Certificate of Calibration
Date of Issue 13 October 2021 Certifieate No + UTJIPIES-MM/211001476
Issued by UCAL TECH’S (M) SDN.BHDycussy) Page 1 of 2
No 8,8A Jalan Shahbandar 2,
‘Taman Ungku Tun Aminah,
81300, Johor Bahru,
Johor Darul Takzim,
Tel: 07-5541466 Fax : 07-5542466
Email :es@ucaltechs.com
‘Submitted by Conmax Resources Sdn Bhd Date Received 13 October 2021
Lot 343-1 Jalan Pegerang Sungai Rengit Job No 4321100312-004/211001476
81620 Kota Tinggi Johor DocumentNo 20RI7:MASO 1 A
A005
Instrument Torque Wrench Model ‘STMX3100
Manufacturer Novatork Serial No. 2 ARO40983
Instrument Condition When Receive
1. Physically in good condition
2.CustomerID-—: NA
Calibration Status :
Calibration Date 13 October 2021
Requested Calibration Due Date 12 October 2022
Recalibration date requested by
this instrument to drift out of
Calibration Venue :-
‘Above Mentioned UCAL TECH'S
Calibration Method :
Instrument Condition When Returned :
1, Calibrated and Test Serviceable
Environment Condition :~
‘Avg Temperature
Ayg Relative Humidity
235°C
554 15%
customer. The user should be aware that there are a number of factors that may caused
libration before that specified calibration interval has expired.
‘Address and refer to Job No.
This Instrument was calibrated using the Calibration Procedure No : CWI-GM 004 (Issue No : 1-0-0 )
BS EN ISO 6789 :2003
Instrument Type Serial No. CaLDue Date Cal.Cert.No
Torque Analyzer UCT-103 10.08.2022 UT21/NS-Ha024 UCALNMIM
Standard Weights, UCT-020-030 24.08.2022 ‘NMIM-2954-M-20 NMIM
‘Thermohygrometer UCT-366 09.03.2022 UTHBAR/21C/006 UCAL/NMIM
The Standard instruments used in this calibration are traceable to either the National Standards maintained at the National
Metrology Laboratory, SIRIM Berhad or other recognized International Standard Laboratories.
Calibration Sticker No. 0312-004/01476
Measurement Uncertainty Refer to calibration data,
Coverage Factor k=2
‘The uncertainty calculation is based on the ISO Guide to the Expression of Uncertainty in measurement.
The expanded uncertainties are based on an estimated confidence probability of approximately 95%
factor of k = 2 unless stated otherwise
~ Technical Manager
> Quality Manager
Approved Signatory and P
Manivannan a/l Subramaniam
Bhavani a/p Gopal Krishnan
and have a coverage
x
LEE
Approved Signatory
Select ()
Zz
‘The uncertainties are for a confidence probability of approximately 95%
‘This cerca is issued in accordance with the measurement capabilly of UCAL TECHS laboratory and is Taceabiiy
to recognized national standards and tothe units of measurement realized atthe corresponding national laboratory.
‘Copyright of his certificate is owned by our laboratory and may not be reproduced other than in full except withthe por
mitten approval ofthe Head ofthe Laboratory.Certificate of Calibration
Date of Issue 13 October 2021 Certificate No UTJIP/ES-MM/211001476
Instrument Torque Wrench Page t.2 of 2
Serial No, £ ARO40983
Calibration Data
Force Measurement Verification -Clockwise
Manufacturer Range: 20 N.m to 100 Nm.
Calibration Range 20.N.m to 100 N.m
Resolution 208
Unit : Nm
ae ‘Applied ] Measured ‘Specified Limit Tnstrumental ‘Measurement
- Value Value Tower Upper Error Uncertainty
20 20 20.5 160 24.0 os ca ol Nm
40 40 40.7 36.0 44.0 07 a Nm
60 60 60.9 56.0 64.0 09 eal Nm
80 80 813 76.0 84.0 13 + 1 Nm
100 100 1018 96.0 104.0 Lg. eo] Nam
Note :- 1) Instrumental Error = Measured Value - Applied Value
2) ‘The specifications / tolerance provided is not regarded as being part of opinions and interpretation
3) The above result reported is only related to this item
4) Customer holds the rights to decide the usage based on their own specification / toleranc
5) Specified limit reference (s) to manufacturer specification
Calibrated By : Firdaus Review By
imployee No : BOSE
Muhammad Syafiq Misban
‘Tho uncertainties are for a confidence probability of approximately 95%
‘Ts cortical is fesued in accordance wit the measurement capably of UCAL TECHS laboratoy and is Waceabaiy
to tecognized national standards and tothe units of measurement realized at the coresponcing national laboratory.
Copyright ofthis cortitcate is owned by our laboratory and may not be reproduced other than in full exoopt withthe prior
writen approval ofthe Head of the Laboratory