Professional Documents
Culture Documents
3. Input from Schlumberger worldwide expertise on sample bottles as close as possible to the gas inlet to reduce
H2S sampling with OH-WFT through Schlumberger’s the exposure area of the flowlines. The pumpout was placed
local coordinator. above the sample chambers and reverse low shock sampling
4. Interaction with ExxonMobil Upstream Research techniques were employed. The sketch of the OH-WFT tool
Company. used in well W-2 is presented as Figure 1.
Based on our findings, it was proposed to obtain downhole
samples from the Khuff zones (Z1-Z4) using OH-WFT and
CH-WFT in the appraisal well W-2. The primary objective for
the proposed OH-WFT/CH-WFT sampling in the W-2
appraisal well was to investigate the feasibility of using OH-
WFT/CH-WFT sampling in future development wells for
accurate H2S and CO2 measurments while the secondary
objective was to obtain representative single phase samples for
fluid characterization studies. To achieve these objectives,
surface samples were also acquired in addition to the
downhole samples collected with both open-hole and cased-
hole formation testers. The H2S and CO2 contents derived
from the three techniques were compared to establish the
accuracy of the OH-WFT/CH-WFT measurements. The CH-
WFT sampling technology was evaluated because it offers the
advantages and flexibility of the OH-WFT sampling
technology at significantly lower risk than the OH-WFT
sampling. Running OH-WFT in high pressure differential
environment is a risky operation, especially in deviated wells,
and may lead to tool or cable sticking.
3. Start pump-out. Monitor the probe response were collected at two depth locations in Z2 & Z3 formations.
(hydrostatic pressure if unset, formation pressure if These two depths were same depth of OH-WFT. The H2S &
set). Observe cleanup process on Fluid Analyzer. CO2 estimates from the CH-WFT were compared to both the
4. Pump-out at least 100 liters (preferably 130-150 OH-WFT and DST sampling results to evaluate the accuracy
liters) of gas at the first station before taking the first of this technique.
sample in the uncompensated bottle. This is to
saturate the flowline and reduce the anticipated H2S
scavenging caused by the flowline.
5. Pump another ~ 5 minutes or until stable flowing
conditions and take the second sample in the
uncompensated bottle.
6. Pump another ~ 5 minutes or until stable flowing
conditions and take the third sample in the pressure
compensated bottle.
7. Wait for 5 minutes, deflate packer. Wait for another 5
minutes to allow for hydrostatic pressure
stabilization.
8. Circulate bottoms-up & condition the mud. Move to
next sampling location.
9. For any subsequent station, pump-out at least 100
liters gas before collecting any sample.
10. Repeat steps 1-7 in another location within the same
reservoir to get good coverage over each reservoir. Figure 2: Cased-hole formation tester string for W-2
11. Repeat steps 1-8 in each of the Khuff reservoirs (Z1
to Z4). Sampling Procedure for CH-WFT
The following sampling guideline decision tree was made to
The dual packer OH-WFT was run on drill pipe with 18 obtain downhole gas samples with cased-hole formation tester
sampling bottles in well W-2. Six stations were planned to in W-2 well:
acquire samples from the Khuff reservoirs. The exact depths
of these stations were selected based on open-hole logs
evaluation as well as OH-WFT pressure measurements
conducted in an earlier run. Whenever possible, the most
permeable interval was selected.
It was planned that 3 samples will be acquired at each
station (two 450cc uncompensated pressure bottles and one
250cc pressure compensated bottle). However, after the first
two stations, the plan was slightly modified to incorporate
operational needs. In the pressure compensated bottles, the
samples were collected at the reservoir pressure and
temperature and then the bottles were pressurized to about
1500-2000 psi above the reservoir pressure to ensure that the
samples remain in a single phase condition until opened for
analysis in the lab. The uncompensated bottles were one air
cushion and one water cushion at each station. In the
uncompensated bottles there was no provision for pressure
compensation and the sample pressure drops down with the
decline of the ambient temperature. The air cushion
uncompensated sample at each depth location was analyzed
on-site to eliminate any H2S scavenging during transportation.
All samples, including the air cushion samples that were
analyzed on-site, were then taken to the lab for accurate
Figure 3: Decision tree for sampling with cased-hole formation
measurement of gas liquid ratio, CGR and NHCG contents. tester
The NHCG results from OH-WFT (on-site and lab), CH- OH-WFT on-
site
OH-WFT Lab OH-WFT Lab
OH-WFT Lab CH-WFT on-
(SPMC) site
CH-WFT Lab CH-WFT Lab CH-WFT Lab DST Lab
WFT (on-site and lab) and DST (on-site) were compared with N2
H2S
N/A
0
-0.22
-0.10
-0.35
-0.24
-0.22
-0.16
-0.22
N/
0A
-0.27
-0.12
-0.28
-0.19
-0.20
-0.13
-0.19
0.00
0.00
the DST (lab) results. The DST lab results are considered as a CO2 -0.84 -0.57 -0.30 -0.35 -0.34 -0.27 -0.24 -0.25 0.00
OH-WFT Lab
4 0.1
3.5 0
3 -0.1
2.5 -0.2
2 -0.3
1.5 -0.4
1 -0.5
0.5 -0.6
0 -0.7
-0.5 -0.8
OH-WFT OH-WFT OH-WFT CH-WFT CH-WFT CH-WFT CH-WFT DST on- DST Lab OH-WFT on-site OH-WFT Lab OH-WFT Lab OH-WFT Lab DST on-site DST Lab
on-site Lab Lab on-site * Lab Lab Lab site (SPMC)
N2 0
N/A -0.07 -0.08 3.86 0.04 0.01 0.02 0
N/A 0.00 N2 0
N/A -0.02 -0.01 -0.01 0
N/A 0
H2S 0.07 0.01 0.10 0.57 0.06 0.06 0.01 0.11 0.00 H2S -0.30 -0.27 -0.17 -0.14 0.09 0.00
CO2 -0.32 -0.29 -0.10 0.43 -0.16 -0.16 -0.15 -0.07 0.00 CO2 -0.74 -0.50 -0.30 -0.28 -0.09 0.00
The following observations are made on Khuff Z3 samples: The following observations are made on Khuff Z4 samples:
1. Khuff Z4 was the first sampling station during the
1. There is good agreement between the H2S contents OH-WFT sampling run. The increasing percentage of
obtained from the OH-WFT sample analyzed on-site the H2S and CO2 measured in the samples with time
and the OH-WFT samples taken to the laboratory. suggests that the OH-WFT tool and its associated
Since Z3 was the second station during the open-hole flowline were still adsorbing H2S and CO2 on its
OH-WFT run, it can be assumed that the OH-WFT surface during the collection of three samples at this
tool flowline has already been saturated with H2S and station, see Figure 8. Had a larger volume of gas been
no more scavenging exist. This gives an indication allowed to flow through the tool before collecting the
that at least 150 liters of gas need to be pumped-out sample, it is likely that we would have measured a
initially to eliminate the effect of scavenging caused higher % of H2S & CO2 closer to that measured by
by the OH-WFT tool. DST. Thus, in future sampling operations with OH-
2. The three CH-WFT samples analyzed on the WFT, larger pump-out volume (150 liters) should be
laboratory showed very consistent H2S content. allowed at the first sampling station before collecting
However, on-site measurement performed on one of samples.
the samples showed higher H2S content and
significantly higher CO2 and N2. This one odd result 2.7 1.20
probably indicates an error in on-site analysis due to
air contamination.
3. DST sample analysis on-site as well as in the
CO2 Mole %
H2S Mole %
laboratory showed practically the same H2S
concentration as measured in OH-WFT and CH-WFT
samples.
4. Excellent match on H2S and NHCG content
measurements exist among the three methods,
indicating the feasibility of OH-WFT and CH-WFT
sampling techniques in measuring the H2S and 1.7 0.00
NHCG content. 0 50 100 150
Cumulative Pump-out Volume, Liters
CO2 H2S
Khuff Z4: Figure 8: Results of H2S and CO2 contents from OH-WFT
Downhole samples were collected with OH-WFT in Z4 in
addition to the surface DST samples. Cased-hole samples were
not collected in this Khuff formation. Figure 7 summarises the 2. The first OH-WFT sample was collected in a sample
NHCG measurement for each sample. bottle with water cushion. It was analyzed on-site for
NHCG measurements before sending to the
laboratory. On-site measurements were repeated at
least three times and showed consistent values.
3. The same water cushion sample along with the other
two samples was sent to the laboratory for further
6 IPTC 10622
DST
and the internal OH-WFT modules. Also, laboratory
measurements showed the presence of some filtrate
water and mud in the samples, another possible
source of scavenging.
4. A total of 110 liters were pumped into the wellbore
prior to collecting samples with open-hole OH-WFT. 0.00
This suggests that even after pumping-out 92 liters 0.00
0.00 1.00
gas through the flow line (18 liters were mud), the OH-WFT
OH-WFT vs. DST H2S Zone 4 Zone 2
scavenging effects did not completely disappear. For
future OH-WFT jobs, it is recommended to pump-out Figure 10: Comparison of results of OH-WFT and DST
longer at high mobility location.
5. The DST on-site measurement showed two levels of
H2S contents corresponding to two different flow Results of PVT Analysis
rates. At the low flow arte, the H2S content was about PVT analysis was done on five samples from OH-WFT and
0.05 mole% lower than the H2S content measured at two samples from CH-WFT. The computed dew point
samples taken at high flow rate. This offers another pressures from all four uncontaminated samples (3 OH & 1
possible explanation for the discrepancy between the CH) were representative and as per expectations while the
OH-WFT sampling results and the DST results. If three mud contaminated samples gave dew points much higher
differences exist in the NHCG of different streaks of than the expected values.
a Khuff zone, then OH-WFT sampling technique The results of detailed composition analysis of OH-WFT
needs to be adjusted to allow collection of samples were compared with DST sample collected over the same
from different streaks/ facies in each zone. This interval, shown below in Figure 11. The results from the two
should provide good approximation for the range of techniques were in good agreement.
H2S expected from a certain zone. A single sample
collected through OH-WFT in a zone may not be able
to accurately estimate the range of NHCG contents.
N 2 O 2 H 2 S C 1 C 2 C 3 i-C 4 -C 4 i- C 5 -C 5 C 6 C 7 C 8 C 9 C 1 0 C 1 1 C 1 2 C 1 3 C 1 4 C 1 5 C 1 6 C 1 7 C 1 8 C 1 9 2 0 +
CH-WFT
C n n C
OH-WFT DST
0.00
Sulfur Species & Mercaptan Analysis
0.00 1.00 The open-hole OH-WFT sampling was also evaluated with
OH-WFT
OH-WFT vs. CH-WFT H2S
respect to sulfur & mercaptan analysis. Samples were
collected in Khuff Z1 of well W-3 with OH-WFT and DST.
Figure 9: Comparison between OH-WFT and CH-WFT Results The analysis of results shows good agreement, in general,
between the two samples. Figure 12 is a mercaptan anlaysis
Figure 10 below presents a comparison of the H2S & CO2 comparison between OH-WFT and DST sample in Khuff Z1.
measurements of OH-WFT and DST samples. As shown, the
results from both techniques are in good agreement.
IPTC 10622 7
Recommendations
Based on detailed analysis of the sampling conditions,
1000 reservoir characteristics and the results of NHCG contents, the
following recommendation are made to improve the quality of
the open-hole/cased-hole samples:
ppm mol
100
• The discrepancy in NHCG measurements between the
ppm
M l M lfid
en
a
C hy pt
er lfi
Et l M ulfi
op bon ulfi
i
f
rt- l M lf
M er p
hy p
C lD p
ro Me p
hy l M ap
-B hyl ap
pt
ph
M isu
Su
te py isu
et rca
l M ca
et ca
yl c a
hy ca
a
S
Thl S
hy yl S
lS
ca
N uty erc
Et py r c
Et erc
e
ut er
et er
hy r
lM e
ar n
til
ob M
C ge
la
Vo locations from each Khuff formation to have better
Is utyl
ro
-P l
5
ut
ar
yd
et
ro
ed
B
B
H
if i
se
Is
nt
de