You are on page 1of 4

2.9.26. Specific surface area by gas adsorption EUROPEAN PHARMACOPOEIA 6.

volume change or sample dilution is needed. Alternatively, between the adsorbate gas molecules and the adsorbent
determine the content of active substance(s) remaining surface of the test powder. The determination is usually
in the gum residue. Carry out the test successively on carried out at the temperature of liquid nitrogen. The
6 medicated chewing gums. amount of gas adsorbed can be measured by a volumetric or
The quantity of active substance(s) dissolved in a specified continuous flow procedure.
time is expressed as a percentage of the content stated on
the label. BRUNAUER, EMMETT AND TELLER (BET) THEORY AND
SPECIFIC SURFACE AREA DETERMINATION
MULTI-POINT MEASUREMENT
The data are treated according to the Brunauer, Emmett and
Teller (BET) adsorption isotherm equation :

(1)

P = partial vapour pressure of adsorbate gas in


equilibrium with the surface at 77.4 K (b.p. of
liquid nitrogen), in pascals,
Po = saturated pressure of adsorbate gas, in pascals,
Va = volume of gas adsorbed at standard temperature
and pressure (STP) [273.15 K and atmospheric
pressure (1.013 × 105 Pa)], in millilitres,
Vm = volume of gas adsorbed at STP to produce an
apparent monolayer on the sample surface, in
millilitres,
C = dimensionless constant that is related to the
Figure 2.9.25.-2 – Funnel
enthalpy of adsorption of the adsorbate gas on
(dimensions in millimetres) the powder sample.
A value of Va is measured at each of not less than 3 values of
P/Po.
Then the BET value

is plotted against P/Po according to equation (1). This


plot should yield a straight line usually in the approximate
relative pressure range 0.05 to 0.3. The data are considered
acceptable if the correlation coefficient, r, of the linear
regression is not less than 0.9975 ; that is, r2 is not less than
0.995. From the resulting linear plot, the slope, which is
equal to (C − 1)/VmC, and the intercept, which is equal to
1/VmC, are evaluated by linear regression analysis. From
these values, Vm is calculated as 1/(slope + intercept), while
C is calculated as (slope/intercept) + 1. From the value of
Vm so determined, the specific surface area, S, in m2·g–1, is
calculated by the equation :

Figure 2.9.25.-3 – Guide (section G-G)


(2)
(dimensions in millimetres)
N = Avogadro constant (6.022 × 1023 mol− 1),
01/2008:20926
a = effective cross-sectional area of one adsorbate
molecule, in square metres (0.162 nm2 for
2.9.26. SPECIFIC SURFACE AREA BY nitrogen and 0.195 nm2 for krypton),
GAS ADSORPTION m = mass of test powder, in grams,
INTRODUCTION 22400 = volume occupied by 1 mole of the adsorbate
The specific surface area of a powder is determined by gas at STP allowing for minor departures from
physical adsorption of a gas on the surface of the solid and the ideal, in millilitres.
by calculating the amount of adsorbate gas corresponding to A minimum of 3 data points is required. Additional
a monomolecular layer on the surface. Physical adsorption measurements may be carried out, especially when
results from relatively weak forces (van der Waals forces) non-linearity is obtained at a P/Po value close to 0.3.

306 See the information section on general monographs (cover pages)


EUROPEAN PHARMACOPOEIA 6.0 2.9.26. Specific surface area by gas adsorption

Because non-linearity is often obtained at a P/Po value Conditions. The outgassing conditions must be
below 0.05, values in this region are not recommended. demonstrated to yield reproducible BET plots, a constant
The test for linearity, the treatment of the data, and the weight of test powder, and no detectable physical or chemical
calculation of the specific surface area of the sample are changes in the test powder.
described above. The outgassing conditions defined by the temperature,
SINGLE-POINT MEASUREMENT pressure and time should be chosen so that the original
Normally, at least 3 measurements of Va each at different surface of the solid is reproduced as closely as possible.
values of P/Po are required for the determination of specific Outgassing of many substances is often achieved by applying
surface area by the dynamic flow gas adsorption technique a vacuum, by purging the sample in a flowing stream of a
(Method I) or by volumetric gas adsorption (Method II). non-reactive, dry gas, or by applying a desorption-adsorption
However, under certain circumstances described below, it cycling method. In either case, elevated temperatures
may be acceptable to determine the specific surface area are sometimes applied to increase the rate at which
of a powder from a single value of Va measured at a single the contaminants leave the surface. Caution should be
value of P/Po such as 0.300 (corresponding to 0.300 mole exercised when outgassing powder samples using elevated
of nitrogen or 0.001038 mole fraction of krypton), using the temperatures to avoid affecting the nature of the surface and
following equation for calculating Vm : the integrity of the sample.
If heating is employed, the recommended temperature
and time of outgassing are as low as possible to achieve
(3) reproducible measurement of specific surface area in an
acceptable time. For outgassing sensitive samples, other
The specific surface area is then calculated from the value of outgassing methods such as the desorption-adsorption
Vm by equation (2) given above. cycling method may be employed.
The single-point method may be employed directly for a series Adsorbate
of powder samples of a given material for which the material The standard technique is the adsorption of nitrogen of
constant C is much greater than unity. These circumstances analytical quality at liquid nitrogen temperature.
may be verified by comparing values of specific surface area For powders of low specific surface area (< 0.2 m2·g− 1) the
determined by the single-point method with that determined proportion adsorbed is low. In such cases the use of krypton
by the multiple-point method for the series of powder at liquid nitrogen temperature is preferred because the low
samples. Close similarity between the single-point values and vapour pressure exerted by this gas greatly reduces error.
multiple-point values suggests that 1/C approaches zero. The use of larger sample quantities where feasible (equivalent
The single-point method may be employed indirectly for a to 1 m2 or greater total surface area using nitrogen) may
series of very similar powder samples of a given material for compensate for the errors in determining low surface areas.
which the material constant C is not infinite but may be All gases used must be free from moisture.
assumed to be invariant. Under these circumstances, the Quantity of sample
error associated with the single-point method can be reduced
Accurately weigh a quantity of the test powder such that the
or eliminated by using the multiple-point method to evaluate
total surface of the sample is at least 1 m2 when the adsorbate
C for one of the samples of the series from the BET plot,
is nitrogen and 0.5 m2 when the adsorbate is krypton.
from which C is calculated as (1 + slope/intercept). Then Vm
is calculated from the single value of Va measured at a single Lower quantities of sample may be used after appropriate
value of P/Po by the equation : validation.
MEASUREMENTS
Because the amount of gas adsorbed under a given pressure
tends to increase on decreasing the temperature, adsorption
(4)
measurements are usually made at a low temperature.
Measurement is performed at 77.4 K, the boiling point of
The specific surface area is calculated from Vm by equation (2) liquid nitrogen.
given above. Method I : the dynamic flow method
Principle
EXPERIMENTAL TECHNIQUES
In the dynamic flow method (see Figure 2.9.26.-1), the
This section describes the methods to be used for the sample recommended adsorbate gas is dry nitrogen or krypton,
preparation, the dynamic flow gas adsorption technique while helium is employed as a diluent gas, which is not
(Method I) and the volumetric gas adsorption technique adsorbed under the recommended conditions.
(Method II). A minimum of 3 mixtures of the appropriate adsorbate gas
SAMPLE PREPARATION with helium are required within the P/Po range 0.05 to 0.30.
Outgassing The gas detector-integrator should provide a signal that is
Before the specific surface area of the sample can be approximately proportional to the volume of the gas passing
determined, it is necessary to remove gases and vapours that through it under defined conditions of temperature and
may have become physically adsorbed onto the surface after pressure. For this purpose, a thermal conductivity detector
manufacture and during treatment, handling and storage. If with an electronic integrator is one among various suitable
outgassing is not achieved, the specific surface area may be types. A minimum of 3 data points within the recommended
reduced or may be variable because an intermediate area range of 0.05 to 0.30 for P/Po is to be determined.
of the surface is covered with molecules of the previously Procedure
adsorbed gases or vapours. The outgassing conditions are A known mixture of the gases, usually nitrogen and helium,
critical for obtaining the required precision and accuracy is passed through a thermal conductivity cell, through the
of specific surface area measurements on pharmaceuticals sample, again through the thermal conductivity cell and
because of the sensitivity of the surface of the materials. then to a recording potentiometer.

General Notices (1) apply to all monographs and other texts 307
2.9.26. Specific surface area by gas adsorption EUROPEAN PHARMACOPOEIA 6.0

Immerse the sample cell in liquid nitrogen, then the sample Procedure
adsorbs nitrogen from the mobile phase. This unbalances
Admit a small amount of dry nitrogen into the sample tube
the thermal conductivity cell, and a pulse is generated on
to prevent contamination of the clean surface, remove the
a recorder chart.
sample tube, insert the stopper, and weigh it. Calculate
Remove from the coolant ; this gives a desorption peak equal the weight of the sample. Attach the sample tube to the
in area and in the opposite direction to the adsorption peak. volumetric apparatus. Cautiously evacuate the sample down
Since this is better defined than the adsorption peak, it is to the specified pressure (e.g. between 2 Pa and 10 Pa).
the one used for the determination. Alternatively, some instruments operate by evacuating to a
defined rate of pressure change (e.g. less than 13 Pa/30 s)
To effect the calibration, inject a known quantity of adsorbate and holding for a defined period of time before commencing
into the system, sufficient to give a peak of similar magnitude the next step.
to the desorption peak and obtain the proportion of gas If the principle of operation of the instrument requires the
volume per unit peak area. determination of the dead volume in the sample tube, for
example, by the admission of a non-adsorbed gas, such as
Use a nitrogen/helium mixture for a single-point helium, this procedure is carried out at this point, followed
determination and several such mixtures or premixing by evacuation of the sample. The determination of dead
2 streams of gas for a multiple-point determination. volume may be avoided using difference measurements, that
Calculation is essentially the same as for the volumetric is, by means of reference and sample tubes connected by a
method. differential transducer. The adsorption of nitrogen gas is
then measured as described below.
Method II : the volumetric method
Raise a Dewar vessel containing liquid nitrogen at 77.4 K
Principle up to a defined point on the sample cell. Admit a sufficient
volume of adsorbate gas to give the lowest desired relative
In the volumetric method (see Figure 2.9.26.-2), the pressure. Measure the volume adsorbed, Va. For multipoint
recommended adsorbate gas is nitrogen which is admitted measurements, repeat the measurement of Va at successively
into the evacuated space above the previously outgassed higher P/Po values. When nitrogen is used as the adsorbate
powder sample to give a defined equilibrium pressure, P, of gas, P/Po values of 0.10, 0.20, and 0.30 are often suitable.
the gas. The use of a diluent gas, such as helium, is therefore
unnecessary, although helium may be employed for other REFERENCE MATERIALS
purposes, such as to measure the dead volume.
Periodically verify the functioning of the apparatus using
Since only pure adsorbate gas, instead of a gas mixture, is appropriate reference materials of known surface area, such
employed, interfering effects of thermal diffusion are avoided as α-alumina, which should have a specific surface area
in this method. similar to that of the sample to be examined.

Figure 2.9.26.-1. — Schematic diagram of the dynamic flow method apparatus

308 See the information section on general monographs (cover pages)


EUROPEAN PHARMACOPOEIA 6.0 2.9.29. Intrinsic dissolution

Figure 2.9.26.-2. — Schematic diagram of the volumetric method apparatus


01/2008:20927 influenced by extrinsic factors (test conditions), such as
hydrodynamics, temperature, viscosity, pH, buffer strength
2.9.27. UNIFORMITY OF MASS and ionic strength of the dissolution medium.
The assessment of intrinsic dissolution rate of a solid
OF DELIVERED DOSES FROM substance involves the preparation of a compact. Assurance
MULTIDOSE CONTAINERS of appropriate compaction properties of the powder to be
tested is needed prior to performing the test.
The following test is intended for oral dosage forms such
The intrinsic dissolution rate is determined by exposing a
as granules, powders for oral use and liquids for oral use,
constant area of the compacted substance to an appropriate
which are supplied in multidose containers provided at
dissolution medium, while maintaining constant stirring
manufacture with a measuring device.
rate, temperature, ionic strength and pH.
Weigh individually 20 doses taken at random from one or The intrinsic dissolution rate is expressed in terms of
more containers with the measuring device provided and dissolved mass of substance per time per exposed area,
determine the individual and average masses. Not more than typically in milligrams per minute per square centimetre
2 of the individual masses deviate from the average mass (mg·min− 1·cm− 2).
by more than 10 per cent and none deviates by more than
20 per cent. APPARATUS
A typical apparatus consists of a punch and die fabricated
out of hardened steel. The base of the die has 3 threaded
01/2008:20929 holes for the attachment of a surface plate made of polished
steel, providing a mirror-smooth base for the compact.
2.9.29. INTRINSIC DISSOLUTION The die has a 0.1-1.0 cm diameter cavity into which a
measured amount of the powder to be tested is placed. The
The test is intended to determine the intrinsic dissolution punch is then inserted in the die cavity and the material is
rate of pure solid substances following compaction. It is compressed, generally using a benchtop hydraulic press. A
carried out under specified experimental conditions such hole through the head of the punch allows insertion of a
that a practical measure of the intrinsic dissolution rate is metal rod to facilitate removal from the die after the test. A
obtained. compact is formed in the cavity with a single face of defined
The intrinsic dissolution rate is a theoretical value referring area exposed on the bottom of the die (Figure 2.9.29.-1). The
to pure solid substances having null porosity, but, practically, bottom of the die cavity is threaded so that at least 50-75 per
intrinsic dissolution rate is determined on substances having cent of the compact can dissolve without falling out of the
a minimal porosity. die. The top of the die has a threaded shoulder that allows
it to be attached to a holder. The holder is mounted on
PRINCIPLE a laboratory stirring device, and the entire die, with the
The intrinsic dissolution rate is defined as the dissolution compact still in place, is immersed in the dissolution medium
rate of pure substances following compaction under the and rotated by the stirring device.
condition of constant surface area. Its assessment is useful
PROCEDURE
in the characterisation of active substances and excipients.
Weigh the material onto a piece of weighing paper. Attach
The dissolution rate of pure substances can be affected by all
the surface plate to the underside of the die, and secure it
the solid state properties such as crystal habit, crystallinity,
with the 3 provided screws. Transfer the sample of powder
amorphism, polymorphism, pseudo-polymorphism, particle
tested into the die cavity. Place the punch into the chamber,
size and specific surface area. In addition, it can also be

General Notices (1) apply to all monographs and other texts 309

You might also like