical fiber cable
45
8300096031 VER.8 System Installation - System Installation
3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3
46
Cabling and Grounding Fig. 44: Filter Panel internal side (Pavilion right side installation example)
47
8300096031 VER.8 System Installation - System Installation
3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3
48
Fig. 47: Electronics Unit front side (Scheme code 910 2681 000 SC Rev.E)
Fig. 48: Electronics Unit left side
49
8300096031 VER.8 System Installation - System Installation
3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3
50
51
8300096031 VER.8 System Installation - System Installation
3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3
52
Fig. 51: Magnet Cabling 1 (scheme code 970 0021 000 SC Rev.7)
Fig. 52: Magnet Cabling 2
53
8300096031 VER.8 System Installation - System Installation
3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3
54
55
8300096031 VER.8 System Installation - System Installation
3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3
56
57
8300096031 VER.8 System Installation - System Installation
3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3
58
59
8300096031 VER.8 System Installation - System Installation
3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3
60
Fig. 60: Magnet Base Covers (the red numbers show the assembling order)
61
8300096031 VER.8 System Installation - System Installation
3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3
62
Fig. 61: Magnet Covers (the red numbers show the assembling order)
Fig. 62: Magnet Covers (the red numbers show the assembling order)
63
8300096031 VER.8 System Installation - System Installation
3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3
64
Magnet Accessories installation ● Take the Bed Brackets (both of them) and remove the Mechanical
Stops placed at the end of them (only rackwork side): if washers
are present take care of their position to be able to correctly
CAUTION From System 5100 included, Magnets are sent re-install them during this procedure
without Bed Brackets ● Insert the Bed Brackets (both of them) into the Magnet Rails (front
and rear) till the stop them on Magnet Translation Brake (to the
● Be sure that the system is correctly installed inside the cage and right side as much as possible) as shown in the following figures
the Electronics Unit and PC Unit are installed, cabled and properly
working
CAUTION BE SURE TO INSERT THE CORRECT BED
● Rotate the Magnet in horizontal position by using the CMR BRACKET SIDE INTO THE MAGNET RAILS:
commands (for more info please refer to the Installation Guide, THE SIDE WITH RACKWORK
Tune-Up chapter, CMR Use and CMR Checks paragraphs)
● Remove the Service Hardware Key if connected to the CMR
● Disassembly the Bed Translation Potentiometer placed on the Fig. 64: Rackwork Inserting
Magnet front side by removing the indicated bolts and then by
disassembling one side of the joining piece that connects the
potentiometer axis with the translation axis
65
8300096031 VER.8 System Installation - System Installation
3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3
66
Fig. 67: Rackwork Joined to the Gear Fig. 68: Bed Bracket Stop
● Place back the Bed Brackets Stops (place back washers as well if
you found them removing the Stops), unlock the Translation Stop
by acting on the CMR in Service modality and push the Bed
Brackets completely to the left side as shown in the following
figures
Fig. 69: Bed Bracket Alignment Fig. 70: Bed Bracket Alignment Check
● Join back the Moving Axis by tighten the bolt you have unscrewed
before in order to be able to slide the Bed Brackets by using the Fig. 71: Bed Bracket Alignment Washer
CMR and do it back and fourth for some times then move the Bed
Brackets completely to the left side and verify their alignment by
using a plastic caliper as shown in the following figures:
– If you have the same quotes on both sides go to the next step
– If you have a different quotes (higher than 3mm difference), put
one or more washers behind the stop of the Bed Brackets with
minor quote and repeat the test till to have the same quotes on
both of them then go to the next step
67
8300096031 VER.8 System Installation - System Installation
3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3
68
Fig. 72: Bed Bracket Alignment Check Fig. 73: Bed Bracket Alignment Check
● Slide the Bed Brackets completely to the left side and mount back
the Translation Potentiometer on the Magnet front side, then move
the Bed Brackets back and fourth some times
● Apply the same procedure on the right Bed Brackets side by
sliding the Bed Brackets completely to the right side
Fig. 74: Magnet Fig. 75: Rail Bolts
69
8300096031 VER.8 System Installation - System Installation
3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3
70
Fig. 76: Base Middle Right and Left Covers Fig. 77: Base Middle Right and Left Covers Mounted
● Move the Bed Table brackets totally on the left side and mount the Fig. 79: Magnet Cavity Lower Right Cover
Magnet Cavity Right (3) Lower cover on the Magnet
71
8300096031 VER.8 System Installation - System Installation
3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3
72
● Remember to join the aluminum plate to the cover as shown in the Fig. 81: Magnet Cavity Lower Cover Mounted
following figures
● Place back the Bed Table Brackets on the Knee position nd rotate
the Magnet 90° then place the Magnet Lower cover (4) on the
Magnet
73
8300096031 VER.8 System Installation - System Installation
3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3
74
Fig. 84: Magnet Lower Cover Mounted ● Open the Magnet Lower cover front opening to join to the magnet
brackets as shown in the following figure
● Set the Magnet 0° and mount the Magnet Cavity Left (5) Lower
cover on the Magnet
Fig. 86: Magnet Cavity Lower Left Cover Fig. 87: Magnet Cavity Lower Cover Mounted
75
8300096031 VER.8 System Installation - System Installation
3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3
76
Fig. 88: Bed Table ● Place the Bed Table on the Patient’s Bed Brackets as described in
the following procedure. In order to easily remove the Bed Table in
case of Shimming correction place a bolt every three holes along
the Bed Table edges and complete the assembling placing all the
bolts after the Shimming correction
– If provided, take the bed cart as shown in the next figure
– Place the Bed on the cart and move it into the RF Cage then lift
the Bed and join it to the Magnet Brackets. When done
disassembly the cart and store it in a safe place: it can be useful
once again during maintenance or fixing
77
8300096031 VER.8 System Installation - System Installation
3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3
78
Fig. 92: Bed Placed on the Cart Fig. 93: Bed Table Bolts
NOTICE Do not put all the Bed table bolts but only one
every three holes because you should have to
remove it to perform the Magnet shimming NOTICE Periodically check the tightening of the screws
adjustment. Complete the bolts installation only present on the teflon rails (upper and lower sides-
after the shimming procedure. lower by sliding the bed laterally). Refer to the
next two images for the screw positions.
Fig. 94: Upper screw positions ● YOU CAN STOP HERE THE COVER ASSEMBLING AND
PROCEED WITH THE SYSTEM TUNING-UP. START FROM
THIS POINT AFTER COMPLETING OF THE TUNE-UP
PROCEDURE
● Mount the finger protection cover on the back side of the bed
upper table as shown in the following figure
● Place the Magnet Cavity Right and Left covers (6 and 7).
Remember to place the Green Stripes on the covers before to
mount them on the Magnet
79
8300096031 VER.8 System Installation - System Installation
3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3
80
Fig. 97: Magnet Cavity Right Cover 6 with green stripe mounted Fig. 98: Aluminum plate
81
8300096031 VER.8 System Installation - System Installation
3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3
82
Fig. 102: Heater Protection Plastic Stripes Fig. 103: Magnet Cavity Upper Cover with green stripes mounted
83
8300096031 VER.8 System Installation - System Installation
3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3
84
– Assembly and mount the switch top cover, then mount it on the
Magnet lower inner right and left covers as shown in the next
figures
85
8300096031 VER.8 System Installation - System Installation
3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3
86
– Close the top cover by keeping it close and screwing the lateral ● Complete the Bed Safety Switch by mounting the two (left and
screw till to meet the metallic parts fixed to the Magnet inner right) lateral protections (to avoid that safe belts can slide into the
lower covers. Then release the cover and verify if it remains Bed Switch during Bed sliding movements) as shown in the next
close and that moving the sliding part to the bed completely figure. Self tapping screws have to be used
inside the Magnet cavity it touches the switch top cover and
consequently close the switch (you have to hear the click noise):
if not adjust the switch height till to get it correctly working
Fig. 110: Swtich Protection Cover (right one) Fig. 111: Magnet Top Cover
● Place the Magnet Top Cover (9) and join it to the Magnet Joke
87
8300096031 VER.8 System Installation - System Installation
3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3
88
Fig. 112: Magnet Top Cover Mounted Fig. 113: Base Cover Bracket
● Before to mount it, place the velcro stripes in the positions shown
in the following figure
89
8300096031 VER.8 System Installation - System Installation
3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3
90
Fig. 116: Base Right Lower with velcro stripes Fig. 117: Base Right Lower Cover Mounted
● Place the base Left Lower cover (12) on the Basement and the
Magnet Base Lower Central cover (13)
Fig. 118: Base Left Lower Cover Fig. 119: Base Left Lower Cover with velcro stripes placed on top of the front
side
91
8300096031 VER.8 System Installation - System Installation
3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3
92
Fig. 120: Base Left Lower Cover with velcro stripe placed Fig. 121: Base Left and Central Lower Covers Mounted
● Place the Base Rear Lower cover (14) on the Magnet Base
Fig. 122: Base Lower Rear Cover Fig. 123: base Lower Rear Cover (internal view)
93
8300096031 VER.8 System Installation - System Installation
3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3
94
Fig. 124: Base Lower Rear Cover Mounted Fig. 125: Base Rear Upper Cover with Bracket and Corner Completion
Covers
● Place the Base Rear Upper cover (15) to the Magnet base and
close it with the Round cover (16). Before to mount the Rear upper
cover, join to it the internal bracket and the corner competition
covers as shown in the following figures
Fig. 126: Base Rear Upper Cover Mounted Fig. 127: Base Rear Upper Cover and Round Cover Mounted
● Place the Magnet Front Lower cover (17) on the Magnet front side
joining it to the Magnet
95
8300096031 VER.8 System Installation - System Installation
3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3
96
Fig. 128: Magnet Front Lower Cover ● Place the Magnet Lower Competition cover (18) on the Magnet
Lower cover
● Place the Bed Termination covers and the Head Protection cover
97
8300096031 VER.8 System Installation - System Installation
3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3
98
Fig. 134: Bed Patient Head Protection Cover Fig. 135: Finger Protection Cover
99
8300096031 VER.8 System Installation - System Installation
3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3
100
● Check if the Patient’s belts are correctly installed on the bed table
as shown in the previous figure
● Place the Stairs in front of the system and place the yellow tape to ● Place the Cushions on the Bad Patient
define the stairs maximum inner position as shown in the next
figures
Fig. 141: Bed Cushions
101
8300096031 VER.8 System Installation - System Installation
3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3
102
103
8300096031 VER.8 System Installation - System Installation
3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3
104
105
8300096031 VER.8 Start Up - Prior to switching on
4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4
106
107
8300096031 VER.8 Start Up - Prior to switching on
4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4
108
Adapting transformer to the line voltage Fig. 147: Insulating Transformer Settings (left Electronics Unit cover off)
109
8300096031 VER.8 Start Up - Prior to switching on
4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4
110
(Ground Fault Interrupter). This device is for breaking a circuit cable can be removed only by means of a special tool (e.g. a
automatically when it senses fraction of an Ampere leakage current screwdriver).
flowing to earth in excess to predetermined amount. Grounding is required. An electric wire with a minimum nominal
The system must be connected, through an UPS system, to a 25A CB cross-section of 6 mm2 must ground the equipment (according to
(Circuit Breaker) that is connected to the previously described RCCB. EN 60601-1).
The CB is a re-settable switch that is tripped either by long term over
current operated by a thermal mechanism or by short term over
● Code 8820061001 it is the IMQ one and must be used in Europe:
current by a magnetic mechanism.
– 10 m long
The CB must comply with the following specifications:
– The yellow-green wire is the grounding wire: it must be
● Clearance distance between different polarity of residual current circuit
connected to the ground terminal of the main line.
breaker must not be less than 2.5 mm
– The brown wire is the phase wire: it must be connected to the
● Creepage distance between different polarity of residual current circuit
breaker must not be less than 4.0 mm
phase terminal of the CB
An UPS systems is always required as, in case of failure of the main – The blue wire is the neutral wire: it must be connected to the
power supply, the magnet needs to be permanently heated to keep its neutral terminal of the CB
temperature and you must be able to move the bed patient in order to – Nominal cross-section of each of its wires is 6 mm2
let the patient comes out from the Magnet.
The UPS system must be selected in order to guarantee at least 3kW CAUTION Be sure to correctly identify the phase and
for 10 minutes (refer to the site planning guide for more info). neutral wires when connecting the equipment
or when wiring the CB.
CAUTION The System has to be connected to the UPS at
the end of the Magnet warming up process. The MTCM and the TPM modules will start to heat the Magnet
automatically after that the System being connected to the main line
Charge the UPS batteries before to connect
and will take approximately 36 hours to reach its rated conditions.
the System to it.
If the modules are warming up the Magnet the TPM leds will be like
shown in the next figure: all the leds on (one central led may on and
These breakers must be approved in according to the international blinking as well) and the middle ones are very bright (that means
and/or national and/or federal and/or local standards in force and they maximum sent power).
must be installed in a wall-mounted sheltered box near the unit.
Wiring between the system and the CB must be via the three-wire
cable supplied with the system itself. It must be cut to the appropriate
length and cable end sleeves must be inserted before inserting them
into the terminal connection. The equipment must be permanent
connected to the Mains (according to EN 60601-1); therefore the
Fig. 150: Correct Led Status at the beginning of the warming up process If the leds are not on check the System Power supply and the
connection of all the cables related to the Heating chain (such as all
the cables connected to the TPM, ACM, Filter Panel and Magnet).
If the magnet temperature is below 0 degrees at the time of
installation the NTC sensor operates as an open circuit and the
MTCM module doesn't provide any power to the heaters. In this case
the warming up will take much longer because the Magnet has to
reach 0°C degrees before that the Electronics warming up can start.
If additional optional lights are installed in the pavilion, you must install
a dedicated separating switch and insert the power cables into a
dedicated duct to avoid electric shocks. The dedicated plug for the
optional lights is placed on the filter panel and has the following
characteristics: 250 V MAX and 5 A MAX.
CAUTION All the main power cables and the cables from
the Electronics unit to the filter panel and from
the filter panel to the magnet must be inserted
under the floor or into a duct that must be fixed
to the floor or the wall.
Spare Plug
If you have to provide power supply inside the Pavilion may be to
power on lights, you have to use the Spare Plug of the filter Panel
running a cable from the Mains to the Filter Panel and from the Filter
If the leds are correctly on, verify that the 220V to power on the Fast Panel to the lights. The cable must be double shielded and with a
Heaters are present on the Magnet Fast heater connectors that are nominal section of 1.5 mm2.
the same points shown before where the blue and brown cables are
connected.
111
8300096031 VER.8 Start Up - Prior to switching on
4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4
112
● Fill in the Installation and Quality Form with the test results
113
8300096031 VER.8 Start Up - Prior to switching on
4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4
114
They are the measurements in normal condition (NC) and single fault
conditions (SFC) of the earth leakage current, and of the patient
leakage current in normal condition (NC) according to the
EN 60601-1:1990, Clause 19.
The patient leakage current measurement coincides with the
enclosure leakage current measurement.
The measured value of the earth leakage current NC shall not exceed
5 mA.
The measured value of the earth leakage current SFC shall not
exceed 10 mA.
The measured value of the patient leakage current NC shall not
exceed 0.1 mA.
● Fill in the Installation and Quality Form with the test results
115
8300096031 VER.8 Start Up - Prior to switching on
4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4
116
Final Checks
117
8300096031 VER.8 Software Update - Introduction
5 5 5 5 5 5 5 5 5 5 5 5 5 5 5 5 5 5
118
General Overview
The calibration parameters of the System must be checked both at
Welcome to the MR System Tune-Up the time of installation and whenever scheduled maintenance is
The tune-up procedure includes a number of adjustments and checks performed. At these times, quality tests will also have to be carried out
for obtaining optimal performance and good image quality from your to guarantee the required standard of performance, especially
MR system: remember to proceed with care even though most of concerning the quality of the resulting images. Final inspection is the
the procedures run automatically. first testing of the program and it will be followed by other checks
performed during service.
An installation protocol must be completed in all of its parts. A copy of
CAUTION Read carefully the complete introduction of
this protocol must be returned to the central service offices.
this chapter before to proceed with the System
Calibrations. All calibration results relating to the various parameters of the System
and those concerning the quality tests are to be recorded in this
protocol to be compared with the corresponding reference values.
CAUTION Unless specified differently, all the phantoms
The entire calibration procedure may be managed from the Operator
are inserted into the gantry in axial position
Interface (OPI). Specifically, some calibration procedures are
and with the bed in horizontal position.
automatic, whilst others require the active participation of the
operator. These latter procedures must be referred to as manual
CAUTION The Instruments used to perform calibrations calibrations.
and or measurements must have an adequate The System's calibration parameters are stored in one file:
accuracy and must be calibrated (refer to the “.proto.proto” (master). This file is copied to the “.predef.proto”
Manufacturer’s user manual). (working copy) file, which is the actual file being used during
acquisitions.
CAUTION After the Coil positioning into the Magnet, Since these operations require the user to know how to operate the
always be aware to run the Coil Cable far from machine, please refer to the relevant sections of the User Manual.
the Coil Body. The phantoms supplied with the machine and the manual calibration
System will be described below.
119
8300096031 VER.8 Tune-Up - General
6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6
120
Scout Acquisition field <Sex> allows two alternative options: <Male> or <Female>
or “Other”. Date-of-birth format is “dd-mm-yyyy”: dd for day,
mm for month and yyyy for year. In this case the acknowledged
During the system calibration several Scout acquisitions have to be characters are digits and hyphens
performed. The Scout is a dedicated sequence necessary to set the
system parameters taking care of the magnet frequency and patient's
load and to identify the patient position during the real examination. Fig. 156: Scout Settings
The Scout sequence must be acquired from the Customer Operator
Interface (OPI), it's not possible to acquire a Scout from the Service
Operator Interface (ARAS). OPI is automatic initialized by the system
when the log on as SERV has been performed.
Scout Setup
121
8300096031 VER.8 Tune-Up - Scout Acquisition
6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6
122
The CMR module is located in the Magnet front upper side and An <Emergency> button is located at the upper section of the control
enables management of the motorized movement - rotation and panel. Selection of this command enables:
right-left movement - of the G-scan system.
● Interruption of all scans in execution/queued
To select a command on the control panel, click on the required
● Rotation of the system to a horizontal position, if at an angle other
button; each button bears a specific icon.
than 0°
● Shutoff of the power supply to amplifiers and motor (the radio
Fig. 159: CMR Buttons frequency impulses and generation of the gradient magnetic field
are locked).
After selection, the system is not ready for a new examination until the
operation configuration is reset. To restore the operation
configuration, follow the instructions in the paragraph “User and
administrator access”, section “System status”
123
8300096031 VER.8 Tune-Up - CMR Use
6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6
124
Tab. 4: Commands The motorized right-left movement of the patient table is only admitted
when no coil is fitted to the connector on the upper section of the
magnet, to avoid damage to the coil connection cable during
Icon Function movement. The maximum duration of right-left movement (from the
position <Ankle/In-Out> to <Shoulder>) is approximately 20
On the control panel, select <Ankle/In-Out> to move the
seconds.
table to the correct position for patient ascent/descent.
To move the patient table in the foot-ankle examinations The patient bed, when not moving, is locked in the current position: it
is only released on activation of one of the right-left travel push
<Knee>, to move the patient table in the knee buttons described above.
examinations
125
8300096031 VER.8 Tune-Up - ARAS VISTA
6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6
126
ARAS Toolbar
Icon Function Icon Function
● The toolbar will appear at the top of the desktop when this option
Create a new text To scroll images back
is active
format file and forth (other arrow)
of the opened series
Fig. 162: ARAS Vista Toolbar Connect to the system Save the file on the local
system
● You can expand the shown tree to visualize the acquired images if
● When done the system will show a end of calibration message
present. Then double click on the Image name to open it
Images are sent without any sensitive data such as patient name,
● To visualize the images sent from OPI (customer images as
age, weight and so on. All these data are replaced by numbers.
explained in the previous paragraph) open the OPI folder and
To visualize the sent images act as described in the next paragraph. select the wanted images
127
8300096031 VER.8 Tune-Up - ARAS VISTA
6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6
128
● If you open all the series sent from OPI you’ll be able to visualize it How to perform measurements
in only one window and to scroll the images by clicking on the
arrows key ● When the images are displayed by the ARAS program, move the
mouse prompt on the image and click on the right button mouse
● If the Serv Tree is open during the image acquisition you will have and the measurement menu will appear then select the wanted
to refresh it at the end of the acquisition to see the obtained feature
image/s. To do that click on the right mouse button on the Modality
node (to refresh all the contained folders) or on the Image type
node (to refresh just the selected folder) and select the Refresh Fig. 166: ARAS Measure menu
button as shown in the next figure
Introduction
All the system calibrations are done using the ARAS software.
To launch it, it's necessary perform the system log on writing SERV in
the user field and the correct password (if you don't know the correct
password, ask it to the ESAOTE headquarter).
No special equipment is necessary: all the hardware and software
tools are already integrated into the system. The main feature of this ● To save the selected variable click on the <Save> icon (floppy)
diagnostic instrument is to display signals sampled by the machine.
● Click on the <Stop> icon (square) to quit from the test
“Diascope” functionality
This instrument consists of an interface embodying the key features of Fig. 168: Choose which variables you want to save
an oscilloscope, a being an extremely useful instrument for analyzing
analog signals. For this reason, this interface was named “Diascope”
that uses a logic channel management (abstract association of a
control variable and an acquisition channel of the multiplexer), the
average signal accumulation mechanism and other features that will
be explained in the next paragraph.
With the help of the “Diascope”, various signals from the system can
be displayed on the monitor screen (e.g.: NMR echo/fid,
varicap-piloting voltage, power generated by heaters, sensor
temperature, magnet frequency, etc.). The physically acquired signal
consists of the two components that are usually called the “real part”
and the “imaginary part”. The two separate signals will be identified
by two physical display channels, A and B (the physical display
channels are not the same as the logic channels associated with the
control variables). Signals are plotted on a grid whose reference axes
are identified by the letters X (for abscissa) and Y (for ordinates).
129
8300096031 VER.8 Tune-Up - ARAS VISTA
6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6
130
● Like using an oscilloscope, it's possible to change some Tab. 6: ARAS Scope Buttons
visualization settings acting on the right column of the ARAS
window (refer to the next figure)
Icon Function Icon Function
Fig. 169: Diascope settings Starts the Shows only the
acquisition B trace
Personalize ARAS
It's possible to set and save your own ARAS settings.
● When the test is running, click the right mouse button and a menu
will appear
● Select the <Properties> function then personalize and save your
own default settings
● Click on the <Save> button then choose or create a new file from
the following window
● Following the same procedure but selecting <Load>, instead of
<Save> in the previous menu, it's possible to load your own
● Other settings are available from the Scope toolbar settings
Quick Tuning Fig. 170: Quick Tuning
131
8300096031 VER.8 Tune-Up - Quick Tuning
6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6
132
133
8300096031 VER.8 Tune-Up - Calibration Procedures
6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6
134
Fig. 173: HIGHER KNEE COIL LOCKING BASE AND SPINE 1 ADAPTING
PLATE
135
8300096031 VER.8 Tune-Up - Calibration Procedures
6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6
136
Fig. 174: FOR MAGNETS WITH HIGHER KNEE COIL LOCKING BASE Fig. 175: FOR MAGNETS WITHOUT HIGHER KNEE COIL LOCKING
Knee Coil in the Spine 1 Position For Calibration (it must be aligned BASE Knee Coil in the Knee Position For Calibration (it must be
with the Coil left anf right directions but also in and out directions) aligned with the Coil left anf right directions but also in and out
directions)
Fig. 176: FOR MAGNETS WITH HIGHER KNEE COIL LOCKING BASE Fig. 177: FOR MAGNETS WITHOUT HIGHER KNEE COIL LOCKING
Geometrical phantom in AXIAL position inside the Magnet (white BASE Geometrical phantom in AXIAL position inside the Magnet
plastic screw in upper right position) (white plastic screw in upper right position)
137
8300096031 VER.8 Tune-Up - Calibration Procedures
6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6
138
Fig. 178: FOR MAGNETS WITH HIGHER KNEE COIL LOCKING BASE Fig. 179: FOR MAGNETS WITHOUT HIGHER KNEE COIL LOCKING
Knee Coil and Small Calibration Spherical Phantom BASE Knee Coil and Small Calibration Spherical Phantom
139
8300096031 VER.8 Tune-Up - Calibration Procedures
6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6
140
Fig. 182: Coil Flex Fig. 184: Coil Flex inside the Magnet
NOTICE The Coil Shoulder DPA must be used with the Left
Support but in the right bed position for adjusting
(Varicap and 180° Pulse) and checks (signal to
noise verification).
141
8300096031 VER.8 Tune-Up - Calibration Procedures
6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6
142
Fig. 187: Coil Neck Fig. 188: Coil Neck inside the Magnet
Fig. 189: Coil Back DPA Fig. 190: Coil Back DPA inside the Magnet
143
8300096031 VER.8 Tune-Up - Calibration Procedures
6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6
144
Fig. 191: Coil Shoulder inside the Magnet Fig. 192: Coil Hand inside the Magnet
Fig. 193: Coil Ankle inside the Magnet
145
8300096031 VER.8 Tune-Up - Calibration Procedures
6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6
146
147
8300096031 VER.8 Tune-Up - Calibration Procedures
6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6
148
Fig. 196: PLEASE COMPLETE THE CALIBRATION FLOWCHART MARKING OFF ALL THE PERFORMED ITEMS AND E-MAIL/FAX IT TO ESAOTE.
Room Temperature Offset Calibration ● Under ARAS select <Test> <Automatic> <Room Temp Offset>
<Run> and the system will warns you about the hardware
The meaning of this test is to provide to the System the correct connection explained in the previous point
hardware offset of the room temperature measurement done by the
automatic and manual temperature test in order to perform a very
accurate measurement. Fig. 198: Room Temp Offset
● Disconnect the temperature sensor connected the FTSENS
connector of the Filter Panel (RF cage internal side) and connect
the provided tool (10K ohm resistor) to the FTSENS connector
149
8300096031 VER.8 Tune-Up - Calibration Procedures
6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6
150
● click on the YES button and the system will calibrate and save the
Room Temperature Offset value
Automatic Test
● Insert a Coil
● Select under ARAS <Test> <Troubleshoot.> <Temp. Check>, ● All the automatic temperature tests (even the tests performed by
then <Run> the Scout) are stored in the Temp.rec file
● The temperature errors (difference between ideal and real – Under Internet Explorer select <Rec Management> <System
temperature) and power values are shown in the logger window Temperatures> <View Rec Folders> and the <Today Folder>,
and in the end of calibration message by scrolling it then look for the “Temp.rec” file and open it
151
8300096031 VER.8 Tune-Up - Calibration Procedures
6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6
152
– Using the scroll bar, go to the end of the file and look at the Temperature Stability Check
calibration data (Exit closing the window)
● Select, under ARAS, <Test> <Troubleshoot.> <Temp. Monitor>
<Run>
Fig. 201: temp.rec ● The System Diascope displays the power and the temperature of
every magnet channel, as shown in the next figure
● It’s possible to modify the trace position changing the trace
Displacement value and zoom it changing the Unit/Div value.
Remember to check the Center Scale box before to modify the
Unit/Div value
● Let the System test run until the temperature and power lines are
almost at the end of the visualization display then click on
<Pause> function in the Diascope bar
● Using the cursor, measure the maximum and minimum values of
the temperature line
Temp Mean = (Temp MAX + Temp MIN) / 2
Temp Deviation = Temp MAX - Temp MIN
● Fill in the Installation Quality Form with the calculated values
● Using the cursor, measure the maximum and minimum values of
the power line
Power Mean = (Power MAX + Power MIN) / 2
Power Deviation = Power MAX - Power MIN
● Fill in the Installation Quality Form with the calculated values
Fig. 202: System Monitor temperature and power Fig. 203: System Monitor Channels
● Select the next channel and repeat the centering and measuring
procedure for every magnet channel
153
8300096031 VER.8 Tune-Up - Calibration Procedures
6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6
154
Procedure
155
8300096031 VER.8 Tune-Up - Calibration Procedures
6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6
156
● If correctly done, on the CMR display, the written shown in the ● Click on the WRIST button to unlock the Bed, it’s now possible to
following figure will appear, otherwise check if the Hardware Key is move the Bed using the buttons SPINE2 (left direction) and
correctly inserted and repeat the previous step. TR and RT SPINE3 (right direction) (hold the button for some seconds to
indicate the digital values respectively of the translation and move the bed)
rotation sensors (if they are disconnected the values will be the ● Click on the button SPINE3 and keep it till to move the Bed
maximum readable - very high value) maximum to the right (at the end of the Bed rails two mechanical
This procedure is starting from the Knee position. stops are present) then lock it pressing the WRIST button
● Click on the PREVIEW button to continue
157
8300096031 VER.8 Tune-Up - Calibration Procedures
6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6
158
Fig. 211: Ankle Position Fig. 212: Correct and Incorrect Bed Stop Positions
● When reached the correct position, lock the bed pressing the
WRIST button and verify that the written ANKLE appears on the
CMR display then click on the ANKLE button to save it, the written
SAVE? will appear on the CMR display: click on the PREVIEW
button to save it (the ABORT button won’t save it).
BEFORE TO SAVE THE BED POSITIONS CHECK IF THE BED
STOP IS CORRECTLY LOCKED (REFER TO THE FOLLOWING
FIGURE), IN CASE OF WRONG BED STOP POSITION MOVE
THE BED A LITTLE BIT IN ORDER TO CORRECTLY ALIGNED
THE GEARS (OF THE BED BRACKET AND OF THE BED
STOP). PERFORM THIS CHECK DURING THE ENTIRE
PROCEDURE FOR ALL THE BED POSITION CALIBRATIONS.
● Click on the WRIST button to unlock the bed and then, using the CMR display then click on the ANKLE button to save it, the written
buttons SPINE2 (to the left) and SPINE3 (to the right), center the SAVE? will appear on the CMR display: click on the PREVIEW
SPINE2 Bed coil position to the Magnet Center Point button to save it (the ABORT button won’t save it)
● Click on the WRIST button to unlock the bed and then, using the
buttons SPINE2 (to the left) and SPINE3 (to the right), center the
Fig. 213: Spine2 Position
KNEE Bed coil base to the Magnet Center Point
● When reached the correct position, lock the bed pressing the
WRIST button and verify that the written KNEE appears on the
CMR display then click on the ANKLE button to save it, the written
SAVE? will appear on the CMR display: click on the PREVIEW
button to save it (the ABORT button won’t save it)
● Click on the WRIST button to unlock the bed and then, using the
buttons SPINE2 (to the left) and SPINE3 (to the right), center the
Wrist Bed coil base to the Magnet Center Point
● When reached the correct position, lock the bed pressing the
WRIST button and verify that the written WRIST appears on the
CMR display then click on the ANKLE button to save it, the written
SAVE? will appear on the CMR display: click on the PREVIEW
button to save it (the ABORT button won’t save it)
● Click on the WRIST button to unlock the bed and then, using the
buttons SPINE2 (to the left) and SPINE3 (to the right), center the
SPINE1 Bed coil base to the Magnet Center Point
● When reached the correct position, lock the bed pressing the
WRIST button and verify that the written SPINE1 appears on the
CMR display then click on the ANKLE button to save it, the written
SAVE? will appear on the CMR display: click on the PREVIEW
button to save it (the ABORT button won’t save it)
● Click on the WRIST button to unlock the bed and then, using the
buttons SPINE2 (to the left) and SPINE3 (to the right), center the
NOTICE The big bottle phantom can be used to easily Elbow Bed coil base to the Magnet Center Point
align the Bed anatomy positions to the Magnet ● When reached the correct position, lock the bed pressing the
Iso Center referring point WRIST button and verify that the written ELBOW appears on the
CMR display then click on the ANKLE button to save it, the written
● When reached the correct position, lock the bed pressing the SAVE? will appear on the CMR display: click on the PREVIEW
WRIST button and verify that the written SPINE2 appears on the button to save it (the ABORT button won’t save it)
159
8300096031 VER.8 Tune-Up - Calibration Procedures
6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6
160
● Click on the WRIST button to unlock the bed and then, using the
buttons SPINE2 (to the left) and SPINE3 (to the right), center the NOTICE After saving of all the bed positions, in case that
Shoulder Bed coil base to the Magnet Center Point (as shown in you want to modify one of them you have just to
the next figure) enter in the service modality and perform only the
steps related to the position you want to modify
Fig. 214: Shoulder Position
● Using the buttons SPINE2 (to the left) and SPINE3 (to the right),
set the Bed in the KNEE position, or close to it, in order to allow
the Magnet rotation
● Click on the ALPHA button to unlock the Magnet, it’s now possible
to rotate the Magnet using the buttons “up arrow” (CCW rotation)
and “down arrow” (CW rotation)
● Click on the button “down arrow” and hold it till to set the Bed in
HORIZONTAL position (end of its movement). Do it even if the
Magnet is already set in Horizontal position
161
8300096031 VER.8 Tune-Up - Calibration Procedures
6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6
162
● Click on the ALPHA button to unlock the Magnet, it’s now possible Fig. 217: CMR Display (Example values for potentiometers)
to rotate the Magnet using the buttons “up arrow” (CCW rotation)
and “down arrow” (CW rotation)
● Click on the button “up arrow” and hold it till to set the Bed in
VERTICAL position (end of its movement)
● When the correct position is reached, click on the ALPHA button
to lock the Magnet and verify that the written 90 appears on the
CMR display then click on the 0° button to save it, the written
SAVE? will appear on the CMR display: click on the PREVIEW
button to save it (the ABORT button won’t save it)
● Set back the Magnet in horizontal position by pressing the ALPHA
button to unlock the Magnet and then click on the button “down
arrow” and hold it till to set the Bed in HORIZONTAL position (end
of its movement)
● Press the ABORT button and the CMR display will show you all
the saved potentiometer values as shown in the following figure
(the values shown are just an example)
● If you have connected the Hardware Key, remove it and store
it in a safe place
● Open Internet Explorer, System Status page and proceed with
the CMR reset procedure to reset the Magnet Rotation and Bed
Translation functions:
– Disconnect the Coil if present into the Gantry
– Click on the RESET button
– Enable the CMR from the OPI, Scout Setup page
– Click on the ANKLE button of the CMR
● Click on the 0° button of the CMR ● Following the User procedure (refer to the User Manual for more
info) try to rotate the magnet with the bed in Ankle, Knee and
Spine positions (according to the cage height, not all these
rotations may be possible): if the system won’t rotate means that
position/s is/are wrongly set (with the gears in wrong positions -
refer to the figure Correct and Incorrect Bed Stop Positions shown at
the beginning of this procedure) and you have to re-adjust it/them
● The procedure is over
Channel Delay Check ● Perform the hardware connection as written in the previous figure:
163
8300096031 VER.8 Tune-Up - Calibration Procedures
6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6
164
– (2) From the filter panel, internal side, disconnect the FBTX1, Fig. 221: Filter Panel Connection
FRF0 and FRF1 cables then connect together the filter panel
connectors using the BNC cables and the BNC adapter, “T”
adapter and the BNC cables (as shown in the next figures)
Fig. 220: Channel Delay Tools (two BNC cables, one T adaptor and one Big
BNC Adaptor)
● When the hardware connection is done, click on the OK button ● If the values are correct, fill in the Installation Quality Form with the
and the system will automatically calibrate and save the receiving calculated values
channel offset, the found values are visible in the logger window ● The corresponding rec file is called channel_delay.rec
and in the end of calibration message by scrolling it
165
8300096031 VER.8 Tune-Up - Calibration Procedures
6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6
166
Introduction Procedure
This calibration is used to adjust the gradients' offsets, that is to say ● Position the Knee Coil 2 with the Small Spherical Phantom 86mm
those field gradient residues that are always on and contribute to the
shimming of the static magnetic field. ● Under ARAS select <Manual> <Homogeneity> then <Run>
The module has a characteristic bell shape. The final result we aim for ● Adjust the <Gain 1>, if necessary, to bring the signal trace in
is a signal intercepting the window's sides at the highest possible higher position on the display
points (obviously, adjusting the receiving gain so that the central peak
is close to the maximum). NOTICE The Gain 1 and 2 are not variables. They modify
When entering the manual plot, the gains must be set so that echo only the trace visualizations. Quitting from the test
peak, in <Module> mode, will be close to 1600 mV. During the SW won’t ask to save them.
calibration, further adjustments may become necessary. Moreover,
when observing the phase of the signal, you must check that ● Adjust the <Fine FREQ.> variable and find a value of this variable
resonance is in place and, if necessary, through small changes in corresponding to a generally horizontal phase (lower signal). If
frequency, the phase must be set as horizontal as possible. necessary (frequency very far from the correct value; e.g.: red line
If the gradients and the shimming in general are not perfect, this has a saw-tooth waveform) you can use the Course FREQ. that
phase has a sinusoidal shape that needs to be “flattened” as much as has a much higher step than the Fine FREQ one: modify it one by
possible by adjusting the gradient offsets. one and verify is the upper signal increase then use the Fine
Find the best gradient offset values in order to obtain as large a FREQ, during this procedure could be necessary to adjust a bit
homogeneity curve as possible (as shown in the following figures). the gradient offsets and then come back to the frequency
Small variations of these values are not a cause for concern. The calibration
most important thing is that the values do not exceed 2000, unless set ● Adjust the <OFFSET X> variable to obtain the maximum “wide” of
during final inspection. the echo signal (blue upper signal)
● Adjust the <OFFSET Y> variable to obtain the maximum “wide” of
the echo signal (blue upper signal)
● Adjust the <OFFSET Z> variable to obtain the maximum “wide” of Fig. 225: Homogeneity test perfectly calibrated
the echo signal (blue upper signal)
167
8300096031 VER.8 Tune-Up - Calibration Procedures
6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6
168
Automatic Calibration
CAUTION Lock the Coil base and place a cushion
between the Spherical Phantom upper side
NOTICE Perform this check only if you didn’t perform the and the Coil to lock it in order to perform the
Manual Homogeneity Test Homogeneity Test also with the Magnet set in
vertical position.
Necessary Tools
– Small Spherical Phantom 86 mm and its support ● From OPI, acquire a scout selecting the district “Other”
● Under ARAS select <Test> <Automatic> <Frequency> <Run>
Procedure
and calibrate the system frequency (the found frequency is
● Insert the Knee Coil 2 into the Magnet then place the holder and automatically saved)
the Small Spherical Phantom (like shown in the next figure).
169
8300096031 VER.8 Tune-Up - Calibration Procedures
6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6
170
TX Phase Adjustment ● Perform the hardware connection as written in the previous figure:
– Open the Electronics Unit
– Disconnect the RFOUT LEFT and RIGHT cables from the RFA
NOTICE This procedure must be performed before the TX modules, disconnect the SB18 cable connected to the SB18
Coil Tuning and Matching checks. connector of the Signal Box
– Connect the BNC cables present inside the Electronics right
● Under ARAS select <Test> <HW Automatic> <TX Phase> side and not connected, of the TX Phase calibration kit provided
<Run> and the system will show you the connection message with the system and present in the Electronics Unit, to the
RFOUT connectors of the RFA modules and the Special
Connector to the SB18 connector of the Signal Box as shown in
Fig. 229: TX Phase Messages
the following figures
171
8300096031 VER.8 Tune-Up - Calibration Procedures
6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6
172
173
8300096031 VER.8 Tune-Up - Calibration Procedures
6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6
174
● Changing the channel, select the SPAN STEP field and set it at Fig. 236: Correct TX Coil (span step to 100 and gain rf to 120)
100 Hz or closer (usually 103)
● Select the GAIN RF channel (from the drop down menu) and set it
to 120. The GAIN RF can be digitally set from 0 (0 Watt) to 255
(900 Watt)
● There are two circuits for the transmission coil: upper and lower.
The peak shown in the monitor upper part is related to the upper
TX circuit while the peak shown in the monitor lower part is related
to the lower TX circuit
● Check if the peaks are in a range of ±5 KHz from the Average
Frequency, otherwise you must adjust the tuning characteristics
175
8300096031 VER.8 Tune-Up - Calibration Procedures
6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6
176
177
8300096031 VER.8 Tune-Up - Calibration Procedures
6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6
178
● Decrease or increase the capacitors by a small amount, then Fig. 243: Compenent Positions of the Lower Transmission Circuit
repeat the test checking if the deepness of the peak is at least
20 dB (or 18dB for the lower one)
● After the adjustment, check the TX Coil peaks with the Bed set
back in SPINE1 position. If the peaks are out of matching repeat
the procedure from the beginning
Gradients Orientation Check – If they have the same orientation fill in the Installation Quality
Form
● Place the geometrical phantom in AXIAL position into the Knee
– If they have a different orientation check the gradient cables
coil taking care to place it with the plastic screw up to the right side
looking for inversions then repeat this test until the correct scout
of the magnet (patient’s head position), then insert the Knee coil
images are obtained
into the magnet (like shown in the next figure)
179
8300096031 VER.8 Tune-Up - Calibration Procedures
6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6
180
● When done click on the Yes button and the system will perform
Channel 4th Polarity Check automatically the polarity check of the Channel 4th looking for
● Under ARAS select <Test> <Troubleshoot.> <4 Ch. Polarity> cable inversion
<Run> and the system will ask to insert the Knee Coil 2 and the
small spherical phantom
Fig. 247: 4th Channel Polarity Check Correct
● If any inversion is present the system will show you the correct
polarity message. In case of inversion the system will show you a
message like this: check the Channel 4th cables connected to the
GRA160 and to the filter panel (internal and external side) looking
for any inversion, then repeat the test
181
8300096031 VER.8 Tune-Up - Calibration Procedures
6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6
182
EC Gr and EC B0 Checks
EC Gr Check
183
8300096031 VER.8 Tune-Up - Calibration Procedures
6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6
184
Fig. 252: Check_SpecifTab.txt.rec ● In the eventually that one or more than one value is not in
specification you must perform the tune procedures (as shown in
the nest figure)
185
8300096031 VER.8 Tune-Up - Calibration Procedures
6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6
186
Fig. 255: Phase Stability Message Fig. 256: EC B0 Check Message: all the values are in specification
187
8300096031 VER.8 Tune-Up - Calibration Procedures
6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6
188
EC Gr Tune
This procedure takes about 1 hour
● Insert the Knee Coil 2 with the Small Spherical phantom in it
● Perform one scout selecting the district “Other”
● Under ARAS select <Test> <HW Automatic> <ECC Tune>
<Run> and the System will ask you to insert the Spherical
phantom inside the Knee Coil 2
● Click on the Yes button and the System will automatically tune the
GRA160 module. The automatic procedure comprehends: the
Regulator calibrations (X, Y, Z and 4th channel), the Phase
Stability check, the phantom position check (axial, sagittal and
coronal images) and the ECC (acronym for eddy current
compensation) tune
● The first automatically step is the Regulator calibration: the Fig. 260: RegulatorsX.rec
system will show the regulator value after every channel
calibration like shown in the following figures
189
8300096031 VER.8 Tune-Up - Calibration Procedures
6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6
190
Fig. 261: Y channel regulator calibration message window Fig. 262: Z channel regulator calibration message window
Fig. 263: IV channel regulator calibration message window
NOTICE The system will show the Phase stability window
message only in case of incorrect value, if the
value is correct the procedure will continue
automatically. In case of Phase Stability problem
perform the magnetic compensation procedure
and repeat the EC Tune
191
8300096031 VER.8 Tune-Up - Calibration Procedures
6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6
192
193
8300096031 VER.8 Tune-Up - Calibration Procedures
6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6
194
● Fill in the Installation Quality Form with the saved values Fig. 269: Tune_IIG_SpecifTab.txt.rec
● It's possible to read the obtained values also selecting the file
auto_eddy_tune_IIG.rec, then scroll the file to read the X, Y, Z
and Channel 4th values. The values of the last performed check
are in the bottom part of the file. If this file is opened during the
calibration the system will not refresh it: quit and re-open the file to
refresh it. Otherwise it’s possible to open the
Tune_IIG_SpecifTab.txt.rec that contains just the result tables
(this file is shown in the next figure)
Gradient Delay Check Fig. 271: Gradient Delay Message
● Insert into the Knee Coil 2 the ghosting phantom (in central
position) using its support. Like shown in the next figure
195
8300096031 VER.8 Tune-Up - Calibration Procedures
6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6
196
Fig. 272: Gradient Delay automatic calibration Fig. 273: Gradient Delay rec
Necessary Tools
– Geometrical Phantom and its supports
197
8300096031 VER.8 Tune-Up - Calibration Procedures
6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6
198
Fig. 274: X Image Opened and Distance Tool Selection Fig. 275: X measurement
199
8300096031 VER.8 Tune-Up - Calibration Procedures
6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6
200
201
8300096031 VER.8 Tune-Up - Calibration Procedures
6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6
202
203
8300096031 VER.8 Tune-Up - Calibration Procedures
6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6
204
Fig. 281: OPI Protocol List – If the values are in specification this procedure is over and you
can proceed to the next test
– If the values are out of specification (as in the following example)
proceed with the Shimming adjusting procedure explained in this
section
– Remove the Upper and Lower Gantry Covers and you’ll have the
CAUTION Disable the GRA160 by switching off the Magnet as shown in the next figures
Electronics Unit before to disconnect the
Gradient Coils or work with Magnet parts.
205
8300096031 VER.8 Tune-Up - Calibration Procedures
6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6
206
Fig. 285: System Without Bed Table Fig. 286: System Without Bed Table
207
8300096031 VER.8 Tune-Up - Calibration Procedures
6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6
208
Fig. 289: Lower Removable Shim Plate (neodymium pieces will face the Fig. 290: Lower Gradient Coil
gradient coils)
209
8300096031 VER.8 Tune-Up - Calibration Procedures
6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6
210
211
8300096031 VER.8 Tune-Up - Calibration Procedures
6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6
212
213
8300096031 VER.8 Tune-Up - Calibration Procedures
6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6
214
● Click on the Next Solution button and the system will provide you Fig. 299: 1st solution to correct the 2.1cos
a new solution that for sure will use a different position that the
one/s you have ticked
● Click on the Next Solution button and the system will provide you ● Let' say that this solution is acceptable and click on the Accept
a new solution that for sure will use a different position that the button to have the first suggested solution for the next out of spec
one/s you have ticked parameter
● When you have accepted a solution for every out of spec
parameter, the system will provide you the complete table. Follow
it plate by plate and line by line!
215
8300096031 VER.8 Tune-Up - Calibration Procedures
6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6
216
the field try to flip it means that is negative, if the field doesn’t
CAUTION Be aware that broken or lost neodymium try to flip it is positive. Set it to the correct polarity (indicated
pieces are attracted by the magnetic field and in the shimming solution tables) and then place it on the
may be forcefully drawn in the direction of the Shimming Plate in the position indicated in the shimming
gantry during shimming maintenance. solution tables. Could be easier place them if the Magnet is
Perform this procedure wearing safety rotated in vertical position so you’ll be able to work inside the
goggles, single use gloves and dust mask and Magnetic field. If you do it be aware that the Upper Plate
using not magnetic tools. becomes the left one and the Lower Plate becomes the right
one
● ONLY FOR SHIMMING ON REMOVABLE PLATES METHOD:
Fig. 302: Shimming Complete Solution Regarding the polarity place the shimming piece in the
defined position and verify its polarity by using the reference
dipole provided with the shimming kit (shown in the next
figure) - attraction means same polarity while rejection
means opposite polarity
● Follow the provided solution line by line to be sure to place all the
shimming pieces that compound this solution. Read the entire
solution by scrolling the table
● ONLY FOR SHIMMING ON POLE METHOD: Regarding the
polarity take the shimming piece you have to position in your
hand and insert it into the Magnet at the Magnet Iso Center: if
● FOR SHIMMING ON REMOVABLE PLATES METHOD: to ● If you close the IE program don’t be worried, you have the entire
correctly place pieces on removable plates, refer to the next shimming solution in the History file by clicking the History button
figures of the Shimming page as shown in the next figure
Fig. 304: How to correctly place pieces on removable shimming plates Fig. 305: Complete Solution Present in the History File
● After the correction, assembly back the System (at least gradient
coils, TX coils and bed table) and repeat the acquisition to verify if
the specifications are meet (may be necessary a second
shimming pass but before it double check the pieces just placed
before to proceed). When reached assembly the System
completely (with all covers and accessories)
217
8300096031 VER.8 Tune-Up - Calibration Procedures
6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6
218
Coil Tuning Adjustment ● Under ARAS select <Tests> <Automatic> <Coil Tuning> then
<Run>. The calibrated value/s is/are automatically saved
Introduction
Varicap calibration is performed for tuning the receiving coils. Every Fig. 306: Automatic procedure
coil has a different varicap value. For every coil two coil tuning checks
must be performed: automatic and manual.
Linear Coils (1, 9 and 6 if present) have one varicap, while DPA (Dual
Phase Array) have two varicaps (one for the Linear channel and the
other for the Saddle channel).
The algorithm used by the automatic calibration procedure (Coil
Tuning) is very simple and consists of changing the value of the
command corresponding to the varicap voltage while observing the
MR signal being picked up. The calibration program considers that the
tuned condition is reached when the maximum acquired signal is
obtained.
The fact that a given varicap voltage value corresponds to each coil
avoids lengthy calibration sessions. The program starts at this value
to search for the maximum signal condition. The comparison with the
default values (refer to the installation protocol) may indicate any
changes in the tuning of the coil being examined.
Necessary Tools:
– Big Spherical Phantom and its support
– Small Homogeneous Phantom and its support
Automatic Procedure ● Look at the message sent by the calibration procedure and repeat
the test two or three times checking if the varicap value is always
● Put the coil with the suitable phantom and perform a Scout the same or different by just a few units
selecting the district “Other”
● If the value/s is/are correct, fill in the Installation Quality Form with Fig. 307: Varicap Rec File
the saved value/s
● Repeat all the procedures for the others coils
If this was not successful, do not continue, but rather check the
“varicap.rec” store file, use the scroll bar to reach the end of the file.
The calibration program sends some information relating to its
compilation date, etc., the list of values of the varicap command that
are being tested and their results in terms of signal level to
“varicap.rec”. Before ending, the program specifies the calculated
varicap value and writes it to the “.predef” temporary file.
Since varicap calibration requires the acquired signal to be viewed, it
must be preceded by calibration of the receiving chain gain, in order
to shift the signal to a visible area (about in the middle of the range).
Abnormal gain values may indicate quantity problems in the picked up
signal; in any case, these will be made apparent in sub-sequent
acquisitions.
219
8300096031 VER.8 Tune-Up - Calibration Procedures
6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6
220
Manual Procedure ● Adjust the <Gain 1> to bring the signal to about 1,8V (inside the
monitor leaving the default visualization settings) as shown in the
Enter in the manual test for troubleshooting, to verify that the signal is following figure
stable and with a correct shape. You can also modify the value/s but
DO NOT SAVE IT/THEM ‘CAUSE THE AUTOMATIC CALIBRATION ● Let the test run for 1 minutes and verify that the trace is stable (not
IS MUCH MORE PRECISE THAN THE MANUAL ONE. jumping) and with a shape like shown in the following figure
During manual calibration, the echo signal is displayed. By adjusting
the <Channel 1>, and <Channel 2> for the DPA coils, variable with Fig. 309: Varicap1 trace
the arrows, it is possible to change the varicap value.
● Position the coil with the suitable phantom (refer to the Quality
Form) into the magnet and perform a scout selecting the district
“Other” then, under ARAS, select <Tests> <Manual> <Coil
Tuning> then <Run>
● For the DPA coils only: change the Channel (from 1 to 2) and
repeat this calibration for the VARICAP 2
● If the trace/s is/are correct quit from the test and go to the next
calibration step. In case of problem refer to the Service Manual
RF 180° pulse gain calibration ● Under ARAS select <Tests> <Automatic> <RF 180 Gain>, then
<Run>. The calibrated value is automatically saved
Introduction
This paragraph provides a detailed description of how to tune the Fig. 310: Automatic 180° Pulse Calibration
180° pulse for each coil. Tuning of the other pulse types will be
automatic, based on the value found for the 180°-pulse. In this case
too, it is clear that, if, for any reason whatsoever, tuning of the
180°-pulse is unsuccessful, tuning of the other pulses will be incorrect
also. The 180°-pulse calibration is performed to adjust for the energy
transferred to the spins by this RF pulse. Every coil has a different
180°-pulse value.
The algorithm used is very simple and consists of changing the
transmission gain value corresponding to 180° and observing the MR
signal picked up. The calibration program considers that the tuned
condition is reached when the maximum acquired signal is obtained.
Since it requires the acquired signal to be viewed, the 180° calibration
must be preceded by calibration of the receiving chain gain to shift the
signal to an area around the middle of the range. Abnormal gain
values may indicate quantity problems in the signal being picked up.
Necessary Tools
– Big Spherical Phantom and its support
– Small Homogeneous Phantom and its support
Automatic Procedure
● Put the coil with the suitable phantom into the magnet and
● Look at the message sent by the calibration procedure and
per-form a Scout selecting the district “Other”
re-peat the test two or three times checking if the 180°-pulse value
is always the same or different by just few units
221
8300096031 VER.8 Tune-Up - Calibration Procedures
6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6
222
● If the value is correct, fill in the Installation Quality Form with the Fig. 311: 180° Pulse Rec File
saved value
● Repeat all the procedures for the other coils
The automatic program sends some information relating to its date of
compilation etc., the list of the 180° transmission gain values
attempted and the relating results in terms of signal level to the
180.rec file.
If this was not successful, do not continue, but rather check the store
file 180.rec. Use the scroll bar to reach the end of the file and record
the 180° pulse value found.
Manual Procedure ● Let the test run for 1 minutes and verify that the trace is stable (not
jumping) and with a shape like shown in the following figure
Enter in the manual test for troubleshooting, to verify that the signal is
stable and with a correct shape. You can also modify the value but DO
NOT SAVE IT BECAUSE THE AUTOMATIC CALIBRATION IS MUCH Fig. 313: 180° Pulse correctly maximized
MORE PRECISE THAN THE MANUAL ONE.
● Put the coil with the suitable phantom into the magnet and
per-form a Scout selecting the district “Other”
● Under ARAS select <Tests> <Manual> <RF 180 Gain>, then
<Run>
● Adjust the <Gain 1> to bring the signal to about 1800 mV
● If the trace is correct quit from the test and go to the next
calibration step. In case of problem refer to the Service Manual
223
8300096031 VER.8 Tune-Up - Calibration Procedures
6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6
224
Introduction
The purpose of this calibration is to adjust the Echo peak generated
by the sequence: an echo is considered “calibrated” when its
amplitude is the maximum obtainable. The G SLICE change (the
range value is between -32767 and + 32767) is necessary to find out
the condition of maximum signal. Write down the value on the form.
Necessary tools
– Geometrical Phantom and its support
Procedure
● Place the Knee Coil 2 with the geometrical phantom in the Axial
position
● Under ARAS select <Test> and <Sequences>
● Select the sequence inside the above mentioned menus and click
on the <Run> button. This selection allows starting up the
“Diascope” program
● Change the <Gain1> and if necessary <Gain2> values to display
the amplitude signal with the peak on 4/5 of the full screen
● Change the <Channel> number, using the arrows, till the
<G SLICE> name is displayed. By clicking the GSLICE arrows
change the <G SLICE> value to reach the highest peak value.
The amplitude of the peak can be read in the information window
at the bottom right corner of the screen under the variable named
<M y>
225
8300096031 VER.8 Tune-Up - Calibration Procedures
6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6
226
227
8300096031 VER.8 Tune-Up - Calibration Procedures
6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6
228
● Click on the YES button and the calibration will start. During the
calibration are shown “End of calibration” messages (related to the
completed part) and/or error messages such as the magnetic
compensation error message (this message will appear and you'll
have to perform very carefully the Magnetic Compensation
procedure and then repeat the calibration)
● At the end of the calibration, if completed without errors, the
system will show the following messages. In this case the
calibration is over and you can proceed to the FSE Rel Calibration
section of this procedure. In case of errors please refer to the
flowchart and to the Problem Example sections to be able to
complete the FSE Calibration step by step
Fig. 318: FSE Calibration Messages: Gradient Delay (1of4) Fig. 320: FSE Calibration Messages: ESP Calibrations (3of4)
Fig. 319: FSE Calibration Messages: Gradient Delay (2of4) Fig. 321: FSE Calibration Messages: End of Calibration (4of4)
229
8300096031 VER.8 Tune-Up - Calibration Procedures
6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6
230
231
8300096031 VER.8 Tune-Up - Calibration Procedures
6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6
232
FSE Calibration Checks by every sequence has the lowest S/N ratio and report this value
on the Quality Form
● Insert the Big Bottle Phantom code 9102242000 into the Knee
Coil and run a Scout (district “Other”)
● When the scout has been acquired, run from OPI all the FSE Fig. 325: FSE Calibration Check S/N Verification from ARAS
Calibration Check sequences (right mouse click on the folder then
select the run function) (refer to the Quality Form related to the
installed SW Release)
● Repeat the same procedure with the others ESP acquired till to
open and report them all
● When all the images have been acquired, open them all (in the
order they have been acquired) from ARAS, then identify which
slice of the three acquired (use the yellow arrows to scroll them)
FSE Rel Calibration Fig. 326: FSE.REC with Rel Calibration at the end
The FSE Rel Calibration folder contains 5 buttons: the first button
(FSE REL Selection Complete) performs the entire calibration (all the
calibrations done by the other buttons step by step); all the other
buttons perform only one ESP calibration (30 ms FC None or 30 ms
FC read or 25 ms or 20 ms)
The entire calibration takes about 2 hours and 30 minutes to be
completed:
● Insert the Big Bottle Phantom code 9102125000 into the Knee
Coil and run a Scout (district “Other”)
● After the scout insert the Big Bottle Phantom code 9102125001
(filled with water instead of the fluid used normally in the ESAOTE
phantoms) into the Knee Coil
● From ARAS select <Tests> <FSE Rel Calibration> then < FSE
REL - Selection Complete> and run it and the system will ask to
confirm the action by clicking on the YES button
● When done, if completed without errors, the system will show the
“End of Calibration” message
● In case of problem follow the message suggestion if present (e.g.:
perform the magnetic compensation) or open the FSE.REC file
and verify in which point the calibration has been interrupted then
repeat the entire calibration or from the not completed one on by
following the folder button order
233
8300096031 VER.8 Tune-Up - Calibration Procedures
6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6
234
FSE Rel Check Fig. 327: FSE Image S/N Check Selection from ARAS
The last part of the FSE Calibration is the FSE Rel image check. It’s
necessary to acquire some images and then compare their S/N ratio.
The FSE Rel image S/N has to be higher than the FSE image S/N. If
not the FSE Rel Calibration must be repeated till to pass this test.
● When the scout has been acquired, remove the phantom and
insert the Big Bottle Phantom code 9102125001 (same shape but
filled with distilled water), when replaced wait for 5 minutes (for
water stabilizing) and then run from OPI all the s/n sequences
(right mouse click on the folder then select the run function)
● When all the images have been acquired, open them all (in the
order they have been acquired) from ARAS, then compare them
as shown in the next figure (every column shows the non rel
image in the bottom line and the corresponding X, Y and Z Rel
images in higher positions)
Fig. 328: FSE Image S/N Check Comparing from ARAS
235
8300096031 VER.8 Tune-Up - Calibration Procedures
6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6
236
Fig. 331: End of Selection Check message Fig. 332: End of Encoding Rewinding Check message
In case of problem such as a wrong calibration, the message will In case of problem such as a wrong calibration, the message will
show a NOT PASSED warning related to the not passed test. In this show a NOT PASSED warning related to the not passed test. In this
case you have to repeat its/their calibration/s by selecting it/them from case you have to repeat its/their calibration/s by selecting it/them from
the <FSE Calibration> folder the <FSE Calibration> folder.
Obviously if more than 2 values failed will be faster to run the
Complete calibration instead to run them one by one by running the
FSE Phase calibration.
237
8300096031 VER.8 Tune-Up - Calibration Procedures
6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6
238
IIG Check
Once you run it the system will remind you to insert the spherical
phantom inside the Knee Coil.
Click on OK button and after about 25 minutes the system will show
the following check message table.
239
8300096031 VER.8 Tune-Up - Calibration Procedures
6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6
240
Acquisition Stop
Method 1
● Insert a Coil and a phantom in it
● From OPI, perform a scout and then put five random sequences in
the queue
● During the acquisition of the second sequence of the queue
● Put the mouse prompt in the Scan area (where the queued
sequences are listed) and click on the right button of the mouse
● From the shown menu select Abort
● Verify that only the running sequence is aborted
Method 2
● Insert a Coil and a phantom in it
● From OPI, perform a scout and then put five random sequences in
the queue
● During the acquisition of the second sequence of the queue
● Put the mouse prompt in the Scan area (where the queued
sequences are listed) and click on the right button of the mouse
● From the shown menu select Abort All
● Verify that all the running sequences are stopped
Part 7 Magnetic Compensation
241
8300096031 VER.8 Magnetic Compensation - Introduction
7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7
242
Mag Probe Positioning according to the magnetic compensation results) as shown in the
next figure
● Take the Mag Probe (that contains two probes: one set in
horizontal position and one set in vertical position - as illustrated
by the label placed on the probe) and join its the holder to it as
shown in the next figure
● Mount the probe right beyond the Magnet (on the Pavilion rear
wall) at about 1,7m from the floor (this value can be modified
DC Compensation Fig. 337: FOR MAGNETS WITH HIGHER KNEE COIL LOCKING BASE
Ghosting Phantom Positioning
● Connect the Vertical Probe MAGN1 cable connector to the
FCMAG1 connector of the Filter Panel and the Horizontal Probe
MAGN2 cable connector to the FCMAG2 connector of the Filter
Panel (the cables from the filter Panel to the Electronics unit must
be connected during the System cabling)
● Insert the Knee Coil 2 with the Ghosting phantom 25 ml placed in
the center hole of its soft support (see the next figure) and perform
a scout selecting the district “Other”
243
8300096031 VER.8 Magnetic Compensation - Mag Probe Positioning
7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7
244
installed (made of iron or aluminum). In case of different shielding Fig. 338: Traces without magnetic compensation (same shape) (external
cage installed then the Pavilion use the Iron values and internal trace scale 1 mG)
245
8300096031 VER.8 Magnetic Compensation - Mag Probe Positioning
7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7
246
Fig. 344: Pick-up coil and its support inside the Magnet
247
8300096031 VER.8 Magnetic Compensation - AC Compensation
7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7
248
NOTICE If you don’t have to perform the DC compensation Fig. 345: Traces not in phase (external and internal trace scale 2 mg)
set the COARSE DC GAIN to zero and verify the
Gain Slow, Mid and Fast and Tau Slow, Mid and
Fast values before to proceed with the AC
compensation.
Fig. 346: Traces in phase (external and internal trace scale 2 mg)
249
8300096031 VER.8 Magnetic Compensation - AC Compensation
7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7
250
Fig. 348: Decreased internal field and again in phase (external trace scale Fig. 349: Correctly internal trace (external trace scale 2 mG while internal
2 mG while internal trace scale 1 mg) trace scale 0,2 mg)
● Increase FINE COMMON GAIN till to reduce the internal field in ● Quit and save all the variables
specification ● Rotate the Magnet in vertical position and repeat this procedure
using the diascope present in the <EFI Vertical> folder
● When done, perform the AC Quality Test contained in the next
chapter. In case of bad results repeat this procedure
Procedure for 16 Hz ● Select <DUAL> to display the internal and external AC fields (blue
and red traces respectively)
● Set the COARSE PROBE GAIN to 1, FINE PROBE GAIN to 1
NOTICE If you have to perform both DC and AC and PHASE PROBE P.U. to 2 (default)
compensations the DC must be performed first.
251
8300096031 VER.8 Magnetic Compensation - AC Compensation
7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7
252
Trigonometric Compensation
Starting from the 9.8 SW release this new features has been added in
the G-SCAN SW. After performing the horizontal and vertical
magnetic compensation, the system will be able to automatically
compensate the magnetic noises when setting the magnet at different
angles than 0° and 90°. To verify if the trigonometric compensation is
properly working at a certain angle (for example if the customer likes
to perform examination with the magnet set at 80°), set the magnet to
the certain angle then open ARAS, the Troubleshooting folder and
select the EFI test corresponding to the compensation you want to
check as shown in the next figure.
When a button has been selected the diascope is opened and the
internal and external magnetic traces are shown. It will be possible
also to modify magnetic compensation values (horizontal and vertical)
but not to save it/them. To save it/them is necessary to enter in the
corresponding magnetic compensation procedure (horizontal and/or
vertical) and save it/them as explained before in this procedure.
The diascope corresponding to the EFI DC Check (50Hz) is shown in
the next figure.
Fig. 351: EFI DC Check (50Hz)
253
8300096031 VER.8 Magnetic Compensation - AC Compensation
7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7
254
Part 8 Quality Assurance
Introduction
All the necessary tests for checking the System image quality will be
described in this chapter:
● Noise Check
● Signal to Noise Check
● Uniformity Check
● Ghost Checks
● Stability Check
● Angle Function and Spatial Resolution Checks
Refer to the Phantoms paragraph for the Coils and Phantoms settings
during the Quality Test.
255
8300096031 VER.8 Quality Assurance - Introduction
8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8
256
Necessary tools
– Homogeneous Phantom 115mm and its support
Operations
● Insert the Knee Coil 1 with the homogeneous phantom in the
magnet center and perform a scout
● From ARAS, start the RX Chain test (Troubleshooting folder)
● When the Diascope is open set:
– FFT Modality
– Average equals to 10
– Gain1 equals to 80 and Gain2 equals to 0
– Horizontal and vertical scale according to the acquired signal
(usually respectively 5k and 64 levels)
● Set the Sampling Rate to 21, 23, 27, 31, 35, 39, 47, 55 and 71 ● Repeat for Channel 2
(reset the Average Acquisition when the new value has been set)
and verify that the acquired signal is never higher than the value
reported in the Quality Form and fill it
Signal to Noise check Fig. 353: Signal to noise measurements
Introduction
The signal-noise ratio test is the most important test and, at the same
time, the most critical since it is affected by the operation of the entire
machine and not by a single parameter. The evaluation is performed
on images obtained with the homogeneous phantoms and the
different coils for the three main views: Axial, Sagittal, and Coronal.
This measurement must be repeated with different scans, so you can
be absolutely certain that the machine is operating correctly.
A dedicated program, in the right button mouse menu, measures this
parameter by evaluating the signal as the average value in a ROI in
the center of the image and four ROI's in the corners of the image.
Necessary tools
– Homogeneous Phantoms and supports
Procedure
● Insert the Coil in Axial position with the suitable phantom and
perform a scout selecting the district “Other”
● Under OPI Click select <Protocol List> then select the correct
sequences following the Installation Quality Form and start them
● Wait for the images are acquired then under ARAS open the
correct image and perform the measurement then write the value
on the Installation Quality Form
● Repeat for the other sequences and coil following the form
257
8300096031 VER.8 Quality Assurance - Noise Check
8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8
258
Introduction
Image uniformity is evaluated on the homogeneous phantom
following the NEMA protocol that defines this parameter as follows:
U% = 100 x [(Smax-Smin) / (Smax+Smin)]
Where Smax and Smin are the maximum and the minimum signals in
a circular ROI centered in the phantom's image.
Following the same procedure explained before for the signal to Noise
test, perform the Uniformity test just for the Knee Coil 2, in axial
orientation.
Necessary tools
– Big Spherical Phantom and its supports
Procedure
● Insert the Knee Coil 2 in Axial position with the suitable phantom
in it and perform a Scout selecting the district “Other”
● Under OPI select <Protocol List> then select the correct
sequences following the Installation Quality Form and start them
● Wait for the images to be obtained
● Under ARAS open the correct image and perform the
measurement then write the value on the Installation Quality Form
Ghost Checks Fig. 356: FOR MAGNETS WITH HIGHER KNEE COIL LOCKING BASE
Phantom and support inserted into the Knee coil
Necessary tools
– Ghosting phantom and its support
259
8300096031 VER.8 Quality Assurance - Noise Check
8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8
260
Necessary tools
● Knee coil number 2
● Homogeneous phantom (axial position)
Procedure
● Place the coil and insert the phantom with its support
● Under OPI Click select <Protocol List> then select the correct
sequences following the Installation Quality Form and start them
● When acquired, under ARAS open the correct image and check if
DC ghost are present also chancing the LUT (e.g.: see the next
figure)
● If the ghost is not present fill the Installation Quality Form; if it's
present check and repeat the magnetic compensation procedure
● If the ghost is still present after the magnetic compensation
procedure, set to 1 the Phase Corr flag (Internet Explorer,
DEBUG page). In this condition some sequences are less
sensitive to external DC magnetic fluctuation but some others
have less signal then before ● In the following image DC noise is visible
261
8300096031 VER.8 Quality Assurance - Noise Check
8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8
262
Necessary tools
● Knee coil number 2
● Spherical phantom and its support
Procedure
● Insert the Knee Coil 2 in Axial position with the suitable phantom
in it and perform a Scout selecting the district “Other”
● Under ARAS, open Test and the Stability Test folder then select
the correct test according to the Magnetic Compensation
procedure performed (DC and or AC) and/or according to the
country power supply (16 or 50 or 60 Hz - that means a possible
ghosting problem related to 16 or 50 or 60 Hz) and start it or them
● Wait for the end of the test then, open the stabilitytest.rec file
(under Internet Explorer) to read the obtained values
– Amp. Stability lower than 3%
– Phase Stability lower then 8°
263
8300096031 VER.8 Quality Assurance - Noise Check
8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8
264
Fig. 362: DC Stability Message ● If the values are larger with respect to the reference values written
on the quality form magnetic fluctuation is not compensated or the
magnetic compensation doesn't work properly. Even if the
magnetic compensation procedure was correct applied but the
stability DC values are out of spec, set to 1 the Phase Corr flag
Internet Explorer, DEBUG page). In this condition some
sequences are less sensitive to external DC magnetic fluctuation.
If enabled the Amplitude Stability values won't be applicable, if the
Phase Stability is still out of spec check the image quality (refer to
the DC Ghost Test).
Fig. 363: AC Stability Rec File Fig. 364: DC Stability Rec File
265
8300096031 VER.8 Quality Assurance - Noise Check
8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8
266
Angle Function and Spatial Resolution Checks Fig. 365: angle Funcition Measurement
Angle Function
In order to check the angle function, it is necessary to evaluate the
angle among the pins of the geometrical phantom as shown in the
following figure.
● Select the SE T1 sequence and open the setting window
● Check and set the sequence parameters as follows:
– TE = 26 ms
– TR = 500 ms
– Number of slices = 1
– Slice thickness = 5 mm
– Average = 1
– FOV = 120 x 120 mm
– Data acquisition matrix = 128 x 128
● Acquire the image
● Select <Angle> icon, click the left mouse button, move the mouse
and draw the first segment till to the angle vertex
● Click the left mouse button and move the mouse from the vertex,
drawing the second segment
● Record the angle value on the Installation and Quality Form
Spatial Resolution Fig. 366: spatial Resolution Measurement
267
8300096031 VER.8 Quality Assurance - Noise Check
8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8
268
Introduction
The Installation Quality Form contains all the calibrations and checks
must be performed during the system installation, maintenance,
software update or after fixing.
The Installation Quality Form code is 8370091100 and the version
related to the SW Release installed can be download from the
Service Restricted Web site together with the Service Information
related to the installed SW Release as well.
269
8300096031 VER.8 Installation Quality Form - Introduction
9 9 9 9 9 9 9 9 9 9 9 9 9 9 9 9 9 9
270
Introduction Procedure
This function allows the creation of a copy of the system calibration
data and user's protocols and configurations, but does not save the NOTICE This procedure must be carried out at the end of
customer's images. To save them insert an empty optical cartridge the System Tune Up and must be repeated after
and use the <Image Backup> feature present under the <Tools> every maintenance or repairing.
button located in the ARAS Program toolbar. Use <Image Restore>
feature to copy the images from the Backup cartridge to the System
HD. ● Log on the system like SERV (typing the correct password) and
launch Internet Explorer
● Insert an empty CD or DVD into the DVD burner
NOTICE It's not possible to use the same optical cartridge
● From the home page, select Configuration Management then
to archive the Customer's images and the Service
Backup & Restore
images.
● Follow the SW indications
Don't create the system back up immediately after the installation but
do it during the first maintenance when the system parameters are
stabilized and the customer created his own protocols.
271
8300096031 VER.8 System Backup - Introduction
10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10
272
273
8300096031 VER.8 Peripheral Settings - Introduction
11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11
274
– In the opened window, select File and New Incoming connection Fig. 369: Incoming connection step 2
– In the opened window, select the “Internet Protocol Version 4 – In the opened window, tip the “Allow callers to access my local
TCP/IP” and than press the Properties button and a new area network”, “Allow calling computer to specify its own IP
window will appear address” and “Specify IP address” and insert the following
addresses: from 169.254.0.1; to 169.254.0.2
– When done press the OK button
275
8300096031 VER.8 Peripheral Settings - Service Modem Settings
11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11
276
Fig. 372: Incoming connection step 5 Fig. 373: Incoming connection step 6
– In the opened window, press the Close button and re-boot the – The procedure is over
system
Part 12 Disassembly & Disposal
277
8300096031 VER.8 Disassembly & Disposal - Disassembly
12 12 12 12 12 12 12 12 12 12 12 12 12 12 12 12 12 12
278
Basic regulations
ÖEating, drinking, smoking, and intentional inhalation of vapors
are not permitted when working with MR phantom fluids.
ÖHands should be thoroughly washed with soap and water after
working with MR phantom fluids.
ÖAs added protection, wear a lab coat whenever working with
MR phantom fluids.
Protective clothing
● Disposable gloves (of a waterproof material, such as latex, PVC,
nitrile rubber, or polyethylene)
● Standard protective goggles
279
8300096031 VER.8 Disassembly & Disposal - Safety guidelines
12 12 12 12 12 12 12 12 12 12 12 12 12 12 12 12 12 12
280
If swallowed
– Drink plenty of water and induce vomiting.
– Contact a physician immediately.
Part 13 Index
281
8300096031 VER.8 Index
13 13 13 13 13 13 13 13 13 13 13 13 13 13 13 13 13 13
282
1 Introduction
2 Safety
3 System Installation
4 Start Up
5 Software Update
6 Tune-Up
7 Magnetic Compensation
8 Quality Assurance
10 System Backup
11 Peripherals Settings
13 Index
14 Tbd