1 Communication Code 2120: Module B1 lost communication 28/12/22 19:37:50 4.8 Lost while module was idle 2 Communication Code 2124: Module B1 communication 29/12/22 08:48:11 4.8 Restored restored 3 Communication Code 2120: Module B1 lost communication 30/12/22 05:35:42 4.8 Lost while module was idle 4 Communication Code 2124: Module B1 communication 30/12/22 09:06:01 4.8 Restored restored 5 Communication Code 2120: Module B1 lost communication 11/03/23 23:16:04 4.8 Lost while module was idle 6 Communication Code 2120: Module B1 lost communication 28/03/23 03:37:28 4.8 Lost while module was idle 7 Communication Code 2120: Module B1 lost communication 03/04/23 19:30:58 4.8 Lost while module was idle 8 Termination error Error 2005: Motion of the syringe drive was 10/07/23 15:04:41 4.8 not detected. Detected motion started at position 854 uL and transferred 745 uL at valve position 589 with pressure 68.1 PSI 9 Communication Code 2120: Module B1 lost communication 01/08/23 21:25:10 4.8 Lost while module was idle 10 Termination error Error 2025: The system failed to find the 14/08/23 15:39:19 4.8 plunger home position. Plunger moved down looking for ADC = 2500. ADC value -12 was detected and stall occurred
1 Communication Code 2120: Module B2 lost communication 28/12/22 19:37:44 4.8 Lost while module was idle 2 Communication Code 2124: Module B2 communication 29/12/22 08:48:12 4.8 Restored restored 3 Communication Code 2120: Module B2 lost communication 30/12/22 05:35:56 4.8 Lost while module was idle 4 Communication Code 2124: Module B2 communication 30/12/22 09:06:02 4.8 Restored restored 5 Termination error Error 2012: An inaccurate valve move to 11/01/23 15:18:44 4.8 position 589 was detected. The valve was detected to stop at position 13
6 Communication Code 2120: Module B2 lost communication 11/03/23 23:15:56 4.8
Lost while module was idle 7 Communication Code 2120: Module B2 lost communication 28/03/23 03:37:44 4.8 Lost while module was idle 8 Communication Code 2120: Module B2 lost communication 03/04/23 19:31:07 4.8 Lost while module was idle 9 Communication Code 2120: Module B2 lost communication 14/05/23 07:25:48 4.8 Lost while module was idle 10 Communication Code 2120: Module B2 lost communication 25/05/23 15:24:51 4.8 Lost while module was idle 11 Communication Code 2120: Module B2 lost communication 01/08/23 21:25:30 4.8 Lost while module was idle
Module Name Instrument S/N Module S/N
B3 810667 736754
# Description Detail Time Version
1 Communication Code 2120: Module B3 lost communication 11/03/23 23:16:05 4.8 Lost while module was idle 2 Communication Code 2120: Module B3 lost communication 28/03/23 03:37:37 4.8 Lost while module was idle 3 Communication Code 2120: Module B3 lost communication 03/04/23 19:31:06 4.8 Lost while module was idle 4 Communication Code 2120: Module B3 lost communication 05/05/23 19:35:06 4.8 Lost while module was idle 5 Communication Code 2120: Module B3 lost communication 14/05/23 07:26:02 4.8 Lost while module was idle
6 Communication Code 2120: Module B3 lost communication 25/05/23 15:24:55 4.8 Lost while module was idle 7 Communication Code 2120: Module B3 lost communication 01/08/23 21:25:25 4.8 Lost while module was idle 8 Communication Code 2120: Module B3 lost communication 06/08/23 09:41:30 4.8 Lost while module was idle
Module Name Instrument S/N Module S/N
B4 810667 677294
# Description Detail Time Version
1 Communication Code 2120: Module B4 lost communication 05/09/22 23:12:08 4.8 Lost while module was idle 2 Termination error Error 2008: Syringe pressure reading of 28/09/22 06:51:17 4.8 120.0 PSI exceeds the protocol limit of 120.0 PSI 3 Communication Code 2120: Module B4 lost communication 30/09/22 19:18:54 4.8 Lost while module was idle 4 Termination error Error 2037: The cartridge integrity test failed 17/10/22 13:33:13 4.8 at valve position 0. The pressure change of 0.3 PSI did not exceed the requirement of 4.0 PSI. The pressure increased from 5.7 PSI to 6.0 PSI during the test 5 Communication Code 2120: Module B4 lost communication 30/10/22 21:26:28 4.8 Lost while module was idle 6 Communication Code 2120: Module B4 lost communication 03/11/22 20:34:15 4.8 Lost while module was idle 7 Communication Code 2120: Module B4 lost communication 12/11/22 20:22:11 4.8 Lost while module was idle 8 Communication Code 2120: Module B4 lost communication 28/12/22 19:37:52 4.8 Lost while module was idle 9 Communication Code 2124: Module B4 communication 29/12/22 08:48:13 4.8 Restored restored 10 Communication Code 2120: Module B4 lost communication 30/12/22 05:36:04 4.8 Lost while module was idle 11 Communication Code 2124: Module B4 communication 30/12/22 09:06:03 4.8 Restored restored 12 Communication Code 2120: Module B4 lost communication 10/01/23 20:50:27 4.8 Lost while module was idle 13 Termination error Error 2012: An inaccurate valve move to 11/01/23 15:18:45 4.8
13 Termination error position 589 was detected. 11/01/23 15:18:45 4.8 The valve was detected to stop at position 8 14 Termination error Error 2005: Motion of the syringe drive was 15/02/23 10:02:16 4.8 not detected. Detected motion started at position 40 uL and transferred 98 uL at valve position 97 with pressure -3.2 PSI 15 Communication Code 2120: Module B4 lost communication 05/03/23 14:03:42 4.8 Lost while module was idle 16 Communication Code 2120: Module B4 lost communication 11/03/23 23:16:03 4.8 Lost while module was idle 17 Communication Code 2120: Module B4 lost communication 28/03/23 03:37:30 4.8 Lost while module was idle 18 Communication Code 2120: Module B4 lost communication 03/04/23 19:31:10 4.8 Lost while module was idle 19 Communication Code 2120: Module B4 lost communication 14/05/23 07:25:59 4.8 Lost while module was idle 20 Communication Code 2120: Module B4 lost communication 25/05/23 15:24:55 4.8 Lost while module was idle 21 Self-test error Error 4012: Heater B operation was not 26/07/23 14:40:51 4.8 verified. Measured temperature changed from 54.0 ºC to 73.0 ºC 22 Self-test error Error 4012: Heater B operation was not 27/07/23 15:49:52 4.8 verified. Measured temperature changed from 52.8 ºC to 70.7 ºC 23 Termination error Error 2005: Motion of the syringe drive was 31/07/23 12:52:05 4.8 not detected. Detected motion started at position 6 uL and transferred 5 uL at valve position 1180 with pressure -2.1 PSI 24 Self-test error Error 4012: Heater B operation was not 01/08/23 14:17:08 4.8 verified. Measured temperature changed from 49.8 ºC to 62.9 ºC 25 Communication Code 2120: Module B4 lost communication 01/08/23 21:25:39 4.8 Lost while module was idle 26 Communication Code 2120: Module B4 lost communication 06/08/23 09:41:30 4.8 Lost while module was idle 27 Run time error Error 1002: The temperature difference of 10/08/23 17:48:32 4.8 11.4 ºC exceeds the limit of 5.0 ºC. The temperatures for heaters A and B are 54.0
27 Run time error ºC and 65.4 ºC 10/08/23 17:48:32 4.8 28 Run time error Error 1002: The temperature difference of 10/08/23 18:08:52 4.8 7.1 ºC exceeds the limit of 5.0 ºC. The temperatures for heaters A and B are 71.1 ºC and 78.2 ºC 29 Run time error Error 1002: The temperature difference of 10/08/23 18:08:57 4.8 7.2 ºC exceeds the limit of 5.0 ºC. The temperatures for heaters A and B are 71.0 ºC and 78.2 ºC 30 Run time error Error 1002: The temperature difference of 10/08/23 18:09:07 4.8 7.3 ºC exceeds the limit of 5.0 ºC. The temperatures for heaters A and B are 71.0 ºC and 78.3 ºC 31 Run time error Error 1002: The temperature difference of 10/08/23 18:10:10 4.8 8.0 ºC exceeds the limit of 5.0 ºC. The temperatures for heaters A and B are 67.6 ºC and 75.6 ºC 32 Run time error Error 1002: The temperature difference of 10/08/23 18:10:46 4.8 8.0 ºC exceeds the limit of 5.0 ºC. The temperatures for heaters A and B are 67.6 ºC and 75.6 ºC 33 Run time error Error 1002: The temperature difference of 10/08/23 18:11:23 4.8 8.0 ºC exceeds the limit of 5.0 ºC. The temperatures for heaters A and B are 67.6 ºC and 75.6 ºC 34 Run time error Error 1002: The temperature difference of 10/08/23 18:12:00 4.8 8.0 ºC exceeds the limit of 5.0 ºC. The temperatures for heaters A and B are 67.6 ºC and 75.6 ºC 35 Run time error Error 1002: The temperature difference of 10/08/23 18:12:37 4.8 8.0 ºC exceeds the limit of 5.0 ºC. The temperatures for heaters A and B are 67.6 ºC and 75.6 ºC 36 Run time error Error 1002: The temperature difference of 10/08/23 18:13:15 4.8 7.8 ºC exceeds the limit of 5.0 ºC. The temperatures for heaters A and B are 67.7 ºC and 75.5 ºC 37 Run time error Error 1002: The temperature difference of 10/08/23 18:13:52 4.8 7.4 ºC exceeds the limit of 5.0 ºC. The temperatures for heaters A and B are 67.9 ºC and 75.3 ºC 38 Run time error Error 1002: The temperature difference of 10/08/23 18:14:29 4.8 7.1 ºC exceeds the limit of 5.0 ºC. The
38 Run time error temperatures for heaters A and B are 68.1 10/08/23 18:14:29 4.8 ºC and 75.2 ºC 39 Run time error Error 1002: The temperature difference of 10/08/23 18:15:06 4.8 7.0 ºC exceeds the limit of 5.0 ºC. The temperatures for heaters A and B are 68.1 ºC and 75.1 ºC 40 Run time error Error 1002: The temperature difference of 10/08/23 18:15:43 4.8 7.0 ºC exceeds the limit of 5.0 ºC. The temperatures for heaters A and B are 68.1 ºC and 75.1 ºC 41 Run time error Error 1002: The temperature difference of 10/08/23 18:16:20 4.8 7.3 ºC exceeds the limit of 5.0 ºC. The temperatures for heaters A and B are 68.0 ºC and 75.3 ºC 42 Run time error Error 1002: The temperature difference of 10/08/23 18:16:57 4.8 7.3 ºC exceeds the limit of 5.0 ºC. The temperatures for heaters A and B are 68.0 ºC and 75.3 ºC 43 Run time error Error 1002: The temperature difference of 10/08/23 18:17:34 4.8 7.2 ºC exceeds the limit of 5.0 ºC. The temperatures for heaters A and B are 68.0 ºC and 75.2 ºC 44 Run time error Error 1002: The temperature difference of 10/08/23 18:18:11 4.8 7.2 ºC exceeds the limit of 5.0 ºC. The temperatures for heaters A and B are 68.0 ºC and 75.2 ºC 45 Run time error Error 1002: The temperature difference of 10/08/23 18:18:48 4.8 7.1 ºC exceeds the limit of 5.0 ºC. The temperatures for heaters A and B are 68.1 ºC and 75.2 ºC 46 Run time error Error 1002: The temperature difference of 10/08/23 18:19:25 4.8 7.3 ºC exceeds the limit of 5.0 ºC. The temperatures for heaters A and B are 68.0 ºC and 75.3 ºC 47 Run time error Error 1002: The temperature difference of 10/08/23 18:20:02 4.8 7.0 ºC exceeds the limit of 5.0 ºC. The temperatures for heaters A and B are 68.1 ºC and 75.1 ºC 48 Run time error Error 1002: The temperature difference of 10/08/23 18:20:39 4.8 7.1 ºC exceeds the limit of 5.0 ºC. The temperatures for heaters A and B are 68.1 ºC and 75.2 ºC 49 Run time error Error 1002: The temperature difference of 10/08/23 18:21:16 4.8
49 Run time error 7.1 ºC exceeds the limit of 5.0 ºC. The 10/08/23 18:21:16 4.8 temperatures for heaters A and B are 68.1 ºC and 75.2 ºC 50 Run time error Error 1002: The temperature difference of 10/08/23 18:21:53 4.8 7.0 ºC exceeds the limit of 5.0 ºC. The temperatures for heaters A and B are 68.1 ºC and 75.1 ºC 51 Run time error Error 1002: The temperature difference of 10/08/23 18:22:30 4.8 7.0 ºC exceeds the limit of 5.0 ºC. The temperatures for heaters A and B are 68.1 ºC and 75.1 ºC 52 Run time error Error 1002: The temperature difference of 10/08/23 18:23:07 4.8 6.9 ºC exceeds the limit of 5.0 ºC. The temperatures for heaters A and B are 68.2 ºC and 75.1 ºC 53 Run time error Error 1002: The temperature difference of 10/08/23 18:23:44 4.8 7.1 ºC exceeds the limit of 5.0 ºC. The temperatures for heaters A and B are 68.1 ºC and 75.1 ºC 54 Run time error Error 1002: The temperature difference of 10/08/23 18:24:22 4.8 7.2 ºC exceeds the limit of 5.0 ºC. The temperatures for heaters A and B are 68.0 ºC and 75.2 ºC 55 Run time error Error 1002: The temperature difference of 10/08/23 18:24:59 4.8 6.9 ºC exceeds the limit of 5.0 ºC. The temperatures for heaters A and B are 68.1 ºC and 75.0 ºC 56 Run time error Error 1002: The temperature difference of 10/08/23 18:25:36 4.8 7.2 ºC exceeds the limit of 5.0 ºC. The temperatures for heaters A and B are 68.1 ºC and 75.2 ºC 57 Run time error Error 1002: The temperature difference of 10/08/23 18:26:13 4.8 7.2 ºC exceeds the limit of 5.0 ºC. The temperatures for heaters A and B are 68.0 ºC and 75.2 ºC 58 Run time error Error 1002: The temperature difference of 10/08/23 18:26:50 4.8 6.9 ºC exceeds the limit of 5.0 ºC. The temperatures for heaters A and B are 68.2 ºC and 75.1 ºC 59 Run time error Error 1002: The temperature difference of 10/08/23 18:27:27 4.8 6.9 ºC exceeds the limit of 5.0 ºC. The temperatures for heaters A and B are 68.2 ºC and 75.1 ºC
60 Run time error Error 1002: The temperature difference of 10/08/23 18:28:04 4.8 7.0 ºC exceeds the limit of 5.0 ºC. The temperatures for heaters A and B are 68.2 ºC and 75.1 ºC 61 Run time error Error 1002: The temperature difference of 10/08/23 18:28:41 4.8 6.9 ºC exceeds the limit of 5.0 ºC. The temperatures for heaters A and B are 68.2 ºC and 75.1 ºC 62 Run time error Error 1002: The temperature difference of 10/08/23 18:32:07 4.8 15.1 ºC exceeds the limit of 5.0 ºC. The temperatures for heaters A and B are 42.2 ºC and 57.3 ºC 63 Run time error Error 1002: The temperature difference of 10/08/23 18:32:53 4.8 9.7 ºC exceeds the limit of 5.0 ºC. The temperatures for heaters A and B are 59.0 ºC and 68.7 ºC 64 Run time error Error 1002: The temperature difference of 10/08/23 18:33:04 4.8 7.2 ºC exceeds the limit of 5.0 ºC. The temperatures for heaters A and B are 70.0 ºC and 77.2 ºC 65 Run time error Error 1002: The temperature difference of 10/08/23 18:33:20 4.8 9.8 ºC exceeds the limit of 5.0 ºC. The temperatures for heaters A and B are 58.8 ºC and 68.6 ºC 66 Run time error Error 1001: The actual temperature 95.3 ºC 10/08/23 18:34:00 4.8 has drifted too far away from the setpoint of 97.0 ºC 67 Self-test error Error 4012: Heater B operation was not 14/08/23 13:44:42 4.8 verified. Measured temperature changed from 56.0 ºC to 75.0 ºC 68 Run time error Error 1002: The temperature difference of 14/08/23 14:20:33 4.8 14.0 ºC exceeds the limit of 5.0 ºC. The temperatures for heaters A and B are 54.2 ºC and 68.2 ºC 69 Run time error Error 1001: The actual temperature 61.7 ºC 14/08/23 14:20:33 4.8 has drifted too far away from the setpoint of 60.0 ºC 70 Run time error Error 1002: The temperature difference of 14/08/23 14:40:53 4.8 5.5 ºC exceeds the limit of 5.0 ºC. The temperatures for heaters A and B are 71.9 ºC and 77.3 ºC
71 Run time error Error 1001: The actual temperature 72.4 ºC 14/08/23 14:40:56 4.8 has drifted too far away from the setpoint of 75.0 ºC 72 Run time error Error 1002: The temperature difference of 19/08/23 16:07:29 4.8 6.3 ºC exceeds the limit of 5.0 ºC. The temperatures for heaters A and B are 56.6 ºC and 62.9 ºC 73 Self-test error Error 4012: Heater B operation was not 25/08/23 14:48:37 4.8 verified. Measured temperature changed from 46.9 ºC to 65.9 ºC 74 Self-test error Error 4012: Heater B operation was not 28/08/23 12:34:46 4.8 verified. Measured temperature changed from 56.0 ºC to 74.2 ºC 75 Self-test error Error 4012: Heater B operation was not 28/08/23 12:36:53 4.8 verified. Measured temperature changed from 61.1 ºC to 79.1 ºC 76 Self-test error Error 4012: Heater B operation was not 02/09/23 08:11:59 4.8 verified. Measured temperature changed from 49.5 ºC to 68.5 ºC
If there is an issue with an instrument, contact Technical Support.