WG -0.36 Pulse Vacuum Autoclave


PHP 2,500,000.00

This series of pulsation vacuum sterilizer uses saturated steam as its sterilization medium. The saturated steam releases a large
amount of latent heated physical properties in the stage of condensation, which presents items to be sterilized with a state of high
temperature and humidity. After a period of insulation in such status, sterilization is achieved. The application of pulsating vacuum
exhausting has successfully eliminated temperature influence from cold air, and helping the sterilization items to be dried by means
of vacuum and dehumidification between the jacketed layers.
Healicom Medical Equipment Co.,Ltd
Add: Add: Room 1720,C2 Kechuangcheng Building, No.30 Fengzhan Road, Yuhuatai Distinct, Nanjing, Jiangsu, China
Tel: +86-25-84452678 Mobile: +86-13951660247
E-mail: Sales@healicom.com www.healicom.com

Structure and Characteristic of The Equipment


Body of the sterilizer:
◎it is designed and manufactured in accordance with GB150 "the Pressure Vessel”, GB 8599 “The Technical Requirements of
Large Steam Sterilizer--Automatic Control Protocol”, and TSG R0004 the “Safety and Technology Specification for Fixed Monitor
Vessels” and related inspection procedures.
◎Tank of the horizontal rectangular double-deck structure is made from S30408.
◎Door sealing: Pneumatic sealing; both doors are interlocked and completely conform to the requirements of the specification.
◎The equipment has standard GMP authentication interface.
Sterilizer piping system: the pipeline is designed with the best configuration. By employing imported parts and quality domestic
brands into main components.
Sterilizer control system: with the adoption of Siemens PLC plus Siemens touch screen, this system is easy to implement the
program selection, parameter setting, equipment operation, report processing and other functions. It has perfect safety measures.
The F0 value and the temperature time gives double guarantee for sterilization. A single parameter control is also available. It has a
perfect sterilization records.

Main Parameters and Utilities

design pressure 0.245MPa working pressure 0.225 MPa


design temperature 139℃. working temperature 105-134℃
vacuum -0.09 MPa temperature equilibrium ≤±1℃
water source pressure 0.15-0.3 MPa compressed Air pressure 0.3-0.7 MPa
steam pressure 0.3-0.7 MPa power AC380V、50HZ
The three-phase and five-wire system
AC380V、50HZ
Healicom Medical Equipment Co.,Ltd
Add: Add: Room 1720,C2 Kechuangcheng Building, No.30 Fengzhan Road, Yuhuatai Distinct, Nanjing, Jiangsu, China
Tel: +86-25-84452678 Mobile: +86-13951660247
E-mail: Sales@healicom.com www.healicom.com

Size & Utilities


Model overall Chamber size Steam Tap water Power Net PRICE
size L×W×H consumptio consumpti (kw) weight
L×W×H (mm) n on ( kg)
(mm) (kg) (kg

WG-0.36 1300× 1350× 1900 1000×600×600 25 KG 40 KG 2+24KW 1000 Kg 2,500,000.00