SUPER INDUCTIVE SYSTEM

COMPETITOR COMPARISON

All you need to know in one place

FOR INTERNAL USE ONLY!


CONTENT

• Introduction
• Observe all things
• Get familiar with competitors
I. INTRODUCTION
CLARIFY FOR YOURSELF

THERE ARE 3 GROUPS OF MEDICAL DEVICES WHICH


USE ELECTROMAGNETIC FIELD TECHNOLOGY

• 1st group – direct competitors using the same technology


like SIS for treatment of the musculoskeletal system

• 2nd group - competitors using the same technology like SIS


for brain stimulation

• 3rd group – devices using low-intensity electromagnetic


field technology for the treatment of the musculoskeletal
system
II. OBSERVE EVERYTHING
WHAT TO OBSERVE

THERE ARE A FEW PARAMETERS YOU SHOULD OBSERVE


IN COMPETITORS

• Max. stimulation intensity (1st and 2nd group)


• Intensity of the electromagnetic field (3rd group)
• Wide frequency range and ability to deliver 100% intensity
• Cooling system

THE EXPLANATION FOLLOWS


MAX. STIMULATION INTENSITY (Max. dB/dT)

IT´S NOT ABOUT TESLAS BUT THE PARAMETER THAT


DETERMINES THE STIMULATION INTENSITY IN THE TISSUE

• The parameter tells how much current you bring to the tissue
• The higher the value, the more intense stimulation
• The current is induced in the tissue based on Faraday´s Law of
Induction
• Easy calculation:
• I = current
• B = intensity of the electromagnetic field (in Teslas)
• π = a constant (value 3.14)
• T = pulse width
• T/s = unit Tesla per second
FREQUENCY RANGE AND 100% INTENSITY

BRING IT TO THE MAXIMUM

• Frequency is one of the parameters determining the therapeutic


effect
• The most important is to have the ability to deliver 100% of
intensity even with higher frequencies
• This is connected also to the ability to work continuously
• A typical characteristic of competitors is the rapid decrease of
intensity
COOLING SYSTEM

COOL DOWN AND REMEMBER

• The cooling system is crucial in high-intensity electromagnetic


technology
• The cooling system is typically connected to the ability of the
device to work continuously
• There are various types of cooling systems such as air-cooling
(forced or passive), oil or water cooling
WHY SUPER INDUCTIVE SYSTEM WINS?

PROUD TO SAY

• The highest max. stimulation intensity


• Delivers 100% intensity even with high frequencies
• Ability to work continuously due to patented Cool-Flow
technology and Intensity Predictor
• Equipped with Pulse Quality Monitor which ensures smooth
operation of the device and monitors each pulse
III. GET FAMILIAR
WITH COMPETITORS
1st GROUP OF COMPETITORS

Korean company Remed offers 3 devices: The same company manufactures device for
Salus Talent, Salus Talent Pro, Salus Talent A Zimmer: EmField Pro

Slovenian company Iskra Medical


offers 2 devices:
Tesla Stym and Tesla Stym Prestige Korean company Biocon offers:
Biocon Pro 1000
OVERVIEW OF PARAMETERS
Parameter Super Salus Salus Salus EmField Tesla Biocon
Inductive Talent Pro Talent A Talent Pro Stym, Pro 1000
System Prestige

Max. dB/dT 28 21 21 17 21 17 8
(kT/s)

Pulse width 280 450 450 450 450 450 390


(microsec)
EM 2.5 up to 3* up to 3* up to 2.5* up to 3* up to 2.5* up to 1*
intensity (T)
Continuous Yes Only up to Only up to Only up to N/A Only up to Only up to
use 40 Hz 40 Hz 40 Hz 50 Hz 50 Hz
Cooling Forced-air Oil cooling Oil cooling Oil cooling Oil cooling Air cooling Air cooling
cooling
Device´s Pulse Quality No No No No No No
feature Monitor,
Intensity
Predictor

* All competitors claim they have “max 3 Tesla” or “up to 3 Teslas” intensity. This means that their real intensity is much
lower and rapidly decreases when the user wants to work continuously or with higher frequencies. See more on slide
16.
ZIMMER TALKS THE TALK AND WALKS THE WALK

DID YOU KNOW THAT?

• Zimmer as a German high-end brand resells cheap Korean


device
• Zimmer offers transformed Salus Talent device in the different
trolley named EmField Pro
• EmField Pro is not even certified on Zimmer
• The device uses oil cooling and has a high number of service
cases
• Almost 100% of applicators need to be replaced during
warranty
ZIMMER TALKS THE TALK AND WALKS THE WALK

Korean company Remed supplies Zimmer with the The labelling from “Zimmer device” clearly
Salus Talent technology, but in a different design states that Remed is a manufacturer
OVERVIEW OF MAX.STIMULATION INTENSITY
OVERVIEW OF INTENSITIES IN RELATION TO
FREQUENCY
EXAMPLE OF REAL MEASUREMENT*
1st GROUP KEY POINTS SUMMARY

SUPER INDUCTIVE SYSTEM & SALUS TALENT FAMILY


• Lower values of max. stimulation intensities
• Limited frequency range and ability to deliver 100% intensity
• Cooling liquid leakage issues

SUPER INDUCTIVE SYSTEM & TESLA STYM FAMILY


• Lower values of max. stimulation intensities
• Limited ability to deliver 100% intensity with frequency higher
than 50 Hz

SUPER INDUCTIVE SYSTEM & BIOCON PRO 1000


• Lower values of max. stimulation intensities
• Limited frequency range and ability to deliver 100% intensity
2nd GROUP OF COMPETITORS

Magventure
Magstim Nexstim

Mag and More


Tamas
Neuro-MS/D Stimulator
OVERVIEW OF PARAMETERS
Parameter Super Magstim MagPro R20 Power Mag Tamas Nexstim Neuro-MS/D
Inductive Stimulator
System

f (Hz) 1 – 150 1 – 100 1 – 20 0.1 – 100 0.5 - 50 0.1 - 50 0.1 - 30

Continuous Yes Max. 3000 Max. Max. 1800 Only up to Max. 1900 Only up to
use pulses 400 – 2000 pulses per 15 Hz pulses 5 Hz
per session pulses per session at 1 Hz
session
Cooling Forced-air External air External Passive Oil cooling Optional air Optional
cooling or liquid liquid cooling cooling cooling silicone oil
cooling unit unit cooling

Device´s Pulse Quality No No No No No No


feature Monitor,
Intensity
Predictor

Device Single unit Multi-unit Multi-unit Multi-unit Single unit Multi-unit Multi-unit
construction
OVERVIEW OF MAGSTIM LIMITATIONS

Frequency Preset protocol

10 Hz Possible to trigger less than 60 trains in time ratio 10s:50s


on 50 %; 80 %; 100 % intensity
25 Hz Possible to trigger less than 30 trains on 50 %
Possible to trigger only 10 trains on 80 %
Possible to trigger only 4 trains on 100% in therapy ratio 5s:25s
1 Hz Possible to trigger 3600 single shots
on 50 %; 80 %; 100 % intensity

Preset protocols
Magstim Air-Cooled Double 70mm Coil System User´s Manual
OVERVIEW OF MAGVENTURE LIMITATIONS

Frequency Intensity

5 Hz 100 %

10 Hz 75 %

15 Hz 40 %

20 Hz 35 %

Maximal intensity in relation to frequency


Magventure MagPro R20 User´s Manual
OVERVIEW OF MAG AND MORE LIMITATIONS

Frequency Intensity

30 Hz 100 %

50 Hz 80 %

100 Hz 70 %

Maximal intensity in relation to frequency


Mag and More User´s Manual
OVERVIEW OF TAMAS LIMITATIONS

Frequency Intensity

15 Hz 100 %

20 Hz 80 %

30 Hz 50 %

50 Hz 30 %

Maximal intensity in relation to frequency


Remed Tamas User´s Manual
OVERVIEW OF NEXSTIM LIMITATIONS

Frequency Intensity

0.1 - 10 Hz 100 %

20 Hz 96 %

30 Hz 77 %

40 Hz 59 %

50 42 %

Maximal intensity in relation to frequency


Nexstim User´s Manual
OVERVIEW OF NEUROSOFT-MS/D LIMITATIONS

Frequency Intensity Intensity


without extra power supply unit with extra power supply unit

5 Hz 100 % 100 %

10 Hz 69 % 100 %

15 Hz 53 % 100 %

20 Hz 43 % 100 %

25 Hz 37 % 89 %

30 Hz 30 % 79 %

Maximal intensity in relation to frequency


Neurosoft User´s Manual of Neuro-MS/D Stimulator
2nd GROUP KEY POINTS SUMMARY

DIFFERENT TECHNOLOGY DUE TO

• Different intended use – diagnostics, brain stimulation


• Limited frequency range and ability to deliver 100% intensity
• Use of more coils during 1 therapy due to cooling issues
• Multi-unit device construction
3rd GROUP OF COMPETITORS

Omnitron S&T Rehatron Alpha

Storz Magnetolith and


Theracell
Papimi
OVERVIEW OF PARAMETERS

Parameter Super Omnitron Storz Storz S&T Papimi


Inductive Magnetolith Theracell 100 Rehatron Alpha
System

EM intensity 2.5 T 20 – 80 mT 80 mT* 40 – 80 mT 40 – 84 mT 50 – 100 mT

f (Hz) 1 – 150 1–3 1-10** 1–3 1–3 1–8

Intensity During the Before the N/A N/A N/A N/A


settings therapy therapy start

Cooling Patented and None Water circuit*** None None None


system maintenance- that requires
free and maintenance
forced-air
cooling
Continuous Yes Max. 20 minutes Max. 15 000 pulses Max. 20 minutes Max. 20 minutes Max. 9 minutes
use and 6 hours**** or 888 pulses

Device´s Pulse Quality No No No No No


feature Monitor,
Intensity
Predictor
STORZ MAGNETOLITH MAINTENANCE
OVERVIEW OF ELECTROMAGNETIC FIELD
INTENSITY
OVERVIEW OF FREQUENCY RANGE
3rd GROUP KEY POINTS SUMMARY

THIS IS NOT A SUPER INDUCTIVE SYSTEM COMPETITOR

• Very low magnetic field intensities (mT)


• Limited frequency range
• Limited continuous use
• Complicated maintenance of water circuit (Storz)
THANK YOU