AerC-3

Auto Hematology Analyzer


Highlight Feature

6 Language 3 Calibration mothods:


Manual,Calibrator,Blood

Automatic re-counting 10.4 Inch touch screen

2 counting modes:Whole
Blood, Pre-diluted Blood
Lyse inside the analyzer,
high space utilization

RBC

PLT

AerC-3 Auto Hematology Analyzer is a clinical test instrument with rich functions and easy
operation. It can provide 21 parameters and 3 histograms. Double channel counting. During
the sample analysis process, automatic blockage removal and automatic re-counting ensure
accurate and stable test results. Widely used in hospital laboratories, clinical departments
and research centers.
Product Advantages

Advanved Parameters
21 Parameters, 3 Histograms
Possible choice of parameter units, Several languages
avaiable
WBC

RBC

PLT
0 50 100 150 200 250 300 fL 0 50 100 150 200 fL 0 5 10 15 25 fL

Block-clearing Abnormal histogram alarm


Soak and wash back and forth. real time monitoring If the histogram of the sample analysis result is
function of high pressure burning and automatic abnormal. The machine will automatically generate
block-clearing a histogram alarm. Prompt R1, R2, R3, R4, Rm, Pm
according to the type of exception

Rm Pm

RBC

PLT
0 50 100 150 200 fL 0 5 10 15 25 30 fL

High Accuracy Cost Efficiency


Double channel counting, no cross-contamination 2 reagent: Diluent and Lyse
between channels and low carryover rate Low reagent consumption
Automatically adjusting voltage of HGB blank, Auto-
Lyse inside the analyzer, high space utilization
matically adjust alarm thresholds of clogging and
bubble of counting apertures

RBC LED

PLT HGB

I II

Automatic re-counting Safety design


Automatically re-counting after unclogging treatment
Circuit and liquid separation, safety protection
while a counting hole is clogged, avoid re-counting
blood Convenient maintenance
Correlation

Comparison result with S brand

y=0.9819x + 3.2725
R²=0.9816
y=0.9924x + 0.0568 180
R²=0.9803
y=1.0094x - 0.0217 7
R²=0.998 160
14
6 140
12

5 120
10

100
8
4

80
6 3
60
4 2
40

2
1 20
0
0 0
0 2 4 6 8 10 12 14
0 1 2 3 4 5 6 7 0 20 40 60 80 100 120 140 160 180
WBC
RBC HGB

y=1.0047x - 0.1691 y=1.0146x - 1.8489


R²=0.9863 500 R²=0.9763
100
90 450

80 400
70 350

60 300

50 250

40 200
30 150
20 100
10 50
0
0
0 10 20 30 40 50 60 70 80 90 100
0 50 100 150 200 250 300 350 400 450 500

MCV
PLT

Routine reagent
Diluent 20 L

Lyse 500 mL

Maintenance reagent
Cleaner 50 mL
Product Specifications

Precision
Parameters Carry-over Linearity Range
Range CV
9
(7.0 ~ 15.0 )× 10 / L ≤2.0%
WBC ≤0.5% 9
9 (0.0 ~ 99.9) ×10 / L
(4.0 ~ 6.9) × 10 / L ≤3.5%
12 12
RBC (3.50 ~ 6.5)× 10 / L ≤1.5% ≤0.5% (0.0 ~ 7) ×10 / L

HGB 100 ~ 180 g/L ≤1.5% ≤0.5% (0.0 ~ 240) g/ L

MCV 70.0~ 110.0 fL ≤0.5% (0.0 ~ 250.0) fL


9
(150 ~ 500) × 10 / L ≤4.0%
PLT ≤1% 9
(0.0 ~ 999) × 10 / L
(100 ~ 149) × 10 9/ L ≤5.0%

Impedance method for cells counting


Principle
Cyanide free colorimetric method for HGB measurement

Counting Modes Whole Blood, Pre-diluted Blood

Throughput 60 samples/hour

Whloe Blood: 9.9μL


Sample Volume
Pre-diluted Blood: ≤20μL

21 Parameters: WBC, Lymph#, Mid#, Gran#, Lymph%, Mid%, Gran%, RBC, HGB,
MCV, MCH, MCHC, RDW-CV, RDW-SD, HCT, PLT, MPV, PDW, PCT, P-LCR, P-LCC
Parameters
3 Histogram: WBC, RBC, PLT

Reagent 2 reagents + 1 maintenance cleanser

10.4-inch touch screen


Input
Mouse, Keyboard

Reference value Settings Preset 5 sets of reference values: Man, Woman, Child, Baby and General

Calibration Manual, Calibrator, Fresh Blood

Data Ports 1 network port, 1 RS-232 port, 4 USB ports

Power AC110/115V or AC220/230V, 50Hz Input power:≤150VA

Operating Temperature 10℃~30℃

Relative Humidity ≤70%

Atmospheric Pressure 70kPa~106kPa


Website: www.aehealthgroup.com