Condition
Software update 20MW42.23_SP3-0-1 (2.2C SP3) is a cumulative service pack release that includes Service Pack 1 (20MW42.23_SP1-0-5) and Service Pack 2
(20MW42.23_SP2-0-0). 2.2C Service Pack 3 may be installed with 20MW42.23 software or later. See Table 6 for applicable software configurations.
At minimum, Revolution CT and Apex 2.2C (20MW42.23) software release (FMI 25498) is required before beginning this procedure, 20MW42.23 (2.2C) and 2.2C SP3
software versions DO NOT support a HP Z800, or System Cabinets I and II. The 20MW42.23 release supports Z420, Z440 and P520 console computers with System
Cabinet III or later. Refer to the FMI 25498 instructions and (20MW42.23) Load From Cold Instructions for 20MW42.23 (2.2C) software installation. It is recommended
to install 20MW42.23_SP3-0-1 (2.2C SP3) after completing FMI 25498.
The 2.2C SP3 software update may be installed by any of the following methods:
1. eDelivery of SP3 software: Direct-to-Device
2. eDelivery of SP3 software: Portal download
3. Installing service pack by USB
Method options 1 and 2 require eDelivery to complete, which is a new process being introduced with this service pack for eligible customers. If not already enabled,
the FE should discuss the Flexera eDelivery process with customer at this point to help determine if they should turn ON Software Download, which requires a Class
M key to enable. Refer to (20MW42.23) Load From Cold Instructions, Software Download Enable section.
Revolution CT ES
Revolution CT with
Apex edition SP3-0-1
1. Remove SSA Service key from USB port on system before starting the service pack load process.
NOTE:
If a SSA service key is plugged into the system and is in its "locked" mode, mouse and/or keyboard lock ups may occur during the system boot up.
2. Remove any USB storage media (external HDD or flash memory drives) before starting the service pack load process.
NOTE:
If a USB storage media is plugged into the system and is in its "locked" mode, mouse and/or keyboard lock ups may occur during the system boot
up.
IMPORTANT:
If the software string does not match any of those listed above, a prior software update is required. DO NOT load this software patch.
4. If the software requires the patch, proceed to installation steps below. If the software is already loaded, stop; no need to install the patch.
1. Follow the eDelivery of Software: Direct-to-Device procedure to install the 2.2C SP3 update (20MW42.23_GEHCServicePack3-0-1.iso.).
c. If the output is indicated above, then the 20MW42.23 SP3 software update has been successfully installed.
d. If the output is NOT indicated as above, contact GE Healthcare to initiate a Request for Service. Refer to the FMI 25501 instructions and eDelivery
procedures for more information.
NOTE:
After successful download or installation, the directory file that will be located in /usr/g/database/HistoryArea or
/usr/g/database/StagingArea will be numbered: 5915027_1.
Step-by-step instructions
1. If applicable software is up, perform Application Shutdown from the Common Service Desktop (CSD).
NOTE:
Current user must have permission to shut down Application. Login as root user if necessary to perform Application shutdown.
5. If not completed yet, confirm software patch is required on the system. (See the Pre-software pack checks section.) If patch is required, continue to Step 6.
6. Insert the USB flash drive into the USB port on the host computer. Type: [root@hostname]# mountUSB
7. Mount service pack image. Type: [root@hostname]# mount -t iso9660 -o loop /USB/20MW42.23_GEHCServicePack3-0-1.iso /USB
NOTE:
Skip this step if using USB PN 5919580. This step is only necessary if ISO file is obtained via eDelivery Portal download.
9. Perform the following steps to install the service pack install commands.
Type: [root@hostname]# ./check_media
System report: ========Starting media check ========= Disk Passed Self Verification **** check media PASSED ****
Type: [root@hostname]# ./install_patch
NOTE:
Skip this step if using USB PN 5919580. This step is only necessary if ISO file is obtained via eDelivery Portal download.
or
IMPORTANT:
If updating the system software for RIB1.x, stationary inverter, or UPS card updates, be aware that Flash Download may fail. Service pack 3 must be
installed and Flash Download may need to be executed.
Step-by-step instructions
1. If necessary, reinstall Scanner Desktop cabinet front cover.
2. Perform the System Scanning Test within Class A Revolution CT Service Manual 5866709-8EN.