Laboratory Micro Precision Calibration India Private Limited,

No.57/65, Sippoy Lane, Avadi Road, Karayanchavadi, Chennai, Tamil Nadu.


Accreditation Standard ISO/IEC 17025: 2005
Certificate Number CC-2490 Page 1 of 7
Validity 19-11-2018 to 18-11-2020 Last Amended on 26-11-2018

Quantity
*Calibration Measurement
Sl. Measured / Range/Frequency Remarks
Capability (±)
Instrument

ELECTRO-TECHNICAL CALIBRATION

I. SOURCE

# Using Calibrator Fluke 5720A


1. DC Voltage 0.5 mV to 1V 0.13% - 0.0012%
1V to 10 V 0.0012% - 0.00072%
(10 to 100) V 0.00072% - 0.00089%
(100 to 1000) V 0.00089% - 0.0011%

#
2. AC Voltage (20 to 40) Hz
(2 to 20) mV 0.35% - 0.068% Using Calibrator Fluke 5720A
(20 to 200) mV 0.068% - 0.046%
200 mV to 2 V 0.046% - 0.039%
(2 to 20) V 0.039%
(20 to 200) V 0.039% - 0.036%

40Hz to 20kHz
2 to 20) mV 0.21% - 0.048%
(20 to 200) mV 0.048% - 0.021%
200 mV to 2 V 0.021% - 0.01%
(2 to 20) V 0.01%
(20 to 200) V 0.01% - 0.012%

(20 to 100) kHz


0.47% - 0.11%
(2 to 20) mV
0.11% - 0.083%
(20 to 200) mV
0.083% - 0.021%
200 mV to 2 V
0.021%
(2 to 20) V
(20 to 200) V 0.021% - 0.023%

100kHz to 1 MHz
(2 to 20) mV 2.4% - 0.72%
(20 to 200) mV 0.72% - 0.47%
200 mV to 2 V 0.47% - 0.33%
(2 to 20) V 0.33% - 0.35%

50 Hz to 1kHz
1000 V 0.13% - 0.11%

# 10 µA to 2 mA 0.011% - 0.006% Using Calibrator Fluke 5720A


3. DC Current
2 mA to 20 mA 0.006% - 0.005%
20 mA to 200 mA 0.005% - 0.007%
200 mA to 2 A 0.007% - 0.011%

Convenor Program Director


Laboratory Micro Precision Calibration India Private Limited,
No.57/65, Sippoy Lane, Avadi Road, Karayanchavadi, Chennai, Tamil Nadu.
Accreditation Standard ISO/IEC 17025: 2005
Certificate Number CC-2490 Page 2 of 7
Validity 19-11-2018 to 18-11-2020 Last Amended on 26-11-2018

Quantity
*Calibration Measurement
Sl. Measured / Range/Frequency Remarks
Capability (±)
Instrument

#
DC Current 2 A to 10 A 0.14% - 0.1% Using Calibrator Fluke 5500A
(cont.) 10A to 18A 0.1% - 0.33%

18 A to 900 A 2.7% - 0.86% Using Calibrator Fluke 5500A


& 5220A with 50 turn Current

#
4. AC Current 10 Hz
200 uA to 2 mA 0.047% - 0.039% Using Calibrator Fluke 5720A
2 mA to 20mA 0.039%
20 mA to 200mA 0.039% - 0.041%
1kHz
10 uA to 2 mA 0.063% - 0.020%
2 mA to 20mA 0.020% - 0.021%
20 mA to 200mA 0.021% - 0.024%
200mA to 2 A 0.024% - 0.071%
10 kHz
200 uA to 2 mA 0.2%
2 mA to 20mA 0.2% - 0.19%
20 mA to 200mA 0.19% - 0.16%
200mA to 2A 0.16% - 0.94%
1 kHz
2 A to 10A 0.88% - 0.41% Using Calibrator Fluke 5500A
50Hz
10 A to 18A 0.59% - 1.8% Using Calibrator Fluke 5500A
& 5220A with 50 turn Current
18A to 900A 2.2% - 0.86% coil

# 240 µΩ Using Calibrator Fluke 5720A


5. Resistance 1Ω
1.9 Ω 190 µΩ
10 Ω 87 µΩ
19 Ω 210 µΩ
100 Ω 658 µΩ
190 Ω 1255 µΩ
1 kΩ 19 mΩ
1.9 kΩ 22 mΩ
10 kΩ 85 mΩ
19 kΩ 181 mΩ
100 kΩ 734 mΩ
190 kΩ 1397 mΩ
1 MΩ 31 Ω
1.9 MΩ 45 Ω
10 MΩ 240 Ω
19 MΩ 387 Ω
100 MΩ 13270Ω

Convenor Program Director


Laboratory Micro Precision Calibration India Private Limited,
No.57/65, Sippoy Lane, Avadi Road, Karayanchavadi, Chennai, Tamil Nadu.
Accreditation Standard ISO/IEC 17025: 2005
Certificate Number CC-2490 Page 3 of 7
Validity 19-11-2018 to 18-11-2020 Last Amended on 26-11-2018

Quantity
*Calibration Measurement
Sl. Measured / Range/Frequency Remarks
Capability (±)
Instrument
#
Resistance
(cont.) 2 Ω to 330 Ω 0.47% - 0.017% Using Calibrator Fluke 5500A
330 Ω to 330 kΩ 0.017% - 0.018%
330 kΩ to117 MΩ 0.018% - 0.63%

#
6. Capacitance 1kHz
0.35 nF to 330 nF 3.7% to 0.56% Using Calibrator Fluke 5500A

1kHz
1.0 µF to 1100 µF 0.33% to 1.2%

7. Temperature
#
Simulation
Thermocouple (-200 to 800) ºC Using Calibrator Fluke 5500A
0.59°C - 0.26°C
Type E
(-200 to 1100 ) ºC 0.36°C - 0.31°C
Type J
(-190 to 1300) ºC 0.42°C - 0.50°C
Type K
(150 to 1700) ºC 0.62°C - 0.76°C
Type S
(-200 to 300) ºC 0.79°C - 0.18°C
Type T
(-200 to 1300) ºC 0.07°C
Type N
(150 to 1700) ºC 0.62°C - 0.76°C
Type R
(600 to 1800) ºC 0.5°C
Type B

RTD 0.037°C – 0.11°C


(-190 to 620) ºC
Pt 385, 1 k
0.059°C – 0.14°C
(-190 to 790) ºC
Pt 385, 100 

#
8. Oscilloscope

7.8% -1.6% Using Calibrator Fluke 5500A


Amplitude 2mV to 50V

Time Base 2 nS to 5S 2.03% - 0.58%

Bandwidth 0.6V, 600MHz 10.5%

Convenor Program Director


Laboratory Micro Precision Calibration India Private Limited,
No.57/65, Sippoy Lane, Avadi Road, Karayanchavadi, Chennai, Tamil Nadu.
Accreditation Standard ISO/IEC 17025: 2005
Certificate Number CC-2490 Page 4 of 7
Validity 19-11-2018 to 18-11-2020 Last Amended on 26-11-2018

Quantity
*Calibration Measurement
Sl. Measured / Range/Frequency Remarks
Capability (±)
Instrument

# Using Calibrator Fluke 5720A


9. Frequency 10 Hz to 1 MHz 131ppm – 22ppm

Using Frequency Standard


10MHz 0.76ppm
GPS HP 58503A

II. MEASURE

# 0.0014% - 0.00096%
1. DC Voltage 100mV to 1V Using Multimeter Agilent
1V to 10 V 0.00096% - 0.00093% 3458A
(10 to 1000) V 0.00093%- 0.0012%

2. # Using HV Probe Fluke 80k-40


DC High Voltage (1 to 20) kV 1.9 % - 6.7 %
with DMM Fluke 115

#
3. DC Current 10 nA to 100 A 1.1 % to 0.58% Using Multimeter Agilent 3458A
100A to 10mA 0.58% - 0.0029%
10mA to 1A 0.0029% - 0.014%
Using Shunt with Mulitmeter
(1 to30) A 0.28%

#
4. AC Voltage 40Hz -1 kHz
(10 to 100) mV 0.043% - 0.042% Using Multimeter Agilent 3458A
100mV to 1V 0.042% - 0.019%
(1 to 10) V 0.019% - 0.037%
(10 to 100) V 0.037% - 0.026%
(100 to 700) V 0.026% 0 -0.10%

20 kHz
(10 to 100) mV 0.056% - 0.019%
100mV to 1V 0.019% - 0.037%
(1 to 10) V 0.037% - 0.095%
(10 to 100) V 0.095% - 0.026%
(100 to 700) V 0.026% - 0.10%

100 kHz
0.59% - 0.037%
(10 to 100) mV
0.037% - 0.099%
100mV to 1V
0.099% - 0.36%
(1 to 10) V
0.36% - 0.043%
(10 to 100) V

1 MHz
0.59% - 0.12%
(10 to 100) mV
0.12%
100mV to 10V

Convenor Program Director


Laboratory Micro Precision Calibration India Private Limited,
No.57/65, Sippoy Lane, Avadi Road, Karayanchavadi, Chennai, Tamil Nadu.
Accreditation Standard ISO/IEC 17025: 2005
Certificate Number CC-2490 Page 5 of 7
Validity 19-11-2018 to 18-11-2020 Last Amended on 26-11-2018

Quantity
*Calibration Measurement
Sl. Measured / Range/Frequency Remarks
Capability (±)
Instrument

#
5. AC High Voltage 50 Hz
(1 to 5) kV 3.2% - 3.8% Using HV Probe Fluke 80k-40
with DMM Fluke 115

# 45 Hz
6. AC Current
(1 to 100) mA 0.093% -0.092% Using Multimeter Agilent 3458A
100mA to 1A 0.092% - 0.12%
1 kHz
(10 to 100) A 0.46% - 0.11%
100 A to 1 mA 0.11% - 0.059%
(1 to 100) mA 0.059% - 0.058%
100mA to 1A 0.058% - 0.14%

# 1 Ω to 100 kΩ 0.058% - 0.0014%


7. Resistance Using Multimeter Agilent
100 kΩ to 10MΩ 0.0014% - 0.0073% 3458A
10MΩ to 100MΩ 0.0073% - 0.0059%

#
8. Frequency 1 Hz to 100 Hz 662ppm – 120ppm Using Multimeter Agilent
3458A

100 Hz to 225MHz 9.5ppm – 116.7ppm Using Frequency counter


HP 53131A

9. Temperature
#
Simulation
Thermocouple
Type E (-200 to 800) ºC Using Multimeter (7 ½ )
0.08 °C
NI PXI 4071
Type J (-200 to 1100 ) ºC 0.08 °C
Type K (-190 to 1300) ºC 0.05 °C
Type S (150 to 1700) ºC 0.14 °C
Type T (-200 to 300) ºC 0.10 °C
Type N (-200 to 1300) ºC 0.09 °C
Type R (100 to 1700) ºC 0.07 °C
Type B (600 to 1800) ºC 0.50 °C
RTD
Pt 385, 100  0.07 °C Using Multimeter Agilent
(-190 to 800) ºC
3458A

Convenor Program Director


Laboratory Micro Precision Calibration India Private Limited,
No.57/65, Sippoy Lane, Avadi Road, Karayanchavadi, Chennai, Tamil Nadu.
Accreditation Standard ISO/IEC 17025: 2005
Certificate Number CC-2490 Page 6 of 7
Validity 19-11-2018 to 18-11-2020 Last Amended on 26-11-2018

Quantity
*Calibration Measurement
Sl. Measured / Range/Frequency Remarks
Capability (±)
Instrument

MECHANICAL CALIBRATION

I. DIMENSION (Basic Measuring Instrument, gauge etc.,)


$
Calipers
1. Using Gauge Blocks and
L.C: 0.01 mm up to 300mm 9.30 µm Caliper Checker as per
300 to 600 mm 9.40 µm IS 3651

$
2. Height Gauges up to 600 mm 8.8 µm Using Gauge Blocks and
L.C: 0.01 mm Caliper Checker as per
IS 2921

3. External Using Gauge Blocks as per


$
Micrometers up to 25 mm 1.40 µm IS 2967
L.C: 0.001 mm 25 to 150 mm 4.50 µm

$
4. Cylindrical Pins 0.5 to 20 mm 1.7 µm Using Laser Scan Micrometer
as per IS 11103

$
5. Plain Plug Gauges 0.5 to 40 mm 1.0 µm Using Gauge Blocks as per
IS 3455, IS 6174

6. Steel Ruler’s / Upto 10 m 580 √L µm Using Measuring Scale & Tape


$ calibration machine
Measuring Tape’s (Where L is meter)
II. PRESSURE INDICATING DEVICES

1. Pressure Gauge / 0 to 20 bar 0.33% of rdg Using Digital Pressure Calibrator


Calibrator / as per DKD R-6-1
Transducer /
Transmitter /
Switch / Compound
#
Gauge

2. Pressure Gauge / 0 to 700 bar 0.30% of rdg Using Digital Pressure Gauge as
Calibrator / per DKD R-6-1
Transducer /
Transmitter /
Switch / Compound
#
Gauge

#
3. Vacuum Gauges -0.68 to 0 bar 0.30% of rdg Using Digital Pressure Calibrator
as per IS 3567

Convenor Program Director


Laboratory Micro Precision Calibration India Private Limited,
No.57/65, Sippoy Lane, Avadi Road, Karayanchavadi, Chennai, Tamil Nadu.
Accreditation Standard ISO/IEC 17025: 2005
Certificate Number CC-2490 Page 7 of 7
Validity 19-11-2018 to 18-11-2020 Last Amended on 26-11-2018

Quantity
*Calibration Measurement
Sl. Measured / Range/Frequency Remarks
Capability (±)
Instrument
III. UTM, TENSION CREEP AND TORSION TESTING MACHINE

1. Universal Testing 1 kN to 50 kN 0.70% Using Precision Load cell and


Machine Digital Load indicator as per
Compression and IS 1828 (Part 1) :2015

Tension ISO 7500-1 : 2004

THERMAL CALIBRATION

1. RTD, Thermocouple, 50°C to400°C 1.9°C Using Standard RTD, Dry block
Thermistor Sensors calibrator and Digital Multimeter
with / without
Indicators,
temperature gauges,
Temperature baths,
controllers and
Transmitters with
#
sensors

2. Analog / Digital/ 15 % RH to 90 %RH 4.9%RH Using Standard Humidity


Thermo hygrometers @ 25 °C Indicator with Sensor and
/Thermo hygrographs Humidity Chamber
$
/ Humidity Sensors 10 °C to 50 °C
@ 50% 0.87°C

3. Temperature -50 °C to 400 °C 1.9°C Using Standard RTD with


Indicator of cold Indicator
chamber, Ovens,
Incubators, Furnaces,
Bath at single position

* Measurement Capability is expressed as an uncertainty () at a confidence probability of 95%


$
Only in Permanent Laboratory

Only for Site Calibration
#
The laboratory is also capable for site calibration however, the uncertainty at site depends on the
prevailing actual environmental conditions and master equipment used.