CR 25.0
Type 5156/105
► Referenced Documents
Document Title
DD+DIS219.06E Chapter 3.2 - Tools and auxiliary means
NOTE:
Dispose the enclosure document after finishing your task
(observe the confidentiality note below).
CONFIDENTIALITY NOTE:
Use, dissemination, distribution or reproduction of this document by unauthorized personnel is not permitted and may be unlawful.
DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Revision 0
11-2011 Printed in Germany Document Node ID: 34192685
Enclosure_e_template_v08
Agfa Company Confidential Copyright © 2011 Agfa HealthCare N.V.
DD+DIS195.11E Enclosure
► Manufacturer
Agfa HealthCare N.V.
Published by
Agfa-Gevaert HealthCare GmbH
Tegernseer Landstraße 161
D - 81539 München
Germany
Copyright 2011 Agfa HealthCare N.V.
All rights reserved.
Technical modifications reserved.
Agfa and the Agfa rhombus are trademarks of Agfa-Gevaert N.V., Belgium,
or its affiliates.
WARNING:
Improper operation or service activities may cause damage or injuries.
INSTRUCTION:
(1) Read the "Generic Safety Directions" document
(see Agfa HealthCare Library > General Info > Agfa HealthCare > Publications >
Service Manual) prior to attempting any operation, repair or maintenance task on
the equipment.
(2) Strictly observe all safety directions within the "Generic Safety Directions" and on
the product.
IMPORTANT:
The installation and service of the product(s) described herein is to be performed
by qualified personnel who are employed by Agfa HealthCare or one of its affiliates
or who are otherwise authorized by Agfa HealthCare or one of its affiliates to provide
such services.
LIST OF CONTENTS
3.3.3 Installing Software c25_3103B on a new Hard Disk or in Case of corrupted Files ................10
4 INSTALLATION VIA FLOPPY DISKS (CR 35-X TYPE 5158 / 100 UP TO SN < 4000, CR
25.0 TYPE 5156 / 105 UP TO SN < 6500) .............................................................................13
4.1 Creating Installation Floppy Disks...........................................................................................13
4.3.2 Installing Software c25_3103B on a new Hard Disk or in Case of corrupted Files ................19
The digitizer software c25_3103B can be used for the following CR applications:
General Radiography (GenRad),
Mammography (Mammo),
General Radiography 50 µm (Extremities) and
Full Leg – Full Spine (FLFS)
NOTE:
The new software is upgradeable from every previous version of c25_xxxx* software.
*XXXX = Version number of the software
.
2 Prerequisites for Installation
REQUIRED TOOLS:
For installation via floppy disk applicable for:
CR 25.0 Type 5156 / 105 up to SN < 6500,
CR 35-X Type 5158 / 100 up to SN < 4000:
o 1 floppy disk for the backup of site specific data
o 7 floppy disks for the installation files
o Service PC with floppy disk drive
NOTE:
USB Memory Stick will be part of delivery in the following digitizers:
CR 25.0 Type 5156 / 105 as of SN ≥ 6500
CR 35-X Type 5158 / 100 as of SN ≥ 4000
9 pins 25 pins
REQUIRED TIME:
Approximately 30 minutes for the software upgrade via floppy disks
Approximately 60 minutes for the installation via floppy disks on
a new hard disk or when corrupted files detected
Approximately 10 minutes for the installation via USB Memory Stick
(software upgrade/blank hard disk)
NOTE:
Register this software installation in the local service management system
as modification, e.g. in SMS as “MOD”.
NOTE:
Following steps are necessary to install software via USB Memory stick:
(1) Download the Digitizer Software c25_3103B (File size approximately 8 MB)
from Agfa HealthCare Library:
Computed Radiography CR Digitizers CR 35-X
Software CR 35-X - Software - Device Software
c25_3103B
or
Computed Radiography CR Digitizers CR 25.0
Software CR 25.0 - Software - Device Software
c25_3103B
(2) Extract all files from the downloaded zip file into a folder on the Service PC.
Among the extracted files is the zip file c25_3103B.zip.
Figure 2
(4) Copy the folder c25_3103B to the USB Memory Stick. The created structure
shown in Figure 3 must be maintained (“Root” is the USB Memory Stick).
Figure 3
NOTE:
C25_<SerialNumber>.bac is optional, for the case, that a backup
has been made previously.
Before performing the software update: Update the USB Memory Stick with the latest
machine specific data.
(1) Switch on and wait until the digitizer shows the ready screen.
(2) Open the right cover of the digitizer to access the USB port.
IMPORTANT:
The digitizer checks light incidence.
Following error message may appear on the screen of the digitizer:
Digitizer locked
RIGHT SIDE PANEL NOT CLOSED.
Close right side panel.
Code 24791
Figure 4
If this error message appears, lean the cover against the right side of the device.
Then continue with next steps.
(3) Enter the Service menu on the digitizer keypad and select:
<3 Save data> <2 Machine specific data>
(4) Follow the instructions on the digitizer display.
NOTE:
If the installation fails repeat installation, see section 3.3.3.
IMPORTANT:
With software version c25_3103B it is not possible to install the software via Service
Menu. Always install the software via boot up as described in section 3.3.1.
(4) Insert the USB Memory Stick into the USB Port of the digitizer.
(5) Press <> to confirm. The digitizer copies, extracts and installs the software.
Operating system “Portex” will not be installed.
(6) After a few moments the following message appears:
“Please remove the storage medium and press <>.”
Remove the USB Memory Stick and press <> to confirm.
(7) After a few moments the following message appears:
“To initialize the new SW a reset is necessary. Press <> to reset now.”
Close the digitizer. The device will restart automatically.
(8) Wait until the digitizer has rebooted.
(9) Click <OK> in the screen “To Confirm this Installation press OK” at the digitizer
display. Message “new software detected” appears in the digitizer display.
(10) After a few moments the following message appears:
“Please insert the backup medium and press <>.”
Press <Cancel>, the backup has to be performed later.
NOTE:
If the installation fails repeat installation, see section 3.3.3.
3.3.3 Installing Software c25_3103B on a new Hard Disk or in Case of corrupted Files
NOTE:
This procedure must be carried out only if the hard disk of the digitizer is corrupted
or has been replaced by a blank one.
NOTE:
Without restoring the device-specific parameters, an operation of the digitizer
is impossible. Use the backup storage medium attached to the Digitizer with
the most recent backup date.
(1) Enter the Service menu on the digitizer keypad and select:
<5 INSTALL data 2 Machine specific data>
(2) Follow the guidelines on the digitizer display.
Use the backup storage medium with the most recent backup date.
(1) Restore the test image files from CD or USB Memory Stick to the directory
E:\testimg on the digitizer. For current version refer to the service
documentation, Chapter 3.2, Tools and auxiliary means.
(2) Switch off and on the digitizer.
(3) Check if the digitizer performs its self test and afterwards changes
its state to Ready.
NOTE:
Between restoring the backup and installing the CPF file, a reset is necessary!
(1) Get the CPF file either from a previous backup or from the Processing Station.
Otherwise generate the CPF file with the QS, the VIPS or the CCM tool.
For details refer to the respective service documentation, Chapter 3.2,
Tools and auxiliary means.
(2) Enter the Service menu on the digitizer keypad and select:
<5 INSTALL data> <6 CPF-File>
(3) Follow the instructions displayed on the digitizer display.
(4) Reconnect the digitizer to the network.
(5) Disconnect the Service PC.
(6) Switch off and switch on the digitizer.
(7) Check if the digitizer performs its self test and afterwards changes
its state to Ready.
If the digitizer does not perform the self test, the software may not be installed
successfully.
(1) Close the front door and switch off and then switch on the digitizer.
(2) Check if the digitizer performs its self test and afterwards changes
its state to Ready.
If the digitizer does not perform the self test, the software may not be installed
successfully.
Create installation floppy disks either using the wizard (see section 4.1.1)
or manually (see section 4.1.2).
(1) Download the Digitizer Software c25_3103B (File size approximately 8 MB) from
the Agfa HealthCare Library:
Computed Radiography CR Digitizers CR 35-X
Software CR 35-X - Software - Device Software
c25_3103B
or
Computed Radiography CR Digitizers CR 25.0
Software CR 25.0 - Software - Device Software
c25_3103B
(2) Extract all files from the downloaded zip file into a folder on the Service PC.
Among the extracted files is the zip file c25_3103B.zip.
Figure 5
(7) When requested, remove floppy disk from the floppy disk drive and label
it with the corresponding floppy name, e.g. C253103B_1_3 for the first disk
of installation files (for a filename overview refer to table below).
(8) Follow the instructions and insert the next floppy disks.
When all 7 floppy disks were copied, the prompt will be closed automatically.
NOTE:
Check the number of floppy disks after finishing the copy process.
There must be 7 floppy disks.
Result The set of floppy disks for installation are labeled with the correct name and contain
the correct files for the installation.
(1) Download the Digitizer Software c25_3103B (File size approximately 8 MB) from
the Agfa HealthCare Library:
Computed Radiography CR Digitizers CR 35-X
Software CR 35-X - Software - Device Software
c25_3103B
or
Computed Radiography CR Digitizers CR 25.0
Software CR 25.0 - Software - Device Software
c25_3103B
(2) Extract all files from the downloaded zip file into a folder on the Service PC.
Among the extracted files is the zip file c25_3103B.zip.
NOTE:
For each of these folders, a floppy disk with the corresponding floppy
name must be created. Follow the instructions below.
Result The set of floppy disks for installation are labeled with the correct name and contain
the correct files for the installation.
IMPORTANT:
The digitizer checks light incidence.
Following error message may appear on the screen of the digitizer:
Digitizer locked
RIGHT SIDE PANEL NOT CLOSED.
Close right side panel.
Code 24791
Figure 7
If this error message appears, lean the cover against the right side of the device.
Then continue with next steps.
(3) Enter the Service menu on the digitizer keypad and select:
<3 Save data> <2 Machine specific data>
(4) Follow the instructions on the digitizer display.
NOTE:
During the installation from floppy disks the digitizer shows no status report.
It is necessary to wait after inserting of a floppy disk, until the LED of the digitizers
floppy disk drive stops flashing.
To see the installation status during the software installation, connect the Service PC
via serial connection cable (RS232) to the digitizer. By using the Hyper Terminal the
several installation steps of the digitizer can be followed.
(1) Insert the floppy disk labeled PORTEX1 into the floppy disk drive of the digitizer.
(2) Press the <Reset> switch on the CPU-Board.
(3) Wait until the control LED of the floppy disk drive is off.
(4) Remove the floppy disk.
(1) Insert the floppy disk labeled PORTEX2 into the floppy disk drive of the digitizer.
(2) Press the <Reset> switch on the CPU-Board.
(3) Wait until the control LED of the floppy disk drive is off.
(4) Wait additionally 1 minute, to be sure that all files are unzipped.
(5) Remove the floppy disk.
NOTE:
Do not insert any other floppy disk!
WAITING TIME:
Wait approximately 2 minutes.
(1) Insert the floppy disk labeled C253103B_1_3 into the floppy disk drive
of the digitizer.
(2) Press the <Reset> switch on the CPU-Board.
(3) Wait until the control LED of floppy disk drive is off.
(4) Remove the floppy disk.
(1) Insert the floppy disk labeled C253103B_2_3 into the floppy disk drive
of the digitizer.
(2) Press the <Reset> switch on the CPU-Board.
(3) Wait until the control LED of floppy disk drive is off.
(4) Remove the floppy disk.
(1) Insert the floppy disk labeled C253103B_3_3 into the floppy disk drive
of the digitizer.
(2) Press the <Reset> switch on the CPU-Board.
(3) Wait until the control LED of floppy disk drive is off.
(4) Wait additionally 1 minute, to be sure that all files are unzipped.
(5) Remove the floppy disk.
NOTE:
Do not insert any other floppy disk!
NOTE:
If the installation fails repeat installation, see section 3.3.3.
4.3.2 Installing Software c25_3103B on a new Hard Disk or in Case of corrupted Files
NOTE:
This procedure must be carried out only if the hard disk of the digitizer has been
replaced by a new one or if corrupted files have been detected on the hard disk.
In this case the hard disk must be formatted using the installation disk HDFORMAT.
(1) Insert the floppy disk labeled HDFORMAT into the floppy disk drive
of the digitizer.
(2) Press the <Reset> switch on the CPU-Board.
(3) Confirm partitioning/formatting of the hard disk on the terminal with <y>.
(4) Wait until the system prompt is displayed on the terminal.
(5) Remove the floppy disk.
(1) Insert the floppy disk labeled PORTEX2 into the floppy disk drive of the digitizer.
(2) Press the <Reset> switch on the CPU-Board.
(3) Wait until the system prompt is displayed on the terminal.
(4) Remove the floppy disk.
NOTE:
Do not insert any other floppy disk!
(1) Insert the floppy disk labeled C25_3103_1_3 into the floppy disk drive
of the digitizer.
(2) Press the <Reset> switch on the CPU-Board.
(3) Wait until the system prompt is displayed on the terminal.
(4) Wait additionally 1 minute, to be sure that all files are unzipped.
(5) Remove the floppy disk.
(1) Insert the floppy disk labeled C25_3103_2_3 into the floppy disk drive
of the digitizer.
(2) Press the <Reset> switch on the CPU-Board.
(3) Wait until the system prompt is displayed on the terminal.
(4) Remove the floppy disk.
(1) Insert the floppy disk labeled C25_3103_3_3 into the floppy disk drive
of the digitizer.
(2) Press the <Reset> switch on the CPU-Board.
(3) Wait until the system prompt is displayed on the terminal.
Wait additionally 1 minute, to be sure that all files are unzipped.
(4) Remove the floppy disk.
(5) Press the <Reset> switch on the CPU-Board.
(6) Wait until the system prompt is displayed on the terminal.
The system extracts the C25_3103B software.
(7) Switch off and then switch on the digitizer.
(8) Wait until the digitizer is ready.
NOTE:
Without restoring the device-specific parameters, an operation of the digitizer
is impossible. Use the backup floppy disk attached to the digitizer with the
most recent backup date.
(1) Enter the Service menu on the digitizer keypad and select:
<5 INSTALL data 2 Machine specific data>
(2) Follow the guidelines on the digitizer display.
Use the backup floppy disk with the most recent backup date.
(1) Restore the test image files from CD or USB Memory Stick to the directory
E:\testimg on the digitizer. For current version refer to the
service documentation, Chapter 3.2, Tools and auxiliary means.
(2) Switch off and on the digitizer.
(3) Check if the digitizer performs its self test and afterwards changes
its state to Ready.
NOTE:
Between restoring the backup and installing the CPF file, a reset is necessary!
(1) Get the CPF file either from a previous backup or from the Processing Station.
Otherwise generate the CPF file with the QS, the VIPS or the CCM tool.
Refer to the respective service documentation, Chapter 3.2, Tools and auxiliary
means, for details.
(2) Enter the Service menu on the digitizer keypad and select:
<5 INSTALL data> <6 CPF-File>
(3) Follow the instructions displayed on the digitizer display.
Have a floppy disk ready for the backup of device-specific parameters.
(4) Reconnect the digitizer to the network.
(5) Disconnect the Service PC.
(6) Switch off and switch on the digitizer.
(7) Check if the digitizer performs its self test and afterwards changes
its state to Ready.
(1) Close the front door and switch off and then switch on the digitizer.
(2) Check if the digitizer performs its self test and afterwards changes
its state to Ready.
If the digitizer does not perform the self test, the software may not be installed
successfully. This can be caused by a corruption of the file system on the hard disk.
(4) Perform the installation according to section 4.3.2, Installing Software c25_3103B
on a new Hard Disk or in Case of corrupted Files and then section 5, Installation
of C25L3103B Language Files.
NOTE:
The messages of the calibration screen have been adapted.
Therefore the C25L3103B language files have to be installed together with
the digitizer software c25_3103B.
NOTE:
With software version c25_3007 the term “floppy disk” has been exchanged with
the term “medium”.
To install C25L3103B language files and to change the language of the digitizer user
interface perform following procedure:
Follow the next steps to check via Telnet, if the software C25_3103B was installed
correctly:
(1) Open a DOS box (e.g. in Windows XP: Click on “Start” in Windows, than click on
“Run” and enter “cmd” in the field.)
(2) Enter "telnet <IP address of the digitizer>” (e.g. “telnet 192.168.3.26”).
(3) Confirm with Return.
(4) Enter login and password.
(5) Enter “liv”.
(6) Confirm with Return.
The Software version “C25_3103B” will be displayed.