2 Copyright / Version / Disclaimer
1Copyright / Version / Disclaimer
Copyright
“© Siemens, 2010“ refers to the copyright of a Siemens entity such as Siemens Aktienge-
sellschaft - Germany, Siemens Mindit Magnetic Resonance Ltd. - China, Siemens Shang-
hai Medical Equipment Ltd. - China, Siemens Medical Solutions USA Inc. - USA and/or
Siemens Healthcare Diagnostics Inc. - USA.
Document Version
Siemens reserves the right to change its products and services at any time.
In addition, manuals are subject to change without notice. The hardcopy documents corre-
spond to the version at the time of system delivery and/or printout. Versions to hardcopy
documentation are not automatically distributed. Please contact your local Siemens office
to order current version or refer to our website http://www.healthcare.siemens.com.
Disclaimer
Siemens provides this documentation “as is“ without the assumption of any liability under
any theory of law.
The installation and service of equipment described herein requires superior understand-
ing of our equipment and may only be performed by qualified personnel who are specially
trained for such installation and/or service.
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Programming in the Memoskop CX . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Backup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Final Work Steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Function test / settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
These instructions are valid for installing a connector for an external monitor at the SIRE-
MOBIL Compact L system, beginning with serial number 11000.
Using the retrofit kit (material number 7142172), an additional connector (SUB-D15) is
installed at the back of the monitor trolley.
A VGA signal (60 Hz, 1280 x 1024 resolution, 4:3 aspect ratio) is transmitted via this con-
nector. The video signal is galvanically isolated.
During the Memoskop CX user set up, you can select whether the video signal from monitor
A or monitor B will be transmitted.
Required parts 0
Required documents 0
Bei einer Gefahr, die bei Nichtvermeidung zum Tod oder zu einer
schweren Körperverletzung führen kann.
Please note:
¹ The product-specific safety notes in these instruc-
tions,
¹ The general safety information in TD00-000.860.01...,
and
¹ The safety notes in accordance with ARTD Part 2.
Follow the instructions in the "Safety Rules for Installation and Repair" (ARTD-002.731.17
...).
The protective conductor resistance must be measured after every intervention in the sys-
tem.
However, documentation of the measured values is required only during periodic safety
checks.
If parts or components that can significantly influence the protective conductor resistance
(e.g., replacement of the power cable, replacement of the power-up module, replacement
of multi-pin connection cables that also create the protective conductor connection
between system parts (e.g., monitor cable or C-arm cable)) are replaced or if protective
conductor connections have been repaired, the protective conductor resistance must be
measured. The values must be documented and assessed in the protective conductor
resistance report.
NOTE For evaluation purposes, the first measured value and the
values documented during previous maintenance or safety
checks must be compared to the measured values. A sudden
or unexpected increase in the measured values may indicate
a defect in the protective conductor connections - even if the
limit value of 0.2 ohms is not exceeded. (Protective conduc-
tor or contacts).
The measurement must be performed in accordance with DIN VDE 0751, Part 1 (refer to
ARTD Part 2). The protective conductor resistance for all touchable conductive parts must
be measured during the normal operating state of the system.
Make sure that control cables or data lines between the components of the system are not
mistaken for protective conductor connections.
During the measurement, the power cable and additional connection cables that also cre-
ate the protective conductor connection between system parts (e.g. monitor cable between
the basic unit and monitor trolley) must be moved section by section to detect cable breaks.
The protective conductor resistance must not exceed 0.2 ohms.
Fig. 2: Measuring circuit for measuring the protective conductor resistance for units that are
disconnected from power, in compliance with DIN VDE 0751-1/2001-10, Fig. C2.
Pos. 1 = System
Pos. 2 = Application part type B (if available)
Pos. 3 = Measurement setup (integrated into measuring device)
• Switch off the SIREMOBIL Compact L and disconnect the power plug.
• Loosen the 4 panel screws for the upper cover at the back of the monitor trolley and
remove the cover (1/Fig. 3 / p. 11).
NOTE (Fig. 3 / p. 11) shows the new adapter plate (with RJ 45 and
SUB-D 15 connectors).
• Install the new adapter plate (fully assembled with connectors (RJ 45 and SUB-D 15)
and VGA video cable) using the 4 existing panel screws.
• Route the VGA video cable in the right monitor slot (viewed from the rear) to the
Memoskop CX and connect to the "VGA Out" connector.
• Reconnect the network cable to the RJ 45 adapter.
Backup 0