Pulse Shaper Cavity Troubleshooting

Perform daily calibration, Click on the image acquisition.


conduct scientific Do you have a pump-
Can you detect probe intensity
experimentation, log all Yes probe signal with a No anywhere on the detector
activity in the 2DIR sample present? array?
logbook

Barely No Yes Barely

Assuming probe intensity is


This means the probe beam is not If pump intensity is
sufficient and absorption of sample
is sufficient, you need to increase hitting the detector. Verify TOPAS sufficiently high, do you
pump-probe signal. You can do this output. Restart QC hardware and have an absorption
by: software. Restart AOM and ensure signal when a sample is This means you may need to
warning indicator light is not on. present? re-route the probe beam to
1) Increasing spatial overlap of the maximize the signal on the
Ensure all shutters entering and
pump-probe beam by slightly detector. Before you do
exiting pulse shaper cavity are Yes No this, verify TOPAS output,
moving the silver pump mirror.
open. Verify that the verify monochromator is at
waveplate/polarizer is allowing the proper wavelength and
2) Increasing temporal overlap by
is set to 30 g/mm. Ensure
moving the probe delay stage, maximum transmission along probe Ensure monochromator
the waveplate and polarizer
motorized pump delay stage, or path. Ensure guide beam/IR is set to 30 g/mm and is
in the probe path are set to
changing the dispersion parameters overlap flip mirror is in the correct set to a proper maximize transmission.
in the phasetech software. wavelength. Ensure the Ensure your sample is fresh
position. Verify probe output
probe beam interacts and interacting with the
using the Pyro detector near the
3) Check your sample, ensure it has with the sample by probe beam by translating
not evaporated and is not expired sample. If pyro output is detected , translating the sample the sample stage as
by undergoing D2O/H2O exchange. it may be necessary to steer the stage as necessary. necessary.
Translate sample stage in X, Y, Z beam onto the detector array using Ensure purge air is
1: Attempt to steer the
direction as necessary. the mirrors. If pyro output is not entering cavity. Check optics located after sample
detected, retrace the probe path your sample; the analyte interaction to determine if
See the Troubleshooting SOP and concentration may be
and determine if the probe beam is signal intensity increases.
perform the “OBTAIN SIGNAL” too low. Check scaling Observe the acquisition on
clipped or not hitting a mirror along
procedure. factor of acquisition. the image tab to determine
the probe path. if the reference beam needs
to change.

2: If step 1 is not sufficient,


you may need to move the
first two silver mirrors in the
pulse shaper cavity to
improve the probe
absorption. This will alter
This means there is probably the pump temporal/spatial
an issue with the pump beam, parameters as well.
either temporally or spatially. Downstream adjustments to
the silver pump mirror and
Perform the “LOSS OF the dispersion and delay
stage will be necessary to
SIGNAL” procedure in the
account for this change.
Troubleshooting SOP