Professional Documents
Culture Documents
-Those that satisfy Class A ASTM precision criteria are stamped with the letter “A”
on the glassware and are preferred for laboratory applications.
5. When the bottom of the meniscus is even with the volume mark,
press your index finger firmly on the top of the pipet so no liquid leaks
out. Pull the pipet out of the solution and touch the tip once to the side of
the container.
6. To transfer the solution, place the tip of the pipet against the wall of
the receiving container at an angle of 10-20 degrees. Slowly allow the
liquid to drain from the pipet. Keep the flow slow so that no droplets
cling to the inside of the pipet.
7. When the solution stops flowing, touch the pipet once to the side of
the receiving container to remove any hanging drops. DO NOT blow out
the remaining solution. The pipet has been calibrated to deliver the
appropriate amount of solution with some remaining in the tip.
Automatic Pipettes
- Are pipetting devices that
permit rapid repetitive
measurement and delivery of
precise measures.
- It utilizes piston to generate
suction, which draws the sample
into a disposable tip
Fixed - vs Variable - volume pipette
● Fixed volume pipette - A pipette associated with only one volume
● In reverse mode pipetting, the purge stroke is used during preparation. During
aspiration, an amount of liquid equal to the amount of purged air is added. This
amount compensates for the liquid that remains inside the tip during dispensing.
Pipettes and Their Calibration
● Volumetric pipettes - are designed to deliver a single, specific volume of liquid
-They are referred to as TD (to deliver) devices.
- Liquid is drawn through the tip of the pipette, filling a bulb in the body of pipette, until the
liquid level reaches the calibration mark.
- The measured liquid is then drained for use. Volumetric pipettes are never to be blown out.
- It has a certificate of calibration from the National Institute of Standards and Technology
(NIST), stating that it is a Class A pipette.
Oswald-Folin pipettes
*** Pipette calibration can be performed using the colored reagents with
known absorbance at specific wavelengths, such as potassium dichromate
or p-nitrophenol****
Procedure
1. Prepare the reagents required
a. In a 100-ml volumetric flask, dissolve 105 mg of high purity p-nitrophenol (PNP) in
deionized water.
b. Fill in according to the specified level and thoroughly mix
c. Prepare various dilutions to the p-nitrophenol stock (PNP)
d. Fill in three 250 ml volumetric flasks halfway with 250 ml 0.01 mol/L NaOH
e. Using a different volumetric pipette, add 1 ml of stock PNP to each flask. Combine
thoroughly
● Pipette
● 10 to 20 pipette tips, if needed
● Balance capable of accuracy and resolution to ±0.1% of dispensed volumetric
weight
● Weighing vessel large enough to hold volume of liquid
● Type I/CLRW (Clinical Laboratory Reagent Water)
● Thermometer and barometer
Procedure
1. Record the weight of the vessel. Record the temperature of the water. It is recommended
that all materials be at room temperature. Obtain the barometric pressure.
2. Place a small volume (0.5 mL) of the water into the container. To prevent effects from
evaporation, it is desirable to loosely cover each container with a substance such as Parafilm.
Avoid handling of the containers.
3. Weigh each container plus water to the nearest 0.1 mg or set the balance to zero.
4. Using the pipette to be tested, draw up the specified amount. Carefully wipe the outside of
the tip. Care should be taken not to touch the end of the tip; this will cause liquid to be wicked
out of the tip, introducing an inaccuracy as a result of technique.
5. Dispense the water into the weighed vessel. Touch the tip to the side.
7. Subtract the weight obtained in step 3 from that obtained in step 6. Record the result.
8. If plastic tips are used, change the tip between each dispensing. Repeat steps 1 to 6 for a minimum of nine
additional times.
9. Accuracy or the ability of the pipette to dispense the Obtain the average or mean of the weight of the
water. Multiply the mean weight by the corresponding density of water at the given temperature and
pressure. At 20°C, the density of water is 0.9982.
10. Determine the accuracy or the ability of the pipette to dispense the expected (selected or stated) volume
according to the following formula:
● The manufacturer usually gives acceptable limitations for a particular pipette,
but they should not be used if the value differs by more than 1.0% from the
expected value. Precision can be indicated as the percent coefficient of
variation (%CV) or standard deviation (SD) for a series of repetitive pipetting
steps.
● Required imprecision is usually ±1 SD. The %CV will vary with the expected
volume of the pipette, but the smaller the %CV value, the greater the
precision. When n is large, the data are more statistically valid.