revision date: 25 April 2008 expiry date: 24 April 2012

Laboratorium Quality Assurance (QA) Infrastructure Research & Development (R&D) PT Telekomunikasi Indonesia, Tbk
Accreditation Number : LK-002-IDN
Address : Jl. Gegerkalong Hilir - Bandung
Telephone : (022)4572322
Facsimile : (022)2012234
Contact Person : Tyas Indah Twi H
Electrical

No. Instrument to be calibrated Measurement Range Best Measurement Capability *) Methods/Specifications Notes

1 DC Voltmeter 0.001 mV ~ 320 mV 60 µV/V + 4.2 µV in-house procedure


320 mV ~ 3.2 V 60 µV/V + 41.6 µV P/PRK/I/K04
3V ~ 32 V 65 µV/V + 416 µV Kalibrasi multimeter
32 V ~ 320 V 65 µV/V + 4.48 mV
320 V ~ 1050 V 60 µV/V + 19.95 mV
2 AC Voltmeter 1 µV ~ 10 mV in-house procedure
f = 40 Hz ~ 3 kHz 0.40 mV/V + 384 µV P/PRK/I/K04
f = 3 kHz ~ 10 kHz 0.40 mV/V + 512 mV Kalibrasi multimeter
f = 10 kHz ~ 20 kHz 0.60 mV/V + 960 mV
10 mV ~ 32 mV
f = 40 Hz ~ 3 kHz 0.40 mV/V + 96 µV
f = 3 kHz ~ 10 kHz 0.40 mV/V + 128 mV
f = 10 kHz ~ 20 kHz 0.60 mV/V + 240 mV
32 mV ~ 320 mV
f = 40 Hz ~ 3 kHz 0.40 mV/V + 19.2 µV
f = 3 kHz ~ 10 kHz 0.40 mV/V + 25.6 µV
f = 10 kHz ~ 20 kHz 0.60 mV/V + 48 µV
0.32 V ~ 3.2 V
f = 40 Hz ~ 3 kHz 0.40 mV/V + 192 µV
f = 3 kHz ~ 10 kHz 0.40 mV/V + 256 µV
f = 10 kHz ~ 20 kHz 0.60 mV/V + 480 µV
3.2 V ~ 32 V
f = 40 Hz ~ 3 kHz 0.40 mV/V + 1.92 mV
f = 3 kHz ~ 10 kHz 0.60 mV/V + 2.56 mV
f = 10 kHz ~ 20 kHz 0.80 mV/V + 4.8 mV
32 V ~ 105 V
f = 40 Hz ~ 3 kHz 0.40 mV/V + 6.3 mV
f = 3 kHz ~ 10 kHz 0.60 mV/V + 8.4 mV
f = 10 kHz ~ 20 kHz 0.80 mV/V + 15.8 mV

1 of 7
revision date: 25 April 2008 expiry date: 24 April 2012
Laboratorium Quality Assurance (QA) Infrastructure Research & Development (R&D) PT Telekomunikasi Indonesia, Tbk
Accreditation Number : LK-002-IDN
Address : Jl. Gegerkalong Hilir - Bandung
Telephone : (022)4572322
Facsimile : (022)2012234
Contact Person : Tyas Indah Twi H

Electrical (cont'd)

No. Instrument to be calibrated Measurement Range Best Measurement Capability *) Methods/Specifications Notes

2 AC Voltmeter (cont'd) 105 V ~ 320 V in-house procedure


f = 40 Hz ~ 1 kHz 0.50 mV/V + 19.2 mV P/PRK/I/K04
f = 1 kHz ~ 3 kHz 0.80 mV/V + 19.2 mV Kalibrasi multimeter
f = 3 kHz ~ 10 kHz 0.80 mV/V + 32 mV
f = 10 kHz ~ 20 kHz 0.12 mV/V + 48 mV
320 V ~ 800 V
f = 40 Hz ~ 1 kHz 0.50 mV/V + 63 mV
f = 1 kHz ~ 3 kHz 0.80 mV/V + 63 mV
f = 3 kHz ~ 10 kHz 0.80 mV/V + 105 mV
f = 10 kHz ~ 20 kHz 0.12 mV/V + 158 mV
800 V ~ 1050 V
f = 40 Hz ~ 1 kHz 0.50 mV/V + 126 mV
f = 1 kHz ~ 3 kHz 0.80 mV/V + 126 mV
f = 3 kHz ~ 10 kHz 0.80 mV/V + 210 mV
f = 10 kHz ~ 20 kHz 0.12 mV/V + 315 mV
3 DC Voltage Source 1 mV ~ 100 mV 9 µV/V + 0.3 µV in-house procedure
100 mV ~ 1V 8 µV/V + 0.3 µV P/PRK/I/K05
1V ~ 10 V 8 µV/V + 0.5 µV kalibrasi DC Power
10 V ~ 100 V 10 µV/V + 30 µV Supply
100 V ~ 1000 V 10 µV/V + 100 µV
4 AC Voltage Source f = 40 Hz ~ 1 kHz
1 µV ~ 10 mV 0.20 mV/V + 1.1 µV
10 mV ~ 100 mV 0.07 mV/V + 2 µV in-house procedure
100 mV ~ 1V 0.07 mV/V + 20 µV P/PRK/I/K27
1V ~ 10 V 0.07 mV/V + 200 µV kalibrasi AC Voltage
10 V ~ 100 V 0.20 mV/V + 2 mV source
100 V ~ 700 V 0.40 mV/V + 20 mV
5 DC Ammeter 1 µA ~ 320 µA 0.14 mA/A + 11 nA in-house procedure
0.32 mA ~ 3.2 mA 0.14 mA/A + 83 nA P/PRK/I/K16
3.2 mA ~ 32 mA 0.14 mA/A + 900 nA Kalibrasi AC/DC
32 mA ~ 320 mA 0.16 mA/A + 9.6 µA current
0.32 A ~ 3.2 A 0.6 mA/A + 118 µA
3A ~ 10.5 A 0.55 mA/A 0.94 mA
10.5 A ~ 20 A 0.55 mA/A + 4.5 mA

2 of 7
revision date: 25 April 2008 expiry date: 24 April 2012
Laboratorium Quality Assurance (QA) Infrastructure Research & Development (R&D) PT Telekomunikasi Indonesia, Tbk
Accreditation Number : LK-002-IDN
Address : Jl. Gegerkalong Hilir - Bandung
Telephone : (022)4572322
Facsimile : (022)2012234
Contact Person : Tyas Indah Twi H

Electrical (cont'd)

No. Instrument to be calibrated Measurement Range Best Measurement Capability *) Methods/Specifications Notes

6 AC Ammeter 1 µA ~ 32 µA
f = 10 Hz ~ 3 kHz 0.7 mA/A + 0.9 µA in-house procedure
f = 3 kHz ~ 10 kHz 1 mA/A + 1.8 µA P/PRK/I/K16
32 µA ~ 320 µA Kalibrasi AC/DC
f = 10 Hz ~ 3 kHz 0.7 mA/A + 0.3 µA current
f = 3 kHz ~ 10 kHz 1 mA/A + 0.6 µA
0.32 mA ~ 3.2 mA
f = 10 Hz ~ 3 kHz 0.7 mA/A + 0.3 µA
f = 3 kHz ~ 10 kHz 1 mA/A + 0.6 µA
3.20 mA ~ 32 mA
f = 10 Hz ~ 3 kHz 0.7 mA/A + 3.2 µA
f = 3 kHz ~ 10 kHz 1 mA/A + 6.4 µA
32 mA ~ 320 mA
f = 10 Hz ~ 3 kHz 0.8 mA/A + 32 µA
f = 3 kHz ~ 10 kHz 1 mA/A + 48 µA
0.32 A ~ 3.2 A
f = 10 Hz ~ 3 kHz 1 mA/A + 0.48 µA
f = 3 kHz ~ 10 kHz 2.5 mA/A + 2.56 µA
3.2 A ~ 10.5 A
f = 10 Hz ~ 3 kHz 2 mA/A + 3 mA
f = 3 kHz ~ 10 kHz 5 mA/A + 10 mA
10.5 A ~ 20 A
f = 10 Hz ~ 3 kHz 2 mA/A + 6.9 mA
f = 3 kHz ~ 10 kHz 5 mA/A + 23 mA
7 DC Current Source 1 µA ~ 100 µA 20 µA/A + 0.8 nA in-house procedure
0.1 mA ~ 1 mA 20 µA/A + 5 nA P/PRK/I/K16
1 mA ~ 10 mA 20 µA/A + 50 nA Kalibrasi AC/DC
10 mA ~ 100 mA 35 µA/A + 500 nA current
0.1 A ~ 1A 110 µA/A + 10 µA
1A ~ 20 A 1 mA/A
8 AC Current Source 1 µA ~ 100 µA in-house procedure
f = 45 Hz ~ 5 kHz 0.6 mA/A + 0.03 µA P/PRK/I/K16
100 µA ~ 1 mA Kalibrasi AC/DC
f = 45 Hz- 100 Hz 0.6 mA/A + 0.2 µA current
f = 100 Hz - 5 kHz 0.3 mA/A + 0.2 µA

3 of 7
revision date: 25 April 2008 expiry date: 24 April 2012
Laboratorium Quality Assurance (QA) Infrastructure Research & Development (R&D) PT Telekomunikasi Indonesia, Tbk
Accreditation Number : LK-002-IDN
Address : Jl. Gegerkalong Hilir - Bandung
Telephone : (022)4572322
Facsimile : (022)2012234
Contact Person : Tyas Indah Twi H

Electrical (cont'd)

No. Instrument to be calibrated Measurement Range Best Measurement Capability *) Methods/Specifications Notes

8 AC Current Source 1 mA ~ 100 mA in-house procedure


(cont'd) f = 45 Hz- 100 Hz 0.6 mA/A + 20 µA P/PRK/I/K16
f = 100 Hz - 5 kHz 0.3 mA/A + 20 µA Kalibrasi AC/DC
0.1 A ~ 1A current
f = 45 Hz- 100 Hz 0.8 mA/A + 0.2 mA
f = 100 Hz - 5 kHz 1 mA/A + 0.2 mA
1A ~ 20 A
f = 45 Hz- 5 kHz 7 mA/A + 2 mA
9 Clamp meter 3.20 A ~ 32 A in-house procedure
(AC Current) f = 10 Hz ~ 100 Hz 2.0 mA/A + 5.5 mA P/PRK/I/K29
f = 100 Hz ~ 400 Hz 7.8 mA/A + 27. mA Kalibrasi clamp meter
32.00 A ~ 160 A
f = 10 Hz ~ 100 Hz 2.0 mA/A + 28 mA
160.00 A ~ 200 A
f = 10 Hz ~ 100 Hz 2.1 mA/A + 0.45 A
9 Clamp meter 20 A ~ 32 A 0.66 mA/A + 1.18 mA in-house procedure
(DC Current) 32 A ~ 105 A 0.6 mA/A + 9.4 mA P/PRK/I/K29
105 A ~ 200 A 0.6 mA/A + 45 mA Kalibrasi clamp meter
200 A ~ 525 A 0.6 mA/A + 47 mA
525 A ~ 1000 A 0.6 mA/A + 225 mA
10 DC Resistor 0.10 mΩ ~ 10 Ω 15.0 µΩ/Ω + 51 µΩ in-house procedure
10 Ω ~ 100 Ω 12.0 µΩ/Ω + 0.5 mΩ P/PRK/I/K09
100 Ω ~ 1 kΩ 10.0 µΩ/Ω + 0.51 mΩ Kalibrasi Resistansi
1 kΩ ~ 10 kΩ 10.0 µΩ/Ω + 5.1 mΩ
10 kΩ ~ 100 kΩ 10.0 µΩ/Ω + 51 mΩ
100 kΩ ~ 1 MΩ 15.0 µΩ/Ω + 2Ω
1 MΩ 10 MΩ 50.0 µΩ/Ω + 100 Ω
10 MΩ ~ 100 MΩ 503.0 µΩ/Ω + 1 kΩ
100 MΩ ~ 1 GΩ 5.0 mΩ/Ω + 10 kΩ
11 DC Ohmmeter 0.1 mΩ ~ 1 mΩ 0.50 mΩ/Ω + 0.012 µΩ
0.1 Ω ~ 100 kΩ 0.10 mΩ/Ω + 0.012 mΩ

0.1 Ω ~ 1111.21 Ω 0.10 mΩ/Ω + 2 mΩ

0.1 Ω ~ 40 Ω 0.50 mΩ/Ω + 20 mΩ in-house procedure


40 Ω ~ 400 Ω 0.15 mΩ/Ω + 20 mΩ P/PRK/I/K04
0.4 kΩ ~ 4 kΩ 0.15 mΩ/Ω + 80 mΩ Kalibrasi Multimeter

4 of 7
revision date: 25 April 2008 expiry date: 24 April 2012
Laboratorium Quality Assurance (QA) Infrastructure Research & Development (R&D) PT Telekomunikasi Indonesia, Tbk
Accreditation Number : LK-002-IDN
Address : Jl. Gegerkalong Hilir - Bandung
Telephone : (022)4572322
Facsimile : (022)2012234
Contact Person : Tyas Indah Twi H

Electrical (cont'd)

No. Instrument to be calibrated Measurement Range Best Measurement Capability *) Methods/Specifications Notes

11 DC Ohmmeter 4 kΩ ~ 40 kΩ 0.15 mΩ/Ω + 800 mΩ


(cont'd) 40 kΩ ~ 400 kΩ 0.18 mΩ/Ω + 8Ω in-house procedure
0.4 MΩ ~ 4 MΩ 0.20 mΩ/Ω + 100 Ω P/PRK/I/K04
4 MΩ ~ 40 MΩ 0.50 mΩ/Ω + 2 kΩ Kalibrasi Multimeter
40 MΩ ~ 400 MΩ 0.60 mΩ/Ω + 40 kΩ
12 Insulation Resistance measurement voltage (DC):
Mega-Ohmmeter 50 V ~ 1000 V
Resistance 100 kΩ ~ 300 kΩ 1.00 mΩ/Ω + 115 mΩ in-house procedure
0.3 MΩ ~ 3 MΩ 1.00 mΩ/Ω + 1.2 Ω P/PRK/I/K20
3 MΩ ~ 30 MΩ 3.00 mΩ/Ω + 12 Ω Kalibrasi Insulation
30 MΩ ~ 300 MΩ 5.00 mΩ/Ω + 120 Ω Tester
0.3 GΩ ~ 2 GΩ 7.00 mΩ/Ω + 1.2 kΩ
Voltage (DC) 10.0 V ~ 100 V 0.20 mV/V + 2 mV
100.0 V ~ 1000 V 0.40 mV/V + 20 mV
13 Continuity Tester max. measuring voltage = 10 V
0.1 Ω ~ 40 Ω 1.00 mΩ/Ω + 50 mΩ in-house procedure
40 Ω ~ 400 Ω 0.35 mΩ/Ω + 100 mΩ P/PRK/I/K04
0.4 kΩ ~ 4 kΩ 0.35 mΩ/Ω + 200 mΩ Kalibrasi Multimeter
14 AC Resistance meter; f = 10 Hz ~ 10 kHz
LCR Meter 10 mΩ 0.013 µΩ in-house procedure
0.1 Ω 0.0087 µΩ P/PRK/I/K17
1Ω 1.5 mΩ kalibrasi LCR Meter
10 Ω 15 mΩ
100 Ω 145 mΩ
1 kΩ 1.5 Ω
10 kΩ 14.5 Ω
100 kΩ 145 Ω
15 Capacitor f = 100 Hz ~ 10 kHz in-house procedure
0.01 pF ~ 10 F 0.6 mF/F P/PRK/I/K10
kalibrasi capacitor
16 Capacitance meter; f = 100 Hz ~ 10 kHz in-house procedure
LCR Meter 100 pF ~ 1 µF 1.0 mF/F P/PRK/I/K17
f = 100 Hz ~ 20 kHz kalibrasi LCR Meter
40 pF ~ 1.2 µF 2.5 mF/F + 3 pF

5 of 7
revision date: 25 April 2008 expiry date: 24 April 2012
Laboratorium Quality Assurance (QA) Infrastructure Research & Development (R&D) PT Telekomunikasi Indonesia, Tbk
Accreditation Number : LK-002-IDN
Address : Jl. Gegerkalong Hilir - Bandung
Telephone : (022)4572322
Facsimile : (022)2012234
Contact Person : Tyas Indah Twi H

Electrical (cont'd)

No. Instrument to be calibrated Measurement Range Best Measurement Capability *) Methods/Specifications Notes

17 Inductance meter; f = 100 Hz ~ 10 kHz in-house procedure


LCR Meter 100 µH 2.5 mH/H P/PRK/I/K17
1.00 mH ~ 10 H 1.0 mH/H kalibrasi LCR Meter
18 Inductor f = 100 Hz ~ 10 kHz in-house procedure
10 nH 99,999 H 0.6 mH/H P/PRK/I/K12
kalibrasi Induktansi
Time and Frequency

No. Instrument to be calibrated Measurement Range Best Measurement Capability *) Methods/Specifications Notes

1 Frequency meter / counter 2 MHz ; 10 MHz; 12 pHz/Hz in-house procedure

P/PRK/I/K08
kalibrasi frequency
200 Hz ~ 10 kHz 12 pHz/Hz
meter/counter

200 Hz ~ 80 MHz 12 pHz/Hz

2 Frequency generator 0.1 Hz ~ 10 Hz 1 mHz in-house procedure


10 Hz ~ 80 MHz 0.1 Hz P/PR/I/K01
80 MHz ~ 11.3 GHz 1 Hz kalibrasi freq. generator
11.3 MHz ~ 20 GHz 2 Hz
3 Audio frequency 20 Hz ~ 100 Hz 40 µHz/Hz + 1 mHz in-house procedure
generator 100 Hz ~ 1 kHz 40 µHz/Hz + 10 mHz P/PR/I/K01
1 kHz ~ 10 kHz 40 µHz/Hz + 0.1 Hz kalibrasi freq. generator
10 kHz ~ 100 kHz 40 µHz/Hz + 1 Hz
4 Audio level generator 50 mV ~ 300 V in-house procedure
f = 20 Hz ~ 20 kHz 0.17 dB P/PR/I/K01
f = 20 kHz ~ 100 kHz 0.36 dB kalibrasi freq. generator
5 RF Power generator f = 10 MHz ~ 20 GHz in-house procedure
100 pW ~ 100 mW 30 mW/W P/PR/I/K01
-70 dBm ~ 30 dBm 0.15 dB kalibrasi freq. generator
6 RF Power meter f = 10 MHz ~ 20 GHz
(without sensor) 3 µW ~ 100 mW 3 mW/W in-house procedure
-25 dBm ~ 20 dBm 0.013 dB P/PR/I/K02
kalibrasi RF Power Meter

6 of 7
revision date: 25 April 2008 expiry date: 24 April 2012
Laboratorium Quality Assurance (QA) Infrastructure Research & Development (R&D) PT Telekomunikasi Indonesia, Tbk
Accreditation Number : LK-002-IDN
Address : Jl. Gegerkalong Hilir - Bandung
Telephone : (022)4572322
Facsimile : (022)2012234
Contact Person : Tyas Indah Twi H

Time and Frequency

No. Instrument to be calibrated Measurement Range Best Measurement Capability *) Methods/Specifications Notes

7 Oscilloscope in-house procedure


Vertical deflection 5 mV ~ 120 mV 20 mV/V + 40 µV P/PRK/I/K06
Horizontal def. (Rise Time) 10 ns ~ 5s 0.25 µs/s kalibrasi osciloscope
Bandwidth source 10 Hz ~ 50 kHz 2.5 mHz/Hz
50 kHz ~ 250 MHz 30 mHz/Hz
250 MHz ~ 600 MHz 50 mHz/Hz
Notes:
Best Measurement Capability is stated as Expanded Uncertainty at approximately 95% confidence level, calculated using
coverage factor of k = 2, includes uncertainty contribution due to reference standards used by the laboratory, measurement process
at the laboratory, and contribution due to typical characteristic of nearly ideal device under calibration respective to the laboratory

The actual characteristic of device under calibration shall also be considered for individual calibration certificate issued
by the laboratory