Professional Documents
Culture Documents
UF50 SM - B Install
UF50 SM - B Install
APPENDIX B INSTALLATION
Table of Contents
Page
B.1 INSTALLING THE FILTER ASSY NO.17 (C-2).................................................. B-1
B.1.1 Pre-installation ................................................................................... B-1
B.1.2 Bubble Removal of the Filter (1)......................................................... B-1
B.1.3 Bubble Removal of the Filter (2)......................................................... B-2
B.1.4 Installation of the Filter ....................................................................... B-3
B.1.5 Bubble Removal inside Filter.............................................................. B-3
B.1.1. Pre-installation
Verity and record that there is no problem in the sheath pressure and UF-Check (Fscw) analyzed
data.
Fscw : Approx.80 ch
Sheath Pressure : Approx. 0.22 - 0.24 kg/cm2
Syringe
20%
ethanol
Syringe
CELLSHEATH
Install the filter filled with CELLSHEATH on Rear Panel of UF-50 Main Unit. At this time,
tubes A is blocked up by forceps.
1) Fix the filter on Rear Panel of UF-50 using screw binding M3x6.
2) Be sure the sheath line tubes (3.6 mm x 1.8 mm) stuck out of rear panel.
3) Block up tubes on both sides (B and C) by the forceps.
4) Remove the tube silicone connected these tubes.
5) Connect the end of tube polyurethane marked as “IN” to the nipple (IN) located at the lower
part of the filter. At this time, verify that the tube is not disconnected from the nipple even if
pulled strongly.
6) Disconnect tube polyurethane (3.6 mm x 1.8 mm) from the nipple “OUT” to the nipple (OUT)
located at the upper part of the filter.
7) Connect the end of the tube polyurethane marked as “OUT” to the nipple (OUT) located at
the upper part of the filter. At this time, verify that the tube is not disconnected from the
nipple even if pulled strongly.
Rear Panel
A Rear Panel C
C
C
C OUT
Silicone
Tube
IN
B B
B
[Condition]
1. RBC, WBC, EC, CAST < 1.0 (count/uL)
2. BACT < 400 (count/uL)
3. Total < 300 (count/uL)
5) Disconnect the connector from the J2 connector of wiring cord No. 262A.
6) Install the ID Unit on the Main Unit chassis using two M3×6 binding screws provided.
J2
ID Unit
M3x6 Binding Screws
7) Connect the ID Unit wiring cord to the J2 connector of the wiring cord No. 262A, and bind the
cords with a clamp.
8) Reinstall the right side cover and front cover B.
9) System Service representative will make the soft ware setting to recognize ID Reader.
10) After turning ON the power, perform following steps 11) and 12).
11) Press [Interfac] in the Service Setting submenu. Select “ 1: Connect” for Barcord setting.
Press [Return] then [Set] to make the setting effective.
12) Perform “Samp. Test” in Maintenance screen to confirm the ID reader reads correctly.
Ball catch
3) Fasten the Sampler Assy (UFSU-1) using four M3×6 binding head screws (provided).
Rack Arrival sensor (Switch Assy No.96) is supplied with UFSU-1 Kit and it should be
fastened using two M3×6 binding head screws (provided). 27
Photointerrupter
5) Fasten two Cover Mounting Plates using four M3×6 binding head screws (provided).
6) Fasten the Sampler Cover-B using two M3×6 binding head screws (provided).
7) Fasten the Sampler Cover-C using two M3×6 binding head screws (provided).
9) Using two M3×6 binding head screws and two M3 toothed washers (provided), fasten the
Sampler Cover and two Cover Mounting Plates.
Binding head
screw
(M3 x 6)
Toothed washers
(M3)
Sampler
1) Position adjustment of UFSU-1 and Sample Aspiration Pipette of UF-50 Main Unit.
Adjustment of aspiration position.
2) Position adjustment of Rack Arrival Sensor.
If a rack is not correctly set, there is a possibility for Tube Detection Sensor to function
incorrectly, causing abnormal Sampler analysis. This sensor avoids this problem.
3) Storage of setting values.
4) Adjustment of Tube Detection Sensor.
5) Verification of adjustment result.
B.4.2. Preparation
(1) Position setting of UFSU-1 and Sample Aspiration Pipette of Main Unit.
Move Sampler Table and Sample Pipette to the aspiration position (tube position 5).
1) Position setting for Sampler Table
a) Move Sampler Table to the Sampler analysis initial position.
In the Service menu.
Press, [Utility] →[ M.Check] → [SV/Motor]
Set as: [Motor No. 4], [Pulses: 3000 pulse], then press [Return] and [OK] key.
Move Sampler Table to CCW Limit position by pressing [M.Drive].
At this time, verify that the indication on the right bottom of the screen changed from
“CCW Limit OFF” to “CCW Limit ON”.
1) Remove the probe support and sample probe from the sample arm.
2) Fasten the sample probe assembly (optional sampler) to the arm using two M3×6 binding head
screws (provided).
Sample arm
Binding head
screw
(M3 x 6)
1) Fasten the mixing unit to the upper part of reagent case using two M3×6 binding head screws
(provided).
2) Connect three solenoid valve cords to the solenoid valve.
3) Connect one hydraulic connector from the hydraulic unit to the one by sandwitching the rubber
packing.
4) Connect one connector from the control unit.
5) Connect the upper part of the sample pipette using two tubes.
6) Fasten the clamp at the upper part of the reagent case.
LCD Unit
(7) Clamp
the tubing
(8) Connect the tubing to the pipette. (11) Clamp the cord