Professional Documents
Culture Documents
PR-9F Procedure For Verification of Pi Tape
PR-9F Procedure For Verification of Pi Tape
1. Table of Content:
1. Table of Content:.............................................................................................1
3. Purpose:......................................................................................................... 3
4. Responsibility:................................................................................................ 3
5. Reference standard:....................................................................................... 3
6. Verification Process:....................................................................................... 3
Sr. No. Issue No. Rev. No. Effective Date Release/Change Comments
1 1 0 01.05.2022 Initial release
3. Purpose:
The purpose of this procedure is to provide specific instructions for the verification of PI Tape. This procedure
also ensures its proper functioning and continued accuracy.
4. Responsibility:
User shall be responsible for verification of PI Tape before use.
5. Reference standard:
Comparison method used
6. Verification Process:
Inspector/Operator/user shall check the PI Tape for:
6.1 Any damage which effect on measurement and general wear of fixed reference points.
6.2 Legibility of Reading on measuring scale
6.3 Take at least one reading on sample PI Tape and compare with Master PI Tape reading and verify the
acceptance criteria. If difference is more than the acceptable error, the equipment shall be replaced with
new one. Defective equipment shall be reported.
6.4 Verification remark shall be recorded in the respective report only.