01/2008:20210
PROCEDURE
2.2.10. VISCOSITY - ROTATING
Select a capillary viscometer of appropriate size to obtain a
minimum flow time of 200 s. VISCOMETER METHOD
Calibration The principle of the method is to measure the force acting on
a rotor (torque) when it rotates at a constant angular velocity
Capillary viscometers are calibrated at regular intervals as (rotational speed) in a liquid. Rotating viscometers are used
defined in the quality management system and dictated by the for measuring the viscosity of Newtonian (shear-independent
frequency of use of the equipment and the application. viscosity) or non-Newtonian liquids (shear dependent
Calibrate the instrument at the temperature used for the viscosity or apparent viscosity). Rotating viscometers can be
measurement by using at least 2 certified reference materials divided in 2 groups, namely absolute and relative viscometers.
matching the viscosity range of the viscometer. In absolute viscometers the flow in the measuring geometry is
(2) The European Pharmacopoeia describes the system proposed by the International Organization for Standardization (ISO)