Att 2 to 00002339 Rev D

Test and Performance Verification for the


Laerdal Suction Unit (LSU)
This document is designed to log information regarding the LSU performance.

SR / WO Number: nmr ____________________________________________________________________________

Model: ☐ BEMIS ☐ Reusable ☐ Serres


LSU Serial No: nmr ____________________________________________________________________________
Test Details: 1) Pressure gauge: Required Accuracy 1.6 %, unit mmHg
2) Flowmeter: Required Accuracy +/- 2.5 %, unit l/min
Conditions of testing: Room Temperature, No special controls
------------------------------------------------------------------------------------
For Laerdal Tech Centers*:
PC with LSU Service Utility Software and Service Utility Adapter

*For Laerdal Tech Centers only: Offload Device Usage Data

Connect external AC power and the Service Utility USB adapter, also connected to your PC.
Start the Service Utility Software and enter the LSU serial number. Select the “Service Data” tab, then
print the page and attach it to this service record.

IMPORTANT: Only open the unit if you are ESD protected

1. Device Test
If Unit Opened: (Perform Test 1- 28), If Unit not opened: (Perform 13 – 25 and 28)

General Inspection, inspect unit for wear and tear PASS FAIL N/A

Internal inspection

1 Power board & AC/DC inlet ☐l ☐l


2 Display ☐l ☐l
3 Battery plug & Battery Board ☐l ☐l
4 MMI Board ☐l ☐l
5 All Electrical & Pneumatic Connections ☐l ☐l
Pump

6 Umbrella Valves ☐l ☐l

Page 1 of 4
Att 2 to 00002339 Rev D

7 Piston Rings ☐l ☐l
8 Piston Arm and bearing ☐l ☐l
9 Vacuum Connector ☐l ☐l
10 Cylinder ☐l ☐l
11 Exhaust Manifold ☐l ☐l
12 Motor & Flywheel ☐l ☐l

General Inspection, inspect unit for wear and tear PASS FAIL N/A

External:

13 Check Enclosure Housing ☐l ☐l


14 Control switch ☐l ☐l
Container:

15 Canister ☐l ☐l
16 Vacuum Seal & Gasket ☐l ☐l
17 Float ball and Filter ☐l ☐l
18 Connectors ☐l ☐l
19 Record battery lot# (wwyy); < three years old: nmr
☐l ☐l ☐l
Functional Tests Using External AC Power
Vacuum Measurement Check and Vacuum Build Test
Use canister for this test. Attach the pressure gauge to the patient tube.
NOTE: Refer to attached table for minimum high vacuum limits at
higher altitudes.
Limits have been adjusted for vacuum gauge accuracy.
Set LSU vacuum selector to 200 mmHg (26.6 kPa).
Check and record that LSU is regulating with external vacuum gauge ☐l ☐l
20 reading between 172 and 228 mmHg: Gauge reading: _________ mmHg
nmr

Check that LSU is indicating 200 mmHg vacuum.


☐l ☐l
Set LSU vacuum selector to 500+ mmHg (66.6+ kPa)
Check that the LSU reaches external vacuum gauge reading of at least ☐l ☐l
21 492 mmHg at sea level. Gauge reading: __________ mmHg
nmr

Check that LSU is indicating at least 475 mmHg vacuum.


(The upper LED may be off to pass, the second can be at half intensity.) ☐l ☐l
Air Flow
Attach the flow meter directly to the suction inlet, without using a
canister. Limits have been adjusted for flow meter accuracy.

Page 2 of 4
Att 2 to 00002339 Rev D
NOTE: Refer to attached table for minimum free flow limits at higher
altitudes.
Free Airflow (FF), at 500+ mmHg setting,
22
Reusable and Serres: > 24.4 l/m Air flow reading: _______________ l/m
nmr
☐l ☐l
Free Airflow (FF), at 500+ mmHg setting
23
Bemis: > 29.2 l/m Air flow reading: _______________ l/m
nmr
☐l ☐l
24 Battery Quality Check; Fully charged bat runs at 500+, FF: > 20 mins) ☐l ☐l ☐l
Failure mode indicator (ref. DFU); Pass if the failure mode is not lit
25
during tests. ☐l ☐l
26 AC Power (charging and functionality of LSU) ☐l ☐l
27 DC Charging (charging and functionality of LSU) ☐l ☐l
28 Device Test (refer to current DFU) ☐l ☐l

2. Overall result: Pass: ☐l Fail: ☐l

IMPORTANT If any of the above Functional Test measurements are failing, the LSU device
must be sent to your local Laerdal Technical Center for further investigation.

Faults Found
nmr _______________________________________________________________________
Comments
nmr _______________________________________________________________________

3. Configuration revision sticker on the LSU:

Yellow: ☐ Green: ☐

Completed by: nmr ______________________________________ Date (dd/mm/yy): nmr ___________________________

Accepted by: nmr _____________________________________ (e)Signature: __________________________

NOTE:

For table of Acceptance Limits for minimum vacuum and free air flow at 500+ Setting please see page 4

Acceptance Limits for minimum vacuum and free air flow at 500+ Setting.

Page 3 of 4
Att 2 to 00002339 Rev D
Altitude Altitude Nominal 500 + Vacuum Free Air Flow Free Air Flow
above above Air Pressure minimum limit minimum limit minimum limit
ft m kPa /mmHg mmHg Bemis Serres and
l/min Reusable
l/min
0 0 101,3 / 760 492 29,2 24,4
1000 300 97,78 / 733 474 28,1 23,3
2000 600 94,2 / 707 455 27,0 22,2
3000 900 90,8 / 681 437 25,9 21,1
4000 1200 87,5 / 656 418 24,9 20,1
5000 1500 84,3 / 632 399 24,0 19,2
6000 1800 81,2 / 609 380 23,0 18,2
7000 2100 78,1 / 586 360 22,0 17,2
8000 2400 75,2 / 565 341 20,9 16,1
10000 3000 69,7 / 523 321 19,9 15,1

Page 4 of 4