The document provides instructions for checking and adjusting sampling flow if a minor alarm occurs. It explains that a FAST alarm means intake water flow is too fast while a LOW alarm means the filling speed is too slow. It then lists the step-by-step procedure for checking the sampling flow by using the service and mode buttons to display the flow rate and for adjusting it by turning the regulator clockwise or counterclockwise depending on if the rate is too high or low.
The document provides instructions for checking and adjusting sampling flow if a minor alarm occurs. It explains that a FAST alarm means intake water flow is too fast while a LOW alarm means the filling speed is too slow. It then lists the step-by-step procedure for checking the sampling flow by using the service and mode buttons to display the flow rate and for adjusting it by turning the regulator clockwise or counterclockwise depending on if the rate is too high or low.
The document provides instructions for checking and adjusting sampling flow if a minor alarm occurs. It explains that a FAST alarm means intake water flow is too fast while a LOW alarm means the filling speed is too slow. It then lists the step-by-step procedure for checking the sampling flow by using the service and mode buttons to display the flow rate and for adjusting it by turning the regulator clockwise or counterclockwise depending on if the rate is too high or low.
please check and adjust the sampling flow as following procedure. The Meaning of Minor Alarm (FAST/LOW) FAST Intake water flow is too fast. LOW The filling speed of sample water to cuvette is too slow.
The procedure for checking the sampling flow.
1. Press the [SERVICE] button 2. Wait for the display to read [HOLD], then press [MODE/EXIT]. Display will show [FLOW] with the number 0. 3. Press either the [▲] or [▼] button. TRO UNIT will pulse in sample water to the cuvette wile a count is displayed on the screen. The display will show one of three massages, HI,LO,Good. The procedure for adjusting the sampling flow. 1. Remove the vinyl cap and loose the locking nut. HI LO 2. Turn the regulator clockwise if the message was LO. Turn the regulator counterclockwise if the message was HI. Press either the [▲]or [▼] button to check the sampling flow every turning the regulator . Adjust the sampling flow while checking it. Please note that only ¼ turn adjustments should be made to the regulator on each attempt.