NX workstation:
Clean Installation Procedure for
All in One PC CyberMed NB19
This enclosure describes the “Clean Installation procedure for the NX workstation
(NX 3.0.21.00 / NX 4.0.21.00 SU1) running on All in One PC CyberMed NB19 of
DR 100e Digital systems”.
The confidentiality note in the document footer classifies this document as confidential.
Dispose of this document when the task is finished.
► Referenced documents
Referenced documents are listed on page 4.
CONFIDENTIALITY NOTE:
Use, dissemination, distribution or reproduction of this document by unauthorized personnel is not permitted and may be unlawful.
DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa Medimg Library. Any printed copy of this document is uncontrolled.
Edition 1, Revision 0
05-2019 Printed in Germany
Enclosure_e_template_v12
Agfa Company Confidential Copyright © 2019 Agfa NV
Document ID: 68524610 Enclosure
► Manufacturer
Agfa NV
Published by
Agfa-Gevaert HealthCare GmbH
Tegernseer Landstraße 161
D - 81539 München
Germany
Copyright 2019 Agfa NV
All rights reserved.
Technical modifications reserved.
Agfa and the Agfa rhombus are trademarks of Agfa-Gevaert N.V., Belgium, or its
affiliates.
All other trademarks mentioned in this document are held by Agfa NV or the respective
owners and are used in an editorial fashion with no intention of infringement.
Nothing contained in this legal notice nor in any text in this document shall be
construed as granting by implication, estoppel or otherwise, any license or right to use
any of the trademarks, service marks, trade names or logos appearing in this document
without the express prior written consent of their respective owner.
WARNING:
Improper operation or service activities may cause damage or injuries.
(1) Read the "Generic Safety Directions" prior to attempting any operation, repair or
maintenance task on the equipment.
Refer to Document ID 11849633.
(2) Strictly observe all safety directions within the "Generic Safety Directions" and on
the product.
IMPORTANT:
The installation and service of the product(s) described herein is to be performed
by qualified personnel who are employed by Agfa NV or one of its affiliates
or who are otherwise authorized by Agfa NV or one of its affiliates to provide such
services.
List of contents
1 CLEAN INSTALLATION SEQUENCE ......................................................................................6
2 WRITE DOWN THE WINDOWS PRODUCT KEY ...................................................................6
3 LOAD THE AGFA CUSTOMIZED IMAGE OF THE OPERATING SYSTEM ...........................7
3.1 Adapt BIOS settings .................................................................................................................. 7
9.2 Load new license from ELMS (Hardware has been exchanged) ........................................... 18
Subject This clean installation instruction is for a DR 100e Digital system in case the All in One
(AIO) PC has to be set up from scratch, e.g. when the operating system is corrupted or
after replacement of the hardware of the All in One (AIO) PC.
REQUIRED TIME:
Approximately 3 hours
30 minutes: Loading operating system and NX software
Up to 1.5 hours: Loading latest hotfixes
30 minutes: Configuration
30 minutes: Calibration
REQUIRED TOOLS :
USB keyboard and USB mouse
REFERENCED DOCUMENTS:
Make sure you have the referenced documents downloaded from Medimg Library
before going on site.
NX 3.0.21.00 / NX 4.0.21.00 Service Manual, Document ID 65880936
DR wireless networking Service Manual, Document ID 58221971
DR 10s / DR 14s Service Manual, Document ID 53659727
DR 10e / 14e / DR 17e Service Manual, Document ID 58669036
FUSE Service Manual, Document ID 23977509
REQUIRED SOFTWARE:
USB flash drive:
DR100E-CYBERMED-NB19-NX21SU1-V1.1.ISO for clean installation DR 100e
CyberMed 19NB Win10
Order number: CM+6232860
The delivery contains the USB flash drive - prepared as bootable device - with the Agfa
customized image and the installation instructions as pdf-file and print-out.
NOTE:
The DR100e Staging Image contains all additional required installers, which were
released at the time of the corresponding NX software release in directory:
Service Software\NX Application NX X.0.XXXX\AdditionalInstallers.
So, if needed, more recent versions of the components must be downloaded from the
Agfa Medimg Library and installed.
Only those installers are started by the activation utility, that have been selected in
ELMS and with which the alf file has been constructed.
REQUIRED BACKUPS:
The following file from the original installation must be available:
Backup of the NX configuration
In case the backup is not available, the complete configuration (X-ray device, detector,
exam tree) has to be redone as described in the NX Service Manual.
Step Chapter
Write down the Windows product key 2
Load the Agfa customized image 3
First boot and login to NX. 4
Connect the AIO PC to local network 5
Set time zone 6
Activate Windows 7
Install latest Hotfixes 8
Load license 9
Network connection 10
Setup the Access Point 11
Install detectors 12
Configure and activate NX 13
Do not connect the external USB flash drive (drive containing the image) yet.
Power on the AIO PC by holding the ON/OFF button for about 2 sec.
IMPORTANT:
If the system shows the “Agfa” logo at the start of booting, the BIOS is already in
accordance with the required default settings. In that case only date and time have to
be adjusted.
If the system boots with the “Cybernet” screen, additional settings in BIOS need to be
adapted. Refer to the annex, chapter 17.
(1) Make sure that the NX dongle is connected to the AIO PC.
(2) If possible, put a small table close to the DR 100e (for mouse and keyboard).
(3) Connect the USB keyboard and USB mouse to the external USB connectors
of the CyberMed AIO PC.
Figure 2
Figure 3
Figure 5
Figure 6
Figure 7
Figure 8
Figure 9
Figure 10
Figure 11
NOTE:
A Message to restart the system after the first login may appear:
Select Restart Later.
Figure 12
Figure 13
Figure 14
5.1 Connect the AIO PC to the local network via LAN cable
(1) Connect the AIO PC via LAN cable to the local network.
Figure 15
(1) Connect the AIO PC wireless to the local network as described in:
Service Manual - DR wireless networking - Document ID 58221971, section
"Connecting a mobile PC to the hospital network".
6 Set timezone
Figure 16
Figure 17
Figure 18
Figure 19
Figure 20
Figure 21
Figure 22
Figure 23
7 Activate Windows
PREREQUISITE: An internet connection is needed.
Use the Fast Update Service FUSE to install the required hotfixes.
For details see NX Service Manual and FUSE Service Manual.
9 Load License
NOTE:
To re-load the license two scenarios are possible:
The existing license file can be installed again. This is the case when the
hardware of the AIO PC has not been changed. See chapter 9.1.
A new license file from ELMS needs to be installed. This is the case when the
hardware of the AIO PC has been exchanged, because the MAC address of the
new PC has to be included in the license file. See chapter 9.2.
Figure 25
(5) Click the Add License ... button.
Figure 26
(9) A message will be shown that the licenses have been updated and the
NX password will be changed after reboot!
9.2 Load new license from ELMS (Hardware has been exchanged)
When replacing the AIO PC, a new NX license file (ALF file) is required.
It is important to follow the "ELMS swap PC workflow" to obtain the new ALF file.
For instructions refer to the latest NX Service Manual (as of NX 22.00), chapter 4,
Installation and Configuration Appendices, section "Obtaining a new ALF file when
replacing a PC (ELMS Swap PC workflow)".
PC Switch DLink
Access Point
192.168.0.10
192.168.0.25
USB
192.168.0.3
IR transceiver
Figure 27
Figure 28
12 Install detectors
NOTE:
Install the DR 10s \ DR 14s detectors according to the DR 10s \ DR 14s Service
Manual. Refer to Document ID 53659727.
As long as the settings of the access point have not been changed, you can skip the
chapter Configuring the Access Point.
NOTE:
Install the DR 14e \ DR 17e detectors according to the DR 14e \ DR 17e Service
Manual. Refer to Document ID 58669036.
As long as the settings of the access point have not been changed, you can skip the
chapter Configuring the Access Point.
13 Configure NX
Figure 29
Figure 30
Figure 31
The Windows Firewall must be turned off for proper operation of NX.
The NX installer turns off Windows Firewall, but during network setup in Windows 10
it is possible that it is turned on automatically.
Figure 32
Figure 33
Figure 34
16 Troubleshooting
16.1 Configure NX
Figure 35
(1) Click on Click here to correct to bring you to the Workflow Management screen.
(2) Click the Identification title bar to close that pane or alternatively scroll down to
the Image Presentation pane.
(3) Within the Musica 2/3 Presentation Taste Settings select Abdomen.
(4) Set the Version drop down to MUSICA3
and click the Store button.
Figure 36
(5) Repeat the procedure for the Chest and Skeleton packages.
Figure 37
Figure 39
Figure 40
Figure 41
IMPORTANT:
Perform this procedure only, if the AGFA-specific BIOS (i.e. “Agfa” is displayed at
start of booting) is not present on the AIO PC. In this case “Cybernet” is displayed at
start of booting.
(1) If possible, put a small table close to the X-ray unit (for mouse and keyboard).
(2) Connect the USB keyboard and USB mouse to the external USB connectors.
Figure 42
Figure 43
Figure 44
Figure 45
Figure 46
Figure 47
Figure 48
Figure 49
Figure 50
Figure 51