Using the Drawer Adjustment Tool

► Purpose of this Document

This enclosure is part of the drawer adjustment tool,


spare part number*: CM+9 5175 6001 2

It describes all necessary steps for usage of the tool.

* The last digit in the spare part number indicates the spare part revision at release of
this document. When ordering, the actual revision of the spare part is delivered.

► Referenced Documents

Document Title
DD+DIS150.06E Service Manual Chapter 3.6, Adjustments and Calibrations

NOTE:
Dispose the enclosure document after finishing your task
(observe the confidentiality note below).

CONFIDENTIALITY NOTE:
Use, dissemination, distribution or reproduction of this document by unauthorized personnel is not permitted and may be unlawful.
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Revision 0
08-2010 printed in Germany Document Node ID: 30488181
Enclosure_e_template_v07
Agfa Company Confidential Copyright © 2010 Agfa HealthCare N.V.
DD+DIS202.10E Enclosure

► Manufacturer
Agfa HealthCare N.V.

Publisher
Agfa-Gevaert HealthCare GmbH
Tegernseer Landstraße 161
D - 81539 München
Germany
Copyright  2010 Agfa HealthCare N.V.
All rights reserved.
Technical modifications reserved.
AGFA and the Agfa-Rhombus are trademarks of Agfa HealthCare N.V.

WARNING:
Improper operation or service activities may cause damage or injuries.

INSTRUCTION:
(1) Read the "Generic Safety Directions" document
(see MEDNET GSO => General Info => Agfa HealthCare => Publications =>
Service Manual) prior to attempting any operation, repair or maintenance task on
the equipment.
(2) Strictly observe all safety directions within the "Generic Safety Directions" and on
the product.

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Revision 0 CR 30-X Page 2 of 11


08-2010 Type 5175 / 100/110 Agfa Company Confidential
DD+DIS202.10E Enclosure

LIST OF CONTENTS

1 PREPARING THE DRAWER UNIT ADJUSTMENT.................................................................5


2 CHECKING AND ADJUSTING THE LATERAL POSITION .....................................................7
3 CHECKING AND ADJUSTING THE HEIGHT ..........................................................................9

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Revision 0 CR 30-X Page 3 of 11


08-2010 Type 5175 / 100/110 Agfa Company Confidential
DD+DIS202.10E Enclosure

Subject Using the Drawer Adjustment Tool

IMPORTANT:
The instructions described here are based on CR 30-X Service Manual chapter 3.6,
Edition 1, Revision 3 (release date: 10-2009).
An update of this chapter may have updated instructions available for usage of the
Drawer Adjustment Tool.
Recommendation: Before using this tool, check whether an updated document is
available under MedNet GSO Library path: Computed Radiography => CR Digitizers
=> CR 30-X => Repair and Service => CR 30-X - Chapter 03.6 - Adjustments and
Calibration

SCOPE OF DELIVERY:

Drawer adjustment tool Installation Instructions


HEALTH CARE Enclosure

Document No: DD+DIS202.10E


This number is needed for production line only:
F7.5175.9023.0 CR 30-X
Type 5175 / 100/110

Using the Drawer Adjustment Tool

Purpose of this Document

This enclosure is part of the drawer adjustment tool,


spare part number*: CM+9.5175.6001.1

It describes all necessary steps for usage of the tool.

* The last digit in the spare part number indicates the spare part revision at release of
this document. When ordering, the actual revision of the spare part is delivered.

Referenced Documents

Document Title
DD+DIS150.06E Service Manual Chapter 3.6, Adjustments and Claibrations

NOTE:
Dispose the enclosure document after finishing your task
(observe the confidentiality note below).

CONFIDENTIALITY NOTE:
Use, dissemination, distribution or reproduction of this document by unauthorized personnel is not permitted and may be unlawful.
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Revision 0
08-2010 printed in Germany Document Node ID: 30488181
Enclosure_e_template_v07
Agfa Company Confidential Copyright © 2010 Agfa HealthCare N.V.

REQUIRED TOOLS:

3 mm

REQUIRED TIME:
Approximately 30 minutes

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Revision 0 CR 30-X Page 4 of 11


08-2010 Type 5175 / 100/110 Agfa Company Confidential
DD+DIS202.10E Enclosure

The drawer unit adjustment is split up in three parts:

# Drawer Unit Adjustment Part Section


1 Preparing the adjustment 1
2 Checking and adjusting the lateral position 2
3 Checking and adjusting the height position 3

1 Preparing the Drawer Unit Adjustment

(1) Switch off the digitizer.


(2) Disconnect the mains cable.
(3) Simultaneously press the two
buttons (A) positioned
underneath the feed table.
(4) Pull out cassette unit (B). B
A
A

Figure 1

CAUTION:
Less reliable IP run possible if IP guide plate is deformed.
Do not bend IP guide plate at removal.

(5) Remove the front cover.


See chapter 3.5.
(6) Remove the erasure unit.
See chapter 3.5.
(7) Slide IP guide plate to the left
and lift it out.

Figure 2

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Revision 0 CR 30-X Page 5 of 11


08-2010 Type 5175 / 100/110 Agfa Company Confidential
DD+DIS202.10E Enclosure

(8) Pull at the toothed belt to open


the cassette opener.

517536ak.cdr

Figure 3

CAUTION:
Misalignment of drawer unit possible when pushing it at the upper plate.
When pushing the drawer unit only do this via drawer unit axle.

(9) Push drawer unit at the axle Upper plate Drawer axle
completely to the front.

Figure 4

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Revision 0 CR 30-X Page 6 of 11


08-2010 Type 5175 / 100/110 Agfa Company Confidential
DD+DIS202.10E Enclosure

2 Checking and Adjusting the lateral Position

(1) Insert the drawer unit adjustment


tool in the cassette unit entry.
(2) Push it completely to the right.
(3) Check, whether plastic finger at
drawer unit is aligned with the
lateral stop of the drawer
adjustment tool. See Figure 6.
Figure 5

NOT OK OK NOT OK

Lateral stop

Notch

Distance exaggerated Drawer adjustment tool Distance exaggerated


517536ah.cdr

Figure 6

IMPORTANT:
When inserting the drawer unit adjustment tool, be sure, that the notch of the tool is
properly inserted in the right hand side of the cassette unit entry. See Figure 6.

Evaluation In case the lateral stop is aligned with the plastic finger at the drawer unit there is no
adjustment required. Continue with section 3.
Otherwise perform lateral adjustment as described in next steps.

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Revision 0 CR 30-X Page 7 of 11


08-2010 Type 5175 / 100/110 Agfa Company Confidential
DD+DIS202.10E Enclosure

LATERAL ADJUSTMENT:
(1) Open the lateral position
fixation screws. See arrows.

517536dd.cdr

Figure 7

(2) Shift the drawer unit upper


plate this way, that:
Plastic finger
 The plastic finger at the
Cassette input
drawer unit is aligned with side wall
the lateral stop
 While the drawer
adjustment tool is shifted to
the cassette input side wall. Lateral stop
517536az.cdr

(3) Fasten the lateral position


fixation screws. Figure 8

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Revision 0 CR 30-X Page 8 of 11


08-2010 Type 5175 / 100/110 Agfa Company Confidential
DD+DIS202.10E Enclosure

3 Checking and Adjusting the Height

(1) Insert the drawer unit adjustment tool in the cassette unit entry.
(2) Check the minimum and maximum height. See also IMPORTANT note below.

Check min. Check min.


height left. height right.

min.
min.
max.

max.
Check max. Check max.
height left. height right.

max. max.

517536an.cdr

min.
min.

Figure 9

IMPORTANT:
Perform the minimum height check Magnet Plastic Finger
between magnet and plastic finger.

min.

Figure 10

Expected result:
 The plate marked with "min."
(= minimum height) has to fit
under the drawer upper plate.
 The plate marked with "max." Min. height
(= maximum height) may not fit Max. height
under the drawer upper plate.
Figure 11

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Revision 0 CR 30-X Page 9 of 11


08-2010 Type 5175 / 100/110 Agfa Company Confidential
DD+DIS202.10E Enclosure

Evaluation In case the height is OK on both sides there is no height adjustment required.
Otherwise proceed with next steps.

HEIGHT ADJUSTMENT
(1) Open the four height
adjustment fixation screws
(see arrows).

517536cd.cdr

Figure 12

(2) Adjust the height with the


height adjustment screws
(see arrows).
(3) Press down the drawer unit
upper plate after each
adjustment cycle (before
checking the height again).
(4) Fasten the four height
adjustment fixation screws.
(5) Again, check the height
adjustment (min. and max.)
with the drawer adjustment tool.
(6) Repeat adjustment if required.
(7) Mount IP guide plate, erasure
unit and front cover.
(8) Connect the mains cable. Figure 13

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Revision 0 CR 30-X Page 10 of 11


08-2010 Type 5175 / 100/110 Agfa Company Confidential
DD+DIS202.10E Enclosure

Verification (1) Switch on the digitizer.


(2) Wait until the green light at the user interface lights up
(takes approximately 1 minute).
(3) Scan one image plate with each cassette format the customer is using.

Result The drawer unit adjustment is performed.

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Revision 0 CR 30-X Page 11 of 11


08-2010 Type 5175 / 100/110 Agfa Company Confidential