You are on page 1of 7

Coccolini et al.

World Journal of Emergency Surgery (2020) 15:25 Page 4 of 7

Table 3 COVID-19 surgical patients’ management must be preemptively prepared in a sterilizable steel wire
Key aspects in COVID-19 surgical patient management basket. Dedicated IRHW containers must be used for in-
All suspected or infected patients must be managed with the maximum fected and sharp disposable instruments. Alcoholic solu-
attention. tion for hand hygiene must always be available. Avoiding
All personnel in contact with the patient must wear PPE. non-strictly necessary commonly used non-disposable
Transfers must be protected. devices is recommended. Disposable material in general
should be preferred, including linen. All operators (i.e.,
Infected patients must be moved as little as possible through the
hospital. surgeon, anesthetist, nurses, technicians) should enter
Transfer routes must be precisely planned and be as short as possible.
the OR timely, aiming to minimize time spent within the
OR itself. Once in the OR, they should not leave until
The COVID operating area should be in a dedicated and possibly
separate area. the operation is completed, and once out they should
not re-enter.
COVID operating room must be dedicated and as close as possible to
the entrance of the theater block.
Personnel dressing
Disposable material should be preferred.
All operators must wear the required PPE before meet-
Minimal material should be used for each intervention.
ing the infected patient. The patient’s receiving
Transport personnel should be the same from transport origin to personnel inside the COA filter area must perform hand
destination.
hygiene and wear full PPE.
Once the patient has entered, the OR doors must be closed.
While taking care of infected patients, gloves should
Operators (i.e., surgeon, anesthetist, nurses, technicians) should enter the be changed immediately after contact with infected ma-
OR in a timely manner to minimize exposure to infected patients.
terial (objects, surfaces, etc.) or if any damage occurs.
Personnel involved in the intervention should not leave the OR during Operator with a beard should exert special attention to
the procedure.
the fit of the mask ensuring adequate protection.
High OR air exchange cycles are recommended (> 25 exchanges/h).
Some procedures likely to generate aerosolized parti-
Clinical documentation must remain outside the OR cles have been associated with increased coronavirus
At the end of each intervention all disposable materials must be transmission: tracheal intubation, non-invasive ventila-
disposed of and all surfaces and electromedical devices accurately tion, tracheostomy, cardiopulmonary resuscitation, and
cleaned and disinfected.
manual ventilation before intubation and bronchoscopy
PPE must be removed and disposed of outside the OR in dedicated
[5, 6]. An FFP3 mask should be therefore worn by oper-
doffing areas ensuring the virus is not transmitted to the healthcare
worker. ators working closer to the patient during these
OR and surrounding donning/doffing areas must be sanitized as soon as procedures.
possible after each procedure. Given the conjunctiva’s susceptibility to viral transmis-
After each procedure, all involved personnel, whenever possible, should sion, it is important to wear visors or goggles to protect
shower. the eyes from potential exposure of viral particles [7].
Recovery phase after surgery must be done in OR, before transfer the
ward/ICU. Anesthesiologic consideration
Careful anesthesiologic planning is recommended to
document consultation is discouraged and should be minimize any infection potentially associated with unex-
minimized. pected complex endotracheal intubation procedures. A
more liberal use of intubation might be justified in pa-
Operating room preparation tients with acute respiratory failure, bypassing non-
Negative pressure ORs would be ideal to minimize infec- invasive ventilation techniques (e.g., CPAP or biPAP) in
tion risk [3, 4]. However, ORs are normally designed to order to minimize the transmission risks [5]. Disposable
have positive pressure air circulation. A high air ex- airway equipment should be preferred. Medical and
change cycle rate (≥ 25 cycles/h) contributes to effect- nursing staff must be equipped with FFP3 filters during
ively reduce the viral load within ORs [2]. Equipment laryngoscopy and intubation [5]. Intubations techniques
kept in each OR must be minimized to what is strictly with the highest chance of first-time success should be
necessary on a case to case basis. Once the operation preferred to avoid repeated airway instrumentation [4,
starts, all efforts must be made to use what is available 5]. Awake intubation techniques should be avoided. At
in the room and minimize staff transiting in and out the the end of these procedures, all staff directly performing
OR, in order to minimize infection risk. Standard the procedure must immediately replace the first pair of
anesthetic trolleys should be replaced with dedicated gloves and other PPEs in case heavy contamination risk
pre-prepared ones with minimal but adequate stock. All exists (i.e., in the event that vomiting, coughing, or else
required surgical material (i.e., stitches, scalpel blades) has occurred). Fiberscope intubation, unless specifically
Coccolini et al. World Journal of Emergency Surgery (2020) 15:25 Page 5 of 7

indicated, should be avoided as it may generate aerosoli- replacing the surgical mask with FFP2 filter and wearing
zation [5]. Rapid sequence intubation (RSI) should be long shoe covers before doing so. All personnel in direct
considered to avoid manual ventilation and potential contact with the patient must wear a double pair of
aerosolization. If manual ventilation is required, small gloves at all times, even while operating. After the pa-
current volumes should be used. If available, a closed tient left the OR, logistics should allow as much time as
suction system should be preferred during airway aspir- possible before the next procedure takes place, to reduce
ation. Disposable covers should be used whenever pos- possible air contamination. This time depends on the
sible to reduce equipment contamination. If a patient is number of air exchanges/hour of the specific room. Air
transferred directly from the intensive care unit, a dedi- exchange cycles should be increased whenever possible
cated transport ventilator should be utilized. In order to to ≥ 25 exchanges/h [2]. After the case, all areas at risk
reduce aerosolization risks, the gas flow should be of contamination must be cleaned and disinfected (Table
turned off and the endotracheal tube clamped with for- 2). Efforts should be made to minimize the contamin-
ceps when switching from the portable device to the OR ation risk associated with specimens sent to the path-
ventilator [4]. When possible, a dedicated ventilator ology department. No data currently exist on COVID-19
should be used in the OR for general anesthesia in posi- viral load in bodily fluids or tissue samples.
tive or suspected positive COVID-19 patients. Invasive
procedures like for example the placement of intercostal PPE undressing/removal
catheters, central venous catheters, or similar should be Staff not directly involved in the patient’s care should
performed at the patient’s bedside, rather than in the leave the OR at the end of the operation and remove all
OR. When a general anesthetic is required, a HEPA PPEs in a dedicated doffing area following the sequence
(high-efficiency particulate air) filter should be con- described below. A clean area should be accessed only
nected to the patient end of the breathing circuit and after the doffing procedure is complete. All used PPEs
another one between the expiratory limb and the must be disposed of through IRHW containers. Scrubs
anesthetic machine [2, 6]. Alternatively, for pediatric pa- must be replaced after each procedure following shower-
tients or other patients in whom additional dead space ing, whenever possible. Personnel responsible for trans-
or the weight of the filter may be problematic, the HEPA ferring the patient away from the operating room must
filter must be placed at the expiratory end of the circuit follow separate access routes and wear PPEs different
(before the exhalation re-enters the ventilator). The gas from the ones worn in the OR.
sampling tube must also be protected by a HEPA filter.
Both HEPA filters and soda lime must be changed after Instructions for PPE removal
each case [4]. At the end of the surgery, during the re- The healthcare professional must take all care not to be-
covery phase, the patient must be assisted directly in the come infected while removing PPE; this must be done
OR until ready to be transferred back to the inpatients through an adequate procedure preventing re-
place of stay. The time patients spend returning to wards contamination of the operator's clothing and hands. The
must be reduced in order to minimize contact between first pair of gloves is likely to be heavily contaminated
COVID-positive patients and the surrounding and must be removed first. All other PPEs must be con-
environment. sidered infected as well and removed with care during
the doffing procedure, especially if the patient had a
Intraoperative management cough. Protective suite, shoe cover, and head cap must
The OR door must be kept closed at all times and clear be subsequently removed. Face mask and glasses must
signs should discourage unnecessarily entering the room. be then removed, taking care to handle the face mask by
Supplying materials to the OR during surgery should the ear laces and without touching its external side. The
also be discouraged. The scout nurse, in collaboration second pair of gloves must be removed as the very last
with the operating surgeon, should anticipate what is PPE and hands disinfection with hydro-alcoholic solu-
needed during the operation before the same starts. Sur- tion must be accurately performed immediately after.
geons should preferably perform the operation with
what is available in the OR once the operation started. Environmental sanitization
Any essential retrieval of necessary equipment should be The OR and surrounding exchange areas must be sani-
done by staff outside the OR. Personnel present in the tized as soon as possible after each procedure, with par-
OR during surgery must not leave the room. Electrome- ticular attention to all objects used when caring for
dical devices (i.e., ultrasound) and surfaces must be used infected patients. Similarly, all areas where COVID pa-
with adequate protective cover and adequately sanitized tients have transited must be carefully sanitized too. All
at the end of the operation. The surgical team will drape personnel must contribute to maintain a clean environ-
the patient according to the surgical procedure, ment including floors and surfaces in general. All
9. Doremalen ND, Bushmaker T, Morris DH, Holbrook MG, Gamble A,
Williamson BN, et al. Aerosol and surface stability of SARS-CoV-2 as
compared with SARS-CoV-1. New Engl J Med. 2020; [Epub ahead of print].

Publisher’s Note
Springer Nature remains neutral with regard to jurisdictional claims in
published maps and institutional affiliations.

You might also like