revision date:9 February 2007 expiry date: 8 February 2011

PT. BERCA HARDAYAPERKASA


Accreditation Number : LK-051-IDN
Address : Jl. Imam Bonjol, Nagoya - BATAM
Telephone : 0778-497912
Facsimile : 0778-427790
Contact Person : Agung Puji Priyantoro, Alexander Deni Wondal
Temperature
Best
No. Instruments to be calibrated Measurement Range Measurement Methods / Specification Notes
Capability *)
1 Temperature indicator without
0
type B 600°C ~ 1820°C 0.8 C EA 10/11: 2000
0
type C 0°C ~ 2316°C 0.8 C
0
type E -250°C ~ 1000°C 0.8 C
0
type J -210°C ~ 1200°C 0.8 C
0
type K -200°C ~ 1372°C 0.8 C
0
type L -200°C ~ 900°C 0.8 C
0
type N -200°C ~ 1300°C 0.8 C
0
type R 0°C ~ 1767°C 0.8 C
0
type S 0°C ~ 1767°C 0.8 C
0
type T -250°C ~ 400°C 0.8 C
0
type U -200°C ~ 600°C 0.8 C
Pressure
Best
No. Instruments to be calibrated Measurement Range Measurement Methods / Specification Notes
Capability *)
1 0.5 bar ~ 60.5 bar 0.10 % BS-EN 837-1: 1998
Pressure Test Gauge (hydraulic)
60.5 bar ~ 1200 bar 0.12 %
2 Digital Manometer (pneumatic) 0 bar ~ 700 kPa 56 kPa EA 10/17: 2002

3 Pressure Gauge (pneumatic) 0 bar ~ 700 kPa 56 kPa BS-EN 837-1: 1998

4 Pressure Gauge (hydraulic) 0 bar ~ 30 bar 0.15 bar


0 bar ~ 70 bar 0.35 bar BS-EN 837-1: 1998
0 bar ~ 100 bar 0.5 bar
0 bar ~ 250 bar 1.25 bar
0 bar ~ 600 bar 3 bar
0 bar ~ 1000 bar 5 bar
0 bar ~ 1200 bar 6 bar
5 0.5 bar ~ 60.5 bar 0.20 % BS-EN 837-1: 1998
Pressure Recorder (hydraulic)
60.5 bar ~ 1200 bar 0.20 % EA 10/17: 2002
( 1 bar = 105 Pa)
Length
Best
No. Instruments to be calibrated Measurement Range Measurement Methods / Specification Notes
Capability *)
1 Outside Micrometer 0 mm ~ 25 mm 0.57 µm
25 mm ~ 100 mm 0.6 µm
2 Vernier Caliper 0 mm ~ 300 mm 6 µm ISO 6906:1984
3 Dial Indicator 0 mm ~ 25 mm 0.8 µm ISO R463: 1965
Electrical
Best
No. Instruments to be calibrated Measurement Range Measurement Methods / Specification Notes
Capability *)
1 DC Voltmeter 0.0001 mV ~ 329.9999 mV 0.29 mV/V In-house procedure WI-E-02
0.33 V ~ 3.299999 V 0.22 mV/V Kalibrasi multimeter
3.3 V ~ 32.99999 V 0.23 mV/V In-hose procedure WI-E-04
33 V~ 329.9999 V 0.28 mV/V Kalibrasi voltmeter
330 V~ 1000 V 0.62 mV/V
2 AC Voltmeter 1 mV ~ 32.999 mV
f = 10 Hz ~ 45 Hz 0.96 mV/V In-house procedure WI-E-02
f = 45 Hz ~ 10 kHz 0.33 mV/V Kalibrasi multimeter
f = 10 kHz ~ 20 kHz 0.31 mV/V In-hose procedure WI-E-04
f = 20 kHz ~ 50 kHz 1.2 mV/V Kalibrasi voltmeter
f = 50 kHz ~ 100 kHz 4.1 mV/V
f = 100 kHz ~ 500 kHz 9.3 mV/V
33 mV ~ 329.999 mV
f = 10 Hz ~ 45 Hz 0.54 mV/V
f = 45 Hz ~ 10 kHz 0.17 mV/V
f = 10 kHz ~ 20 kHz 0.19 mV/V
f = 20 kHz ~ 50 kHz 0.41 mV/V
f = 50 kHz ~ 100 kHz 0.93 mV/V
f = 100 kHz ~ 500 kHz 2.3 mV/V
0.33 V ~ 3.29999 V
f = 10 Hz ~ 45 Hz 0.51 mV/V
f = 45 Hz ~ 10 kHz 0.39 mV/V
f = 10 kHz ~ 20 kHz 0.42 mV/V In-house procedure WI-E-02
f = 20 kHz ~ 50 kHz 0.63 mV/V Kalibrasi multimeter
f = 50 kHz ~ 100 kHz 0.97 mV/V In-hose procedure WI-E-04
f = 100 kHz ~ 500 kHz 2.8 mV/V Kalibrasi voltmeter
3.3 V ~ 32.9999 V
f = 10 Hz ~ 45 Hz 0.35 mV/V
f = 45 Hz ~ 10 kHz 0.18 mV/V
f = 10 kHz ~ 20 kHz 0.28 mV/V
f = 20 kHz ~ 50 kHz 0.41 mV/V
f = 50 kHz ~ 100 kHz 1.1 mV/V
33 V ~ 329.999 V
f = 10 Hz ~ 45 Hz 0.23 mV/V
f = 45 Hz ~ 10 kHz 0.24 mV/V
f = 10 kHz ~ 20 kHz 0.29 mV/V
f = 20 kHz ~ 50 kHz 0.35 mV/V
f = 50 kHz ~ 100 kHz 2.3 mV/V
330 V ~ 1020 V
f = 45 Hz ~ 1 kHz 0.35 mV/V
f = 1 kHz ~ 5 kHz 0.29 mV/V
f = 5 kHz ~ 10 kHz 0.35 mV/V
Electrical (cont'd)
Best
No. Instruments to be calibrated Measurement Range Measurement Methods / Specification Notes
Capability *)
3 DC Ammeter 0.001 mA~ 329.999 mA 0.18 mA/A
0.33 mA~ 3.29999 mA 0.12 mA/A
3.29999 mA~ 32.9999 mA 0.12 mA/A
32.9999 mA~ 329.999 mA 0.12 mA/A In-house procedure WI-E-02
0.329999 A ~ 1.09999 A 0.25 mA/A Kalibrasi Multimeter
1.1 A ~ 2.99999 A 0.45 mA/A In-house procedure WI-E-03
3 A ~ 10.9999 A 0.59 mA/A Kalibrasi Ammeter
11 A ~ 20.5 A 1.2 mA/A
4 AC Ammeter 29.00 mA ~ 329.99 mA
f = 10 Hz ~ 20 Hz 2.3 mA/A In-house procedure WI-E-02
f = 20 Hz ~ 45 Hz 1.7 mA/A Kalibrasi Multimeter
f = 45 Hz ~ 1 kHz 1.5 mA/A In-house procedure WI-E-03
f = 1 kHz ~ 5 kHz 3.5 mA/A Kalibrasi Ammeter
f = 5 kHz ~ 10 kHz 9.3 mA/A
f = 10 kHz ~ 30 kHz 19 mA/A
0.33 mA ~ 3.2999 mA
f = 10 Hz ~ 20 Hz 2.3 mA/A
f = 20 Hz ~ 45 Hz 1.5 mA/A
f = 45 Hz ~ 1 kHz 1.2 mA/A
f = 1 kHz ~ 5 kHz 2.3 mA/A
f = 5 kHz ~ 10 kHz 5.8 mA/A
f = 10 kHz ~ 30 kHz 12 mA/A
3.3 mA ~ 32.999 mA
f = 10 Hz ~ 20 Hz 2.1 mA/A In-house procedure WI-E-02
f = 20 Hz ~ 45 Hz 1.1 mA/A Kalibrasi Multimeter
f = 45 Hz ~ 1 kHz 0.52 mA/A In-house procedure WI-E-03
f = 1 kHz ~ 5 kHz 0.96 mA/A Kalibrasi Ammeter
f = 5 kHz ~ 10 kHz 2.3 mA/A
f = 10 kHz ~ 30 kHz 4.6 mA/A
33 mA ~ 329.99 mA
f = 10 Hz ~ 20 Hz 2.1 mA/A
f = 20 Hz ~ 45 Hz 1.1 mA/A
f = 45 Hz ~ 1 kHz 0.52 mA/A
f = 1 kHz ~ 5 kHz 1.2 mA/A
f = 5 kHz ~ 10 kHz 2.3 mA/A
f = 10 kHz ~ 30 kHz 4.6 mA/A
0.33 A ~ 1.09999 A
f = 10 Hz ~ 45 Hz 2.1 mA/A
f = 45 Hz ~ 1 kHz 0.8 mA/A
f = 1 kHz ~ 5 kHz 7 mA/A
f = 5 kHz ~ 10 kHz 29 mA/A
1.1 A ~ 2.99999 A
f = 10 Hz ~ 45 Hz 2.1 mA/A
f = 45 Hz ~ 1 kHz 0.73 mA/A
f = 1 kHz ~ 5 kHz 7 mA/A
f = 5 kHz ~ 10 kHz 29 mA/A
3 A ~ 10.9999 A
f = 45 Hz ~ 100 Hz 0.9 mA/A
f = 100 Hz ~ 1 kHz 1.3 mA/A
f = 1 kHz ~ 5 kHz 35 mA/A
Electrical (cont'd)
Best
No. Instruments to be calibrated Measurement Range Measurement Methods / Specification Notes
Capability *)
4 AC ammeter (cont'd) 11 A ~ 20.5 A
f = 45 Hz ~ 100 Hz 1.4 mA/A
f = 100 Hz ~ 1 kHz 1.8 mA/A
f = 1 kHz ~ 5 kHz 35 mA/A
5 Ohmmeter 0.0001 Ω ~ 10.9999 Ω 0.53 mΩ/Ω
11 Ω ~ 32.9999 Ω 0.18 mΩ/Ω
33 Ω ~ 109.9999 Ω 0.06 mΩ/Ω
110 Ω ~ 329.9999 Ω 0.04 mΩ/Ω In-house procedure WI-E-05
330 Ω ~ 1.099999 kΩ 0.06 mΩ/Ω kalibrasi ohmmeter
1.1 kΩ ~ 3.299999 kΩ 0.04 mΩ/Ω
3.3 kΩ ~ 10.99999 kΩ 0.06 mΩ/Ω
11 kΩ ~ 32.99999 kΩ 0.04 mΩ/Ω
33 kΩ ~ 109.9999 kΩ 0.06 mΩ/Ω
110 kΩ ~ 329.9999 kΩ 0.04 mΩ/Ω
330 kΩ ~ 1.099999 MΩ 0.53 mΩ/Ω
1.1 MΩ ~ 3.299999 MΩ 0.19 mΩ/Ω
3.3 MΩ ~ 10.99999 MΩ 0.16 mΩ/Ω
11 MΩ~ 32.99999 MΩ 0.34 mΩ/Ω
33 MΩ ~ 109.9999 MΩ 1.1 mΩ/Ω
110 MΩ ~ 329.9999 MΩ 3.5 mΩ/Ω
330 MΩ ~ 1100 MΩ 17 mΩ/Ω
6 DC Voltage Source 10 nV ~ 0.1 V 0.6 mV/V In-house procedure WI-E-09
0.1 V - 1 V 0.2 mV/V Kalibrasi Voltage Source
1 V - 10 V 0.2 mV/V
10 V - 100 V 0.2 mV/V
100 V - 1000 V 0.2 mV/V
7 AC Voltage Source 10 mV
f = 1 Hz ~ 20 kHz 6.2 mV/V In-house procedure WI-E-09
f = 20 kHz ~ 50 kHz 6.2 mV/V Kalibrasi Voltage Source
f = 50 kHz ~ 100 kHz 8.5 mV/V
f = 100 kHz ~ 300 kHz 47 mV/V
100 mV ~ 10 V
f = 1 Hz ~ 50 kHz 2 mV/V
f = 50 kHz ~ 100 kHz 2.2 mV/V
100 mV ~ 10 V
f = 100 kHz ~ 300 kHz 4.1 mV/V
f = 300 kHz ~ 1 MHz 12 mV/V
f = 1 MHz ~ 2 MHz 18 mV/V
100 V
f = 1 Hz ~ 50 kHz 2 mV/V
f = 50 kHz ~ 100 kHz 2.4 mV/V
f = 100 kHz ~ 300 kHz 5.1 mV/V
f = 300 kHz ~ 1 MHz 18 mV/V
1000 V
f = 1 Hz ~ 20 kHz 2.1 mV/V
f = 20 kHz ~ 50 kHz 2.5 mV/V
f = 50 kHz ~ 100 kHz 4 mV/V
8 DC Current Source 100 nA 0.54 mA/A In-house procedure WI-E-10
1 mA 0.22 mA/A Kalibrasi Current Source
10 mA 0.2 mA/A
100 mA 0.2 mA/A
1 mA 0.21 mA/A
Electrical (cont'd)
Best
No. Instruments to be calibrated Measurement Range Measurement Methods / Specification Notes
Capability *)
8 DC Current Source (cont'd) 10 mA 0.2 mA/A
100 mA 0.21 mA/A
1A 0.3 mA/A
9 AC Current Source 100 mA
10 Hz ~ 20 Hz 7 mA/A In-house procedure WI-E-10
20 Hz ~ 45 Hz 5.4 mA/A Kalibrasi Current Source
45 Hz ~ 5 kHz 5.1 mA/A
1 mA ~ 100 mA
10 Hz ~ 20 Hz 7 mA/A
20 Hz ~ 45 Hz 5.4 mA/A
45 Hz ~ 20 kHz 5.1 mA/A
20 kHz ~ 50 kHz 7.1 mA/A
50 kHz ~ 100 kHz 9.5 mA/A
1A In-house procedure WI-E-10
10 Hz ~ 20 Hz 9 mA/A Kalibrasi Current Source
20 Hz ~ 45 Hz 8 mA/A
45 Hz ~ 100 Hz 7.7 mA/A
100 Hz ~ 5 kHz 7.8 mA/A
5 kHz ~ 20 kHz 8.5 mA/A
20 kHz ~ 50 kHz 14 mA/A
10 LCR Meter
Inductance 100 mH ~ 1 mH 23 mH/H In-house procedure WI-E-07
1 mH ~ 100mH 19 mH/H Kalibrasi LCR meter
100 mH ~ 1 H 11 mH/H
1 H ~ 11.11 H 11 mH/H
Capacitance 1 pF - 1.11111 mF 8 mF/F
Resistance 100 Ω ~ 20 MΩ 0.83 mΩ/Ω
20 MΩ - 1100 MΩ 17 mΩ/Ω
11 Decade Resistor 10 Ω - 100 MΩ 1.2 mΩ/Ω In-house procedure WI-E-08
Standard Resistor 100 MΩ - 1 GΩ 5.9 mΩ/Ω Kalibrasi resistor
12 Osciloscope Time / Div 5 s - 2ns 1 ms/s In-house procedure WI-E-11
Volt / Div 1.8 mV - 105 V p-p 2 mV/V Kalibrasi Osciloscope
13 Distortion Analyzer Range level 50 mV - 300 V 4.3 mV/V In-house procedure WI-E 12

Frekuensi 20 Hz - 100 kHz 0.1 µHz/Hz Kalibrasi Distortion Analyzer

14 Audio Analyzer Level 50 mV ~ 300 V In-house procedure WI-E-13


f = 20 Hz ~ 20 kHz 4.3 mV/V Kalibrasi Audio Analyzer
f = 20 kHz ~ 100 kHz 4.3 mV/V
Level 0.3 mV - 50 mV
f = 20 kHz ~ 100 kHz 4 mV/V
f= 20 Hz - 100 kHz 0.1 mV/V
15 Function Generator Level 50 mV ~ 300 V In-house procedure WI-E-14
(Audio) f = 20 Hz ~ 20 kHz 4.3 mV/V Kalibrasi Audio
f = 20 kHz ~ 100 kHz 4.3 mV/V Function Generator
Level 0.3 mV - 50 mV
f = 20 kHz ~ 100 kHz 4 mV/V
f = 20 Hz - 100 kHz 0.1 mV/V
Time and Frequency
Best
No. Instruments to be calibrated Measurement Range Measurement Methods / Specification Notes
Capability *)
1 Function Generator Frekuensi 10 Hz - 20 GHz 0.84 nHz/Hz In-house procedure WI-E-15
(High Frequency) RF Power -30 dBm - +20 dBm 2.4 dB Kalibrasi Function Generator HF
pada f = 10 MHz - 18 GHz
2 Function Generator Frekuensi 10 Hz - 20 GHz 0.84 nHz/Hz In-house procedure WI-E-16

(µ Wave Frequency) RF Power -30 dBm - +20 dBm 2.4 dB Kalibrasi Function Generator µ-wave

pada f = 10 MHz - 18 GHz


3 Signal Generator RF Frekuensi 150 kHz – 1.3 GHz 820 nHz/Hz In-house procedure WI-E-18
Level -20 dBm ~ -10 dBm 0.13 dB Kalibrasi Signal Generator RF
Level -10 dBm ~ +30 dBm 0.05 dB
AM 0 - 99 % depth modulation 3.2 %
FM 0 - 400 kHz deviation 6.3 %
4 Signal Generator 10 Hz - 20 GHz 0.84 nHz/Hz In-house procedure WI-E-17

(µ Wave) RF Power -30 dBm - +20 dBm 2.4 dB Kalibrasi Signal Generator µ-wave

f = 10 MHz - 18 GHz
1.7 dB In-house procedure WI-E-20
5 Power Meter RF Power -143 dBm - +13 dBm
Kalibrasi Power Meter
6 Spectrum Analyzer 10 MHz - 20 GHz 8.3 nHz/Hz In-house procedure WI-E-19
100 kHz to 1.04 GHz 27 nHz/Hz Kalibrasi Spectrum Analyzer
RF Power output -30 dBm - +20 dBm 2.4 dB
RF Power input -143 dBm - +13 dBm 1.7 dB
f = 10 MHz - 18 GHz
7 Frequency Counter Frekuensi 10 Hz - 20 GHz
w/ internal ref. 0.84 nHz/Hz In-house procedure WI-E-21
w/out internal ref. 8.3 nHz/Hz Kalibrasi Frequency Counter

Notes:
Best Measurement Capability is stated as Expanded Uncertainty at approximately 95% confidence level, calculated using
coverage factor of k = 2, includes uncertainty contribution due to reference standards used by the laboratory, measurement process
at the laboratory, and contribution due to typical characteristic of nearly ideal device under calibration respective to the laboratory

The actual characteristic of device under calibration shall also be considered for individual calibration certificate issued
by the laboratory