Professional Documents
Culture Documents
版本号:20161024
CONTENTS
01
PART ONE Monthly Maintenance 02
PART TWO Needle position
03
PART THREE Washer lift position 04
PART FOUR Chamber cleaning
05
PART FIVE Starter injector 06
PART SIX
pump volume
07
PART SEVEN
<Background> 08
PART EIGHT
<Plateau>
Note: When adjusting the needle position, it is required to confirm the position from the front and
the left or right side of the cuvette.
Fig. Second sampling after wash Fig right pipetting position (for
Maglumi 1000 and above)
Fig “Inject” position: all the inject needles shall be at the opening of the cuvette
•Before removal of
chamber cover, check the
“Background” in service
software making sure that
there is no light leakage.
Fig Make sure the starter tubes are 1-2 mm out of the starter
injector and stable during the movement of chamber lift. In
case of adjustment needed, adjust the screws marked 1 or/and 2
Fig Background
FOCUS MAKES PROFESSION P28
07
PART Seven
Content
Background
Steps Note
Step 1 Click <Initialization> in the main menu to
initialize the analyzer.
Step 2 Click the <Globals> — <background> to
open the “background” dialog.
Fig Plateau
FOCUS MAKES PROFESSION P31
08
PART Eight
Plateau
Content Steps Note
Step 1 Click <Globals>—<Plateau> to open the
“Plateau” dialog.
Step 2 Input “300000” in “Max RLU”, choose
“50000” in “Scal.”, and click the <Draw
1. Please ensure
diagr.> button. Tick “Reference LED on”.
that the
Step 3 Click the <Start> button to draw the D/A
“initialization” has
value curve. It should be smooth as shown in
been done in the
the figure below. If not, contact SNIBE after-
first place
sales specialist.
PMT 2. Choose the
Step 4 After getting the smooth curve, note down the
Voltage’s right value of the
“Suggest D/A highvolt”. Click <Start>
setting “D/A Ref Led” to
button for another 4 times to record the
make sure the
“Suggest D/A highvolt”. The difference
value of the “Ref.
between these 5 values shall be no more than
LED on” is
5. If not, adjust the value of the “D/A Ref
between 140,000
Led” and repeat Step 2 to Step 4.
and 230,000
Step 5 When we get all the values of “Suggest D/A
highvolt”. Input the smallest one in “D/A
highvolt” in the “Global” interface and click
“Writeparam”.
FOCUS MAKES PROFESSION P32
08
PART Eight
Plateau
Fig The curve shall be smooth and among the five times of tests,
“Suggest D/A Highvolt” shall not exceed 170.
FOCUS MAKES PROFESSION P33
Thanks
FOCUS MAKES PROFESSION