TI TECHNO TESTING SERVICES PVT, LID,
(NABL Accredited Testing & Calibration Facilities)
(PAN : AADCT89420, TIN : 33561667359, Corporate ID No : U74900TN2011PTC126536 & ST : AADCT89420SD001, GSTIN : 33AADCTB94201ZN)
CALIBRATION CERTIFICATE Page 1 of |
ULR No:CC279621400005864F
CERTIFICATE, NO : TH210313-8-9
[Customer Name & Addres ‘SRF No 2 Ti210313-8
IM/s.Anpam Engineering, SRF Date : 13-Mar-2021
19/1, Kumaramangalam Industrial Estate, Calibration Date 2 13-Mar-2021
(Mathur, Pudukkottai. Calibration Due Date: 13-Mar-2022
Details of Test Instrument:
[Description + Digital DC Ammeter Range 70 to 200A/ 75mV
Make Lianya Operating Range :
Model No. Resolution
Mic Serial No. ABJET/21/109 Mic ID :
IID No / Tag No Location ie
[Calibrated at Site Condition On Receipt _: Good
[Work Instruction No TSPLIWVETCAL006 _Basedon Standard ___: 1S 1248(Part I to 9)-2003/Rev
[Environmental Details: ‘Temperature: 27.9 °C Relative Humidit 32_%RH
[Details of Master Used [Master
[Description [Multifunction Calibrator
[SLNo. 1978909
(Certificate No. [20799-01
[Valid up to 3-Sep-2021
ELECTROTECHNICAL CALIBRATION
CALIBRATION DATA
DC CURRENT
Range Master Equivalent [Test Reading Mean] Deviation Expanded
A mV) A A A A
15.00 40 40 0 0.59
30.00 80 81 1 0.60
200 45.00 120 121 1 0.60
60.00 160 161 1 0.61
75.00 200 202 2 0.61
Uncertainty of measurement is at 95.45% confidence level(k=2)
Note :
1 This certificate refers to the measurements for the instrument in as received condition and under
stated environment,
2.Certificate shall not be reproduced except in full without the written approval of Director.
3.*The recalibration interval - as advised by the customer.
4,The standard used for calibration is traceable to Accredited laboratory.
‘Team Member Authorised Signatory
z wie
‘Senior Engineer - Calibratior | Manager - Calibration