SECTION 1
INTRODUCTION
1.1. PURPOSE OF THE SERVICE MANUAL
This manual includes servicing and maintenance information for NB5, NB9 and NB20
water baths. It is only to be used by technicians who were formerly trained by Nuve.
This manual informs the technicians about the operating principles, diagnosing and
repairing methods and spare part replacing.
In any problem which is not identified in this manual is determined, please contact to
Nuve Servicing Team.
Factory
Esenboga Yolu 22. Km.
Akyurt 06287 ANKARA
Tel: 00 90 312 399 28 30 (2 lines)
Fax: 00 90 312 399 21 97
e-mail: service@nuve.com.tr
The heating function is obtained by of the sheet heaters placed onto the outer sides
and the outer bottom surface of the stainless steel tank. The microprocessor
controlled main PCB operates on “PID” controlling system. Triple insulation
consisting of glasswool, an aluminum layer and an air gap gives the temperature
uniformity and stability.
2
SECTION 2
OPERATING PRINCIPLES
The microprocessor controlled PCB operates on “PID” controlling system. All control
elements are connected to main PCB. These are;
1. Thermocouple
2. Keyboard and Display PCB
3. Power supply terminals of heaters
3
2.2.3. Keyboard and Display PCB
The illustration below shows the keyboard and display PCB’s functions. For more
information, please see the microprocessor user’s manual.
6 7 8 9 10
1 2 3 4 5
4
SECTION 3
SERVICING
Caution: Before servicing the instrument, please take the necessary precautions for
your health. Please respect to the warnings on the unit!!
The elements on the PCB must not be replaced even the failure is found to be caused by the
main PCB . In this case, please send the failed PCB to factory servicing team along with a
note on which the failure explanations are written.
Before replacing the PCB or any other control element, please make sure that the failure is
not caused by weak wire and terminal connections.
5
3.2. GENERAL FAILURE
1 The On / Off switch is ON Power supply inlet failure. Check that mains voltage is supplied to unit.
but it does not lighten and Check that the terminal connections and
the display is blank. power supply socket connections are not bad.
The glass fuse has blown. Replace the glass fuse and check the short
circuit that causes blowing of the fuse.
2 The On / Off switch is ON The main PCB / Display Remove this connection cable and place
and lightens but the The PCB connection cable is it carefully.
display is blank or some not properly connected.
segments are blank.
The main PCB / display Replace the connection cable.
connection cable is defective.
3 The On / Off switch is OFF The On / Off switch cable Check the connections and
but it lightens. connections are in reverse. correct them.
6
3.3. TEMPERATURE REGULATION OPERATING FAILURE
1 The temperature does not The safety thermostat is Set the safety thermostat temperature value
reach to the set value. adjusted to a temperature Higher than the set temperature
lower than the set temp.
The heaters are defective. Check the heater's resistance value with a
multimeter.
Replace it if they are defective.
The main PCB does not Check the outlet with a multimeter.
outlet to solid state relay. If 12 V DC does not exist, replace
(SSR) the main PCB.
2 The temperature of the The main PCB outlets Check the outlet with a multimeter.
water exceeds the set continuously to the solid Replace the main PCB if continuous
temperature. state relay.(SSR) outlet (12 VDC) exists.
3 Actual temperature in the The thermocouple measures Check if the thermocouple touches the
tank is different than the incorrectly. samples.
displayed one.
Change the thermocouple
7
SECTION 4
CALIBRATION
4.1 HOW TO MEASURE THE ACTUAL TEMPERATURE
• Press temperature set button continuously until PAS appears on the temperature display.
• Press the time increasing button until 277 appears on the time display. (277 is the
password)
• Press the temperature set button once, 0 appears on the temperature display.
• Press temperature increasing button four times, 4 appears on the temperature display, 4th
parameter is the calibration parameter.
• Change the value on the time display according to difference between the equipment’s
display and the external sensor.
If the set temperature is 37.5°C and the external sensor measures 36.9°C, increase the
value, which appears on the time display, by adding 0.6.
• After applying the above calibration procedure please press temperature set button to
return to stand-by mode.
8
SECTION 5
• Remove two screws on the panel sheet and pull the panel sheet away.
• Disconnect all cable terminals which are connected to the main PCB.
• Disconnect the keyboard and display PCB / main PCB connection cable from
main PCB.
• Remove the 4 connection clipses and then pull the main PCB away.
• Place a new PCB on the connection clipses and make the terminal connections
according to the electric diagram.
• Connect the display PCB / main PCB connection cable.
• Disconnect the display PCB / main PCB connection cable from display PCB.
• Remove the 4 connection screws and remove the PCB.
• Place a new PCB.
• Check that the display keys and leds meet their places on the plastic panel.
• Connect the connection cable.
9
5.6. REPLACING HEATER
10
SECTION 6
DRAWINGS AND DIAGRAMS
6.1. ELECTRICAL CIRCUIT DIAGRAM
11
SECTION 7
2 10
11
3
12
4
9
8
12
Spare Parts List
NO NAME OF MATERIALCODE/QUANTITY
MATERIAL NB 5 QTY NB 9 QTY NB 20 QTY
1 Shelf Z11 R 01 069 1 Z11 R01 067 1 Z11 R01 067 1
2 Tank Z11 T 04 020 1 Z11 T04 019 1 Z11 T04 021 1
3 Thermocouple Z19 I01 051 1 Z19 I01 051 1 Z19 I01 051 1
4 Heater set Z11 I02 118 1 Z11 I 02 117 1 Z11 I02 121 1
5 Panel Z15 P15 151 1 Z15 P 15 151 1 Z15 P15 151 1
6 SSR Z12 R07 013 1 Z12 R 07 013 1 Z12 R07 013 1
7 EMC filter Z15 F06 020 1 Z15 F 06 020 1 Z15 F06 020 1
8 Main PCB Z15 T09 097 1 Z15 T09 097 1 Z15 T09 097 1
9 Display PCB Z15 G05 020 1 Z15 G05 020 1 Z15 G05 020 1
10 On/Off switch Z12 A 03 018 1 Z12 A03 018 1 Z12 A03 018 1
11 Power socket Z12 S13 010 1 Z12 S13 010 1 Z12 S13 010 1
12 Safety thermostat Z11 T 50 001 1 Z11 T50 001 1 Z12 T50 001 1
13