You are on page 1of 7

The Properties of Damaged and Undamaged Suture Used in

Metal and Bioabsorbable Anchors: An In Vitro Study

Patrick B. Wright, M.D., Jeffrey E. Budoff, M.D., Ming Long Yeh, Ph.D., Zachary S. Kelm,
and Zong-Ping Luo, Ph.D.

Purpose: The purpose of this study was to determine the mechanical properties of undamaged and
damaged sutures in metal and bioabsorbable suture anchors. Methods: Undamaged and damaged
FiberWire (Arthrex, Naples, FL), Tevdek (Deknatel, Mansfield, MA), and PDS (Ethicon, Somerville,
NJ) sutures were tested by a single pull to failure while being pulled parallel to the axis of either a
metal or bioabsorbable suture anchor. Sutures were damaged by use of a razor blade incorporated into
a custom-designed jig. The friction of the sutures through the anchor eyelets was also tested. Results:
For both anchor types, FiberWire was the strongest suture studied. Undamaged PDS had a signifi-
cantly greater load to failure than Tevdek. Although all sutures lost significant strength when
damaged, PDS lost the most, with damaged PDS becoming significantly weaker than damaged
Tevdek. Damaged FiberWire was significantly stronger in metal anchors compared with bioabsorb-
able anchors, with failure of the bioabsorbable suture eyelet preventing testing of undamaged
FiberWire. Neither undamaged nor damaged PDS or Tevdek had a significant difference in strength
between metal and bioabsorbable anchors. However, in metal anchors the mechanical properties of
undamaged Tevdek were inferior to those of the other undamaged sutures tested. For undamaged or
damaged sutures through either anchor type, PDS suture had the highest coefficient of friction,
significantly higher than FiberWire and Tevdek. All sutures, undamaged or damaged, had signifi-
cantly less friction in bioabsorbable anchors compared with metal anchors. Conclusions: The
FiberWire-anchor construct is significantly weaker when bioabsorbable anchors are used instead of
metal anchors. For Tevdek and PDS sutures, the anchor type does not affect the strength of the
construct, as the suture is the limiting factor. When used with suture anchors, PDS has the most
friction of the sutures tested, potentially leading to suture damage, which disproportionately weakens
PDS compared with the other sutures tested. For both undamaged and damaged sutures, bioabsorb-
able anchors lead to less friction than do metal anchors, which may lessen suture damage in vivo.
Clinical Relevance: The mechanical properties of damaged suture are important to all surgeons who
use suture arthroscopically. Key Words: Suture—Damage—Biomechanics—Suture anchor.

S utures are an integral part of arthroscopic shoulder


repairs and reconstructions. Especially in the
shoulder, they are routinely passed through and along-
side sharp instruments that may damage them. In
addition, they are routinely used in conjunction with
suture anchors, which may be composed of either
metal or a bioabsorbable polymer. The in vivo envi-
ronment exposes the suture to cyclic loading and
abrasion against the anchor’s eyelet during arthro-
From the Department of Orthopaedic Surgery, Baylor College of
Medicine, Houston, Texas, U.S.A. scopic knot tying and during early postoperative re-
The authors report no conflict of interest. habilitation.1 The properties of damaged suture have
Address correspondence and reprint requests to Jeffrey E.
Budoff, M.D., Department of Orthopaedic Surgery, Baylor College
been examined.2 The influence of the type of anchor
of Medicine, 6620 Main St, Suite 1300, Houston, TX 77030, U.S.A. on suture/anchor friction, which is a factor in the
E-mail: Jebudoff@yahoo.com occurrence and propagation of suture damage, as well
© 2007 by the Arthroscopy Association of North America
0749-8063/07/2306-6357$32.00/0 as on the mechanical properties of the suture once
doi:10.1016/j.arthro.2007.01.008 damaged, is important for the arthroscopic surgeon to

Arthroscopy: The Journal of Arthroscopic and Related Surgery, Vol 23, No 6 (June), 2007: pp 655-661 655
656 P. B. WRIGHT ET AL.

be aware of when deciding on his or her choice of


suture or anchor. The purpose of this study was to
determine the mechanical properties of undamaged
and damaged sutures in metal and bioabsorbable su-
ture anchors. The null hypothesis is that there is no
significant difference in mechanical properties for any
of the conditions tested.

METHODS
Three types of suture were tested: Tevdek (Deknatel,
Mansfield, MA), FiberWire (Arthrex, Naples, FL),
and PDS (Ethicon, Somerville, NJ). The anchors used
were the metal Arthrex 5.0-mm Corkscrew and the
bioabsorbable Arthrex 5.0-mm Bio-Corkscrew. Simi-
lar to other studies,2,3 preliminary testing did not show
differences in failure behavior between dry and wet
suture material. Therefore, to be consistent with pre-
vious studies,1-5 testing of all sutures was performed
dry at room temperature. The manufacturers have FIGURE 1. Diagram of suture anchor U test. (Reprinted with
indicated that the difference between room and body permission.3)
temperature would be too small to make a significant
difference in the biomechanical properties of the su-
tures. it, toward the center of the eyelet. All tests were
Twenty-four sutures of each type were used for the performed at a displacement rate of 60 mm/min on an
load-to-failure/failure stress test; 12 were used with MTS 858 Bionix machine (MTS, Eden Prairie, MN).
metal anchors and 12 with bioabsorbable anchors. The anchor was rotated 180° after each test to reduce
Each of these groups was subdivided into 6 sutures to any error resulting from manufacturing variance. Any
be tested undamaged and 6 to be tested damaged. To damaged anchors were discarded. Any test that re-
simulate potential damage incurred at the anchor site sulted in anchor failure was discarded so that only data
during surgical procedures, sutures were then dam- where suture strength was the limiting factor were
aged by cutting approximately 33% of their width analyzed. The load to failure (LTF) was recorded as
with a razor blade placed in a custom jig. This per- the maximal load.
centage of suture damage was chosen after a pilot Technically, the term ultimate tensile stress is only
study so that the newer polyethylene suture used for tests that solely measure material properties,
(FiberWire) could be tested in bioabsorbable anchors such as the straight-line pull test, and this term is
without routinely destroying the anchor’s eyelet. The technically incorrect for a test involving an interaction
goal was to minimize anchor breakage as a confound- with another object (i.e., a suture anchor). Therefore,
ing factor. The total width of the suture and the cut for the tensile failure test, we use the term failure
depth were measured by use of a light microscope stress (FS) to denote the material property of the
attached to an electric micrometer to determine total suture, as well as its interaction with the suture eyelet.
remaining cross-sectional area. FS was calculated as LTF divided by cross-sectional
For all tests, the suture anchors were clamped be- area, which had the decreased area corrected for in
tween the flanges of a vise. The suture was passed damaged sutures. Because the purpose of this study is
through the suture anchor eyelet. The free ends of the to evaluate the biomechanical properties of suture and
suture were wound around a metallic hook to avoid not the anchor, only trials that failed because of suture
weakening the suture with knot tying. For single-pull, breakage were included in the analysis.
load-to-failure testing, the load was applied to both The friction of damaged and undamaged sutures
free ends of the suture parallel to the axis of the anchor was measured while being pulled through a suture
insertion, as in the U test2,3 (Fig 1). For tests involving anchor eyelet at a 45° angle, a different angle than was
damaged suture, the damaged area was placed against used for the LTF suture anchor test. An additional 24
the superior rim of the anchor loop, facing away from sutures of each type were used for the friction test; 12
SUTURES AND METAL AND BIOABSORBABLE ANCHORS 657

recorded on a computer. The quotient of Fa/Fp


(Fig 2) is an indicator for the frictional resistance of
the suture in the anchor eyelet. Without friction, Fp
would be equal to Fa, and the quotient (Fa/Fp) would
equal 1.3 A pilot study confirmed that suture pulled
straight through a suture anchor had mechanical
properties similar to suture without an anchor
pulled to failure.
One-way analysis of variance with the post hoc
Tukey test was used to test the differences among the
sutures. The Student t test was used to compare cut
versus uncut sutures and metal versus bioabsorbable
anchors for each kind of suture. The difference was
considered significant at P ⬍ .05.

RESULTS
Suture Characteristics
Despite the fact that all sutures tested were No. 2
sutures, their cross-sectional areas varied. The cross-
FIGURE 2. Photograph of suture anchor U test. The body of the sectional area of uncut Tevdek was 0.26 ⫾ 0.03 mm2,
suture anchor is securely clamped between the flanges of a vise that of uncut PDS was 0.29 ⫾ 0.02 mm2, and that of
in such a way as to avoid damaging the anchor or altering the uncut FiberWire was 0.39 ⫾ 0.12 mm2, significantly
eyelet.
greater than both Tevdek and PDS (P ⬍ .05). The
percentage of suture cut for each test was very repro-
ducible within 3.1%, ranging from 33.50% to 36.60%.
were used with metal anchors and 12 with bioabsorb-
able anchors. Each of these groups was subdivided
into 6 sutures to be tested undamaged and 6 to be
tested damaged. The bodies of the suture anchors were
securely clamped between the flanges of a vise in such
a way as to avoid damaging the anchor or altering the
eyelet (Fig 2). The technique has been previously
described.2,3 The sutures were attached to an MTS
machine’s cross-head by use of Capstan grips, with a
passive load cell (range, 0 to 1,000 N). All tests were
performed at a displacement rate of 60 mm/min. One
suture branch was passed through the anchor eyelet
and pulled straight in line with the suture anchor axis
by an MTS machine, and the force at failure (Fa) was
recorded. The resulting force on the second suture
branch with an angle of 45° to the suture anchor axis
(Fp) was measured simultaneously by use of a second
electronic load cell (Fig 3).
For damaged sutures, the cut area was placed on the
suture branch held at 45° from the suture anchor axis,
1 cm from the anchor eyelet, facing toward the supe-
rior rim of the anchor loop, to ensure that the damaged
area would pass against the anchor loop. Loads were
FIGURE 3. Diagram of friction test. The testing machine pulls one
electronically transformed to an analog voltage de- branch (Fa, active) while the tensile load on the second branch (Fp,
vice, processed by an analog/digital converter, and passive) is recorded simultaneously. (Reprinted with permission.3)
658 P. B. WRIGHT ET AL.

Tensile Strength Testing difference in the FS between undamaged FiberWire


and PDS. We could not evaluate undamaged
The results for LTF and FS of cut and uncut sutures FiberWire using a bioabsorbable anchor, because it
through metal and bioabsorbable anchors are summa- consistently destroyed the anchors. There was no sig-
rized in Table 1. For undamaged sutures through a nificant difference between the FS of undamaged PDS
metal anchor, FiberWire had the highest LTF, signif-
and Tevdek (P ⫽ .38).
icantly higher than the other sutures (P ⫽ .01). Un-
As for LTF, when damaged, PDS lost a significantly
damaged FiberWire could not be evaluated with bio-
greater amount of FS than the other sutures. For
absorbable anchors, because it consistently destroyed
damaged sutures with either anchor, PDS had a
the anchors. Damaged FiberWire broke 2 bioabsorb-
significantly lower FS than FiberWire and Tevdek
able anchor eyelets; the results of these trials were
(P ⫽ .00001). For metal anchors, damaged FiberWire
excluded from data analysis. Neither Tevdek nor PDS
resulted in any anchor failures. The eyelets of the had a significantly higher FS than damaged Tevdek
bioabsorbable anchors routinely broke at the top. (P ⫽ .0007). For bioabsorbable anchors, there was no
There was no evidence of anchor slippage from the significant difference between the FS of damaged
vise during testing. FiberWire and damaged Tevdek.
For damaged sutures with either metal or bioabsorb- Damaged FiberWire had a significantly higher LTF
able anchors, FiberWire had the highest LTF, signif- (P ⫽ .004) and FS (P ⫽ .009) with metal anchors
icantly stronger than the other sutures (P ⫽ .0001). compared with bioabsorbable anchors. Again, undam-
Failure routinely occurred at the cut in the suture. aged FiberWire could not be tested with bioabsorbable
Undamaged sutures also failed at the eyelet, at the anchors. Neither undamaged nor damaged PDS or
“kink” in the suture caused by the eyelet. Tevdek had a significant difference in strength (LTF
For both metal and bioabsorbable anchors, undam- or FS) between metal and bioabsorbable anchors
aged PDS had a significantly greater LTF than Tevdek (P ⬎ .05).
(P ⫽ .0004 for metal anchors and P ⫽ .01 for bioab- For both types of anchors, the LTF was signifi-
sorbable anchors). PDS lost a significant amount of cantly lower for both damaged Tevdek and PDS
LTF when damaged and became significantly weaker (P ⫽ .004) compared with undamaged sutures. For
than damaged Tevdek (P ⫽ .00001 for metal anchors metal anchors, the LTF of undamaged FiberWire
and P ⫽ .00001 for bioabsorbable anchors). was significantly greater than the LTF of damaged
For undamaged sutures with a metal anchor, FiberWire (P ⫽ .0001). As previously mentioned,
FiberWire and PDS both had a significantly greater FS undamaged FiberWire could not be tested with bio-
than Tevdek (P ⫽ .01). There was no significant absorbable anchors. However, this suggests that the

TABLE 1. LTF and FS of Cut and Uncut Tevdek, PDS, and FiberWire in Metal and Bioabsorbable Anchors
Tevdek PDS FiberWire Significance

LTF (N)
Undamaged suture
Metal anchor 202.80 ⫾ 7.70 248.37 ⫾ 16.00 365.88 ⫾ 15.57 FW ⬎ PDS ⬎ Tevdek
Bioabsorbable anchor 200.68 ⫾ 17.39 240.72 ⫾ 21.24 NA PDS ⬎ Tevdek
Damaged suture
Metal anchor 135.02 ⫾ 10.67 55.08 ⫾ 2.77 262.88 ⫾ 30.35 FW ⬎ Tevdek ⬎ PDS
Bioabsorbable anchor 138.41 ⫾ 22.54 53.45 ⫾ 5.16 218.63 ⫾ 8.93 FW ⬎ Tevdek ⬎ PDS
FS (MPa)
Undamaged suture
Metal anchor 833.94 ⫾ 21.65 1,110.50 ⫾ 193.26 1,033.08 ⫾ 109.83 FW ⫽ PDS ⬎ Tevdek
Bioabsorbable anchor 916.57 ⫾ 150.25 1,028.36 ⫾ 223.21 NA PDS ⫽ Tevdek
Damaged suture
Metal anchor 837.46 ⫾ 57.45 376.04 ⫾ 100.13 1,208.87 ⫾ 164.87 FW ⬎ Tevdek ⬎ PDS
Bioabsorbable anchor 1,011.94 ⫾ 262.43 359.44 ⫾ 100.37 839.27 ⫾ 118.08 FW ⫽ Tevdek ⬎ PDS

NOTE. A “greater than” sign indicates significantly greater, whereas an “equals” sign indicates that there is no significant difference
between the two sutures’ values.
Abbreviation: NA, not applicable.
SUTURES AND METAL AND BIOABSORBABLE ANCHORS 659

same relationship holds true, because the bioabsorb- icant change in friction when damaged (P ⫽ .17).
able anchor’s eyelet was strong enough to allow Tevdek had a significant decrease in friction when
testing of 6 of 8 damaged FiberWire sutures but no damaged (P ⫽ .014). When being pulled through a
undamaged FiberWire sutures. bioabsorbable anchor, no suture had a significant
For metal anchors, there was no significant differ- change in friction when damaged (P ⫽ .52).
ence in FS for FiberWire and Tevdek between undam-
aged and damaged sutures (P ⬎ .05). For bioabsorb-
able anchors, Tevdek had no significant change in FS DISCUSSION
when damaged (P ⬎ .05). Damaged PDS had a sig- In this study we tested the strength (LTF and FS)
nificantly lower FS compared with undamaged PDS and friction of undamaged and damaged sutures in
for both metal (P ⫽ .0001) and bioabsorbable anchors metal and bioabsorbable anchors. For metal anchors,
(P ⫽ .002). failure occurred exclusively by suture breakage. Dam-
Friction Testing aged FiberWire was found to be stronger (higher LTF
and FS) in metal anchors than in bioabsorbable an-
The results of friction through the anchor are sum- chors because in bioabsorbable anchors, FiberWire’s
marized in Table 2. For undamaged sutures through high LTF shifts the failure mechanism from suture
either a metal or bioabsorbable anchor, PDS suture failure to eyelet failure. This is similar to the findings
had the highest coefficient of friction, significantly of Bardana et al.,1 who reported that constructs using
higher than the other sutures (P ⫽ .0003). There was bioabsorbable anchors had a mixture of suture and
no significant difference in friction between undam- anchor eyelet failures. The metal Corkscrew anchor
aged Tevdek and FiberWire through either type of and its eyelet are cast from the same piece of metal,
anchor (P ⫽ .44). leading to significant anchor eyelet strength. The bio-
For damaged sutures through either anchor, PDS absorbable anchor’s eyelet is made of No. 3 Tevdek
suture again had the highest coefficient of friction, suture, which is molded into the anchor’s polymer. It
significantly greater than the other sutures (P ⫽ .0001 is therefore significantly weaker than the metal an-
for metal anchors and P ⫽ .0003 for bioabsorbable chor’s eyelet.
anchors). Damaged FiberWire had significantly Although all tests that resulted in anchor damage
greater friction than damaged Tevdek with metal an- were not used for the purposes of the results, the fact
chors (P ⫽ .004), but there was no significant differ- that 2 of 8 bioabsorbable anchor eyelets were de-
ence between these 2 sutures with bioabsorbable an- stroyed by damaged FiberWire lowers the mean
chors. strength of the construct by invalidating those trials
All sutures, damaged and undamaged, had signifi- in which the strength of the damaged FiberWire
cantly less friction in bioabsorbable anchors compared was above that of the bioabsorbable anchor’s eye-
with metal anchors (P ⫽ .028 for undamaged suture, lets. Given that the tensile strength of undamaged
P ⫽ .0016 for damaged FiberWire and PDS, and P ⫽ FiberWire was higher than that of the bioabsorbable
.05 for damaged Tevdek). When being pulled through anchor’s eyelet, the bioabsorbable anchors themselves
a metal anchor, FiberWire had a significant increase in limited the strength of the FiberWire-anchor con-
friction when damaged (P ⫽ .03). PDS had no signif- struct, and it was not possible to test the mechanical

TABLE 2. Friction of Uncut and Cut Tevdek, PDS, and FiberWire in Metal and Bioabsorbable Anchors
Fa/Fp

Tevdek PDS FiberWire Significance

Uncut suture
Metal anchor 0.32 ⫾ 0.01 0.37 ⫾ 0.02 0.31 ⫾ 0.02 PDS ⬎ FW ⫽ Tevdek
Bioabsorbable anchor 0.29 ⫾ 0.01 0.33 ⫾ 0.01 0.29 ⫾ 0.01 PDS ⬎ FW ⫽ Tevdek
Cut suture
Metal anchor 0.29 ⫾ 0.01 0.39 ⫾ 0.01 0.33 ⫾ 0.02 PDS ⬎ FW ⬎ Tevdek
Bioabsorbable anchor 0.29 ⫾ 0.01 0.34 ⫾ 0.01 0.29 ⫾ 0.02 PDS ⬎ FW ⫽ Tevdek

NOTE. A “greater than” sign indicates significantly greater, whereas an “equals” sign indicates that there is no significant difference
between the 2 sutures’ values.
660 P. B. WRIGHT ET AL.

properties of uncut FiberWire in this setting. For decreased so greatly when damaged. Damaging a
Tevdek and PDS, the sutures themselves were the braided suture does not increase the stress concentra-
limiting factor of the suture– bioabsorbable anchor tion as much because there is no leading edge, given
construct. that there are still intact bundles on the edge of the cut.
Undamaged FiberWire had a significantly higher PDS was one of the first sutures used for arthro-
LTF than PDS with both metal and bioabsorbable scopic shoulder reconstructions.6-8 PDS is an absorb-
anchors, and both FiberWire and PDS had a signifi- able monofilament suture composed of the polyester
cantly higher LTF than Tevdek. With metal anchors, poly(p-dioxanone). Surgeons then began using nonab-
there was no significant difference between the FS for sorbable sutures, such as Tevdek. Tevdek is a braided
FiberWire and PDS, implying that with its greater polyester suture impregnated with polytetrafluoroeth-
cross-sectional area adjusted for, the undamaged com- ylene. Newer hybrid sutures have been developed and
posite FiberWire material had no increased strength are now popular for shoulder reconstructions.9,10
compared with undamaged PDS. For metal anchors, FiberWire has an ultrahigh–molecular weight polyeth-
both undamaged FiberWire and PDS had a signifi- ylene core with a braided polyethylene terephthalate
cantly higher FS than undamaged Tevdek. For bioab- polyester ultrahigh–molecular weight polyethylene
sorbable anchors, there was no significant difference jacket.
in FS between undamaged PDS and Tevdek. This sug- Under cyclic loading, sutures may abrade on their
gests that when interacting with a metal eyelet, Tevdek anchor’s eyelet. The friction created depends on the
has inferior mechanical (structural and material) proper- design of the anchor eyelet, the roughness of the
ties compared with both PDS and FiberWire sutures. matching surfaces, and the arc of contact between
Undamaged PDS may have a significantly greater the suture material and the anchor eyelet. In the study
FS than Tevdek through metal anchors, but not of Bardana et al.,1 sutures performed significantly
through bioabsorbable anchors, because Tevdek may better when cycled at an angle of pull in line with the
be more sensitive to the sharper edges of the metal anchors. To compare metal and bioabsorbable an-
eyelet. As noted in a previous study,2 PDS lost sig- chors, we chose to use these optimum conditions. We
nificantly more strength (LTF and FS) when damaged noted that, for all undamaged and damaged sutures,
than did the other sutures. For both anchor types, friction was less in bioabsorbable anchors compared
damaged PDS had a significantly lower strength (LTF with metal anchors. In addition, when being pulled
and FS) than did Tevdek and FiberWire. For metal through a bioabsorbable anchor, no suture had a sig-
anchors, damaged FiberWire had a greater strength nificant change in friction when damaged. This agrees
(LTF and FS) than did Tevdek. For bioabsorbable with the findings of previous authors who noted that
anchors, although damaged FiberWire had a higher metallic anchors tend to have rough eyelet surfaces
LTF than damaged Tevdek, there was no significant with sharp edges, which may damage sutures, leading
difference between their FS. This implies that, with its to failure,1,5 whereas the eyelets of bioabsorbable im-
greater cross-sectional area accounted for, the dam- plants have smoother edges, decreasing suture abra-
aged composite FiberWire material did not have sion under cyclic loading.11
greater strength than cut Tevdek. It is possible that PDS (undamaged or damaged, in either metal or
Tevdek may be relatively stronger in bioabsorbable bioabsorbable anchors) had significantly more friction
anchors because of the lack of sharp edges on their than the other sutures tested, which may predispose it
eyelets. to damage, which then weakens it significantly. In
All sutures lost significant LTF when damaged. metal anchors FiberWire’s friction increased signifi-
Neither FiberWire nor Tevdek had a significant loss of cantly when damaged and was significantly greater
FS when damaged, implying that for these sutures, FS than cut Tevdek. This may be due to the fact that
is, in fact, truly a material property. However, PDS did FiberWire is stiffer than Tevdek,2 allowing its cut
have a significant loss of FS when damaged. This may edge to catch the metal eyelet with greater force.
be because of the fact that PDS is a monofilament and As is the case for all in vitro studies, ours has
not a braided suture (made up of bundles of individual several limitations. The amount of damage studied
fibers), such as FiberWire and Tevdek. When mono- here (33% of suture diameter) may over- or under-
filament suture is damaged, there is an increase of represent the damage present in common clinical sit-
stress concentration at the leading edge of the cut, uations, and a razor blade may not exactly simulate
which greatly decreases the force required to cause damage that occurs in vivo. However, it is difficult to
failure. This may be why both the LTF and FS of PDS classify the damage that occurs to a suture in the
SUTURES AND METAL AND BIOABSORBABLE ANCHORS 661

“typical” in vivo situation, because suture damage CONCLUSIONS


may vary with the procedure performed, equipment
used, and surgeon’s skill level. The strength of the eyelet of the metal suture anchor
Sutures may act differently in an in vivo environ- used in this study exceeded the suture strength for all
ment compared with an in vitro experiment. Specifi- sutures tested in vitro. Bioabsorbable suture eyelet
cally, friction between a suture and the anchor eyelet strength was less than undamaged FiberWire suture
may be different in a fluid environment. Furthermore, strength but stronger than Tevdek and PDS via this
the friction-testing model—where one suture end is testing model. The undamaged or damaged FiberWire-
pulled straight in line with the suture anchor axis and anchor construct is significantly weaker when bioabsorb-
the second suture end is pulled at a 45° angle—may able anchors are used instead of metal anchors. For
not accurately reflect the resulting pull of the sutures Tevdek and PDS sutures, the anchor type does not
in the human shoulder. affect the strength of the construct, because the suture
We used static tensile failure rather than cyclic is the limiting factor. When used with suture anchors,
loading to failure, which may better represent the PDS has the most friction of the sutures tested, poten-
clinical situation. This study is an initial attempt to tially leading to suture damage, which disproportion-
examine the biomechanical properties of undamaged ately weakens PDS compared with the other sutures
and damaged sutures in metal and bioabsorbable an- tested. For both undamaged and damaged sutures,
chors, and further study is needed to fully understand bioabsorbable anchors lead to less friction than do
suture interactions in the in vivo environment. metal anchors under dry in vitro conditions.
Because damaged FiberWire destroyed 2 bioab-
sorbable anchor eyelets, 2 of this suture’s strongest REFERENCES
runs had to be disqualified, potentially lowering the
mean strength of damaged FiberWire noted in this 1. Bardana DD, Burks RT, West JR, Greis PE. The effect of
suture anchor design and orientation on suture abrasion: An in
study. Therefore the reported results probably under- vitro study. Arthroscopy 2003;19:274-281.
estimate the strength of damaged FiberWire. Alterna- 2. Wright PB, Budoff JE, Yeh ML, Kelm ZS, Luo ZP. Strength of
tively, because visually undamaged anchors were reused, damaged suture: An in vitro study. Arthroscopy 2006;22:1270-
1275.e3. Available online at www.arthroscopyjournal.org.
it is possible that the anchor eyelets that failed were 3. Meyer DC, Nyffeler RW, Fucentese SF, Gerber C. Failure of
subtly weakened, in which case the strength of damaged suture material at suture anchor eyelets. Arthroscopy 2002;18:
FiberWire would not have been underestimated. 1013-1019.
4. Barber FA, Herbert MA, Richards DP. Sutures and suture
Ideally, sutures should be strong enough to with- anchors: Update 2003. Arthroscopy 2003;19:985-990.
stand the forces applied to them and should show little 5. Rupp S, Georg T, Gauss C, Kohn D, Seil R. Fatigue testing of
creep during the healing process. Although newer suture anchors. Am J Sports Med 2002;30:239-247.
6. Wolf EM. Arthroscopic capsulolabral repair using suture an-
sutures generally have greater strength and stiffness chors. Orthop Clin North Am 1993;24:59-69.
than older sutures, it is unknown whether this strength 7. Wolf EM, Pennington WT, Agrawal V. Arthroscopic rota-
is needed for clinical success or whether the increased tor cuff repair: 4- to 10-year results. Arthroscopy 2004;20:
5-12.
stiffness of these new sutures might lead to the sutures 8. McIntyre LF, Caspari RB, Savoie FH. The arthroscopic treat-
cutting through the soft tissues or the cortical margin ment of anterior and multidirectional shoulder instability. Instr
of the bone over which they are tensioned.11,12 Other Course Lect 1996;45:47-56.
9. Budoff JE. Tendinopathy of the rotator cuff and proximal
important properties such as knot security, suture cut- biceps. In: Trumble TE, Budoff JE, Cornwall R, eds. Core
out through soft tissues, and biocompatibility were knowledge in orthopaedics: Hand, elbow and shoulder. Phil-
also not examined in our study. Therefore our results adelphia: Elsevier Sciences, 2005;573-592.
10. Budoff JE. Glenohumeral instability, adhesive capsulitis and
should be extrapolated with care. superior labral anteroposterior lesions. In: Trumble TE, Budoff
Although other new hybrid sutures are now being JE, Cornwall R, eds. Core knowledge in orthopaedics: Hand,
marketed, we tested only one of each suture class elbow and shoulder. Philadelphia: Elsevier Sciences, 2005;
593-612.
(absorbable PDS, nonabsorbable Tevdek, and hybrid 11. McFarland EG, Park HB, Keyurapan E, Gill HS, Selhi HS. Suture
FiberWire) to provide insight into the biomechanical anchors and tacks for shoulder surgery, part 1: Biology and
differences between these different suture classes. The biomechanics. Am J Sports Med 2005;33:1918-1923.
12. Bynum CK, Lee S, Mahar A, Tasto J, Pedowitz R. Failure
results of this study may be updated with other more mode of suture anchors as a function of insertion depth. Am J
recently developed sutures at a future time. Sports Med 2005;33:1030-1034.

You might also like