6'-0"
3'-8"
14 FINISHED CEILING HEIGHT 9'-6"
11 4'-4"
Camera Below Finished Ceiling Note:
Please note that the finished ceiling height is above the maximum camera installation height. Extender
4'-2" kit 5821337 is required to install the Junction Plate.
17'-0"
1'-8"
For Accessory Sales: (866) 281-7545 Options 1, 2, 1, 2 or mail to: gehcaccessorysales@ge.com
2'-9" 2 HCA Florida West Marion Hospital A part of HCA Florida Ocala Hospital
Ocala, FL
United States of America
EXST 7'-9"
12 1 John Proctor
3
5'-10" GE Healthcare (678) 977-4672
2'-11"
7
3'-11"
John.proctor@ge.com
6
EXAM ROOM ALL INFORMATION IN
REVOLUTION ASCEND THIS DRAWING SET IS
PRELIMINARY, SUBJECT
16 PRELIMINARY STUDY TO CHANGE!
EXST 6'-11"
18'-5"
DELIVERY DISCLAIMER
THE CUSTOMER/CONTRACTOR SHOULD: This drawing is a preliminary drawing. Site conditions and/or equipment configuration may have a significant
impact on room layout and site preparation. Final study must be done before installation of the GE equipment. GE
· Provide an area adjacent to the installation site for delivery and unloading of the GE equipment. cannot accept any responsibility for errors due to lack of information.
· Ensure that the dimensions of all doors, corridors, ceiling heights are sufficient to accommodate the
movement of GE equipment from the delivery area into the definitive installation room. The room dimensions used to create the equipment layout may originate from a previous layout and may not be
· Ensure that access routes for equipment will accommodate the weights of the equipment and any accurate as they may not have been verified on site. GE cannot take any responsibility for errors due to lack of
transportation, lifting and rigging equipment. information.
· Ensure that all necessary arrangements for stopping and unloading on public or private property not
belonging to the customer have been made. It is the responsibility of the customer to prepare the site in accordance with the specifications stated in the final
drawings. These drawings are not to be used for actual construction purposes. The company cannot take
responsibility for any damage resulting therefrom.
DIMENSIONS OF DELIVERY WITH DOLLY TRANSPORT EQUIPMENT
The customer must ensure the floor strength is sufficient to support the fixings as required. A qualified structural
EQUIPMENT DIMENSIONS WEIGHT engineer must be consulted and all work carried out according to his specifications.
LENGTH 2910 mm 115 in
GANTRY WIDTH 1290 mm 51 in 2035 kg 4486 lbs Suitable radiological protection must be determined by a qualified radiological physicist in conformation with local
HEIGHT 2000 mm 79 in regulations. GE does not take responsibility for the specification or provision of radio-protection.
LENGTH 2489 mm 98 in
GT1700V TABLE WIDTH 762 mm 30 in 576 kg 1270 lbs
HEIGHT 1143 mm 45 in
THE UNDERSIGNED, HEREBY CERTIFIES THAT I HAVE READ AND APPROVED THE PLANS IN THIS DOCUMENT.
HCA Florida West Marion Hospital A part of HCA Florida Ocala Hospital REVOLUTION ASCEND CT-M316512-PRE-00-A.DWG Rev A Date 08/Jun/2022 Details 02/02