You are on page 1of 1

revision date: August 1st, 2008 expiry date: July 31st, 2012

UPTD BPSMB Disperindagkop Prov Kaltim


Accreditation Number : LK 100 IDN
Address : Jl. MT Haryono no. 45 Samarinda 75213
Telephone : 0541 733731
Facsimile : 0541 733731
Contact Person : Hj. Erna Astuti

Temperature

No Instrument to be calibrated Measurement Range Best Measurement Capability Method/Specification Note

1 Enclosure
Oven 30 oC ~ 200 oC 1.2 oC AS 2853 1986
Muffle Furnace 400 oC ~ 900 oC 1.3 oC
Waterbath 30 oC ~ 100 oC 0.9 oC

Pressure

No Instrument to be calibrated Measurement Range Best Measurement Capability Method/Specification Note

1 Pressure Gauge (pneumatik) 0 kPa ~ 4000 kPa 120 kPa BS EN 837 : 1998

Volumetric

No Instrument to be calibrated Measurement Range Best Measurement Capability Method/Specification Note

1 Volumetric glassware 1 ml ~ 5 ml 0.003 ml ASTM E 542 01


5 ml ~ 10 ml 0.004 ml
10 ml ~ 25 ml 0.01 ml
25 ml ~ 50 ml 0.02 ml

Notes:
Best Measurement Capability is stated as Expanded Uncertainty at approximately 95% confidence level, calculated using
coverage factor of k = 2, includes uncertainty contribution due to reference standards used by the laboratory, measurement process
at the laboratory, and contribution due to typical characteristic of nearly ideal device under calibration respective to the laboratory

The actual characteristic of device under calibration shall also be considered for individual calibration certificate issued
by the laboratory