Photo 1: “Local Operating Interface.”

Step 1.

Using Local Operating Interface connect USB side to you PC ( USB port) and the other
side to Coriolis flowmeter connection port (see below Photo 2).

! IMPORTANT: Please be careful when connecting/disconnecting the cable on


flowmeters side. Connection port of the flowmeter is very sensitive can be damaged.

2|Page
Photo 2: “Coriolis Flowmeter Connection Port.”

Step 2.

Start “HTT-Coriolis Master” program on your PC.

3|Page
Step 3.

Click icon “CONNECT”

Photo 3: “HTT - CoriolisMaster Interface.”

Step 4.
Once you are connected to flowmeter go to tab “dynamic chart”

Photo 3: “HTT - CoriolisMaster Connected.”

4|Page
Step 5.

Select the checkboxes of the values that you want to enable.

Also, on the right window select the following values to be displayed (Massflow,
Temperature, Sensor Signal A, Sensor Signal B, Sensor Driver Output, Density)

Photo 4: “HTT - CoriolisMaster dynamic chart tab.”

Step 6.

Then press button “Start”


Please take a screenshot of the current window and send it to us.

Photo 5: “HTT - CoriolisMaster Measurements.”

5|Page
Step 7.

Go to tab “Common Diagnosis” and press “Get Diagnosis” button and take a
screenshot of the current window.

Photo 6: “HTT - CoriolisMaster Diagnosis.”

Step 8.

Go to tab “Modbus” and press button “Read” take a screenshot of the current window.

Photo 7: “HTT - CoriolisMaster Modbus.”

6|Page
Step 9.

Press button Stop and disconnect from flowmeter.

Step 10.
Follow the same instructions for Outlet Coriolis Flowmeter and revert with results.

Step 11.
Take a photo of the RVG200 screen and forward to us.