Professional Documents
Culture Documents
Abstract— The temporal response of CdZnTe (CZT) detector Fast temporal response ability is important for high flux
to γ-ray is experimentally studied in this article. Infrared (IR) X-ray application (108 X/cm2 /s or even higher). Limited by the
illumination is utilized to improve the performance of CZT high densities of impurities and defects, excessive space charge
detector. The results show that with no IR illumination operated
on the CZT detector, the output signal of the CZT detector accumulation will occur under high flux operation (also called
appears as a short overshoot process (sharp peak) at the begin- polarization). The charged carrier dynamic process between
ning of stable γ-ray irradiation. After the sharp peak, a decay defect occupation and charge injection from X-ray will last
process follows, and finally, a stable output of CZT detector is several milliseconds before the signal stabilizes [1], [2]. This
achieved. IR illumination can eliminate the instability of CZT unstable process before signal stability in the CZT detector
detector under γ-ray irradiation, such as the output overshoot
of detector (sharp peak) during the transition state. With the will greatly limit its further application.
increase of IR illumination intensity, the sharp peak disappears It has been reported that the electric field distribution
gradually and the temporal response of CZT detector follows in CZT crystal can be probed by means of the Pockels
the intensity of γ-ray irradiation without transition state. The effect and the internal electric field can be controlled by
fitting time constant of the sharp peak is in good consistency infrared (IR) illumination [3]–[10]. From the Pockels images,
with the typical detrapping lifetime on the order of seconds.
The stable current of CZT γ-ray detector can be improved by the electric field between the cathode and anode planar con-
8.2% due to IR illumination, which can also be explained by tacts can be acquired using the crossed polarizer method [3].
the prefilling effects of IR illumination on the defect level. The Rejhon et al. [4] demonstrated that deep levels are responsible
sharp peak in the temporal response of CZT detector is probably for the polarization of CZT detector under high photon fluxes,
caused by the deep level, related to the Te antisite-related deep including E V + 0.41 eV, E C − 0.75 eV, and E C − 0.58 eV.
donors. The current uncertainty of CZT γ-ray detector under
IR illumination is smaller than 1.4%, which is small enough to Xu et al. [5] and Zhang et al. [6] confirmed the existence of
be ignored. The flux rate response of CZT detector can also be the deep donor level of E C −0.58 eV (or 0.57 eV) by the ther-
improved by IR illumination. This study will provide significant mally stimulated current (TSC) and the photo-induced current
supports for the further application of CZT material in radiation transient spectroscopy (PICTS), respectively. Teague et al. [7]
measurements. confirmed that the increase in bulk current during IR illumi-
Index Terms— γ-ray, CdZnTe (CZT), infrared (IR) illumina- nation is attributed to the detrapping of the charge carriers
tion, photocurrent, temporal response, uncertainty. and the increase during above bandgap light illumination is
I. I NTRODUCTION largely attributed to the generation of electron-hole pairs.
Dedic et al. [8], Franc et al. [9], Pekarek et al. [10], and
T HE CdZnTe (CZT) crystal possesses high resistivity,
wide bandgap (∼1.52 eV), high γ/X-ray stopping ability,
high mobility-lifetime product, and so on. Due to all these
Washington et al. [11] verified the feasibility of depolarization
on CZT detector by IR illumination.
Since IR illumination can depolarize the CZT detector,
properties, the CZT detector can be operated at room temper-
it might also improve the temporal response of CZT detector
ature with a low dark current, and the whole system can be
under γ-ray irradiation. In this article, careful experiments
simplified within centimeter-scale space.
are carried out to study the effect of IR illumination on the
Manuscript received March 26, 2021; accepted September 3, 2021. Date unstable charged carrier dynamic process in the CZT detector
of publication September 9, 2021; date of current version October 18, 2021. exposed by high flux rate γ-ray.
This work was supported in part by the National Natural Science Foundation
of China under Grant 11535010 and in part by the State Key Laboratory of
Intense Pulsed Radiation Simulation and Effect under Grant SKLIPR1805. II. T HEORY
Xiang Chen is with China Institute of Atomic Energy, Beijing 102413, When the CZT crystal is irradiated by high γ-ray flux rate,
China, on leave from the Northwest Institute of Nuclear Technology,
Xi’an 710024, China (e-mail: chenxiang@nint.ac.cn). large amounts of free electron–hole pairs are generated and
Zhao-Hui Song, Yi Lu, Yan Ma, and He-Tong Han are with the uniformly distributed in the whole volume. Free electrons
Northwest Institute of Nuclear Technology, Xi’an 710024, China (e-mail: are drifted toward the anode and holes are drifted to the
songzhaohui@nint.ac.cn; luyi@nint.ac.cn; mayan@nint.ac.cn; hanhetong@
nint.ac.cn). cathode. During the transport process of carriers, it will be
Long Hou is with China Institute of Atomic Energy, Beijing 102413, China affected by defects in CZT crystal, such as trapping, detrap-
(e-mail: houl_401@sohu.com). ping, and recombination. Since deep levels are responsible
Color versions of one or more figures in this article are available at
https://doi.org/10.1109/TNS.2021.3111090. for the polarization of CZT detector under high photon flux
Digital Object Identifier 10.1109/TNS.2021.3111090 rates, our concentration is focused on the effects of deep
0018-9499 © 2021 IEEE. Personal use is permitted, but republication/redistribution requires IEEE permission.
See https://www.ieee.org/publications/rights/index.html for more information.
Authorized licensed use limited to: Indian Institute of TechnologyDharwad. Downloaded on March 15,2022 at 07:36:39 UTC from IEEE Xplore. Restrictions apply.
2534 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 68, NO. 10, OCTOBER 2021
Fig. 1. Six main transit processes for n-type and p-type defects in CZT
crystal. Fig. 2. Layout of system.
TABLE I
T YPICAL D EFECT PARAMETERS IN CZT C RYSTAL Since deep levels are responsible for the polarization of
CZT detector under high photon flux rates, the detrapping
lifetimes for deep defect levels on carriers are estimated based
on the typical parameters below. The electron trapping cross
section is about 6 × 10−13 cm2 , and the carrier recombination
cross section between trapped electron and free hole is about
10−14 cm2 [2], [6]. Nc is at the magnitude of 1017 cm−3 .
Nt is at the magnitude of 1013 cm−3 . n t is at the magnitude
of 104 cm−3 . The product of k and T (300 K) is 0.026 eV.
defects on carriers [4]. There are two types of defects in E c − E et is set as 0.58 eV. The thermal velocity is about
CZT crystal, n-type and p-type. For each type of defect, 107 cm/s. Based on these parameters, the estimated detrapping
three transit processes exist, including trapping, detrapping, lifetimes for T1, T2, T3, and T4 defects are at the magnitude
and recombination. Six main transit processes can be taken of 0.1 ns, 1 ns, 100 ns, and 1 s, respectively. The estimated
to describe the effects of defects (both n-type and p-type) detrapping lifetime of the deep level is in accordance with the
on carriers, as shown in Fig. 1. The lifetimes for six transit results in [6], which is also at the magnitude of second.
processes can be described by the following equations [2]:
III. E XPERIMENT
1
τet = (1) In-doped CZT crystal with the dimension of 5 mm ×
(Nt − n t )σet v th
5 mm × 2 mm is grown with the vertical Bridgman (VB)
1
τed = (2) method by Northwestern Polytechnic University, China. Elec-
NC σet v th exp(−(E C − E et )/kT ) trodes are deposited on the 5 mm × 5 mm surfaces by gold
1 deposition after chemical polishing. Typical defect parameters
τer = (3)
pt σer v th are listed in Table I [12], [13]. The temporal response of
τht =
1
(4) CZT detector to γ-ray is calibrated on a 60 Co γ-ray radiation
(Pt − pt )σht v th instrument at the Northwest Institute of Nuclear Technology
1 (NINT), Xi’an, China. The activity of 60 Co source is about
τhd = (5)
NV σht v th exp(−(E ht − E V )/kT ) 2600 Ci. The output current of CZT detector is recorded by
1 a micro electrometer (Keithley 6517A), which can acquire
τhr = (6) current signals smaller than pA. The working bias voltage
n t σhr v th
(600 V) is supplied by a high-voltage source (Stanford PS350).
where Nt is the total electron trapping state density with The peak wavelength of IR LED is 940 nm and the spectral
an energy level of E et , n t is the trapped electron density at bandwidth is 45 nm. It is supplied by a current source
this state, Pt is the total hole trapping state density with an (Keysight U8032A). The layout of the system is shown
energy level of E ht , pt is the trapped hole density at this state, in Fig. 2.
v th is the thermal velocity, Nc is the available state density Since the activity of 60 Co source is about 2600 Ci, it is
for the conduction band, Nv is the available state density for surrounded by thick lead walls. Only a small hole remains
the valence band, σet is the electron capture cross section, in the wall to utilize γ-ray. A lead shutter is set at the front
σer is the recombination cross section between the free electron of the hole. Radiation and nonradiation states are achieved by
and the trapped holes, σht is the hole trapping cross section, lifting or putting down the shutter. To test the duration time of
σhr is the hole recombination cross section with free electrons, the lifting process, a photoelectric tube (named GD) is used.
k is Boltzmann’s constant, and T is the Kelvin temperature. GD possesses a fast time response of ns [14]. The testing result
The typical defect levels of In-doped CZT crystals (from is shown in Fig. 3, in which the black line is the responding
Northwestern Polytechnical University, China) are listed curve from GD and the red line is from the CZT detector.
in Table I, which is experimentally tested by the method of About 1.7 s is needed for the lifting process of the shutter.
TSC and thermoelectric effect spectroscopy (TEES) [5]. It verifies that the linear increases of the current (both from
Authorized licensed use limited to: Indian Institute of TechnologyDharwad. Downloaded on March 15,2022 at 07:36:39 UTC from IEEE Xplore. Restrictions apply.
CHEN et al.: IMPROVEMENT ON TEMPORAL RESPONSE OF CZT γ-RAY DETECTOR BY IR ILLUMINATION 2535
IV. D ISCUSSION
A. Shape Changes Caused by IR Illumination
1) Sharp Peaks and Stable Current Value: With no IR
illumination and γ-ray irradiation, the output current of CZT
detector is attributed by the stable dynamic process between
carrier generation (such as carrier detrapping and electrode
injection) and carrier disappearance (trapping, recombination,
and so on). When γ-ray is exposed, the stable dynamic process
Fig. 4. Temporal responses of CZT detector to γ-ray under different IR
is broken and a new stable state will be built under γ-ray
photocurrents. (a) Effects of IR photocurrent on Ibase . (b) Effects of IR irradiation. As shown in Fig. 4(b), the built-up process consists
photocurrent on sharp peak, and Ibase values are deleted. of the linear increase of current, a sharp peak existing at
the beginning of the irradiation, a stable state, and a decay
process from the sharp peak to the stable state. The linear
the GD and CZT detector) are related to the lifting process increase of current is related to the lifting process of the
of the shutter and the sharp peak is only related to the CZT shutter, as shown in Fig. 3. The sharp peak might be due to
detector. the excessive carrier traps at the beginning of γ-ray irradiation,
During the experiment, IR illumination is always irradiated which would detrap the redundant carriers before the stable
on the surface of CZT crystal before and after γ-ray, as shown state. The decay process is the built-up of a new charged
in Fig. 2. IR photocurrent refers to the current generated by carrier dynamic equilibrium between the defect occupation and
IR illumination. With the change of IR illumination intensity, the charge injection from γ-ray irradiation.
the IR photocurrents from the CZT detector are different, IR illumination can help the transit of electron from valence
as shown in Fig. 4(a). Ibase is the sum between the intrinsic to the defect level. This transit process can be taken as a
dark current (2.8 nA) and IR photocurrent of CZT detector prefilling process for the defect level, which will reduce the
before γ-ray irradiation. Isignal is the output current of CZT number of trapping carriers generated by γ-rays. As shown
Authorized licensed use limited to: Indian Institute of TechnologyDharwad. Downloaded on March 15,2022 at 07:36:39 UTC from IEEE Xplore. Restrictions apply.
2536 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 68, NO. 10, OCTOBER 2021
Authorized licensed use limited to: Indian Institute of TechnologyDharwad. Downloaded on March 15,2022 at 07:36:39 UTC from IEEE Xplore. Restrictions apply.
CHEN et al.: IMPROVEMENT ON TEMPORAL RESPONSE OF CZT γ-RAY DETECTOR BY IR ILLUMINATION 2537
Fig. 7. Current uncertainties of CZT detector under different IR Fig. 8. Flux rate responses of CZT detector under 5.403 and 2.590 Gy/h.
photocurrents.
Authorized licensed use limited to: Indian Institute of TechnologyDharwad. Downloaded on March 15,2022 at 07:36:39 UTC from IEEE Xplore. Restrictions apply.
2538 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 68, NO. 10, OCTOBER 2021
Authorized licensed use limited to: Indian Institute of TechnologyDharwad. Downloaded on March 15,2022 at 07:36:39 UTC from IEEE Xplore. Restrictions apply.