810667 RSUD Kab Lombok Utara-Lombok-NTB 04/09/23 09:43:52

System Log Report

- Selection Criteria -
Date Range: All

Modules: Currently Connected Modules


Module B1,B2,B3,B4.

Show: Errors Only

User: laboratorium

Module Name Instrument S/N Module S/N


B1 810667 210079174

# Description Detail Time Version


1 Communication Code 2120: Module B1 lost communication 28/12/22 19:37:50 4.8
Lost while module was idle
2 Communication Code 2124: Module B1 communication 29/12/22 08:48:11 4.8
Restored restored
3 Communication Code 2120: Module B1 lost communication 30/12/22 05:35:42 4.8
Lost while module was idle
4 Communication Code 2124: Module B1 communication 30/12/22 09:06:01 4.8
Restored restored
5 Communication Code 2120: Module B1 lost communication 11/03/23 23:16:04 4.8
Lost while module was idle
6 Communication Code 2120: Module B1 lost communication 28/03/23 03:37:28 4.8
Lost while module was idle
7 Communication Code 2120: Module B1 lost communication 03/04/23 19:30:58 4.8
Lost while module was idle
8 Termination error Error 2005: Motion of the syringe drive was 10/07/23 15:04:41 4.8
not detected.
Detected motion started at position 854 uL
and transferred 745 uL at valve position 589
with pressure 68.1 PSI
9 Communication Code 2120: Module B1 lost communication 01/08/23 21:25:10 4.8
Lost while module was idle
10 Termination error Error 2025: The system failed to find the 14/08/23 15:39:19 4.8
plunger home position.
Plunger moved down looking for ADC =
2500. ADC value -12 was detected and stall
occurred

GeneXpert® Dx System Version 4.8 Page 1 of 9


810667 RSUD Kab Lombok Utara-Lombok-NTB 04/09/23 09:43:52

System Log Report

Module Name Instrument S/N Module S/N


B2 810667 210079320

# Description Detail Time Version


1 Communication Code 2120: Module B2 lost communication 28/12/22 19:37:44 4.8
Lost while module was idle
2 Communication Code 2124: Module B2 communication 29/12/22 08:48:12 4.8
Restored restored
3 Communication Code 2120: Module B2 lost communication 30/12/22 05:35:56 4.8
Lost while module was idle
4 Communication Code 2124: Module B2 communication 30/12/22 09:06:02 4.8
Restored restored
5 Termination error Error 2012: An inaccurate valve move to 11/01/23 15:18:44 4.8
position 589 was detected.
The valve was detected to stop at position 13

6 Communication Code 2120: Module B2 lost communication 11/03/23 23:15:56 4.8


Lost while module was idle
7 Communication Code 2120: Module B2 lost communication 28/03/23 03:37:44 4.8
Lost while module was idle
8 Communication Code 2120: Module B2 lost communication 03/04/23 19:31:07 4.8
Lost while module was idle
9 Communication Code 2120: Module B2 lost communication 14/05/23 07:25:48 4.8
Lost while module was idle
10 Communication Code 2120: Module B2 lost communication 25/05/23 15:24:51 4.8
Lost while module was idle
11 Communication Code 2120: Module B2 lost communication 01/08/23 21:25:30 4.8
Lost while module was idle

Module Name Instrument S/N Module S/N


B3 810667 736754

# Description Detail Time Version


1 Communication Code 2120: Module B3 lost communication 11/03/23 23:16:05 4.8
Lost while module was idle
2 Communication Code 2120: Module B3 lost communication 28/03/23 03:37:37 4.8
Lost while module was idle
3 Communication Code 2120: Module B3 lost communication 03/04/23 19:31:06 4.8
Lost while module was idle
4 Communication Code 2120: Module B3 lost communication 05/05/23 19:35:06 4.8
Lost while module was idle
5 Communication Code 2120: Module B3 lost communication 14/05/23 07:26:02 4.8
Lost while module was idle

GeneXpert® Dx System Version 4.8 Page 2 of 9


810667 RSUD Kab Lombok Utara-Lombok-NTB 04/09/23 09:43:52

System Log Report

# Description Detail Time Version


6 Communication Code 2120: Module B3 lost communication 25/05/23 15:24:55 4.8
Lost while module was idle
7 Communication Code 2120: Module B3 lost communication 01/08/23 21:25:25 4.8
Lost while module was idle
8 Communication Code 2120: Module B3 lost communication 06/08/23 09:41:30 4.8
Lost while module was idle

Module Name Instrument S/N Module S/N


B4 810667 677294

# Description Detail Time Version


1 Communication Code 2120: Module B4 lost communication 05/09/22 23:12:08 4.8
Lost while module was idle
2 Termination error Error 2008: Syringe pressure reading of 28/09/22 06:51:17 4.8
120.0 PSI exceeds the protocol limit of 120.0
PSI
3 Communication Code 2120: Module B4 lost communication 30/09/22 19:18:54 4.8
Lost while module was idle
4 Termination error Error 2037: The cartridge integrity test failed 17/10/22 13:33:13 4.8
at valve position 0.
The pressure change of 0.3 PSI did not
exceed the requirement of 4.0 PSI.
The pressure increased from 5.7 PSI to 6.0
PSI during the test
5 Communication Code 2120: Module B4 lost communication 30/10/22 21:26:28 4.8
Lost while module was idle
6 Communication Code 2120: Module B4 lost communication 03/11/22 20:34:15 4.8
Lost while module was idle
7 Communication Code 2120: Module B4 lost communication 12/11/22 20:22:11 4.8
Lost while module was idle
8 Communication Code 2120: Module B4 lost communication 28/12/22 19:37:52 4.8
Lost while module was idle
9 Communication Code 2124: Module B4 communication 29/12/22 08:48:13 4.8
Restored restored
10 Communication Code 2120: Module B4 lost communication 30/12/22 05:36:04 4.8
Lost while module was idle
11 Communication Code 2124: Module B4 communication 30/12/22 09:06:03 4.8
Restored restored
12 Communication Code 2120: Module B4 lost communication 10/01/23 20:50:27 4.8
Lost while module was idle
13 Termination error Error 2012: An inaccurate valve move to 11/01/23 15:18:45 4.8

GeneXpert® Dx System Version 4.8 Page 3 of 9


810667 RSUD Kab Lombok Utara-Lombok-NTB 04/09/23 09:43:52

System Log Report

# Description Detail Time Version


13 Termination error position 589 was detected. 11/01/23 15:18:45 4.8
The valve was detected to stop at position 8
14 Termination error Error 2005: Motion of the syringe drive was 15/02/23 10:02:16 4.8
not detected.
Detected motion started at position 40 uL
and transferred 98 uL at valve position 97
with pressure -3.2 PSI
15 Communication Code 2120: Module B4 lost communication 05/03/23 14:03:42 4.8
Lost while module was idle
16 Communication Code 2120: Module B4 lost communication 11/03/23 23:16:03 4.8
Lost while module was idle
17 Communication Code 2120: Module B4 lost communication 28/03/23 03:37:30 4.8
Lost while module was idle
18 Communication Code 2120: Module B4 lost communication 03/04/23 19:31:10 4.8
Lost while module was idle
19 Communication Code 2120: Module B4 lost communication 14/05/23 07:25:59 4.8
Lost while module was idle
20 Communication Code 2120: Module B4 lost communication 25/05/23 15:24:55 4.8
Lost while module was idle
21 Self-test error Error 4012: Heater B operation was not 26/07/23 14:40:51 4.8
verified.
Measured temperature changed from 54.0
ºC to 73.0 ºC
22 Self-test error Error 4012: Heater B operation was not 27/07/23 15:49:52 4.8
verified.
Measured temperature changed from 52.8
ºC to 70.7 ºC
23 Termination error Error 2005: Motion of the syringe drive was 31/07/23 12:52:05 4.8
not detected.
Detected motion started at position 6 uL and
transferred 5 uL at valve position 1180 with
pressure -2.1 PSI
24 Self-test error Error 4012: Heater B operation was not 01/08/23 14:17:08 4.8
verified.
Measured temperature changed from 49.8
ºC to 62.9 ºC
25 Communication Code 2120: Module B4 lost communication 01/08/23 21:25:39 4.8
Lost while module was idle
26 Communication Code 2120: Module B4 lost communication 06/08/23 09:41:30 4.8
Lost while module was idle
27 Run time error Error 1002: The temperature difference of 10/08/23 17:48:32 4.8
11.4 ºC exceeds the limit of 5.0 ºC. The
temperatures for heaters A and B are 54.0

GeneXpert® Dx System Version 4.8 Page 4 of 9


810667 RSUD Kab Lombok Utara-Lombok-NTB 04/09/23 09:43:52

System Log Report

# Description Detail Time Version


27 Run time error ºC and 65.4 ºC 10/08/23 17:48:32 4.8
28 Run time error Error 1002: The temperature difference of 10/08/23 18:08:52 4.8
7.1 ºC exceeds the limit of 5.0 ºC. The
temperatures for heaters A and B are 71.1
ºC and 78.2 ºC
29 Run time error Error 1002: The temperature difference of 10/08/23 18:08:57 4.8
7.2 ºC exceeds the limit of 5.0 ºC. The
temperatures for heaters A and B are 71.0
ºC and 78.2 ºC
30 Run time error Error 1002: The temperature difference of 10/08/23 18:09:07 4.8
7.3 ºC exceeds the limit of 5.0 ºC. The
temperatures for heaters A and B are 71.0
ºC and 78.3 ºC
31 Run time error Error 1002: The temperature difference of 10/08/23 18:10:10 4.8
8.0 ºC exceeds the limit of 5.0 ºC. The
temperatures for heaters A and B are 67.6
ºC and 75.6 ºC
32 Run time error Error 1002: The temperature difference of 10/08/23 18:10:46 4.8
8.0 ºC exceeds the limit of 5.0 ºC. The
temperatures for heaters A and B are 67.6
ºC and 75.6 ºC
33 Run time error Error 1002: The temperature difference of 10/08/23 18:11:23 4.8
8.0 ºC exceeds the limit of 5.0 ºC. The
temperatures for heaters A and B are 67.6
ºC and 75.6 ºC
34 Run time error Error 1002: The temperature difference of 10/08/23 18:12:00 4.8
8.0 ºC exceeds the limit of 5.0 ºC. The
temperatures for heaters A and B are 67.6
ºC and 75.6 ºC
35 Run time error Error 1002: The temperature difference of 10/08/23 18:12:37 4.8
8.0 ºC exceeds the limit of 5.0 ºC. The
temperatures for heaters A and B are 67.6
ºC and 75.6 ºC
36 Run time error Error 1002: The temperature difference of 10/08/23 18:13:15 4.8
7.8 ºC exceeds the limit of 5.0 ºC. The
temperatures for heaters A and B are 67.7
ºC and 75.5 ºC
37 Run time error Error 1002: The temperature difference of 10/08/23 18:13:52 4.8
7.4 ºC exceeds the limit of 5.0 ºC. The
temperatures for heaters A and B are 67.9
ºC and 75.3 ºC
38 Run time error Error 1002: The temperature difference of 10/08/23 18:14:29 4.8
7.1 ºC exceeds the limit of 5.0 ºC. The

GeneXpert® Dx System Version 4.8 Page 5 of 9


810667 RSUD Kab Lombok Utara-Lombok-NTB 04/09/23 09:43:52

System Log Report

# Description Detail Time Version


38 Run time error temperatures for heaters A and B are 68.1 10/08/23 18:14:29 4.8
ºC and 75.2 ºC
39 Run time error Error 1002: The temperature difference of 10/08/23 18:15:06 4.8
7.0 ºC exceeds the limit of 5.0 ºC. The
temperatures for heaters A and B are 68.1
ºC and 75.1 ºC
40 Run time error Error 1002: The temperature difference of 10/08/23 18:15:43 4.8
7.0 ºC exceeds the limit of 5.0 ºC. The
temperatures for heaters A and B are 68.1
ºC and 75.1 ºC
41 Run time error Error 1002: The temperature difference of 10/08/23 18:16:20 4.8
7.3 ºC exceeds the limit of 5.0 ºC. The
temperatures for heaters A and B are 68.0
ºC and 75.3 ºC
42 Run time error Error 1002: The temperature difference of 10/08/23 18:16:57 4.8
7.3 ºC exceeds the limit of 5.0 ºC. The
temperatures for heaters A and B are 68.0
ºC and 75.3 ºC
43 Run time error Error 1002: The temperature difference of 10/08/23 18:17:34 4.8
7.2 ºC exceeds the limit of 5.0 ºC. The
temperatures for heaters A and B are 68.0
ºC and 75.2 ºC
44 Run time error Error 1002: The temperature difference of 10/08/23 18:18:11 4.8
7.2 ºC exceeds the limit of 5.0 ºC. The
temperatures for heaters A and B are 68.0
ºC and 75.2 ºC
45 Run time error Error 1002: The temperature difference of 10/08/23 18:18:48 4.8
7.1 ºC exceeds the limit of 5.0 ºC. The
temperatures for heaters A and B are 68.1
ºC and 75.2 ºC
46 Run time error Error 1002: The temperature difference of 10/08/23 18:19:25 4.8
7.3 ºC exceeds the limit of 5.0 ºC. The
temperatures for heaters A and B are 68.0
ºC and 75.3 ºC
47 Run time error Error 1002: The temperature difference of 10/08/23 18:20:02 4.8
7.0 ºC exceeds the limit of 5.0 ºC. The
temperatures for heaters A and B are 68.1
ºC and 75.1 ºC
48 Run time error Error 1002: The temperature difference of 10/08/23 18:20:39 4.8
7.1 ºC exceeds the limit of 5.0 ºC. The
temperatures for heaters A and B are 68.1
ºC and 75.2 ºC
49 Run time error Error 1002: The temperature difference of 10/08/23 18:21:16 4.8

GeneXpert® Dx System Version 4.8 Page 6 of 9


810667 RSUD Kab Lombok Utara-Lombok-NTB 04/09/23 09:43:52

System Log Report

# Description Detail Time Version


49 Run time error 7.1 ºC exceeds the limit of 5.0 ºC. The 10/08/23 18:21:16 4.8
temperatures for heaters A and B are 68.1
ºC and 75.2 ºC
50 Run time error Error 1002: The temperature difference of 10/08/23 18:21:53 4.8
7.0 ºC exceeds the limit of 5.0 ºC. The
temperatures for heaters A and B are 68.1
ºC and 75.1 ºC
51 Run time error Error 1002: The temperature difference of 10/08/23 18:22:30 4.8
7.0 ºC exceeds the limit of 5.0 ºC. The
temperatures for heaters A and B are 68.1
ºC and 75.1 ºC
52 Run time error Error 1002: The temperature difference of 10/08/23 18:23:07 4.8
6.9 ºC exceeds the limit of 5.0 ºC. The
temperatures for heaters A and B are 68.2
ºC and 75.1 ºC
53 Run time error Error 1002: The temperature difference of 10/08/23 18:23:44 4.8
7.1 ºC exceeds the limit of 5.0 ºC. The
temperatures for heaters A and B are 68.1
ºC and 75.1 ºC
54 Run time error Error 1002: The temperature difference of 10/08/23 18:24:22 4.8
7.2 ºC exceeds the limit of 5.0 ºC. The
temperatures for heaters A and B are 68.0
ºC and 75.2 ºC
55 Run time error Error 1002: The temperature difference of 10/08/23 18:24:59 4.8
6.9 ºC exceeds the limit of 5.0 ºC. The
temperatures for heaters A and B are 68.1
ºC and 75.0 ºC
56 Run time error Error 1002: The temperature difference of 10/08/23 18:25:36 4.8
7.2 ºC exceeds the limit of 5.0 ºC. The
temperatures for heaters A and B are 68.1
ºC and 75.2 ºC
57 Run time error Error 1002: The temperature difference of 10/08/23 18:26:13 4.8
7.2 ºC exceeds the limit of 5.0 ºC. The
temperatures for heaters A and B are 68.0
ºC and 75.2 ºC
58 Run time error Error 1002: The temperature difference of 10/08/23 18:26:50 4.8
6.9 ºC exceeds the limit of 5.0 ºC. The
temperatures for heaters A and B are 68.2
ºC and 75.1 ºC
59 Run time error Error 1002: The temperature difference of 10/08/23 18:27:27 4.8
6.9 ºC exceeds the limit of 5.0 ºC. The
temperatures for heaters A and B are 68.2
ºC and 75.1 ºC

GeneXpert® Dx System Version 4.8 Page 7 of 9


810667 RSUD Kab Lombok Utara-Lombok-NTB 04/09/23 09:43:52

System Log Report

# Description Detail Time Version


60 Run time error Error 1002: The temperature difference of 10/08/23 18:28:04 4.8
7.0 ºC exceeds the limit of 5.0 ºC. The
temperatures for heaters A and B are 68.2
ºC and 75.1 ºC
61 Run time error Error 1002: The temperature difference of 10/08/23 18:28:41 4.8
6.9 ºC exceeds the limit of 5.0 ºC. The
temperatures for heaters A and B are 68.2
ºC and 75.1 ºC
62 Run time error Error 1002: The temperature difference of 10/08/23 18:32:07 4.8
15.1 ºC exceeds the limit of 5.0 ºC. The
temperatures for heaters A and B are 42.2
ºC and 57.3 ºC
63 Run time error Error 1002: The temperature difference of 10/08/23 18:32:53 4.8
9.7 ºC exceeds the limit of 5.0 ºC. The
temperatures for heaters A and B are 59.0
ºC and 68.7 ºC
64 Run time error Error 1002: The temperature difference of 10/08/23 18:33:04 4.8
7.2 ºC exceeds the limit of 5.0 ºC. The
temperatures for heaters A and B are 70.0
ºC and 77.2 ºC
65 Run time error Error 1002: The temperature difference of 10/08/23 18:33:20 4.8
9.8 ºC exceeds the limit of 5.0 ºC. The
temperatures for heaters A and B are 58.8
ºC and 68.6 ºC
66 Run time error Error 1001: The actual temperature 95.3 ºC 10/08/23 18:34:00 4.8
has drifted too far away from the setpoint of
97.0 ºC
67 Self-test error Error 4012: Heater B operation was not 14/08/23 13:44:42 4.8
verified.
Measured temperature changed from 56.0
ºC to 75.0 ºC
68 Run time error Error 1002: The temperature difference of 14/08/23 14:20:33 4.8
14.0 ºC exceeds the limit of 5.0 ºC. The
temperatures for heaters A and B are 54.2
ºC and 68.2 ºC
69 Run time error Error 1001: The actual temperature 61.7 ºC 14/08/23 14:20:33 4.8
has drifted too far away from the setpoint of
60.0 ºC
70 Run time error Error 1002: The temperature difference of 14/08/23 14:40:53 4.8
5.5 ºC exceeds the limit of 5.0 ºC. The
temperatures for heaters A and B are 71.9
ºC and 77.3 ºC

GeneXpert® Dx System Version 4.8 Page 8 of 9


810667 RSUD Kab Lombok Utara-Lombok-NTB 04/09/23 09:43:52

System Log Report

# Description Detail Time Version


71 Run time error Error 1001: The actual temperature 72.4 ºC 14/08/23 14:40:56 4.8
has drifted too far away from the setpoint of
75.0 ºC
72 Run time error Error 1002: The temperature difference of 19/08/23 16:07:29 4.8
6.3 ºC exceeds the limit of 5.0 ºC. The
temperatures for heaters A and B are 56.6
ºC and 62.9 ºC
73 Self-test error Error 4012: Heater B operation was not 25/08/23 14:48:37 4.8
verified.
Measured temperature changed from 46.9
ºC to 65.9 ºC
74 Self-test error Error 4012: Heater B operation was not 28/08/23 12:34:46 4.8
verified.
Measured temperature changed from 56.0
ºC to 74.2 ºC
75 Self-test error Error 4012: Heater B operation was not 28/08/23 12:36:53 4.8
verified.
Measured temperature changed from 61.1
ºC to 79.1 ºC
76 Self-test error Error 4012: Heater B operation was not 02/09/23 08:11:59 4.8
verified.
Measured temperature changed from 49.5
ºC to 68.5 ºC

If there is an issue with an instrument, contact Technical Support.

GeneXpert® Dx System Version 4.8 Page 9 of 9