Professional Documents
Culture Documents
Radiotherapy [NAHU]-001
Revision 0.0
Dated [2011-06-09]
SPECIFICATION
Approved
__________________________________________________
Division of NAHU --- Date 08.07.2011 (MTPS)
Page 1 of 6
Linear accelerator for IAEA Specification
Radiotherapy [NAHU]-001
Revision 0.0
Dated [2011-06-09]
1. Scope
This specification describes the requirements for a medical linear accelerator for the
treatment of patients in radiation oncology.
The radiotherapy centre in Parirenyatwa hospital has 3 high energy treatment rooms,
one of which contains a linear accelerator. The layout of the bunker that previously
housed the Brown-Boveri Dynaray-CH (6 MV) linear accelerator, is included
(Annexure I). This is the bunker in which the linear accelerator specified in this
document, is to be installed.
2. Applicable Documents
The following documents shall be applicable for this Specification to the extent
specified hereinafter:
“International Basic Safety Standards for Protection against Ionizing Radiation and for
the Safety of Radiation Sources”, IAEA, Vienna (1996) (http://www-
pub.iaea.org/MTCD/publications/PDF/Pub996_EN.pdf)
In the event of conflict between the documents listed above and the content of this
Specification, the content of this Specification shall take precedence to the extent of
the conflict.
The following definitions, acronyms, and abbreviations shall apply throughout this
Specification unless defined otherwise hereinafter:
MV: megavoltage
cm: centimeters
mu: monitor units
Gy: Gray
CCTV: closed circuit television
Page 2 of 6
Linear accelerator for IAEA Specification
Radiotherapy [NAHU]-001
Revision 0.0
Dated [2011-06-09]
4. Requirements
The System shall meet the following functional and performance requirements:
4.2.1. Single high energy photon beam linear accelerator (4 – 6 MV) machine,
consisting of
4.2.2. A range of output from at least 50 mu/min to 200 mu/min providing a dose rate of 1
Gy/100 mu in water at the depth of maximum dose at the isocenter in a field size of 10
cm × 10 cm.
4.2.3. A gantry for treatments at SAD = 100 cm,
4.2.4. A patient table providing motorized or hydraulic vertical movement, longitudinal and
lateral motions with isocentric rotation. It should be equipped with a transparent window
for non-attenuated posterior treatment and a centre spine for posterior oblique portals
4.2.5. Collimators providing a maximum field size of 40 cm x 40 cm with at least one set of
asymmetric collimators capable of shielding half the beam.
4.2.6. A set of wedge filters (or a universal or soft wedge). Note that wedging shall be
possible in the direction of the asymmetric collimation.
4.2.7. A comprehensive set of standard shielding blocks with a set of at least 100 coded
block trays,
4.2.8. An optical backpointer
4.2.9. An in-room monitor
4.2.10. A cassette holder with four cassettes for megavoltage portal imaging, including a
tabletop film processor (alternatively an integrated portal imaging system)
4.2.11. a set of 3 lasers for patient centring,
4.2.12. an audio-visual radiation room monitor,
4.2.13. 2 CCTV systems; one with pan, tilt and zoom capability
4.2.14. A patient intercommunication system,
Page 3 of 6
Linear accelerator for IAEA Specification
Radiotherapy [NAHU]-001
Revision 0.0
Dated [2011-06-09]
Page 4 of 6
Linear accelerator for IAEA Specification
Radiotherapy [NAHU]-001
Revision 0.0
Dated [2011-06-09]
5. Marking
The System shall have all safety markings in the English language.
6. Packing
The System, for the shipment by air to the End-User, shall be packed in accordance
with international standards that are applicable for the shipment by air of this kind of
equipment.
7. Quality Requirements
7.1. The System shall be manufactured, shipped and installed in accordance with the ISO
9000 quality assurance system or an equivalent quality assurance system.
7.2. The Contractor shall document the compliance with this quality assurance system.
The System, prior to shipment, shall be tested for conformance of the System with
manufacturer’s performance specifications and the minimum requirements specified
herein.
The System, after installation, shall be tested by the Contractor together with the
hospital medical physicist to demonstrate that the performance meets the
manufacturer’s performance specifications and the minimum requirements specified
herein as determined by the IAEA and the End-User.
The results of the testing of the System shall be documented by the Contractor in an
acceptance protocol that shall be signed by the hospital medical physicist.
The Contractor shall install the System at the Radiotherapy Centre within the
Parirenyatwa Group of Hospitals, Mazowe Street, Harare, Zimbabwe. Training of a
local hospital service engineer in front-line maintenance shall be included, preferably during
the installation and acceptance testing process.
The Contractor shall provide three days’ training for up to three staff of the End-User
in the operation of the System at the End-User’s location immediately after the
installation and acceptance testing of the System.
Page 5 of 6
Linear accelerator for IAEA Specification
Radiotherapy [NAHU]-001
Revision 0.0
Dated [2011-06-09]
The Contractor shall provide two complete sets of operation and servicing manuals
and technical drawings in the English language.
ANNEXURE I
Page 6 of 6