Padcal Assay Laboratory Balance Verification Version. No./Date 01/9-1-2018
PHILEX MINING CORPORATION Log Sheet Prepared by E. A. PEROCILLO
Padcal, Tuba, Benguet Approved by V. J. ANDALESIO

MONITORING LOG SHEET


(Balances)

The following procedure should be adopted and data recorded as appropriate.

- Use a fresh line for each item, use an ink colour other than red to record data.
- Perform the weight verification before the use of balances.
- When the balance is not in use, leave it at standby mode. Do not unplug the
balances.
- Newly plugged/turned on balance should be warmed up by at least 30 minutes
before use.
- Perform internal balance calibration if the unit is newly installed or turned on.
- Zero/tare the balance before the measurements.
- Place the calibrated weight on the pan - allow the balance to stabilize and record
the weight.
- Repeat using a second calibration weight - of different mass.

Acceptable Tolerances

Calibration weight 5 decimal 4 decimal 3 decimal 2 decimal 1 decimal

1 to 200mg +/- 0.005mg +/- 0.05mg - - -


250 to 500mg +/- 0.01mg +/- 0.1mg - - -
0.1 to 1.0g +/- 0.00010 +/- 0.0010 +/- 0.003 +/- 0.02 -
2.0 to 10.0g +/- 0.00015 +/- 0.0015 +/- 0.005 +/- 0.05 +/- 0.1
20.0 to 100.0g +/- 0.00020 +/- 0.0020 +/- 0.010 +/- 0.10 +/- 0.5
0.5 kg to 20kg +/- 0.00010 +/- 0.0010 +/- 0.005 +/- 0.10 +/- 0.5

- Sign for your work - legibly.


- When completed, this worksheet will be stored for archiving. New worksheets will
be used and should be kept till needed.

(Note: Errors should be crossed out and initiated with the date.)

- If the calibration weight happens to fall outside the tolerance, then;


i) Check the balance level
ii) Remove any residual dust from around the plate
iii) Contact the supervisor or Laboratory Manager

Note : The balance should not be used if the check indicates the calibration as being out.

Balance No. BAL-008 Acceptable Tolerances: + 0.0490 mg


For Assay Department use only
Reference No. FORM-LAB-036
Padcal Assay Laboratory Balance Verification Version. No./Date 01/9-1-2018
PHILEX MINING CORPORATION Log Sheet Prepared by E. A. PEROCILLO
Padcal, Tuba, Benguet Approved by V. J. ANDALESIO

Test weight: 1.0000 m grams Range: 0.9510 to 1.0490 mg


Test weight: 2.0000 m grams Range: 1.9510 to 2.0490 mg

Weight 1, Weight 2,
Date Verified By Remarks
(mg) (mg)
01 OCT. 2021
02 OCT. 2021
03 OCT.2021
04 OCT.2021
05 OCT.2021
06 OCT.2021
07 OCT. 2021
08 OCT. 2021
09 OCT.2021
10 OCT. 2021
11 OCT. 2021
12 OCT.2021
13 OCT. 2021
14 OCT. 2021
15 OCT 2021
16 OCT.T 2021
17 OCT. 2021
18 OCT. 2021
19 OCT. 2021
20 OCT. 2021
21 OCT. 2021
22 OCT. 2021
23 OCT.2021
24 OCT.T021
25 OCT.2021
26 OCT. 2021
27 OCT. 2021
28 OCT. 2021
29 OCT. 2021
30 OCT. 2021
31 OCT.2021