Imaging Services Enclosure

Document ID: 68524610


This number is needed for production line only:
A6232870 DR 100e
Type 6012/003/004

NX workstation:
Clean Installation Procedure for
All in One PC CyberMed NB19

► Purpose of this document

This enclosure describes the “Clean Installation procedure for the NX workstation
(NX 3.0.21.00 / NX 4.0.21.00 SU1) running on All in One PC CyberMed NB19 of
DR 100e Digital systems”.

The confidentiality note in the document footer classifies this document as confidential.
Dispose of this document when the task is finished.

► Referenced documents
Referenced documents are listed on page 4.

CONFIDENTIALITY NOTE:
Use, dissemination, distribution or reproduction of this document by unauthorized personnel is not permitted and may be unlawful.
DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa Medimg Library. Any printed copy of this document is uncontrolled.

Edition 1, Revision 0
05-2019 Printed in Germany
Enclosure_e_template_v12
Agfa Company Confidential Copyright © 2019 Agfa NV
Document ID: 68524610 Enclosure

► Manufacturer
Agfa NV

Published by
Agfa-Gevaert HealthCare GmbH
Tegernseer Landstraße 161
D - 81539 München
Germany
Copyright  2019 Agfa NV
All rights reserved.
Technical modifications reserved.
Agfa and the Agfa rhombus are trademarks of Agfa-Gevaert N.V., Belgium, or its
affiliates.
All other trademarks mentioned in this document are held by Agfa NV or the respective
owners and are used in an editorial fashion with no intention of infringement.
Nothing contained in this legal notice nor in any text in this document shall be
construed as granting by implication, estoppel or otherwise, any license or right to use
any of the trademarks, service marks, trade names or logos appearing in this document
without the express prior written consent of their respective owner.

WARNING:
Improper operation or service activities may cause damage or injuries.
(1) Read the "Generic Safety Directions" prior to attempting any operation, repair or
maintenance task on the equipment.
Refer to Document ID 11849633.
(2) Strictly observe all safety directions within the "Generic Safety Directions" and on
the product.

IMPORTANT:
The installation and service of the product(s) described herein is to be performed
by qualified personnel who are employed by Agfa NV or one of its affiliates
or who are otherwise authorized by Agfa NV or one of its affiliates to provide such
services.

DOCUMENT CONTROL NOTE:


The controlled version of this document is available from the Agfa Medimg Library. Any printed copy of this document is uncontrolled.

Edition 1, Revision 0 DR 100e Page 2 of 30


05-2019 Type 6012/003/004 Agfa Company Confidential
Document ID: 68524610 Enclosure

List of contents
1 CLEAN INSTALLATION SEQUENCE ......................................................................................6
2 WRITE DOWN THE WINDOWS PRODUCT KEY ...................................................................6
3 LOAD THE AGFA CUSTOMIZED IMAGE OF THE OPERATING SYSTEM ...........................7
3.1 Adapt BIOS settings .................................................................................................................. 7

3.2 Boot from the USB Flash Drive ................................................................................................. 8

3.3 Restore the disc image ............................................................................................................. 9

4 FIRST BOOT AND LOGIN TO NX .........................................................................................10


5 CONNECT THE AIO PC TO THE LOCAL NETWORK ..........................................................12
5.1 Connect the AIO PC to the local network via LAN cable ........................................................ 12

5.2 Connect the AIO PC wireless to the local network ................................................................. 12

6 SET TIMEZONE .....................................................................................................................13


7 ACTIVATE WINDOWS ...........................................................................................................16
8 INSTALL LATEST HOTFIXES................................................................................................16
9 LOAD LICENSE......................................................................................................................17
9.1 Load existing license (hardware was not exchanged) ............................................................ 17

9.2 Load new license from ELMS (Hardware has been exchanged) ........................................... 18

10 NETWORK CONNECTION OVERVIEW ................................................................................19


11 CONFIGURE ACCESS POINT ..............................................................................................20
12 INSTALL DETECTORS ..........................................................................................................20
13 CONFIGURE NX ....................................................................................................................21
14 CHECK THAT WINDOWS FIREWALL IS DISABLED ...........................................................22
15 CREATE BACKUP OF NX CONFIGURATION DATA ............................................................23
16 TROUBLESHOOTING............................................................................................................23
16.1 Configure NX........................................................................................................................... 23

16.2 Configure COM Port for Generator Connection...................................................................... 25

16.3 (Re)-Install the Hot Virtual Keyboard ...................................................................................... 26

17 ANNEX: EXTENDED PROCEDURE TO ADAPT BIOS SETTINGS ......................................27

DOCUMENT CONTROL NOTE:


The controlled version of this document is available from the Agfa Medimg Library. Any printed copy of this document is uncontrolled.

Edition 1, Revision 0 DR 100e Page 3 of 30


05-2019 Type 6012/003/004 Agfa Company Confidential
Document ID: 68524610 Enclosure

Subject This clean installation instruction is for a DR 100e Digital system in case the All in One
(AIO) PC has to be set up from scratch, e.g. when the operating system is corrupted or
after replacement of the hardware of the All in One (AIO) PC.

It applies to following components:


 AIO PC CyberMed NB19 with operating system Win 10 IOT
 NX software version NX 3.0.21.00 / NX 4.0.21.00 SU1

REQUIRED TIME:
Approximately 3 hours
 30 minutes: Loading operating system and NX software
 Up to 1.5 hours: Loading latest hotfixes
 30 minutes: Configuration
 30 minutes: Calibration

REQUIRED TOOLS :
 USB keyboard and USB mouse

REFERENCED DOCUMENTS:
Make sure you have the referenced documents downloaded from Medimg Library
before going on site.
 NX 3.0.21.00 / NX 4.0.21.00 Service Manual, Document ID 65880936
 DR wireless networking Service Manual, Document ID 58221971
 DR 10s / DR 14s Service Manual, Document ID 53659727
 DR 10e / 14e / DR 17e Service Manual, Document ID 58669036
 FUSE Service Manual, Document ID 23977509

DOCUMENT CONTROL NOTE:


The controlled version of this document is available from the Agfa Medimg Library. Any printed copy of this document is uncontrolled.

Edition 1, Revision 0 DR 100e Page 4 of 30


05-2019 Type 6012/003/004 Agfa Company Confidential
Document ID: 68524610 Enclosure

REQUIRED SOFTWARE:
USB flash drive:
DR100E-CYBERMED-NB19-NX21SU1-V1.1.ISO for clean installation DR 100e
CyberMed 19NB Win10
Order number: CM+6232860

The delivery contains the USB flash drive - prepared as bootable device - with the Agfa
customized image and the installation instructions as pdf-file and print-out.

NOTE:
The DR100e Staging Image contains all additional required installers, which were
released at the time of the corresponding NX software release in directory:
Service Software\NX Application NX X.0.XXXX\AdditionalInstallers.
So, if needed, more recent versions of the components must be downloaded from the
Agfa Medimg Library and installed.
Only those installers are started by the activation utility, that have been selected in
ELMS and with which the alf file has been constructed.

REQUIRED BACKUPS:
The following file from the original installation must be available:
Backup of the NX configuration
In case the backup is not available, the complete configuration (X-ray device, detector,
exam tree) has to be redone as described in the NX Service Manual.

DOCUMENT CONTROL NOTE:


The controlled version of this document is available from the Agfa Medimg Library. Any printed copy of this document is uncontrolled.

Edition 1, Revision 0 DR 100e Page 5 of 30


05-2019 Type 6012/003/004 Agfa Company Confidential
Document ID: 68524610 Enclosure

1 Clean installation sequence

Step Chapter
Write down the Windows product key 2
Load the Agfa customized image 3
First boot and login to NX. 4
Connect the AIO PC to local network 5
Set time zone 6
Activate Windows 7
Install latest Hotfixes 8
Load license 9
Network connection 10
Setup the Access Point 11
Install detectors 12
Configure and activate NX 13

Check that Windows Firewall is disabled 14

Create backup of NX configuration data. 15

2 Write down the Windows product key

(1) Write down the Windows product


key that can be found on the
Microsoft sticker on the rear side
of the AIO PC.
You will need this key when
activating Windows later.
Figure 1

DOCUMENT CONTROL NOTE:


The controlled version of this document is available from the Agfa Medimg Library. Any printed copy of this document is uncontrolled.

Edition 1, Revision 0 DR 100e Page 6 of 30


05-2019 Type 6012/003/004 Agfa Company Confidential
Document ID: 68524610 Enclosure

3 Load the Agfa customized image of the operating system

Do not connect the external USB flash drive (drive containing the image) yet.
Power on the AIO PC by holding the ON/OFF button for about 2 sec.

IMPORTANT:
If the system shows the “Agfa” logo at the start of booting, the BIOS is already in
accordance with the required default settings. In that case only date and time have to
be adjusted.
If the system boots with the “Cybernet” screen, additional settings in BIOS need to be
adapted. Refer to the annex, chapter 17.

3.1 Adapt BIOS settings

(1) Make sure that the NX dongle is connected to the AIO PC.
(2) If possible, put a small table close to the DR 100e (for mouse and keyboard).
(3) Connect the USB keyboard and USB mouse to the external USB connectors
of the CyberMed AIO PC.

(4) Start the AIO PC: At startup


enter BIOS (press DEL key
repeatedly).
(5) Adjust system date and time in
the BIOS before restoring the
image.

Figure 2

DOCUMENT CONTROL NOTE:


The controlled version of this document is available from the Agfa Medimg Library. Any printed copy of this document is uncontrolled.

Edition 1, Revision 0 DR 100e Page 7 of 30


05-2019 Type 6012/003/004 Agfa Company Confidential
Document ID: 68524610 Enclosure

(6) Go to Save & Exit tab and


select Save Changes and
Reboot.
Alternatively press <F10>.
(7) Confirm with Yes.

Figure 3

3.2 Boot from the USB Flash Drive

(1) Switch off the AIO PC.


(2) Insert the Restore USB Flash Drive for clean installation DR100e CyberNet
19NB Win10 in an USB port of the AIO PC.
(3) Start the AIO PC.
(4) During system boot press
repeatedly <F11> to enter the
boot menu.
(5) As boot device select the
UEFI USB Restore Stick
(Partition 2).
(6) Press Enter.
The monitor might display distorted
images for about 15 seconds. Figure 4

DOCUMENT CONTROL NOTE:


The controlled version of this document is available from the Agfa Medimg Library. Any printed copy of this document is uncontrolled.

Edition 1, Revision 0 DR 100e Page 8 of 30


05-2019 Type 6012/003/004 Agfa Company Confidential
Document ID: 68524610 Enclosure

3.3 Restore the disc image

(1) Click the Restore button at the


TOP of the Window.

Figure 5

(2) Click Continue.

Figure 6

The hard disk will now be partitioned


and formatted. This takes around one
minute.

Figure 7

After that the partitions will be


populated with the correct files. This
takes around 7 minutes.

Figure 8

(3) Click Reboot.

Figure 9

DOCUMENT CONTROL NOTE:


The controlled version of this document is available from the Agfa Medimg Library. Any printed copy of this document is uncontrolled.

Edition 1, Revision 0 DR 100e Page 9 of 30


05-2019 Type 6012/003/004 Agfa Company Confidential
Document ID: 68524610 Enclosure

4 First boot and login to NX

When the AIO PC boots the first time,


it will execute some automatic scripts.
Do not stop this. This will result in a
second and third boot.
After less than 2 minutes the script
finishes.

Figure 10

Two automatic reboots will follow.

Figure 11

(1) Login with the crservice account.

NOTE:
A Message to restart the system after the first login may appear:
Select Restart Later.

The NX Splash screen appears.

Figure 12

DOCUMENT CONTROL NOTE:


The controlled version of this document is available from the Agfa Medimg Library. Any printed copy of this document is uncontrolled.

Edition 1, Revision 0 DR 100e Page 10 of 30


05-2019 Type 6012/003/004 Agfa Company Confidential
Document ID: 68524610 Enclosure

(2) Then a NX Startup message


about incorrect screen
resolution appears. Click OK.

Figure 13

(3) Stop NX:


Go to NX Control Center 
Stop NX

Figure 14

DOCUMENT CONTROL NOTE:


The controlled version of this document is available from the Agfa Medimg Library. Any printed copy of this document is uncontrolled.

Edition 1, Revision 0 DR 100e Page 11 of 30


05-2019 Type 6012/003/004 Agfa Company Confidential
Document ID: 68524610 Enclosure

5 Connect the AIO PC to the local network

5.1 Connect the AIO PC to the local network via LAN cable

(1) Connect the AIO PC via LAN cable to the local network.

When the AIO PC is connected for


the first time to a public network (a
network with Internet Access), the
following dialog pops up:
Do you want to allow your PC to be
discoverable by other PCs and
devices on this network?
(2) Click Yes.
If you would click No, the
Windows Firewall would be
turned on, which is not allowed.

Figure 15

5.2 Connect the AIO PC wireless to the local network

(1) Connect the AIO PC wireless to the local network as described in:
Service Manual - DR wireless networking - Document ID 58221971, section
"Connecting a mobile PC to the hospital network".

DOCUMENT CONTROL NOTE:


The controlled version of this document is available from the Agfa Medimg Library. Any printed copy of this document is uncontrolled.

Edition 1, Revision 0 DR 100e Page 12 of 30


05-2019 Type 6012/003/004 Agfa Company Confidential
Document ID: 68524610 Enclosure

6 Set timezone

(1) Go to Start  Settings.

(2) Select button Time & language

Figure 16

(3) Change the Time zone.

Figure 17

(4) Select Region & language.

Figure 18

DOCUMENT CONTROL NOTE:


The controlled version of this document is available from the Agfa Medimg Library. Any printed copy of this document is uncontrolled.

Edition 1, Revision 0 DR 100e Page 13 of 30


05-2019 Type 6012/003/004 Agfa Company Confidential
Document ID: 68524610 Enclosure

(5) Select Additional date, time &


regional settings.

Figure 19

(6) Select Region.

Figure 20

(7) Check the country in the Format tab.


It must be set to English.

The language setting here is used as


display language for user "crservice".
By default, this has to be in English.

(8) Click OK.

Figure 21

(9) Select Language.

Figure 22

DOCUMENT CONTROL NOTE:


The controlled version of this document is available from the Agfa Medimg Library. Any printed copy of this document is uncontrolled.

Edition 1, Revision 0 DR 100e Page 14 of 30


05-2019 Type 6012/003/004 Agfa Company Confidential
Document ID: 68524610 Enclosure

(10) Select a language and move it to the


top of the list.
If the required language is not listed,
click Add a language, select the
proper language and click OK.
The selection of the language will also
set the keyboard type.

Figure 23

(11) Close the window.

(12) Reboot the AIO PC, if the system asks to.

DOCUMENT CONTROL NOTE:


The controlled version of this document is available from the Agfa Medimg Library. Any printed copy of this document is uncontrolled.

Edition 1, Revision 0 DR 100e Page 15 of 30


05-2019 Type 6012/003/004 Agfa Company Confidential
Document ID: 68524610 Enclosure

7 Activate Windows
PREREQUISITE: An internet connection is needed.

(1) Open the Control panel.

(2) Change view to Small Icons.

(3) Open System.

(4) At the bottom of the dialog, change the


Windows Product ID with the actual
Product ID (= Windows product key),
which you have noted down in
chapter 2.
Figure 24

8 Install latest Hotfixes

Use the Fast Update Service FUSE to install the required hotfixes.
For details see NX Service Manual and FUSE Service Manual.

DOCUMENT CONTROL NOTE:


The controlled version of this document is available from the Agfa Medimg Library. Any printed copy of this document is uncontrolled.

Edition 1, Revision 0 DR 100e Page 16 of 30


05-2019 Type 6012/003/004 Agfa Company Confidential
Document ID: 68524610 Enclosure

9 Load License

NOTE:
To re-load the license two scenarios are possible:
 The existing license file can be installed again. This is the case when the
hardware of the AIO PC has not been changed. See chapter 9.1.
 A new license file from ELMS needs to be installed. This is the case when the
hardware of the AIO PC has been exchanged, because the MAC address of the
new PC has to be included in the license file. See chapter 9.2.

9.1 Load existing license (hardware was not exchanged)

(1) Login to NX as crservice.

(2) Ensure that the license dongle is inserted.

(3) Copy the license key file to: C:\temp

(4) Start up the Agfa license manager:


NX Control Center  Service 
License manager

Figure 25
(5) Click the Add License ... button.

(6) Browse to the C:\AGFA\Healthcare\NX\Configuration\Licensing folder.

DOCUMENT CONTROL NOTE:


The controlled version of this document is available from the Agfa Medimg Library. Any printed copy of this document is uncontrolled.

Edition 1, Revision 0 DR 100e Page 17 of 30


05-2019 Type 6012/003/004 Agfa Company Confidential
Document ID: 68524610 Enclosure

(7) Select the license file.

(8) A message might appear.


Click Yes.

Figure 26

(9) A message will be shown that the licenses have been updated and the
NX password will be changed after reboot!

(10) Close the tool and reboot the AIO PC.


Note that the Service unique password is active now.
If the Service unique password is not known, refer to:
NX 8900 SB 12, Unique password per NX for user crservice as of NX 8900 SU2,
Document ID 51679365.

9.2 Load new license from ELMS (Hardware has been exchanged)

When replacing the AIO PC, a new NX license file (ALF file) is required.
It is important to follow the "ELMS swap PC workflow" to obtain the new ALF file.
For instructions refer to the latest NX Service Manual (as of NX 22.00), chapter 4,
Installation and Configuration Appendices, section "Obtaining a new ALF file when
replacing a PC (ELMS Swap PC workflow)".

DOCUMENT CONTROL NOTE:


The controlled version of this document is available from the Agfa Medimg Library. Any printed copy of this document is uncontrolled.

Edition 1, Revision 0 DR 100e Page 18 of 30


05-2019 Type 6012/003/004 Agfa Company Confidential
Document ID: 68524610 Enclosure

10 Network connection overview


These are the settings that will be applied by production.
(1) Adapt the default network settings if needed.

Detector DR 10s \ DR 14s (Trixell) with default settings:

PC Switch DLink
Access Point
192.168.0.10
192.168.0.25

USB

DR 10s /14s detector

192.168.0.3
IR transceiver

Figure 27

Detector DR 10e/DR 14e/DR 17e (Innolux) with default settings:

Figure 28

DOCUMENT CONTROL NOTE:


The controlled version of this document is available from the Agfa Medimg Library. Any printed copy of this document is uncontrolled.

Edition 1, Revision 0 DR 100e Page 19 of 30


05-2019 Type 6012/003/004 Agfa Company Confidential
Document ID: 68524610 Enclosure

11 Configure Access Point


No action is required, as long as no access point settings have been changed.
If any settings at the access point have to be adapted, setup the access point as
described in the DR wireless networking Service Manual.
Make a backup of the configuration. Refer to Document ID 58221971.

12 Install detectors

Install a DR 10s \ DR 14s detector (Trixell)

NOTE:
Install the DR 10s \ DR 14s detectors according to the DR 10s \ DR 14s Service
Manual. Refer to Document ID 53659727.
As long as the settings of the access point have not been changed, you can skip the
chapter Configuring the Access Point.

Install a DR 14e \ DR 17e detector (Innolux)

NOTE:
Install the DR 14e \ DR 17e detectors according to the DR 14e \ DR 17e Service
Manual. Refer to Document ID 58669036.
As long as the settings of the access point have not been changed, you can skip the
chapter Configuring the Access Point.

DOCUMENT CONTROL NOTE:


The controlled version of this document is available from the Agfa Medimg Library. Any printed copy of this document is uncontrolled.

Edition 1, Revision 0 DR 100e Page 20 of 30


05-2019 Type 6012/003/004 Agfa Company Confidential
Document ID: 68524610 Enclosure

13 Configure NX

(1) Stop NX:


Go to NX Control Center  Stop NX.

(2) Start NX Configurator:


Go to NX Control Center  NX 
Start NX configurator

Figure 29

(3) Click Load a configuration from an


external file.
If you have exported the NX
configuration beforehand, load it at this
step.
If not, start from another export file.
Alternatively, you can start the default
configuration from:
C:\temp\DR100e factory
config_2.0.xml

Figure 30

(4) Ignore the warning icon, if it should


appear.

Figure 31

DOCUMENT CONTROL NOTE:


The controlled version of this document is available from the Agfa Medimg Library. Any printed copy of this document is uncontrolled.

Edition 1, Revision 0 DR 100e Page 21 of 30


05-2019 Type 6012/003/004 Agfa Company Confidential
Document ID: 68524610 Enclosure

(5) Continue configuring NX, including exam tree and detector(s).


Note that following Trigger mode should be set:
 For a DR 10s / DR 14s detector: ExposyncNetwork
 For a DR 14e / DR 17e detector: ExposyncNetwork

(6) When ready, activate the NX configuration :


Go to: File  Activate configuration  Continue  Activate

(7) Close the configuration tool (Close without any changes).

14 Check that Windows Firewall is disabled

The Windows Firewall must be turned off for proper operation of NX.
The NX installer turns off Windows Firewall, but during network setup in Windows 10
it is possible that it is turned on automatically.

(1) Click the Windows start button and click


Settings.
(2) In the Windows settings panel, type
Firewall and click Windows Firewall.

Figure 32

(3) Within the Firewall section, choose


Turn Windows Firewall on or off.

Figure 33

DOCUMENT CONTROL NOTE:


The controlled version of this document is available from the Agfa Medimg Library. Any printed copy of this document is uncontrolled.

Edition 1, Revision 0 DR 100e Page 22 of 30


05-2019 Type 6012/003/004 Agfa Company Confidential
Document ID: 68524610 Enclosure

(4) Check that Windows Firewall is turned


off.
If it is turned on: Select Turn off
Windows Firewall twice, and click OK
to confirm.

Figure 34

15 Create backup of NX configuration data


After the clean installation, it is recommended to create a backup of the NX
configuration data.
Details are described in the applicable NX Service Manual, chapter Installation and
Configuration. Refer to Document ID 65880936.

16 Troubleshooting
16.1 Configure NX

The factory file DR100e factory


config_2.0.xml is intended for systems
that do not have Musica 3 licenses. If
the license file contains Musica 3
licenses, the activation of the
configuration after loading the factory
file can fail because of the errors below
(depending on available licenses).

Figure 35

DOCUMENT CONTROL NOTE:


The controlled version of this document is available from the Agfa Medimg Library. Any printed copy of this document is uncontrolled.

Edition 1, Revision 0 DR 100e Page 23 of 30


05-2019 Type 6012/003/004 Agfa Company Confidential
Document ID: 68524610 Enclosure

(1) Click on Click here to correct to bring you to the Workflow Management screen.
(2) Click the Identification title bar to close that pane or alternatively scroll down to
the Image Presentation pane.
(3) Within the Musica 2/3 Presentation Taste Settings select Abdomen.
(4) Set the Version drop down to MUSICA3
and click the Store button.

Figure 36

(5) Repeat the procedure for the Chest and Skeleton packages.

The Version in the overview table


should now be equal to MUSICA3 for
the Abdomen, Chest and Skeleton
packages.

Figure 37

DOCUMENT CONTROL NOTE:


The controlled version of this document is available from the Agfa Medimg Library. Any printed copy of this document is uncontrolled.

Edition 1, Revision 0 DR 100e Page 24 of 30


05-2019 Type 6012/003/004 Agfa Company Confidential
Document ID: 68524610 Enclosure

16.2 Configure COM Port for Generator Connection

The Generator Connection is by default


configured for COM Port 2. The COM
Port number is automatically assigned
by Windows. If you get a Generator
Connection Error (see screenshot)
when selecting an empty thumbnail for
a DR exposure, the Generator
Connection has to be reconfigured to
COM Port 1.
Figure 38

Change the configured COM Port to Port 1:


(1) Stop NX.
(2) Browse to the folder: C:\Agfa\Healthcare\XRDI\Components\THX\Configuration
(3) Open the following file in notepad: THXConfig.xml
(4) In the section
<GeneratorConfig>
<COMPort>2</COMPort>
</GeneratorConfig>
replace “2” by “1”, resulting in
<GeneratorConfig>
<COMPort>1</COMPort>
</GeneratorConfig>
(5) Save the file and close notepad.

DOCUMENT CONTROL NOTE:


The controlled version of this document is available from the Agfa Medimg Library. Any printed copy of this document is uncontrolled.

Edition 1, Revision 0 DR 100e Page 25 of 30


05-2019 Type 6012/003/004 Agfa Company Confidential
Document ID: 68524610 Enclosure

16.3 (Re)-Install the Hot Virtual Keyboard

After restoring the system from the USB


flash drive, there is no more need to run
the NX Activation Utility.
However, if the NX Activation Utility is –
for some reason -executed again, also the
Hot Virtual Keyboard has to be reinstalled.
The installer of the Hot Virtual Keyboard –
VK4NX 1.0.010.exe - is located in the
folder
C:\Agfa\Healthcare\NX\Installers\VK4NX.

Figure 39

(1) Click Install.

(2) Wait for the installation to finish and


click Next >

Figure 40

(3) Click Finish to reboot the system.

Figure 41

DOCUMENT CONTROL NOTE:


The controlled version of this document is available from the Agfa Medimg Library. Any printed copy of this document is uncontrolled.

Edition 1, Revision 0 DR 100e Page 26 of 30


05-2019 Type 6012/003/004 Agfa Company Confidential
Document ID: 68524610 Enclosure

17 Annex: Extended procedure to adapt BIOS settings

IMPORTANT:
Perform this procedure only, if the AGFA-specific BIOS (i.e. “Agfa” is displayed at
start of booting) is not present on the AIO PC. In this case “Cybernet” is displayed at
start of booting.

(1) If possible, put a small table close to the X-ray unit (for mouse and keyboard).
(2) Connect the USB keyboard and USB mouse to the external USB connectors.

(3) At startup: enter BIOS (press


DEL key repeatedly).
(4) Adjust system date and time in
the BIOS before restoring the
image.

Figure 42

(5) Go to Save & Exit tab and


select Load Optimized
Defaults.
Alternatively press <F9>.
(6) Select Yes.

Figure 43

DOCUMENT CONTROL NOTE:


The controlled version of this document is available from the Agfa Medimg Library. Any printed copy of this document is uncontrolled.

Edition 1, Revision 0 DR 100e Page 27 of 30


05-2019 Type 6012/003/004 Agfa Company Confidential
Document ID: 68524610 Enclosure

(7) Go to the Advanced tab.


(8) Disable Intel Virtualization
Technology.
(9) Disable Intel VT-d.

Figure 44

(10) In Advanced tab, enter AMT


Configuration almost at the
end of the list.

Figure 45

(11) Disable Intel AMT.


(12) Press <Esc> to go back to the
previous list.

Figure 46

DOCUMENT CONTROL NOTE:


The controlled version of this document is available from the Agfa Medimg Library. Any printed copy of this document is uncontrolled.

Edition 1, Revision 0 DR 100e Page 28 of 30


05-2019 Type 6012/003/004 Agfa Company Confidential
Document ID: 68524610 Enclosure

(13) Enter Windows 8/8.1/10


Configuration at the end of the
list.

Figure 47

(14) Enable Windows 8/8.1/10 Logo


Requirement.

Figure 48

(15) Enter Secure Boot.


(16) Disable Secure Boot.
(17) Press <Esc> to go back to
Windows 8/8.1/10
Configuration.
(18) Press <Esc> again to go back
to Advanced.

Figure 49

(19) Go to Power tab.


(20) Set Restore after AC Power
Loss to Power On.

Figure 50

DOCUMENT CONTROL NOTE:


The controlled version of this document is available from the Agfa Medimg Library. Any printed copy of this document is uncontrolled.

Edition 1, Revision 0 DR 100e Page 29 of 30


05-2019 Type 6012/003/004 Agfa Company Confidential
Document ID: 68524610 Enclosure

(21) Go to Save & Exit tab and


select Save Changes and
Reboot.
Alternatively press <F10>.
(22) Confirm with Yes.

Figure 51

DOCUMENT CONTROL NOTE:


The controlled version of this document is available from the Agfa Medimg Library. Any printed copy of this document is uncontrolled.

Edition 1, Revision 0 DR 100e Page 30 of 30


05-2019 Type 6012/003/004 Agfa Company Confidential