CALIBRATION CERTIFICATE Page | of 1
ULR No:CC279621400005866F
CERTIFICATE NO : Tii210313-8-11
[Customer Name & Address: SRF No PTH210313-8
IM/s.Anpam Engineering, SRF Date : 13-Mar-2021
9/1, Kumaramangalam Industrial Estate, Calibration Date 2 13-Mar-2021
Mathur, Pudukkottai, Calibration Due Date _: 13-Mar-2022
[Details of Test Instrument:
[Description + Digital DC Ammeter Range £0 to 400A7 75miV
Make : Lianya Operating Range
[Model No. Resolution 1A
IM/c Serial No. : ABET/21/111 Mic ID i
{ID No / Tag No = Location ie
[Calibrated at Site Condition On Receipt _: Good
[Work Instruction No TSPLWIETCAL006 Based on Standard: IS 1248(Part 1 10 9):2003/Rev
[Environmental Details 27.5 °C Relative Humidity 33% RH
Details of Master Used
[Description [Multifunction Calibrator
SLNo. 1978909
[Certificate No. 20799-01
[Valid up to 3-Sep-2021
ELECTROTECHNICAL CALIBRATION
CALIBRATION DATA.
DC CURRENT
‘Range Master Equivalent [Test Reading Mean] Deviation Expanded
A mV A A A A
15.00 80 81 T 0.60
30.00 160 163 3 0.60
400 45.00 240 243 3 0.60
60.00 320 323 3 0.61
75.00 400 404 4 0.62
Uncertainty of measurement is at 95.45% confidence level(k=2)
Note:
1.This certificate refers to the measurements for the instrument in as received condition and under
stated environment.
2.Certificate shall not be reproduced except in full without the written approval of Director.
3.*The recalibration interval - as advised by the customer.
4.The standard used for calibration is traceable to Accredited laboratory.
Team Member Authorised Signatory
: Cs
Gay D
Senior Engineer - Calibration, Manager - Calibration