You are on page 1of 16

20/04/2021

COURSE OUTLINE:

VII. Quality Assurance and Maintenance Issues


Quality Assurance
Quality Assurance and 1. Technologist responsibilities
▪ It is the planned and systematic action that a
1.1. Image quality control
Maintenance Issues 1.1.1 Exposure indicator appropriateness
diagnostic x-ray facility will produce a
consistently high-quality images with minimum
1.1.2. Image Accuracy
exposure of the patients and personnel.
1.2. Plate Maintenance
1.2.1. Cleaning and inspection
Ms. Danica P. Pardalis, RRT. Quality Control
1.2.2. Erasure
Lecturer, B.S. Radiologic Technology Program ▪ It deals with the technique used in the monitoring
La Consolacion University Philippines 1.3. Reject analysis
(or testing) and maintenance of the technical
2. Service engineer or medical physicist
elements or components of an X-ray system.
responsibilities

1 2 3

1
20/04/2021

(CR) Computed Radiography Basic Components of CR CR WORKFLOW


Preview & ID Station
ID Tablet
Cassette with
Digital way of doing general radiography Imaging Plate
with Conventional X-ray machines
Rx Exposure
Printing

It is a process of capturing radiographic data Network

from a conventional X-ray machine and processing Digitizer


Processing Server Laser Camera
the data digitally to produce crisp and high quality
radiographic images.
Processing server
Identification Digitizer

4 5 6

2
20/04/2021

INTRODUCTION to QA QC of CR and Recommended routine QC tests


Digital Radiography
DR for DR systems
Frequency Test Section

Direct digital radiography (DR) and Daily Checks on acquisition and reporting monitors
Daily System check
▪ Performed by digital X-ray machines computerized radiography (CR) systems are Daily Printer checks using test pattern

with flat panel detectors now being used routine for radiography Weekly Check of contrast-to-noise ratio
Weekly Image quality tests
Screening Programme. Weekly Uniformity check
Monthly AEC thickness check
Monthly Test of reporting monitors
Routine quality control (QC) is essential to Monthly Mechanical safety and function checks
As required Detector flat-field calibration
ensure that the equipment is performing as As required Repeat analysis
expected meets standards. As required Printer checks following software upgrade

7 8 9

3
20/04/2021

Recommended routine QC tests Artifacts in Computed


for CR systems Artifacts
Radiography
Frequency Test Section
Daily Checks on acquisition and reporting monitors
CR - Dust on the IP, scraches/cracks on IP, dirt on reader,
Daily System check
roller marks ▪ Hardware artifacts
Daily Printer checks using test pattern
Weekly Check of contrast-to-noise ratio DR - Faulty individual pixel, cluster of pixel, loss of resolution
Weekly Image quality tests in one or both direction
▪ Soft ware artifacts
Weekly Uniformity check
Monthly AEC thickness check
Monthly Test of reporting monitors MONITOR – CRT (distortion, possibly due to interference
Monthly Mechanical safety and function checks from other electrical devices)
Six-monthly Image plate matching and artifact check
As required Image plate erasure PRINTER - if artifacts not seen on hard copy, then it is
As required Image plate cleaning caused by printer, bandling, streaking, fine line in direction of
As required Repeat analysis film appear
11
As required Printer checks following software upgrade

10 11 12

4
20/04/2021

Damaged laser beam head in CR


Hard ware artifacts
reader
▪ dust particle & scratches
▪ Damage to the laser beam head of CR system
REASONS APPEARANCE Appearance
Appears as multiple linear white line
REMEDY:
▪ Head must be cleaned by Camel hairbrush by vendor engineer ▪ dust particle & scratches
▪ Damage to the laser
beam head of CR system
Remedy
REMEDY:
▪ Head must be cleaned by Laser beam head
Camel hairbrush by Has to be replaced
vendor engineer Dust on the laser beam head
appear as an irregular
white lines 13

13 14 15

5
20/04/2021

Software artifacts
Computed Radiography and MONITOR CHECKS
Excessive edge enhancement
Quality Control
A B
▪ Computed radiography still uses conventional ▪ Monitor checks should be performed on both
radiographic equipment so basic radiography quality the acquisition and reporting* monitors
control testing is still needed. This would include:
under recommended working conditions. It
 Calibration of the generator
 Collimation
should be noted that cathode ray tube (CRT)
 Beam Alignment
and flat panel displays (FPD) may have
 Linearity of exposure different types of problems.
 Spatial resolution (focal spot)

Artifact simulate acute phase of Paget's disease as a result of


excessive edge enhancement. (B) normal E.E 14

16 17 18

6
20/04/2021

Monthly test of reporting


Daily checks on acquisition and reporting
monitors monitors SMPTE
▪ Using the SMPTE Test Pattern
This organization developed a test pattern that can be used to
Method
determine grey scale image quality on videos and computer
display monitor.
• Check for obvious faults such as flicker, gross distortion,
artifacts.
• Check general condition. Other parameter which can be evaluated are:
• Clean if necessary (follow the supplier’s instructions). ➢ Spatial resolution
• Look at the text and lines on the screen – are they sharp and ➢ Brightness
straight?
➢ Contrast Aliasing
• Keep a record of all checks, note any problems and take
action to get them corrected.

19

19 20 21

7
20/04/2021

SMPTE SMPTE Diagnostic video monitor luminance &


uniformity

▪ There should be no blurring or bleeding of bright ▪ The gray scale is shown as a series of squares in
display areas into dark areas or aliasing of the the center of the image that range from black For Digital Systems:
spatial imaging patterns. (0%) to white (100%).
• Display SMPTE test pattern to evaluate initial quality of monitor
▪ All monitors used for interpretation should be ▪ It may be difficult to get the monitor to show ▪ Determine manufacturers recommended settings
tested at least monthly. both of these inner squares perfectly. ▪ Ensure monitor is set accordingly

▪ For the dynamic range test, both the 5% and 95% ▪ Most monitors do better showing the 95%
contrast areas should be visible and distinctly square than the 5% square. • If no guidance material available follow:
different than the adjacent 0% and 100% areas. – Video Monitor Manual Brightness & Contrast Setting Procedure
▪ However, you might try adjusting the ambient ▪ Diagnostic quality monitors typically include this
▪ Therefore: The brightness and contrast of the light to improve the visibility of the 5% square.
monitor is adequate if the 5% squares at both measurement in calibration check
ends of the gray scale are visible. ▪ Perform test following vendor recommendations

21

22 23 24

8
20/04/2021

Diagnostic video monitor luminance & Diagnostic video monitor luminance &
uniformity uniformity Monitor type Maximum luminance
Color CRT > 80 Cd/m2
Color LCD > 100Cd/m2
•Determine monitor type
• Use the TG-18 UNL -80 or SMPTE test Grey scale CRT secondary > 150 Cd/m2
•Calculate % difference of each of corners to centre
pattern ,or set an entirely white display
• Use a photo timer to measure Grey scale diagnostic >200 Cd/m2
luminance
Grey scale LCD 350 Cd/m2
-Measure maximum luminance in
the center and four corners of Minimum luminance
display CRT <1.0 Cd/m2
-Measure minimum luminance in LCD <1.5 Cd/m2
center of display % diff
All With in 15%
22 23 24

25 26 27

9
20/04/2021

Image quality uniformity & Image Quality – High Contrast Image Quality – Low Contrast
artifacts (Spatial) Resolution Resolution (detectability)

For photo stimulable phosphor For photo stimulable phosphor systems:


systems:
▪ Erase cassettes
▪ Determine standard set up & technical ▪ Obtain several contact images of a low contrast phantom
factors - do once process using a high contrast algorithm & edge
▪ low kVp, Extended SID enhancement algorithm.
▪ Use a lead –bar resolution test pattern
phantom
▪ slightly angle phantom
▪ Minimize moiré & line pairing pattern
process with high contrast algorithm
25

28 29 30

10
20/04/2021

Image Quality – Low Contrast Weekly check of contrast-to-


Reader Reboot
Resolution (detectability) noise ratio (CNR)
If region of interest (ROI) facility is available: ▪ Reboot weekly or as
required.
Method
• Use the image of the test object from the daily test ▪ Follow the schedule as
• Draw two ROIs prescribed by the
• Record the mean (M1) and standard deviation (SD) of the pixel value in ROI 1.
• Record the mean (M2) of the pixel value in ROI 2. manufacture.
• Subtract M2 from M1 and divide by SD to calculate the signal-to-noise ratio
(CNR):
▪ Daily wiping of the insertion
slot reduces the chance of
CNR = M1- M2
SD
dust getting into the
CNR baseline ± 20% readout chamber.
- If this level is exceeded then take action to correct the problem.

28

31 32 33

11
20/04/2021

Weekly uniformity check Monthly AEC thickness check Monthly AEC thickness check
▪ If region of interest (ROI) facility is available
▪ Uniformity should be visually checked on a weekly basis ▪ Measure SNR and compare with baseline values
using the image of the test object or plain Perspex from the ▪ Set up test just as done for film.
daily test. ▪ Measure CNR and compare with baseline values
▪ Make 1st exposure on the thinner phantom.
▪ expose thicker phantom at the same setting
▪ By setting a narrow grey scale window areas of non- ▪ record the mAs ▪ Remedial levels
uniformity will be seen. ▪ Detector dose indicator baseline for that thickness ±
▪ If a printer is available , print both images and
measure optical densities. 10%
▪ Magnify or zoom the image electronically and inspect it in a
systematic fashion to look for artefacts such as faulty clusters ▪ If using the on-screen tool , select same size ▪ SNR baseline for that thickness ± 20%
of pixels or areas of unusually low noise (where the ▪ CNR baseline for that thickness ± 20%
region-of-interest (ROI) at same location on both
background mottle appears blurred or smoother than other ▪ No disturbing artefacts should be visible.
images and record average measurement.
areas of the image). ▪ If any of the levels are exceeded then take action to
correct the problem.
▪ This is particularly important for some DR systems.

34 35 36

12
20/04/2021

DETECTOR FLAT-FIELD CALIBRATION DETECTOR FLAT-FIELD CALIBRATION IMAGE PLATE (IP) CHECKS
(DR SYSTEMS ONLY) (DR SYSTEMS ONLY) (CR SYSTEMS ONLY)
▪ Some system require the user to carry out this flat-
▪ Some DR detectors may have a non-uniform Image plate erasure
response (due to variations in sensitivity, faulty fielding process periodically, and it is therefore
pixels etc.). Also, there are non-uniformities in the x-
ray beam due to the anode heel effect and x-ray included here although it is not strictly a QC test. ▪ It is recommends that any cassettes left in
beam divergence. ▪ On other systems this is carried out by the service the x-ray room, or left un-used for a few
▪ Some DR systems correct for these inherent non- engineer at routine service visits. days (e.g., over a weekend)
uniformities by a process of flat-fielding. Flat-field
correction maps are obtained using a standard beam
attenuator (usually a Perspex block) for one or more Method
▪ It should be erased using the image reader
exposure conditions.
▪ (e.g., different target/filter combinations and focal spot ▪ Carry out the flat field calibration according to ‘secondary erasure’ cycle before use, to
sizes) remove any ‘fog’ from the IP.
manufacturer’s protocols.
▪ Record and initial that the procedure has been
performed.

37 38 39

13
20/04/2021

Six-monthly image plate matching Six-monthly image plate matching


• Record the S value.
and artifact check and artifact check
• View image with a narrow display window and check for
Method artefacts.
▪ This procedure should also be carried out on new • Repeat for each IP.
image plates. Then:
Using each IP in turn:
▪ Image plates and cassettes should be matched in • Calculate the mean (average) mAs value for all IPs.
both their sensitivity (S value per unit exposure) • Conduct primary erasure before starting. • Calculate the mean S value for all IPs.
and in the mAs derived under automatic
exposure control. • Place 4 cm thickness of Perspex on the table (if more
convenient, other thicknesses such as 4.5 or 5 cm would also be Remedial levels:
▪ Full size Perspex blocks are preferable for these suitable). ▪ mAs value mean ± 5%
tests so that the whole IP can be checked for • Place IP cassette in the bucky. ▪ S value mean ± 10% (this allows for variation due to mAs and
artefacts. variation in IP sensitivity).
• Operate the unit in a fixed kV, automatic mAs mode –
select kV in normal use ▪ Plates showing scratches or marks should be cleaned and the test
repeated. Plates with permanent scratches or marks should be
• Make an exposure and record the mAs value. removed from service
• Read the IP after a standard delay time (e.g., 1 minute)

40 41 42

14
20/04/2021

MONTHLY MECHANICAL SAFETY AND


FUNCTION CHECKS PRINTER CHECKS
▪ Daily printer checks using test pattern
▪ Printer checks should be carried out using standard viewing
▪ Check the safety and function of the system. ▪ The printer is set up on installation when the conditions each day that the printer is used. Print
▪ It is recommended that a local checklist is drawn up engineer should ensure that the hard copy ▪ SMPTE test pattern or the manufacturer supplied test
for each system to identify relevant features to be pattern and perform the following checks:
checked (e.g., items that are safety-critical or areas
matches the soft copy image.
known to be prone to faults). • Geometrical distortion – check image is printed without
▪ The local system, for example: geometrical distortion; borders should be completely visible
▪ environmental checks (some digital systems are and straight lines should be straight.
particularly sensitive to environmental conditions
such as temperature and humidity) • Contrast visibility – SMPTE test pattern, the 5% and 95%
▪ checks relating to the reporting workstation . squares should be clearlyvisible.
▪ Keep a record of all checks, note any problems and
take action to get them corrected

43 44 45

15
20/04/2021

Questions?
REFERENCES:
• Printer artefacts – check test pattern for printer
artefacts no disturbing artefacts should
be visible.
• If a densitometer is available, measure densities 1. Bushberg, J., Seibert, J.A., Leidholdt, Jr., E., Boone, J.
and compare with baseline values (2012). The Essential Physics of Medical Imaging. 2nd
. Edition. Philadelphia Lippincott Williams & Wilkins.
▪ Printer checks following software upgrade 2. Bushong, Stewart C. (2013). Radiologic Science for
▪ After software changes or an upgrade, it may be Technologist: Physics, Biology, and Protection. 10th
advisable to print both a test pattern and a clinical Edition. Mosby.
image to confirm that the hard copy remains
similar to the soft copy display 3. Thakur, Anju (2014). Department Of Radiodiagnosis
and Imaging Pgimer, Chandigarh

46 47 48

16

You might also like