Quantity
*Calibration Measurement
Sl. Measured / Range/Frequency Remarks
Capability (±)
Instrument
ELECTRO-TECHNICAL CALIBRATION
I. SOURCE
#
2. AC Voltage (20 to 40) Hz
(2 to 20) mV 0.35% - 0.068% Using Calibrator Fluke 5720A
(20 to 200) mV 0.068% - 0.046%
200 mV to 2 V 0.046% - 0.039%
(2 to 20) V 0.039%
(20 to 200) V 0.039% - 0.036%
40Hz to 20kHz
2 to 20) mV 0.21% - 0.048%
(20 to 200) mV 0.048% - 0.021%
200 mV to 2 V 0.021% - 0.01%
(2 to 20) V 0.01%
(20 to 200) V 0.01% - 0.012%
100kHz to 1 MHz
(2 to 20) mV 2.4% - 0.72%
(20 to 200) mV 0.72% - 0.47%
200 mV to 2 V 0.47% - 0.33%
(2 to 20) V 0.33% - 0.35%
50 Hz to 1kHz
1000 V 0.13% - 0.11%
Quantity
*Calibration Measurement
Sl. Measured / Range/Frequency Remarks
Capability (±)
Instrument
#
DC Current 2 A to 10 A 0.14% - 0.1% Using Calibrator Fluke 5500A
(cont.) 10A to 18A 0.1% - 0.33%
#
4. AC Current 10 Hz
200 uA to 2 mA 0.047% - 0.039% Using Calibrator Fluke 5720A
2 mA to 20mA 0.039%
20 mA to 200mA 0.039% - 0.041%
1kHz
10 uA to 2 mA 0.063% - 0.020%
2 mA to 20mA 0.020% - 0.021%
20 mA to 200mA 0.021% - 0.024%
200mA to 2 A 0.024% - 0.071%
10 kHz
200 uA to 2 mA 0.2%
2 mA to 20mA 0.2% - 0.19%
20 mA to 200mA 0.19% - 0.16%
200mA to 2A 0.16% - 0.94%
1 kHz
2 A to 10A 0.88% - 0.41% Using Calibrator Fluke 5500A
50Hz
10 A to 18A 0.59% - 1.8% Using Calibrator Fluke 5500A
& 5220A with 50 turn Current
18A to 900A 2.2% - 0.86% coil
Quantity
*Calibration Measurement
Sl. Measured / Range/Frequency Remarks
Capability (±)
Instrument
#
Resistance
(cont.) 2 Ω to 330 Ω 0.47% - 0.017% Using Calibrator Fluke 5500A
330 Ω to 330 kΩ 0.017% - 0.018%
330 kΩ to117 MΩ 0.018% - 0.63%
#
6. Capacitance 1kHz
0.35 nF to 330 nF 3.7% to 0.56% Using Calibrator Fluke 5500A
1kHz
1.0 µF to 1100 µF 0.33% to 1.2%
7. Temperature
#
Simulation
Thermocouple (-200 to 800) ºC Using Calibrator Fluke 5500A
0.59°C - 0.26°C
Type E
(-200 to 1100 ) ºC 0.36°C - 0.31°C
Type J
(-190 to 1300) ºC 0.42°C - 0.50°C
Type K
(150 to 1700) ºC 0.62°C - 0.76°C
Type S
(-200 to 300) ºC 0.79°C - 0.18°C
Type T
(-200 to 1300) ºC 0.07°C
Type N
(150 to 1700) ºC 0.62°C - 0.76°C
Type R
(600 to 1800) ºC 0.5°C
Type B
#
8. Oscilloscope
Quantity
*Calibration Measurement
Sl. Measured / Range/Frequency Remarks
Capability (±)
Instrument
II. MEASURE
# 0.0014% - 0.00096%
1. DC Voltage 100mV to 1V Using Multimeter Agilent
1V to 10 V 0.00096% - 0.00093% 3458A
(10 to 1000) V 0.00093%- 0.0012%
#
3. DC Current 10 nA to 100 A 1.1 % to 0.58% Using Multimeter Agilent 3458A
100A to 10mA 0.58% - 0.0029%
10mA to 1A 0.0029% - 0.014%
Using Shunt with Mulitmeter
(1 to30) A 0.28%
#
4. AC Voltage 40Hz -1 kHz
(10 to 100) mV 0.043% - 0.042% Using Multimeter Agilent 3458A
100mV to 1V 0.042% - 0.019%
(1 to 10) V 0.019% - 0.037%
(10 to 100) V 0.037% - 0.026%
(100 to 700) V 0.026% 0 -0.10%
20 kHz
(10 to 100) mV 0.056% - 0.019%
100mV to 1V 0.019% - 0.037%
(1 to 10) V 0.037% - 0.095%
(10 to 100) V 0.095% - 0.026%
(100 to 700) V 0.026% - 0.10%
100 kHz
0.59% - 0.037%
(10 to 100) mV
0.037% - 0.099%
100mV to 1V
0.099% - 0.36%
(1 to 10) V
0.36% - 0.043%
(10 to 100) V
1 MHz
0.59% - 0.12%
(10 to 100) mV
0.12%
100mV to 10V
Quantity
*Calibration Measurement
Sl. Measured / Range/Frequency Remarks
Capability (±)
Instrument
#
5. AC High Voltage 50 Hz
(1 to 5) kV 3.2% - 3.8% Using HV Probe Fluke 80k-40
with DMM Fluke 115
# 45 Hz
6. AC Current
(1 to 100) mA 0.093% -0.092% Using Multimeter Agilent 3458A
100mA to 1A 0.092% - 0.12%
1 kHz
(10 to 100) A 0.46% - 0.11%
100 A to 1 mA 0.11% - 0.059%
(1 to 100) mA 0.059% - 0.058%
100mA to 1A 0.058% - 0.14%
#
8. Frequency 1 Hz to 100 Hz 662ppm – 120ppm Using Multimeter Agilent
3458A
9. Temperature
#
Simulation
Thermocouple
Type E (-200 to 800) ºC Using Multimeter (7 ½ )
0.08 °C
NI PXI 4071
Type J (-200 to 1100 ) ºC 0.08 °C
Type K (-190 to 1300) ºC 0.05 °C
Type S (150 to 1700) ºC 0.14 °C
Type T (-200 to 300) ºC 0.10 °C
Type N (-200 to 1300) ºC 0.09 °C
Type R (100 to 1700) ºC 0.07 °C
Type B (600 to 1800) ºC 0.50 °C
RTD
Pt 385, 100 0.07 °C Using Multimeter Agilent
(-190 to 800) ºC
3458A
Quantity
*Calibration Measurement
Sl. Measured / Range/Frequency Remarks
Capability (±)
Instrument
MECHANICAL CALIBRATION
$
2. Height Gauges up to 600 mm 8.8 µm Using Gauge Blocks and
L.C: 0.01 mm Caliper Checker as per
IS 2921
$
4. Cylindrical Pins 0.5 to 20 mm 1.7 µm Using Laser Scan Micrometer
as per IS 11103
$
5. Plain Plug Gauges 0.5 to 40 mm 1.0 µm Using Gauge Blocks as per
IS 3455, IS 6174
2. Pressure Gauge / 0 to 700 bar 0.30% of rdg Using Digital Pressure Gauge as
Calibrator / per DKD R-6-1
Transducer /
Transmitter /
Switch / Compound
#
Gauge
#
3. Vacuum Gauges -0.68 to 0 bar 0.30% of rdg Using Digital Pressure Calibrator
as per IS 3567
Quantity
*Calibration Measurement
Sl. Measured / Range/Frequency Remarks
Capability (±)
Instrument
III. UTM, TENSION CREEP AND TORSION TESTING MACHINE
THERMAL CALIBRATION
1. RTD, Thermocouple, 50°C to400°C 1.9°C Using Standard RTD, Dry block
Thermistor Sensors calibrator and Digital Multimeter
with / without
Indicators,
temperature gauges,
Temperature baths,
controllers and
Transmitters with
#
sensors