* The last digit in the spare part number indicates the spare part revision at release of
this document. When ordering, the actual revision of the spare part is delivered.
► Referenced Documents
Document Title
DD+DIS150.06E Service Manual Chapter 3.6, Adjustments and Calibrations
NOTE:
Dispose the enclosure document after finishing your task
(observe the confidentiality note below).
CONFIDENTIALITY NOTE:
Use, dissemination, distribution or reproduction of this document by unauthorized personnel is not permitted and may be unlawful.
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Revision 0
08-2010 printed in Germany Document Node ID: 30488181
Enclosure_e_template_v07
Agfa Company Confidential Copyright © 2010 Agfa HealthCare N.V.
DD+DIS202.10E Enclosure
► Manufacturer
Agfa HealthCare N.V.
Publisher
Agfa-Gevaert HealthCare GmbH
Tegernseer Landstraße 161
D - 81539 München
Germany
Copyright 2010 Agfa HealthCare N.V.
All rights reserved.
Technical modifications reserved.
AGFA and the Agfa-Rhombus are trademarks of Agfa HealthCare N.V.
WARNING:
Improper operation or service activities may cause damage or injuries.
INSTRUCTION:
(1) Read the "Generic Safety Directions" document
(see MEDNET GSO => General Info => Agfa HealthCare => Publications =>
Service Manual) prior to attempting any operation, repair or maintenance task on
the equipment.
(2) Strictly observe all safety directions within the "Generic Safety Directions" and on
the product.
LIST OF CONTENTS
IMPORTANT:
The instructions described here are based on CR 30-X Service Manual chapter 3.6,
Edition 1, Revision 3 (release date: 10-2009).
An update of this chapter may have updated instructions available for usage of the
Drawer Adjustment Tool.
Recommendation: Before using this tool, check whether an updated document is
available under MedNet GSO Library path: Computed Radiography => CR Digitizers
=> CR 30-X => Repair and Service => CR 30-X - Chapter 03.6 - Adjustments and
Calibration
SCOPE OF DELIVERY:
* The last digit in the spare part number indicates the spare part revision at release of
this document. When ordering, the actual revision of the spare part is delivered.
Referenced Documents
Document Title
DD+DIS150.06E Service Manual Chapter 3.6, Adjustments and Claibrations
NOTE:
Dispose the enclosure document after finishing your task
(observe the confidentiality note below).
CONFIDENTIALITY NOTE:
Use, dissemination, distribution or reproduction of this document by unauthorized personnel is not permitted and may be unlawful.
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Revision 0
08-2010 printed in Germany Document Node ID: 30488181
Enclosure_e_template_v07
Agfa Company Confidential Copyright © 2010 Agfa HealthCare N.V.
REQUIRED TOOLS:
3 mm
REQUIRED TIME:
Approximately 30 minutes
Figure 1
CAUTION:
Less reliable IP run possible if IP guide plate is deformed.
Do not bend IP guide plate at removal.
Figure 2
517536ak.cdr
Figure 3
CAUTION:
Misalignment of drawer unit possible when pushing it at the upper plate.
When pushing the drawer unit only do this via drawer unit axle.
(9) Push drawer unit at the axle Upper plate Drawer axle
completely to the front.
Figure 4
NOT OK OK NOT OK
Lateral stop
Notch
Figure 6
IMPORTANT:
When inserting the drawer unit adjustment tool, be sure, that the notch of the tool is
properly inserted in the right hand side of the cassette unit entry. See Figure 6.
Evaluation In case the lateral stop is aligned with the plastic finger at the drawer unit there is no
adjustment required. Continue with section 3.
Otherwise perform lateral adjustment as described in next steps.
LATERAL ADJUSTMENT:
(1) Open the lateral position
fixation screws. See arrows.
517536dd.cdr
Figure 7
(1) Insert the drawer unit adjustment tool in the cassette unit entry.
(2) Check the minimum and maximum height. See also IMPORTANT note below.
min.
min.
max.
max.
Check max. Check max.
height left. height right.
max. max.
517536an.cdr
min.
min.
Figure 9
IMPORTANT:
Perform the minimum height check Magnet Plastic Finger
between magnet and plastic finger.
min.
Figure 10
Expected result:
The plate marked with "min."
(= minimum height) has to fit
under the drawer upper plate.
The plate marked with "max." Min. height
(= maximum height) may not fit Max. height
under the drawer upper plate.
Figure 11
Evaluation In case the height is OK on both sides there is no height adjustment required.
Otherwise proceed with next steps.
HEIGHT ADJUSTMENT
(1) Open the four height
adjustment fixation screws
(see arrows).
517536cd.cdr
Figure 12