Summary of the accident: SYSTEM ID: 211

Date of occurrence 04.11.2023 (at 11:00 am)

PES Boards – Both CB and PB (version C)

CB: 206891160001021923879152

PB: 206891160002021944818953

Saga comments: After assembly issues were sorted on April 06, 2023, the system was put for POT.
After igniting the system went into run state accelerating to 190krpm when it stopped with an error “
low rotor speed”. Follow up attempts had similar results with other error being “data comparison”.

On 11 April, attempts were renewed to run the system in non POT mode. When the system was
resetting at the first attempt of the day, the main fuse was tripped likely due to error in PS2.11. with
the condense pump checked (as similar tripping issues were found due to incorrect condense pump
cabling) and confirmed ok the next attempt showed “gid inverter failed to initialize”. Which was
followed by attempts with Marko.

Marko comments: After the error was cleared from the GRID INVERTER in full control in LabVIEW, I
tried to initialize only the grid inverter after 60 seconds connection time, the Grid Inverter Switches
did not initialized and a bang sound with a smell from the control/power board was noticed.

Marko comments: After the error was cleared from the GRID INVERTER in full control in LabVIEW, I
tried to initialize only the grid inverter after 60 seconds connection time, the Grid Inverter Switches
did not initialized and a bang sound with a smell from the control/power board was noticed.

Marko comments: After the error was cleared from the GRID INVERTER in full control in LabVIEW, I
tried to initialize only the grid inverter after 60 seconds connection time, the Grid Inverter Switches
did not initialized and a bang sound with a smell from the control/power board was noticed.