revision date: February 27 2004 expiry date: February 26, 2008

BALAI BESAR PENELITIAN DAN PENGEMBANGAN BAHAN DAN BARANG TEKNIK (B4T)
Accreditation Number : LK - 022 - IDN
Address : Jl. Sangkuriang No, 14 - Bandung
Telephone : 022 - 250 2027
Facsimile : 022 - 250 2027
Contact Person : Ir. Budi Susanto; Drs. Tatto Bustomi
Temperature
Best
No. Instrument to be calibrated Measurement Range Measurement Methods/Specifications Notes
Capability *)
1 Liquid in glass thermomter -25 °C ~ 150 °C 0.07 °C ASTM E77-1998

2 Temperature sensor with display unit -25 °C ~ 145 °C 0.4 °C JIS Z 8710 - 1993

3 Temperature indicator without sensor -150 °C ~ 2000 °C 0.25 °C simulasi

4 Dial type-expansion thermometer -25 °C ~ 145 °C 0.4 °C BS 5235: 1992


5 Temperature enclosure
oven, incubator, waterbath 25 °C ~ 400 °C 2.5 °C
furnace 25 °C ~ 1200 °C 2.5 °C AS 2853: 1986
Mass
Best
No. Instrument to be calibrated Measurement Range Measurement Methods/Specifications Notes
Capability *)
1 Weights 0.001 g ~ 0.5 g 0.07 mg The Calibration
(conventional mass) 1g ~ 5g 0.08 mg of Balance
10 g 0.09 mg CSIRO-NML: 1995
20 g 0.11 mg
50 g 0.12 mg
100 g 0.17 mg
200 g 0.31 mg
500 g 5 mg
1 kg 5 mg
2 kg 9 mg
5 kg 23 mg
10 kg 46 mg
20 kg 93 mg
2 Balance 0g ~ 100 g 0.13 mg The Calibration
(electronic, mechanic) 100 g ~ 200 g 0.24 mg of Balance
200 g ~ 500 g 0.84 mg CSIRO-NML: 1995
0.5 kg ~ 1 kg 1.3 mg
1 kg ~ 2 kg 7.0 mg
2 kg ~ 5 kg 11 mg
5 kg ~ 10 kg 16 mg
10 kg ~ 20 kg 88 mg
20 kg ~ 50 kg 135 mg
50 kg ~ 100 kg 300 mg
100 kg ~ 150 kg 1g
Volume
Best
No. Instrument to be calibrated Measurement Range Measurement Methods/Specifications Notes
Capability *)
1 Volumetric measures; 1 ml ~ 25 ml 0.02 ml
Volumetric Glassware 25 ml ~ 50 ml 0.04 ml
50 ml ~ 100 ml 0.08 ml
100 ml ~ 250 ml 0.1 ml ASTM E 542: 1999

1 of 3
Volume (cont'd) LK - 022 - IDN
Best
No. Instrument to be calibrated Measurement Range Measurement Methods/Specifications Notes
Capability *)
1 Volumetric measures; 250 ml ~ 500 ml 0.2 ml ISO 4787: 1984
Volumetric Glassware 500 ml ~ 1000 ml 0.3 ml
(cont'd) 1000 ml ~ 2000 ml 0.5 ml
Pressure
Best
No. Instrument to be calibrated Measurement Range Measurement Methods/Specifications Notes
Capability *)
Pressure gauge;
1 1 bar ~ 120 bar 0.2 % BS EN 837-1: 1998
Pressure recorder (hydraulic)
5
( 1 bar = 10 Pa )
Force
Best
No. Instrument to be calibrated Measurement Range Measurement Methods/Specifications Notes
Capability *)
1 Tensile testing machine 50 kN ~ 1000 kN 0.2 %
2 Compression testing machine 50 kN ~ 3000 kN 0.2 % ISO 7500-1: 1999
3 Universal testing machine 50 kN ~ 1000 kN 0.2 %
4 Impact testing machine (charpy) 15 joule ~ 300 joule 0.2 % BS EN 10045-2:1993
5 Proving Ring 50 kN ~ 1000 kN 0.2 % BS 1377-1: 1990
6 Hydraulic Jack (compression) 50 kN ~ 3000 kN 0.35 % BS 1377-1: 1990
7 Load Cell (compression) 50 kN ~ 3000 kN 0.2 % BS 1377-1: 1990
8 Concrete Test Hammer 0.5 kg/mm² ~ 500 kg/mm² 2.1 % Operation Manual

Hardness
Best
No. Instrument to be calibrated Measurement Range Measurement Methods/Specifications Notes
Capability *)
1 Rockwell Hardness Tester 22.5 HRB ~ 81.6 HRB 1.3 %
21.6 HRC ~ 60.6 HRC 1.3 %
76.4 HR15T ~ 86.3 HR15T 1.3 %
52.4 HR30T ~ 75.1 HR30T 1.3 % ISO 6508-2: 19999
21.6 HR45T ~ 64.2 HR45T 1.3 %
69.2 HR15N ~ 90.1 HR15N 1.3 %
50 HR30N ~ 73.8 HR30N 1.3 %
20 HR45N ~ 66.8 HR45N 1.3 %
2 Brinel Hardness Tester 216 HBW ~ 624 HBW 1.7 %
10 / 300
97.5 HBW 1.7 %
10 / 500 ISO 6505-2: 1999
219 HBS ~ 358.5 HBS 1.7 % ISO 6506-1: 1999
10 / 500
96 HBS ~ 97.5 HBS 1.7 %
10 / 500
4 Vickers Hardness Tester 101.9 HV0.3 ~ 805.5 HV0.3 1.1 % ISO 6507-2: 1997
193.8 HV10 ~ 1558 HV10 1.1 %
5 Shore Hardness Tester 32.1 HS ~ 96.8 HS 0.5 % JIS B 7727: 2000
Length
Best
No. Instrument to be calibrated Measurement Range Measurement Methods/Specifications Notes
Capability *)

1 Dial indikator 0 mm ~ 50 mm 8 mm JIS B 7503: 1992

2 of 3
Analytical instruments LK - 022 - IDN
Best
No. Instrument to be calibrated Measurement Range Measurement Methods/Specifications Notes
Capability *)
1 pH meter 0 pH ~ 14 pH 0.03 pH ASTM D1293:1990
2 Optical vacuum emission
Spectrometer
C 2.28 % ~ 3.84 % 0.01 % ASTM E 415: 1989
Si 0.82 % ~ 2.82 % 0.01 % ASTM E 1086: 1989
P 0.002 % ~ 1.77 % 0.001 %
S 0.001 % ~ 0.24 % 0.001 %
Mn 0.11 % ~ 1.19 % 0.001 %
3 Conductivity meter 0 µS ~ 200 µS 0.3 µS ASTM D 1125: 1991
200 µS ~ 200 mS 0.3 mS
ASTM E 1024: 1984;
4 Atomic Absorption Spectrophotometer 0 Abs ~ 2 Abs 0.04 Abs
ASTM E 1330: 1990
Notes:
Best Measurement Capability is stated as Expanded Uncertainty at approximately 95% confidence level, calculated using
coverage factor of k = 2, includes uncertainty contribution due to reference standards used by the laboratory, measurement process
at the laboratory, and contribution due to typical characteristic of nearly ideal device under calibration respective to the laboratory

The actual characteristic of device under calibration shall also be considered for individual calibration certificate issued
by the laboratory