Anyscan 720

User’s Manual
Urine Analyzer
Contents
1. General information on ANYSCAN720 ---------------------------------------- 3
1.1 Introduction ------------------------------------------------------------------- 3
1.2 System specification --------------------------------------------------------- 4
1.3 Name of components -------------------------------------------------------- 5
2. System installation ----------------------------------------------------------------- 7
2.1 Components ------------------------------------------------------------------- 7
2.2 Connection to PC ------------------------------------------------------------- 7
2.3 Brief operating instruction -------------------------------------------------- 8
3. Components and their functions ------------------------------------------------- 8
3.1 Keyboard ---------------------------------------------------------------------- 8
3.2 LCD ---------------------------------------------------------------------------- 9
3.3 Printer -------------------------------------------------------------------------- 9
3.4 Waste box --------------------------------------------------------------------- 9
3.5 Strip loading sensor ----------------------------------------------------------10
4. How to operate ANYSCAN 720------------------------------------------------- 10
4.1 System initiation -------------------------------------------------------------- 11
4.2 System calibration------------------------------------------------------------- 11
4.3 Measurement ------------------------------------------------------------------ 12
4.4 Loading the printing paper -------------------------------------------------- 14
5. System setup menu ---------------------------------------------------------------- 15
5.1 System сlock setting --------------------------------------------------------- 16
5.2 System configuration -------------------------------------------------------- 16
5.3 Registration ------------------------------------------------------------------- 18
5.4 Database (how to use the saved database) -------------------------------- 19
5.5 Database block control (transferring and reprinting saved data)------- 20
6. Maintenance of ANYSCAN 720------------------------------------------------ 21
6.1 Daily check-up --------------------------------------------------------------- 21
6.2 Storage precautions ---------------------------------------------------------- 21
6.3 Errors -------------------------------------------------------------------------- 21
6.4 Error messages and troubleshooting -------------------------------------- 23
7. Disassembly/assembly of ANYSCAN 720 for cleaning --------------------- 26
7.1 Front panel removal---------------------------------------------------------- 26
7.2 Push bar and strip loading plate removal --------------------------------- 26
7.3 Cleaning the strip transfer belt --------------------------------------------- 27
7.4 Strip loading plate and push bar installation ----------------------------- 27
7.5 Front panel installation ------------------------------------------------------ 28
8. Important ---------------------------------------------------------------------------- 29

CDP-EA720-(R) 2 2022-04-7
1. General information on ANYSCAN 720

1.1 Introduction
Urine test strips simplify laboratory diagnosis through ease of use, high sensitivity and specificity.
These benefits allow you to identify pathological changes in the urine quick and reliably. The addition
of the Urine Analyzer to the urinalysis process allows for standardization and efficiency of urine test
by eliminating potential sources of error associate with the visual reading of test strips, such as
improper lighting at the workplace, or different color discrimination by the user or different timing
when the values are read.

ANYSCAN 720 is a urine chemistry analyzer which can be easily operated by a medical doctor or a
laboratory technician.
• ANYSCAN 720 can only be used with Self-Stik strips (2~11) manufactured by CHUNGDO
PHARM. CO., LTD.
• Maximum 720 tests per hour
• Sample (patient)’s ID input via keyboard or barcode reader.
• Memory capacity of maximum 100,000 tests.
• Test parameters for ANYSCAN 720
Urobilinogen Glucose Ketones
Bilirubin Protein Nitrite
pH Occult Blood S.G. (Specific Gravity)
Leukocytes Ascorbic Acid

CDP-EA720-(R) 3 2022-04-7
1.2 System specification
Supply Voltage DC 12V 3.33A
Input 100-240V AC, 50/60Hz, 1.2A
Power adapter
Output 12V 3.33A
Depth: 260 mm (10.2 in)
Size Width: 320 mm (12.6 in)
Height: 178 mm (7.0 in)
Weight 2.4 Kg (5.3 lb.)
Maximum 720 tests per hour
Technical
ID input via keyboard, PC or barcode reader
characteristics
Possibility of changing the user interface to any language
Memory capacity 100,000 test results
Print Internal thermal printer
LCD 320x240 QVGA Color
RS-232C interface
- COM1: Communication with PC.
- COM2: Communication with barcode reader
USB Interface
Interface
- A Type USB: External keyboard or USB type barcode reader can
be used
- B Type USB: Connect with PC to upgrade software, manage
result files, etc.
Indoor use only: operate at ambient temperature
Temperature 5℃ to 40℃, maximum relative humidity 80%
decreasing linearly to 50% for temperature up to 30℃ relative
Storage and
at 40℃.
Handling
Shipping & storage temperature: 5℃ to 40℃
conditions
Altitude up to 2000 m
Pollution degree 2
OVERVOLTAGE CATEGORY II

CDP-EA720-(R) 4 2022-04-7
1.3 Name of components
LCD Display
Keyboard
Printer Paper
Paper Lever

Front Cover with Strip Plate


Push Bar
LED indicator of strip loading
Waste Box
[Fig. 1.1] Top View

CDP-EA720-(R) 5 2022-04-7
[Fig. 1.2] Rear View

CDP-EA720-(R) 6 2022-04-7
2. System installation

2.1 Components
After unpacking ANYSCAN 720, check that all the items in the list below are available in the
package
Item Quantity
1. ANYSCAN 720 1
2. Thermal print paper 1 Roll (50mm x 51mm)
3. Power Cable & Adapter 1
4. RS-232C Cable 1
5. User Manual CD 1
6. Push Bar 1
7. Calibration strips 1 Bottle (50T)
8. USB Cable (A/B type) 1
[Table 2.1] Components of an ANYSCAN 720 set

[Fig. 2.1] RS-232C Cable [Fig. 2.2] USB Cable (A/B type)

2.2 Connection to PC
• When using RS-232C Cable, connect one end of the RS-232C cable to the PC and the other end to
COM1 of the main body.
• When using USB, after connecting the B type connector of the USB cable to the USB of the device,
connect the A type connector of the USB cable to the PC.
• There are two types of barcodes available.
1. To use a serial (RS-232C) type barcode reader, connect it to COM2.
(Baud Rate: 9600 bps barcode reader must be used)
2. To use a USB type barcode reader, connect it to the A type USB port located in the back of the
device.

CDP-EA720-(R) 7 2022-04-7
[Fig. 2.2] Connection diagram between Anyscan 720 and host computer

2.3 Brief operating instruction

1) Connect the AC adapter to the port in the back of the device and turn the power on.

2) Perform White Calibration (Refer to 4.2 System Setting)


3) Set up system configuration (Refer to 5.2 System Configuration)
Set up the required environment such as language, display unit, enable/disable printer etc.
4) Perform measurement (Refer to 4.3 Measurement)
After dipping the strip in the urine sample and removing excessive urine, lay the strips one
by one. Test result comes out in 100 seconds after applying a strip.
5) Review test results stored in database (maximum 100,000 tests).

3. Components and their functions


3.1 Keyboard
The keyboard consists of 16 keypads and is used to input data and set up each function.

CDP-EA720-(R) 8 2022-04-7
[Fig. 3.1] Keyboard
1) ENT Key
ENT Key is used to initiate testing or select a mode at each step.
2) ESC Key
ESC Key is used to terminate testing or leave from each step. In Standby mode, press
ESC key twice, then System Setup Mode will be displayed.

3) ◀▶▲▼ Keys (Direction keys)

Direction keys are used to move to the left, right, up and down, respectively, in each
menu or to increase or decrease numbers instead of using Numeric Keys.
4) 0~9 Keys (Numeric keys)
Numeric Keys are used to select a mode at each step or input data such as patient’s ID,
lot number, etc.

3.2 LCD
LCD (Liquid Crystal Display) displays a current operating status of the system and process
steps.

3.3 Printer

• ANYSCAN 720 test results can be displayed (viewed in real time) as well as printed out.

• Optionally, you can enable or disable printer mode in the System Configuration menu.

• When a red line appears on the edge of the paper, the paper needs to be replaced.

3.4 Waste box


• After testing is completed, the strips automatically fall into the waste box.

CDP-EA720-(R) 9 2022-04-7
• When the waste box is full, the LCD will display “Tray Open” and beep. Empty the waste box
after removing it by pulling out the knob as shown on [Fig 3.3].
• After emptying the waste box, push it back into place so that “Tray Open” warning disappears
from the display.

[Fig. 3.2] Displayed tray warning [Fig. 3.3] Strip waste box

3.5 Strip loading sensor


Place the test strip onto the strip loading plate and push it in as far as it will go. The push bar
will automatically move the test strip and the strip loading sensor will change from green to red.

[Fig. 3.4] Strip Loading Sensor

CDP-EA720-(R) 10 2022-04-7
4. How to operate ANYSCAN 720
Brief user manual
• Turn on the power of the device. After the device checks the system, the initial screen will
appear on the display.
• If necessary, set up the current time and date (Refer to 5.1 System Clock Setting).
• Configure the system (refer to 5.2 System Configuration).
• If calibration is required, select it from the configuration menu and perform calibration.
• Choose the desired user interface language, strip type, analysis units, and printer mode.
• Enter the patient/sample ID (refer to 5.3 Registration).
• Perform the sample measurement (refer to 4.3 Measurement).
• Review the sample measurement results stored in the database (refer to 5.4 Database)

4.1 System initiation


Turn on the power of ANYSCAN 720. The system will enter Standby Mode as shown below
[Fig. 4.1].

[Fig. 4.1] Standby Mode

4.2 System calibration


• In Stanby Mode [Fig. 4.1] press ESC twice to enter System Setup Mode. Select System

Configuration by pressing ② key. Then move to White Calibration item using up/down

(▲/▼) keys.
• Selecting “Yes” and press ESC key to activate calibration [Fig. 4.2]. Press ESC to exit to
the Standby Mode.
• Calibration will start immediately after a white calibration strip is loaded in Standby

CDP-EA720-(R) 11 2022-04-7
Mode. R, G, B values of 330±10 mean normal interpretation of the data.
* For accurate results, white calibration should be performed each time before testing.
* Use only calibration strips manufactured by CHUNGDO PHARM. CO., LTD.

[Fig. 4.2] White Calibration Mode

4.3 Measurement
• Sample measurement must start from the initial screen.
• Immerse the strip in a fresh urine sample so that all reagent pads are covered.
• Remove the excessive urine by touching the strip edge on a tissue or clean, soft cloth.
Carefully place the test strip on the loading plate and push it all the way up. The strip load
sensor will detect the test strip and the LED will change from green to red. Do not place the
next strip on the tray while the LED is red. After the push bar moves the test strip into the device,
the LED color will change to green. Then you can load the next test strip.
• When a strip is detected, the main screen counts down for about 5 seconds [Fig. 4.3], and then
the push bar is activated to push the strip into the device. Removing the strip within 5 seconds
will cancel the test.
• The LCD displays the number of strips that have entered the device for testing and the number
of strips that have been tested [Fig. 4.4].

CDP-EA720-(R) 12 2022-04-7
[Fig. 4.3] Countdown before strip loading

Number of loaded strips


Number of tested strips
Progress of strip testing

[Fig. 4.4] Screen after loading the strips

• To analyze more than one sample, repeat the above process.


• If the strips were loaded continuously, the analysis results of the first strip will be displayed on
the LCD after about 2 minutes.
• After the analysis results of the first strip are displayed on the screen and printed out, the
analysis results of subsequent strips will be displayed every 5 seconds [Fig. 4.5].
• Press ESC key twice to return to Standby Mode.

ⓐ Time of measurement
ⓑ Strip type
ⓒ Tested sample’s number saved in the database

[Fig. 4.5] Display of test results

CDP-EA720-(R) 13 2022-04-7
[Fig. 4.6] Printed result sheet
Note:
1) Use only strips manufactured by CHUNGDO PHARM. CO., LTD.
2) Do not touch the reagent zones (pads) of the strip.
3) Do not use bent strips.
4) Use fresh urine samples.
5) Lay the strips on the loading plate one at a time.
6) As with all laboratory tests, the definitive diagnostic or therapeutic decisions should not be
based on any single test result or method.

4.4 Loading the printing paper


• Insert the printer arm into the holes in the paper roll. Slide the edge of the paper into the slot
above the guide roller.
• Turning the lever so that the rubber roller is facing you, align the central projections of the
lever with the holes in the device as shown in [Fig. 4.7].
• Pull the printer lever towards you and snap it into the printer head.
• After 2 seconds, the device will recognize the paper and automatically adjust the paper
position.

CDP-EA720-(R) 14 2022-04-7
[Fig. 4.7] Loading the printing paper

5. System setup menu


• To enter the system setup, press ESC key twice in the standby mode [Fig. 5.1].

[Fig. 5.1] System setup mode

CDP-EA720-(R) 15 2022-04-7
5.1 System clock setting

1) In system setup mode [Fig. 5.1] press ① to enter the system clock setting mode [Fig. 5.2].

2) Move between menu items by pressing the ▲▼ (up, down) keys. Adjust the time using ◀▶

(left/right) keys or number keys.


3) After setting the time and date, press ENT. The day of the week (Day) will be set
automatically after the date is set.
4) To exit the system clock setting mode, press ESC and the device will return to system setup
mode.

[Fig. 5.2] System clock setting mode

5.2 System сonfiguration

In the system setup mode, press ② to enter the system configuration mode [Fig. 5.3]. Variables

are set using the ▲▼◀▶ keys.

CDP-EA720-(R) 16 2022-04-7
[Fig. 5.3] System configuration mode
• Audio Enable:
Select YES to enable beep sound while operating the device.
• Set Bps (PC)
Select speed rate for transferring data between the device and PC. The default value is 115.2.
• Display Unit
Select the units of measurement, mg or mol.
• Strip Type
Select the strip type, CD-02, 04, 4SG, 05, 06, 06SG, 07, 08, 8SG, 09, 10, 11.
• Print Enable
Enable or disable printing by selecting YES or NO. If YES is selected, the test results will
be displayed on the LCD and printed out on paper. If NO is selected, test results will only
be displayed on the LCD.
• Translate
Translates to the specified language when connected to a computer.
• Audio Volume
Adjusts the volume of the beep.
• Printing Intensity
Adjusts the print density.
• White Calibration
Select white calibration mode.

To return to the menu screen after changing the settings, press ESC. A window will appear asking if
you want to save the changes, if any. To accept or cancel changes, select YES or NO and press ENT
to return to the menu screen.

CDP-EA720-(R) 17 2022-04-7
[Fig. 5.4] System configuration confirmation window
5.3 Registration

In System setup mode [Fig.5.1], press ③ to enter registration mode [Fig.5.3]. Use the ◀▶

keys to select menu values.

[Fig. 5.5] Registration Mode


ⓐ: Operator’s ID
ⓑ: Lot number of the strip
ⓒ: Patient/sample’s ID; enter using the numeric keys or a barcode reader (maximum 13 digits).
ⓓ: Color and turbidity of urine
(The Self-Stik test-strips manufactured by Chungdo Pharm contain a compensation pad for determining the color and turbidity
of urine which allows the analyzer adjust the results accordingly; therefore, this function is not needed)
ⓔ: Block of data that will be stored in database after measurement
ⓕ: Serial number of the data that will be stored in database after measurement
ⓖ: A serial number of the sample you are going to test

• The operator ID and test strip lot fields can be left blank if not needed.
• If patient/sample ID numbers are consecutive, enter only the first number, the others will be
entered automatically. In this case, each subsequent number will increase by one. The
numbering will be retained even if the system is restarted after a power failure.
• If no patient ID was entered, this field will appear blank when printed.
• If the ID number contains letters, each new number must be entered manually (automatic
numbering is only possible when the ID consists entirely of numbers).

• After entering the registration information manually or using a barcode scanner, press the ▶

(right) or ENT, then the entered information will be saved and the next patient/sample
information can be entered.

CDP-EA720-(R) 18 2022-04-7
• If you need to edit the entered information, use the left/right keys to move to the desired data
number, and then use the up/down keys to enter the information again.
• Information about the color or turbidity of urine can be entered optionally by manually selecting
from the following color options (yellow, dark yellow, straw, amber, red, orange, green, and
other) and turbidity options (clear, slightly cloudy, cloudy, turbid, and others) from the menu.
The selection is made by pressing the left/right keys.

5.4 Database (how to use the saved database)

• In system setup mode [Fig.5.1], press ④ to enter database mode [Fig. 5.6].

• Search among the data stored in the instrument's memory by pressing the ▲▼ keys.

[Fig. 5.6] Database searching mode

- Go End: Go to the latest saved sample record


- Go Mem: Go to the record by inputting its sample number
- Go ID: Go to the registered ID number

• A total of 100,000 test results can be stored in the database. Datasets of 20,000 results are
stored in memory under the file name DB* (DB1-DB5 are generated sequentially). If there
is data in the file, DB* will be displayed in blue. If not, then DB* will be displayed in red.
• All database files displayed in blue are searchable. The desired database file is selected using
the left/right keys, after which it will be displayed in green and data search in this file will
become possible. (In [Fig. 5.6], you can search for data in the range 1 to 20000 stored in a
DB3 file).
• After selecting the search method using the up/down keys, enter the search number and press
ENT to display the search result on the LCD. Press ENT again to print the search results.

CDP-EA720-(R) 19 2022-04-7
• If multiple entries with the same ID number are found via Go ID, their total number will be
displayed on the screen.
• If there is no record with the given sequential or ID number, the message "Not Found
Matched..." will be displayed on the screen.
• To clear the database, type "44444" in the search result display mode and press ENT [Fig.
5.7]. This will delete all existing records from the database without the possibility of
further recovery (please proceed with this step cautiously).

[Fig. 5.7] Data result Mode

5.5 Database block control (transferring and reprinting saved data)

• In system setup mode [Fig.5.1], press ⑤ to enter database block control mode [Fig. 5.8].

[Fig. 5.8] Database block control

• Transferring the data to a computer (Upload to PC)

CDP-EA720-(R) 20 2022-04-7
The saved test results can be transferred to a computer after connecting the computer to the

analyzer with an RS-232C cable. Use the numeric keys or the ◀▶ and ENT to select the

start and end numbers of the data block.


• Printing of the data
There are two types of printing available.
“Print normal” allows you to print all the data in the selected block.
“Print abnormal” allows you to print only positive results in the selected block.

Use the numeric keys or the ◀▶ and ENT to select the start and end data block numbers.

The data will be printed in blocks of 50 tests.


The choice of data transmission mode or printout mode is carried out by the ▲▼ keys. In
case of mode change (data transmission ↔ printout), the start and end data block numbers
will be automatically changed to 0001.
• If the database is empty, the error message "No Data in Data Base" will appear in the
center of the display and the device will return to the previous mode. When data
processing is complete, the message “OK Finish!” will appear.

6. Maintenance of ANYSCAN 720


6.1 Daily check-up
1) Keep the device free of dust.
2) Store and operate the device indoors with no sudden changes in temperature.
3) Remove the strip loading plate and rinse it under running water. Reinstall after complete
drying. Also check the waste box, empty if necessary.

6.2 Storage precautions


1) Avoid dust accumulation on the device.
2) Do not leave the device unattended during severe temperature fluctuations.
3) Avoid inclined or vibrating storage surfaces.
4) Avoid any shock to the device when moving it.
5) Avoid the places with high temperature, high humidity, direct sunlight and air containing dust,
salt and sulfur.

6.3 Errors
1) Checking the waste box

CDP-EA720-(R) 21 2022-04-7
If the waste box is installed properly, the display will show ‘TRAY’. Install the waste box
properly and restart the device.
2) Checking the condition of the strip
A bent test strip may become lodged inside the strip loading plate. Remove the front panel
and remove the stuck strips.
3) Checking the front panel
Errors can occur due to an open or missing front panel. Test only with the front panel in
place. After installing the panel, wipe the sensor surface with a damp, clean cloth and restart
the device.
4) Checking the strip loading plate
Accumulation of urine sediment or other foreign matter on the strip loading tray may cause
the strip loading sensor to malfunction due to light reflection. In this case, the “Push bar”
error will be displayed on the screen.

6.4 Error messages and troubleshooting

Cause: The sensor does not recognize the test strip, or the strip is incorrectly placed
“Misplaced
on the plate.
strip”
Action: Repeat the measurement whit a new strip.
Cause: When the power is turned on, internal calibration is performed automatically.
“ERR Load If the strip loading plate is not inserted correctly or does not work properly,
Over Cnt” this error message is displayed.
Action: Restart the device. If this error occurs repeatedly, contact technical support.

Cause/symptom No.&
Malfunction Possible cause or related symptoms
solution
A-1 The device does 1. The power cable is not connected properly. 1. Power off the device. Make
not turn on even sure the power cable is securely
though power is on. connected to the device and to an
AC outlet, and then turn on the
power.(AC adapter: 100-
240V/12VDC 3A)

2. The electronic system of the device is damaged. 2. Contact technical support.

CDP-EA720-(R) 22 2022-04-7
B-1 The device 1. During testing one or more of the following 1,2: Perform the testing again,
always shows a conditions have been met: handling the test strips
false positive test a) The order (sequence) of the test parameters on the carefully according to their
result. printed results is different from the label on the instructions for use and the
strip vial. (The tested strip model is not suitable for operating manual for the
B-2 Visual this analyzer). analyzer.
comparison with b) The color(s) of the reagent pad on the strip after
the color chart testing is very different from the color chart label
gives a negative on the strip vial.
result. c) Testing was performed with a dry strip (not
immersed in urine) or a strip from another
B-3 Other manufacturer that is not suitable for this analyzer
reference tests model.
show a negative d) Testing was performed without placing a strip on
result. the strip plate.
e) The strip was placed upside down or was otherwise
placed incorrectly.
f) The test strip was not completely immersed in urine
covering all reagent pads.
g) The test strip was not placed on the tray
immediately after dipping into urine and removing
excess urine.
f) A bent strip was used.

2. Wrong strip type was selected.


(Select strip type: System setup /2. System
configuration/Strip type)

3. Urine residue has accumulated on the calibration 3.4: Perform calibration again
strip. with a new calibration strip.

4. RGB values during calibration are out of normal


range (333±10) (See page 11 of the given manual).

CDP-EA720-(R) 23 2022-04-7
Cause/symptom No.&
Malfunction Possible cause or related symptoms
solution
C-1 Test results are not printed. 1. Printer mode is disabled. (Enable printer 1. Enable printer mode.
mode: System setup /2. System
configuration/Enable printer).
2. Test results are not displayed on the LCD 2. The printer module needs
monitor. replacement; contact technical
support.

C-2 Paper does not exit the The sound of the printer motor can be heard. Remove the paper from the
printer properly. printer and reinstall it. If the
problem persists after this,
C-3 Small characters are contact technical support.
printed on paper and there is an
unusual sound.

C-4 Printed characters are The printer paper is not installed correctly. Install the printer paper
unreadable or not clear. (Results should be printed on the shiny side of properly. If the problem
the paper). persists after this, contact
C-5 Characters are printed technical support.
partially.

C-6 Non-standard characters The test results displayed on the LCD monitor Restart the device. If the
are printed on the paper. also contain non-standard characters. problem persists after this,
contact technical support.

D-1 An error message The strip is incorrectly positioned on the strip Perform testing again with the
"Misplaced strip" is displayed. plate. correct strip position.

CDP-EA720-(R) 24 2022-04-7
Cause/symptom
Malfunction Possible cause or related symptoms
No.& solution

E-1 The strip plate 1. An unsuitable power supply has been used (12V, 1. Use the correct power
does not move. 3A). supply.
E-2 The strip plate is
stuck in pulled out 2. The strip plate is damaged. 2. Replace the strip
position. plate.
E-3 The strip plate 3. The strip plate sensor is damaged.
stops unexpectedly 3,4 Contact technical
during operation. 4. The motor is damaged. support.

F-1 Key(s) does not 1. Some keys do not work, while all other functions 1. The keyboard module
work. (strip detection, printing etc.) work properly. needs replacement;
contact technical
2. Other functions (strip detection, printing etc.) do not support.
work properly.
2. Contact technical
support.

G-1 The LCD 1. The error happens immediately after the device is 1. The LCD monitor
monitor shows started, all other functions (strip detection, printing module needs
nothing (black) or etc.) work properly. replacement; contact
displays incorrect technical support.
information. 2.Other functions (strip detection, printing etc.) do not
work properly.
2. Contact technical
support.

CDP-EA720-(R) 25 2022-04-7
7. Disassembly/assembly of ANYSCAN 720 for cleaning
7.1 Front panel removal
1) Lift up the front cover panel [Fig. 7.1].
2) Pull the front panel in the direction of the arrow and remove it [Fig.7.2].

[Fig. 7.1] [Fig. 7.2]

7.2 Push bar and strip loading plate removal


1) Pull out the push bar to remove it (magnetic holder) [Fig. 7.3]
2) Pull the loading plate to the left to remove it [Fig. 7.4]

[Fig. 7.3]

[Fig. 7.4]

CDP-EA720-(R) 26 2022-04-7
7.3 Cleaning the strip transfer belt
• Wash the removed push bar, strip loading plate and front panel with water and dry with a
soft cloth. Clean the inside of the belt with a wet swab [Fig.7.5].
• After drying completely, reinstall the push bar, strip loading plate and front panel (see
below).

[Fig. 7.5] Cleaning area inside the device

7.4. Strip loading plate and push bar installation


1) Holding the strip loading tray, insert it into the holes on the ANYSCAN 720 plate as shown
in [Fig.7.6], and then slide it to the right [Fig.7.7].

CDP-EA720-(R) 27 2022-04-7
[Fig. 7.6] [Fig.7.7]

2) To install the push bar, slide it into place as shown in [Fig. 7.8].

[Fig. 7.8]

7.5. Front panel installation


1) Place the front panel as shown in [Fig. 7.9].
2) Lower the front panel down and press lightly until it is fully seated [Fig. 7.10].

[Fig. 7.9] [Fig. 7.10]

Important! Be careful when assembling. Careless assembly may result in device malfunction.

CDP-EA720-(R) 28 2022-04-7
IMPORTANT
1. To prevent electric shock, make sure the power cord is connected to ground.
2. If you notice smells or noises while using the device, immediately turn off the device, unplug
the power cable, and contact technical support.
3. This device only works with urine test strips manufactured by Chungdo Pharm.
4. Follow the instructions for correct operation, avoiding damage to the device, and obtaining
accurate results.
5. This device can only be used with the supplied transformer. Do not use other transformers. The
use of other transformers may damage the instrument and cause dangerous situations.
6. If the supplied power cord does not meet local requirements, contact your local distributor.
7. Operate and store the device on a level surface.
8. In the event of a malfunction, do not attempt to repair the device yourself. In case of problems
with the device, please contact the technical support of the local distributor or the technical
support of Chungdo Pharm (contact information is provided on the last page of this manual).
9. Do not allow water or other liquids to get inside the device.
10. After dipping the strip into the urine sample, remove excess urine from the strip. Performing
a test with excess urine may result in a buildup of urine residue inside the device, which may
result in a malfunction.
11. Do not reuse test strips.
12. Symbol information:

Symbol Referent Symbol Referent


Caution, consult accompanying Collection for electrical and electronic
documents equipment

In vitro diagnostic medical device Hazardous voltage

Consult instructions for use Manufacturer


CE mark

13. If the device is used in a manner not specified by the manufacturer, the warranty provided by
the manufacturer may be void.

14. Warranty for the device is 1 year.

CDP-EA720-(R) 29 2022-04-7
♣ Contact information:

202, HI-TECH VENTURE TOWN, 32, Soyanggang-ro


Chuncheon, Gangwon-do 200-160, Korea
T : +82-10-2840 4002 F : +82-33-258 6115
E-mail : ichungdo@gmail.com
Website : www.ichungdo.com

CDP-EA720-(R) 30 2022-04-7