Diso Opus Operational Procedure With Parameter List
Diso Opus Operational Procedure With Parameter List
1
↓
Over range 110%
↓
Press enter key
↓
Mech. Factor 3.000
↓
Press enter key
↓
Adjustment factor 1.000
↓
Press enter key
↓
Related capacity 22000Kg
↓
Press enter key
↓
No. of L/C 1
↓
Press enter key
↓
L/C –C (mv/v) 2.85
↓
Press enter key
↓
Deadweight 6.230t (Automatically calculate, manually not possible)
↓
Press enter key
↓
L.f.t. NO
↓
Press OK Key
After few seconds again press OK Key and system return back from parameter
section.
And System showed 441: Calibration para.
↓
Then press (→) right arrow key 4 times & System shown 445: CALIBRATION
↓
Press down arrow key (↓),
↓
System shown 4451: Deadweight
↓
Press enter key
2
↓
System Bin’s Taring job automatically done and send to Deadweight data in 441:
Calibration para Dead weight section.
↓
Press Right arrow key (→) 2 times,
↓
System shown 4453: Range
↓
Press enter key
↓
Then Put the Actual Value of Known weights by manually type through keypad in
Range Block, then Press Enter Key.
After few seconds system shown Rage Ok and return back by Return key (↑) for
normal weighing display G: 0.00 T
3
4