CALIBRATION CERTIFICATE Page 1 of 1
ULR No:CC279621400005859F
CERTIFICATE NO : Tii210313-8-4
[Customer Name & Address: ‘SRF No : TH210313-8
IM/s.Anpam Engineering, SRF Date : 13-Mar-2021
19/1, Kumaramangalam Industrial Estate, Calibration Date 13-Mar-2021
Mathur, Pudukkottai. Calibration Due Date _: 13-Mar-2022
[Details of Test Instrument:
Description Digital DC Ammeter ‘Range 70 to 400A 75mV
Make Lianya Operating Range
Model No. - Resolution
Mic Serial No. : AB/ET/21/104 Mic ID :
ID No / Tag No - Location -
(Calibrated at Site) ‘Condition On Receipt: Good
[Work Instruction No : TSPLIWHETCALI006 Based on Standard: IS 1248(Part I to 9):2003/Rev
[Environmental Details: Temperature: 28.2 °C Relative Humidity: 49 %RH.
[Details of Master Used [Master
[Description Multifunction Calibrator
SLNo. 1978909
[Certificate No. [20/99-01
[Valid up to 3-Sep-2021
ELECTROTECHNICAL CALIBRATION
CALIBRATION DATA
DC CURRENT
‘Range Master Equivalent [Test Reading Mean] Deviation Expanded
A mV A A A A
15.00 80 81 1 0.60
30.00 160 162 2 0.60
400 45.00 240 242 2 0.60
60.00 320 322 2 0.61
75.00 400 403 3 0.62
Uncertainty of measurement is at 95.45% confidence level(k=2)
1 This certificate refers to the measurements for the instrument in as received condition and under
stated environment.
2.Certificate shall not be reproduced except in full without the written approval of Director.
3.*The recalibration interval - as advised by the customer.
4,The standard used for calibration is traceable to Accredited laboratory.