You are on page 1of 6

IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 68, NO.

10, OCTOBER 2021 2533

Improvement on the Temporal Response of CZT


γ-Ray Detector by Infrared Illumination
Xiang Chen , Zhao-Hui Song, Yi Lu, Yan Ma, Long Hou, and He-Tong Han

Abstract— The temporal response of CdZnTe (CZT) detector Fast temporal response ability is important for high flux
to γ-ray is experimentally studied in this article. Infrared (IR) X-ray application (108 X/cm2 /s or even higher). Limited by the
illumination is utilized to improve the performance of CZT high densities of impurities and defects, excessive space charge
detector. The results show that with no IR illumination operated
on the CZT detector, the output signal of the CZT detector accumulation will occur under high flux operation (also called
appears as a short overshoot process (sharp peak) at the begin- polarization). The charged carrier dynamic process between
ning of stable γ-ray irradiation. After the sharp peak, a decay defect occupation and charge injection from X-ray will last
process follows, and finally, a stable output of CZT detector is several milliseconds before the signal stabilizes [1], [2]. This
achieved. IR illumination can eliminate the instability of CZT unstable process before signal stability in the CZT detector
detector under γ-ray irradiation, such as the output overshoot
of detector (sharp peak) during the transition state. With the will greatly limit its further application.
increase of IR illumination intensity, the sharp peak disappears It has been reported that the electric field distribution
gradually and the temporal response of CZT detector follows in CZT crystal can be probed by means of the Pockels
the intensity of γ-ray irradiation without transition state. The effect and the internal electric field can be controlled by
fitting time constant of the sharp peak is in good consistency infrared (IR) illumination [3]–[10]. From the Pockels images,
with the typical detrapping lifetime on the order of seconds.
The stable current of CZT γ-ray detector can be improved by the electric field between the cathode and anode planar con-
8.2% due to IR illumination, which can also be explained by tacts can be acquired using the crossed polarizer method [3].
the prefilling effects of IR illumination on the defect level. The Rejhon et al. [4] demonstrated that deep levels are responsible
sharp peak in the temporal response of CZT detector is probably for the polarization of CZT detector under high photon fluxes,
caused by the deep level, related to the Te antisite-related deep including E V + 0.41 eV, E C − 0.75 eV, and E C − 0.58 eV.
donors. The current uncertainty of CZT γ-ray detector under
IR illumination is smaller than 1.4%, which is small enough to Xu et al. [5] and Zhang et al. [6] confirmed the existence of
be ignored. The flux rate response of CZT detector can also be the deep donor level of E C −0.58 eV (or 0.57 eV) by the ther-
improved by IR illumination. This study will provide significant mally stimulated current (TSC) and the photo-induced current
supports for the further application of CZT material in radiation transient spectroscopy (PICTS), respectively. Teague et al. [7]
measurements. confirmed that the increase in bulk current during IR illumi-
Index Terms— γ-ray, CdZnTe (CZT), infrared (IR) illumina- nation is attributed to the detrapping of the charge carriers
tion, photocurrent, temporal response, uncertainty. and the increase during above bandgap light illumination is
I. I NTRODUCTION largely attributed to the generation of electron-hole pairs.
Dedic et al. [8], Franc et al. [9], Pekarek et al. [10], and
T HE CdZnTe (CZT) crystal possesses high resistivity,
wide bandgap (∼1.52 eV), high γ/X-ray stopping ability,
high mobility-lifetime product, and so on. Due to all these
Washington et al. [11] verified the feasibility of depolarization
on CZT detector by IR illumination.
Since IR illumination can depolarize the CZT detector,
properties, the CZT detector can be operated at room temper-
it might also improve the temporal response of CZT detector
ature with a low dark current, and the whole system can be
under γ-ray irradiation. In this article, careful experiments
simplified within centimeter-scale space.
are carried out to study the effect of IR illumination on the
Manuscript received March 26, 2021; accepted September 3, 2021. Date unstable charged carrier dynamic process in the CZT detector
of publication September 9, 2021; date of current version October 18, 2021. exposed by high flux rate γ-ray.
This work was supported in part by the National Natural Science Foundation
of China under Grant 11535010 and in part by the State Key Laboratory of
Intense Pulsed Radiation Simulation and Effect under Grant SKLIPR1805. II. T HEORY
Xiang Chen is with China Institute of Atomic Energy, Beijing 102413, When the CZT crystal is irradiated by high γ-ray flux rate,
China, on leave from the Northwest Institute of Nuclear Technology,
Xi’an 710024, China (e-mail: chenxiang@nint.ac.cn). large amounts of free electron–hole pairs are generated and
Zhao-Hui Song, Yi Lu, Yan Ma, and He-Tong Han are with the uniformly distributed in the whole volume. Free electrons
Northwest Institute of Nuclear Technology, Xi’an 710024, China (e-mail: are drifted toward the anode and holes are drifted to the
songzhaohui@nint.ac.cn; luyi@nint.ac.cn; mayan@nint.ac.cn; hanhetong@
nint.ac.cn). cathode. During the transport process of carriers, it will be
Long Hou is with China Institute of Atomic Energy, Beijing 102413, China affected by defects in CZT crystal, such as trapping, detrap-
(e-mail: houl_401@sohu.com). ping, and recombination. Since deep levels are responsible
Color versions of one or more figures in this article are available at
https://doi.org/10.1109/TNS.2021.3111090. for the polarization of CZT detector under high photon flux
Digital Object Identifier 10.1109/TNS.2021.3111090 rates, our concentration is focused on the effects of deep
0018-9499 © 2021 IEEE. Personal use is permitted, but republication/redistribution requires IEEE permission.
See https://www.ieee.org/publications/rights/index.html for more information.

Authorized licensed use limited to: Indian Institute of TechnologyDharwad. Downloaded on March 15,2022 at 07:36:39 UTC from IEEE Xplore. Restrictions apply.
2534 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 68, NO. 10, OCTOBER 2021

Fig. 1. Six main transit processes for n-type and p-type defects in CZT
crystal. Fig. 2. Layout of system.

TABLE I
T YPICAL D EFECT PARAMETERS IN CZT C RYSTAL Since deep levels are responsible for the polarization of
CZT detector under high photon flux rates, the detrapping
lifetimes for deep defect levels on carriers are estimated based
on the typical parameters below. The electron trapping cross
section is about 6 × 10−13 cm2 , and the carrier recombination
cross section between trapped electron and free hole is about
10−14 cm2 [2], [6]. Nc is at the magnitude of 1017 cm−3 .
Nt is at the magnitude of 1013 cm−3 . n t is at the magnitude
of 104 cm−3 . The product of k and T (300 K) is 0.026 eV.
defects on carriers [4]. There are two types of defects in E c − E et is set as 0.58 eV. The thermal velocity is about
CZT crystal, n-type and p-type. For each type of defect, 107 cm/s. Based on these parameters, the estimated detrapping
three transit processes exist, including trapping, detrapping, lifetimes for T1, T2, T3, and T4 defects are at the magnitude
and recombination. Six main transit processes can be taken of 0.1 ns, 1 ns, 100 ns, and 1 s, respectively. The estimated
to describe the effects of defects (both n-type and p-type) detrapping lifetime of the deep level is in accordance with the
on carriers, as shown in Fig. 1. The lifetimes for six transit results in [6], which is also at the magnitude of second.
processes can be described by the following equations [2]:
III. E XPERIMENT
1
τet = (1) In-doped CZT crystal with the dimension of 5 mm ×
(Nt − n t )σet v th
5 mm × 2 mm is grown with the vertical Bridgman (VB)
1
τed = (2) method by Northwestern Polytechnic University, China. Elec-
NC σet v th exp(−(E C − E et )/kT ) trodes are deposited on the 5 mm × 5 mm surfaces by gold
1 deposition after chemical polishing. Typical defect parameters
τer = (3)
pt σer v th are listed in Table I [12], [13]. The temporal response of
τht =
1
(4) CZT detector to γ-ray is calibrated on a 60 Co γ-ray radiation
(Pt − pt )σht v th instrument at the Northwest Institute of Nuclear Technology
1 (NINT), Xi’an, China. The activity of 60 Co source is about
τhd = (5)
NV σht v th exp(−(E ht − E V )/kT ) 2600 Ci. The output current of CZT detector is recorded by
1 a micro electrometer (Keithley 6517A), which can acquire
τhr = (6) current signals smaller than pA. The working bias voltage
n t σhr v th
(600 V) is supplied by a high-voltage source (Stanford PS350).
where Nt is the total electron trapping state density with The peak wavelength of IR LED is 940 nm and the spectral
an energy level of E et , n t is the trapped electron density at bandwidth is 45 nm. It is supplied by a current source
this state, Pt is the total hole trapping state density with an (Keysight U8032A). The layout of the system is shown
energy level of E ht , pt is the trapped hole density at this state, in Fig. 2.
v th is the thermal velocity, Nc is the available state density Since the activity of 60 Co source is about 2600 Ci, it is
for the conduction band, Nv is the available state density for surrounded by thick lead walls. Only a small hole remains
the valence band, σet is the electron capture cross section, in the wall to utilize γ-ray. A lead shutter is set at the front
σer is the recombination cross section between the free electron of the hole. Radiation and nonradiation states are achieved by
and the trapped holes, σht is the hole trapping cross section, lifting or putting down the shutter. To test the duration time of
σhr is the hole recombination cross section with free electrons, the lifting process, a photoelectric tube (named GD) is used.
k is Boltzmann’s constant, and T is the Kelvin temperature. GD possesses a fast time response of ns [14]. The testing result
The typical defect levels of In-doped CZT crystals (from is shown in Fig. 3, in which the black line is the responding
Northwestern Polytechnical University, China) are listed curve from GD and the red line is from the CZT detector.
in Table I, which is experimentally tested by the method of About 1.7 s is needed for the lifting process of the shutter.
TSC and thermoelectric effect spectroscopy (TEES) [5]. It verifies that the linear increases of the current (both from

Authorized licensed use limited to: Indian Institute of TechnologyDharwad. Downloaded on March 15,2022 at 07:36:39 UTC from IEEE Xplore. Restrictions apply.
CHEN et al.: IMPROVEMENT ON TEMPORAL RESPONSE OF CZT γ-RAY DETECTOR BY IR ILLUMINATION 2535

detector irradiated by γ-ray and IR illumination together. The


difference between Isignal and Ibase is the net current induced by
γ-ray, which is also called the stable state current. As shown
in Fig. 4(a), the stable state current is about 23.5 nA under
different IR photocurrents. During the analysis of the shape
changes, the IR photocurrent of CZT detector is deducted
from Isignal . However, IR photocurrent is used as the key
parameter to describe the effect of IR illumination on the
CZT detector. This is due to the effect of IR illumination,
which transits electrons from the valence to the defect level
to improve the defect occupation.
After the deduction of IR photocurrent, the temporal
responses of CZT detector to γ-ray under different IR pho-
Fig. 3. Testing results on the duration time for the lifting process of the tocurrents are aligned at the same Ibase . The effects of IR
shutter. photocurrent on the sharp peaks are shown in Fig. 4(b). The
flux rate at the position of CZT detector is measured by a
flux rate instrument (named UNIDOS), which is 5.403 Gy/h.
As shown in Fig. 4(b), while no IR illumination is operated on
CZT detector, a sharp peak appears at the beginning of γ-ray
irradiation. After the sharp peak, a stable state is achieved after
a decay process. The time scale of the sharp peak is at the
magnitude of second. Since the switch-on time of irradiation
and RC constant of the circuit are both shorter than 10 ms, this
sharp peak and the decay process are not related to the circuit
factors. With the increase of IR photocurrent, the amplitude
of the sharp peak decreases and nearly disappears when the
IR photocurrent is larger than 40 nA. The stable state current
increases slightly with the increase of IR photocurrents. Under
different IR photocurrents, the temporal responses of CZT
detector to γ-ray irradiation are analyzed, including shape
changes (such as the sharp peak, the stable current value, and
the decay process), current uncertainty, and flux rate response.

IV. D ISCUSSION
A. Shape Changes Caused by IR Illumination
1) Sharp Peaks and Stable Current Value: With no IR
illumination and γ-ray irradiation, the output current of CZT
detector is attributed by the stable dynamic process between
carrier generation (such as carrier detrapping and electrode
injection) and carrier disappearance (trapping, recombination,
and so on). When γ-ray is exposed, the stable dynamic process
Fig. 4. Temporal responses of CZT detector to γ-ray under different IR
is broken and a new stable state will be built under γ-ray
photocurrents. (a) Effects of IR photocurrent on Ibase . (b) Effects of IR irradiation. As shown in Fig. 4(b), the built-up process consists
photocurrent on sharp peak, and Ibase values are deleted. of the linear increase of current, a sharp peak existing at
the beginning of the irradiation, a stable state, and a decay
process from the sharp peak to the stable state. The linear
the GD and CZT detector) are related to the lifting process increase of current is related to the lifting process of the
of the shutter and the sharp peak is only related to the CZT shutter, as shown in Fig. 3. The sharp peak might be due to
detector. the excessive carrier traps at the beginning of γ-ray irradiation,
During the experiment, IR illumination is always irradiated which would detrap the redundant carriers before the stable
on the surface of CZT crystal before and after γ-ray, as shown state. The decay process is the built-up of a new charged
in Fig. 2. IR photocurrent refers to the current generated by carrier dynamic equilibrium between the defect occupation and
IR illumination. With the change of IR illumination intensity, the charge injection from γ-ray irradiation.
the IR photocurrents from the CZT detector are different, IR illumination can help the transit of electron from valence
as shown in Fig. 4(a). Ibase is the sum between the intrinsic to the defect level. This transit process can be taken as a
dark current (2.8 nA) and IR photocurrent of CZT detector prefilling process for the defect level, which will reduce the
before γ-ray irradiation. Isignal is the output current of CZT number of trapping carriers generated by γ-rays. As shown

Authorized licensed use limited to: Indian Institute of TechnologyDharwad. Downloaded on March 15,2022 at 07:36:39 UTC from IEEE Xplore. Restrictions apply.
2536 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 68, NO. 10, OCTOBER 2021

Fig. 5. Characteristic current versus IR photocurrent of CZT detector.

Fig. 6. Bi-exponential fitting curves of the decay process in the temporal


in Fig. 5, with the increase of IR photocurrent, the sharp peak response curves of CZT detector under different IR photocurrents. (a) IR pho-
values decrease and the stable current values increase. The tocurrent = 0 nA. (b) IR photocurrent = 10 nA. (c) IR photocurrent = 18 nA.
(d) IR photocurrent = 40 nA.
decrease of sharp peak value represents the suppression of
the excessive carrier traps. The increase of the stable current
TABLE II
values refers to the effects of the prefilling process. When IR
B I -E XPONENTIAL F ITTING T IME C ONSTANTS OF CZT D ETECTOR
photocurrent reaches 50 nA, the sharp peak disappears and U NDER D IFFERENT IR P HOTOCURRENTS
the stable current value keeps constant. The stable current
value is improved from 23.2 nA (IR photocurrent = 0 nA)
to 25.1 nA (IR photocurrent = 50 nA). This increased stable
current value refers to the prefilling effects on the carrier
trapping process. Thus, more carriers generated by γ-ray are
collected and the collection efficiency of the generated carriers
in the CZT detector is improved by 8.2%.
The effect of IR illumination is to transit electrons from
the valence to the defect level and from the defect level to
the conduction. The 50-nA photocurrent means that about
When IR photocurrent is over 18 nA, the bi-exponential
3.1 × 1011 free carriers (free holes or free electrons) are
fitting time constants become to the same value within the time
generated within 1 s. The corresponding generated free carrier
of the red lines in Fig. 6. Single exponential fitting method
concentration in the 5 mm × 5 mm × 2 mm volume is
is used and the fitting results are listed in Table II. With
6.2 × 1012 cm−3 . This carrier concentration is comparable
the increase of IR photocurrent, τ1 gets smaller and finally
to the T4 defect concentration in Table I, which is reported
disappears when the IR photocurrent is over 18 nA, as shown
as 5.82 × 1013 cm−3 by TEES and TSC methods [5]. Before
in Fig. 6. τ2 also gets smaller from 1.85 to 0.47 s, while
the IR photocurrent reaches 150 nA (see Fig. 5), the stable
the IR photocurrent increases from 0 to 40 nA. This result
current value of CZT detector keeps constant and decreases
means that the influence of deep defect level on the temporal
with the further increase of IR photocurrent. This decrease
response of CZT γ-ray detector is at the magnitude of second,
of the stable current value might be related to the stronger
which is in good accordance with the estimated detrapping
depletion layer near the cathode caused by the excess car-
lifetime of deep defect level. The deep level dominates the
rier injection. To keep the stability of CZT γ-ray detector,
unstable dynamic process in the temporal response of CZT
the IR photocurrent of CZT γ-ray detector should be limited
detector. For the experimental CZT crystal, the main deep
to 50–150 nA.
donor defect exists at the level of E C − 0.58 eV, related to the
2) Decay Process: With no IR illumination, a sharp Te antisite-related deep donors. IR illumination can actually
peak appears at the beginning of the irradiation, as shown impress the effect of the defect occupation and shorten the
in Fig. 4(b). With the increase of IR photocurrent, the sharp unstable dynamic process to a stable state under high flux rate
peak disappears gradually. Deep levels are reported to be γ-ray irradiation.
responsible for the polarization of CZT detector under high
photon fluxes [4]. IR illumination has a “prefilling” effect on
the deep level. This sharp peak is probably related to the deep B. Current Uncertainty Analysis
defect level. Assuming that the detrapping process follows The net current uncertainty of CZT γ-ray detector induced
exponential decay order, a bi-exponential fitting method is by different IR photocurrents is analyzed in Fig. 7. Under no
used to analyze the decay process, as shown in Fig. 6. The IR illumination, the current uncertainty is smaller than 0.2%.
fitting parameters are listed in Table II. With the increase of IR photocurrent, the current uncertainty

Authorized licensed use limited to: Indian Institute of TechnologyDharwad. Downloaded on March 15,2022 at 07:36:39 UTC from IEEE Xplore. Restrictions apply.
CHEN et al.: IMPROVEMENT ON TEMPORAL RESPONSE OF CZT γ-RAY DETECTOR BY IR ILLUMINATION 2537

Fig. 7. Current uncertainties of CZT detector under different IR Fig. 8. Flux rate responses of CZT detector under 5.403 and 2.590 Gy/h.
photocurrents.

also increases. When the IR photocurrent is within 50–150 nA,


the current uncertainty keeps a stable value of about 0.84%. IR illumination can prefill the deep level and suppress the
It is understandable that the increase of current uncertainty effects of deep levels on the sensitivity of CZT γ-ray detector.
is related to IR photocurrent. When IR illumination is turned With the increase of IR illumination, the sensitivity difference
off, the net current of CZT detector is about 24 nA. Compared of CZT detector becomes lower. The flux rate response of CZT
with this net current, IR photocurrent is much larger. The detector can be improved by IR illumination.
current uncertainty generated by IR photocurrent cannot be
ignored for the net current of CZT detector with no IR V. C ONCLUSION
illumination. This is the reason for the increase of current
In this article, the temporal response of CZT detector to
uncertainty under IR illumination. When the IR photocurrent
γ-ray is experimentally studied. IR illumination is used to
of CZT detector is smaller than 200 nA, the current uncertainty
improve the performance of CZT detector, including the output
is lower than 1.4%. Under this condition, the influence of
stability, the current uncertainty, and the flux rate response.
current uncertainty induced by IR illumination can be ignored
The stability of CZT γ-ray detector refers to the current shape,
in the temporal response of CZT detector.
such as the sharp peak, the decay process, and the stable state.
The results show that the following conditions hold.
C. Flux Rate Response 1) IR illumination can improve the stability of CZT detec-
Flux rate response of CZT detector under different γ-ray tor under γ-ray irradiation. With the increase of IR pho-
intensities is also investigated. Near the hole in the lead wall, tocurrent, the sharp peak disappears gradually and the
γ-ray irradiation does not distribute uniformly. To assure the temporal response of CZT detector follows the intensity
uniform γ-ray irradiation on CZT detector, the CZT detector of γ-ray irradiation without transition state. The fitting
is set at the place, 2.2 and 3.1 m away from 60 Co source. time constant of the decay process is at the magnitude of
The flux rate at the position of 2.2 and 3.1 m is measured second, which is in good accordance with the estimated
by UNIDOS, which are 5.403 and 2.590 Gy/h, respectively. detrapping lifetime of deep level. The stable current of
The sensitivity of CZT detector to γ-ray is defined as the ratio CZT detector to γ-ray irradiation can be improved by
between the stable current and the flux rate. Sensitivity versus 8.2% due to the prefilling effects of IR illumination on
IR photocurrent curves of CZT detector are shown in Fig. 8. the defect level.
When no IR is illuminated on CZT detector, the absolute 2) While IR photocurrent is smaller than 200 nA, the influ-
sensitivity difference is 9.7%. Under different IR photocurrents ence of current uncertainty induced by IR illumination
(90, 100, and 110 nA), the absolute response differences are is smaller than 1.4% and can be ignored in the temporal
5.2%, 3.9%, and 2.6%, respectively, much lower than that with response of CZT detector.
no IR illumination. 3) The flux rate response of CZT detector can also be
This difference is mainly related to the carrier com- improved by IR illumination, which is attributed to
bination under different γ-ray intensities. Based on (3), the prefilling effect of IR illumination on the deep
the lifetime of electron recombination τer is reciprocal to the level.
trapped hole density pt . With the increase of γ-ray intensity, 4) The deep level is verified to dominate the unstable
pt also increases. The increase of pt results in the decrease dynamic process in the temporal response of CZT
of τer , meaning that the probability of electron recombination detector, which is related to the Te antisite-related deep
becomes higher. More electrons will be trapped and recom- donors.
bined by the deep level. With the increase of γ-ray intensity, This work solves the instability problems during the sen-
the sensitivity of CZT detector becomes lower, and this is sitivity calibration process of CZT γ-ray detector, which
the reason for the sensitivity difference of 9.7% with no IR will provide significant experimental supports for the further
illumination under 5.403 and 2.590 Gy/h. application of CZT material.

Authorized licensed use limited to: Indian Institute of TechnologyDharwad. Downloaded on March 15,2022 at 07:36:39 UTC from IEEE Xplore. Restrictions apply.
2538 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 68, NO. 10, OCTOBER 2021

R EFERENCES [8] V. Dědič, J. Franc, M. Rejhon, R. Grill, J. Zazvorka, and


P. J. Sellin, “De-polarization of a CdZnTe radiation detector by
[1] M. Prokesch, D. S. Bale, and C. Szeles, “Fast high-flux response of pulsed infrared light,” Appl. Phys. Lett., vol. 107, no. 3, 2015,
CdZnTe X-ray detectors by optical manipulation of deep level defect Art. no. 032105.
occupations,” IEEE Trans. Nucl. Sci., vol. 57, no. 4, pp. 2397–2399, [9] J. Franc et al., “Control of electric field in CdZnTe radiation detec-
Aug. 2010. tors by above-bandgap light,” J. Appl. Phys., vol. 117, no. 16, 2015,
[2] Y. F. Du, J. LeBlanc, G. E. Possin, B. D. Yanoff, and S. Bogdanovich,
Art. no. 165702.
“Temporal response of CZT detectors under intense irradiation,” IEEE
[10] J. Pekarek et al., “Infrared LED enhanced spectroscopic CdZnTe detec-
Trans. Nucl. Sci., vol. 50, no. 4, pp. 1031–1035, Aug. 2003.
tor working under high fluxes of X-rays,” Sensors, vol. 16, no. 10,
[3] A. Cola, I. Farella, A. M. Manicini, and A. Donati, “Electric field
p. 1591, 2016.
properties of CdTe nuclear detectors,” IEEE Trans. Nucl. Sci., vol. 54,
no. 4, pp. 868–872, Aug. 2007. [11] A. L. Washington, L. C. Teague, M. C. Duff, A. Burger, M. Groza,
[4] M. Rejhon, J. Franc, V. Dědič, J. Kunc, and R. Grill, “Analysis of and V. Buliga, “Effect of sub-bandgap illumination on the internal
trapping and de-trapping in CdZnTe detectors by Pockels effect,” J. Phys. electric field of CdZnTe,” J. Appl. Phys., vol. 110, no. 7, pp. 1–6,
D, Appl. Phys., vol. 49, no. 37, pp. 1–8, 2016. 2011.
[5] L. Xu, X. Fu, G. Zha, T. Wang, Y. Xu, and W. Jie, “Effects of IR irradi- [12] L. Xu et al., “Effects of sub-bandgap illumination on electrical properties
ation on the transport properties through deep-level defects in CdZnTe and detector performances of CdZnTe:In,” Appl. Phys. Lett., vol. 104,
crystals studied by thermally stimulated current,” J. Cryst. Growth, no. 23, Jun. 2014, Art. no. 232109.
vol. 468, pp. 749–752, Jun. 2017, doi: 10.1016/j.jcrysgro.2016.09.065. [13] L. Xu et al., “Axial distribution of deep-level defects in as-
[6] J. Zhang et al., “Effect of point defects trapping characteristics on grown CdZnTe:In ingots and their effects on the material’s
mobility-lifetime (μτ ) product in CdZnTe crystals,” J. Cryst. Growth, electrical properties,” J. Cryst. Growth, vol. 409, pp. 71–74,
vol. 519, pp. 41–45, Aug. 2019. Jan. 2015.
[7] L. C. Teague, A. L. Washington, II, M. C. Duff, M. Groza, and [14] X. Chen, Z.-C. Zhang, K. Zhang, X.-Y. Guan, X.-F. Weng, and
A. Burger, “Photo-induced currents in CdZnTe crystals as a function H.-T. Han, “Study on the time response of a barium fluoride scintillation
of illumination wavelength,” J. Phys. D, Appl. Phys., vol. 45, no. 10, detector for fast pulse radiation detection,” IEEE Trans. Nucl. Sci.,
2012, Art. no. 105101. vol. 67, no. 8, pp. 1893–1898, Aug. 2020.

Authorized licensed use limited to: Indian Institute of TechnologyDharwad. Downloaded on March 15,2022 at 07:36:39 UTC from IEEE Xplore. Restrictions apply.

You might also like