FESEM utilizes high vacuum conditions and uses electron beam to form an
image, the sample must be as per the specifications given below:
➢ Sample should be in dry condition and it can be powder, film, pellet, coating
etc.
➢ All water/volatile contents must be removed from the sample because water
would vaporize in vacuum.
➢ In case of magnetic samples, the details about the nature of the sample should
be provided to the operator before submitting the sample for FESEM analysis.
➢ Samples which are not compatible with high vacuum and samples which are
not stable for electron beam exposure cannot be analysed.
➢ For a better analysis basic details about the sample should be given.
➢ Samples received for the analysis will not be returned back. Please provide only
the required quantity of the sample for analysis (for example 200 mg of
powder).
➢ If you need the samples back after analysis, please inform us during the sample
submission itself.