You are on page 1of 2

Test Procedure

Raw Water Jar Test

JAR TESTING Jar tests should be benchmarked using a


Raw water characteristics and treatment plant’s current treatment program. This
processes vary widely, but water treat- allows the testor to adjust fast/slow/
ment experts agree that the chemicals settling times to correlate with full scale
needed to accomplish treatment goals can performance. The actual dose in use
be screened by jar testing. in the plant at the time of the tests is set
as the Benchmark for the series of tests.
However, the chemicals and dosages
needed for optimal coagulation and
flocculation are bestdetermined in FAST MIX
practice by full-scale evaluation. To The time and energy of the rapid mix
accomplish appropriate product selection portion of the test is chosen to approxi-
at the bench, jar testing techniques are mate plant conditions. Usually, a fast mix
used to simulate the of 20 sec to 3 minutes is used. Mixing
treatment process energy is at relatively high rpm, about
Standard jar test mixing parameters and full-scale 200 rpm or 70-100% of gang stirrer
1) FAST MIX — to disperse the chemical and form pin floc performance. maximum speed. Primary coagulant such
via charge neutralization of suspended solids Parameters of as alum, ULTRION, etc. is added at the
chemical type, beginning of the fast mix. Since mixing is
2) SLOW MIX — to allow particle to particle collisions for chemical dosage, important, all jars should have coagulant
additional floc formation and growth mixing energy, and added in exactly the same manner each
duration of mixing time. The syringe should be at the same
3) SETTLING TIME — to simulate sedimentation processes are typically varied angle, with chemical going into the same
in bench tests part of the vortex every time. Avoid
known as jar tests. splashing chemical, getting a lower than
A multiple station paddle mixer, or gang expected dose. If flocculant is being used,
stirrer, is typically used for testing. this is normally added just before the
end of the fast mix which is a compro-
The duration of each mixing step must be mise between thoroughly mixed and
determined for the individual application. minimally sheared flocculant.
For example, if alum is added at the raw
water intake one-half mile away from the
clarifier, then the fast mix should be SLOW MIX
longer than if alum is added only 20 feet
This part of the process simulation is
before the clarifier. A plant that is
generally accomplished in much less time
hydraulically overloaded suggests a
but at higher energy than the actual
shorter slow mix time be used. Some-
plant operation. After the fast mix, the
times, appropriate jar testing parameters
stirrers are turned down to a slower
can be learned from experienced plant
speed (15 to 40 rpm) to allow particle
personnel. Also, they can be determined
to particle contact without shearing of
empirically by varying the mixing param-
the floc. Slow mixing is continued for a
eters and measuring performance.
period of time varying from about 5 to
NOTE: Jar testing only simulates 20 minutes. If the sample water has high
sedimentation and cannot simulate turbidity, the rate of particle collisions
filtration through a sludge blanket or may be high enough to shorten the slow
sand filter. mix time. Relatively clean water (low

(Continued on Reverse Side)

Form 646
turbidity) may require a longer slow SAMPLING
mix to get a sufficient number of
Settled water turbidity
collisions for floc growth.
measurements are generally
Floc size of the average particle formed used to make quantitative
in the flocculation stage is usually comparison of the jar tests as
ranked versus the benchmark under a function of the principal
the same mixing parameters. Other parameter of the test. A
factors such as water clarity between syringe or pipette can be
floc particles, floc shape, tightness, etc. used for sampling the
should be noted for comparison. supernatant water in the jars
at a depth where the settled
solids will not be disturbed.
Some jars may have built-in
SETTLING sampling ports. Water is
After stopping the stirrer, the stirring usually sampled approxi-
paddles are removed and the treated mately 1 cm below the
water is allowed to settle undisturbed surface and is usually the Figure 1 — Typical dosage curve for an organic coagulant.
for sufficient time to settle solids and cleanest possible water.
clarify the supernatant. This usually is Sampling can also be done 5 to 10 cm
done in 5 to 30 minutes, but may take below the surface to determine if slow • Measure water volume — consistently!
over an hour. Supernatant is then settling floc is a problem. This would be
withdrawn from the jars, and turbidity important in hydraulically overloaded • Inject into vortex — avoid splashing.
is measured in a turbidimeter. water treatment plants. In many plants,
• Treat a dosage RANGE to avoid a false
other parameters are important, such as
minimum.
pH or alkalinity, so these tests may also be
performed. • If a plant has a Jar Test Procedure, use
PRODUCT MAKE-UP it, especially to finalize product
Coagulants can be diluted for easy selection.
dosing; aqueous solutions concentra-
• Dosage range should start at 25%
tions are usually 1, 5, or 10%. PRODUCT PERFORMANCE increase to 150% of current program’s
Flocculants require lower solution COMPARISON dosage and adjusted accordingly.
concentration of about 0.1 percent to
Product performance is usually compared • Always benchmark versus the current
allow for adequate mixing in the jar
to the current customer treatment program.
test. Emulsion polymers must be
program. Settled solids volume, supernatant
properly inverted prior to dilution. • Coagulation curves initially decrease
turbidity and program cost (i.e., $/volume
Actual product dosages required turbidity, but overdosed coagulant
water treated) can be used to establish the
depend on the application. tends to resuspend particles. This is
cost effectiveness of various treatment
programs which meet customer perfor- especially true of organic coagulants.
mance requirements. Figure 1 is a typical dosage curve for an
TEMPERATURE organic coagulant. The second, ‘false’
Water temperatures will increase minima occurs presumably due to
rapidly in indoor jar tests and changing coagulant polymer forming multiple
water temperature may change test
RAW WATER
layers on a particle. Because of such
results. Use water of the same temper-
JAR TESTING TIPS potential behavior, dosage ranges should
ature as the plant raw water in the jar • Jar testing cannot simulate a sand filter be tested.
tests. To minimize any temperature or a sludge blanket!
increase, try turning off the lights • Try to treat every jar the same.
illuminating the jars until needed in
the settling tests.

NALCO COMPANY OPERATIONS


North America: 1601 West Diehl Road • Naperville, Illinois 60563-1198 • USA
Europe: Ir.G.Tjalmaweg 1 • 2342 BV Oegstgeest • The Netherlands
Asia Pacific: 2 International Business Park • #02-20 The Strategy Tower 2 • Singapore 609930
Latin America: Av. das Nações Unidas 17.891 • 6° Andar 04795-100 • São Paulo • SP • Brazil

www.nalco.com

ULTRION, NALCO and the logo are Registered Trademarks of Nalco Company
©1996, 2004 Nalco Company All Rights Reserved 6-04

You might also like