Professional Documents
Culture Documents
COURSE OUTLINE:
1 2 3
1
20/04/2021
4 5 6
2
20/04/2021
Direct digital radiography (DR) and Daily Checks on acquisition and reporting monitors
Daily System check
▪ Performed by digital X-ray machines computerized radiography (CR) systems are Daily Printer checks using test pattern
with flat panel detectors now being used routine for radiography Weekly Check of contrast-to-noise ratio
Weekly Image quality tests
Screening Programme. Weekly Uniformity check
Monthly AEC thickness check
Monthly Test of reporting monitors
Routine quality control (QC) is essential to Monthly Mechanical safety and function checks
As required Detector flat-field calibration
ensure that the equipment is performing as As required Repeat analysis
expected meets standards. As required Printer checks following software upgrade
7 8 9
3
20/04/2021
10 11 12
4
20/04/2021
13 14 15
5
20/04/2021
Software artifacts
Computed Radiography and MONITOR CHECKS
Excessive edge enhancement
Quality Control
A B
▪ Computed radiography still uses conventional ▪ Monitor checks should be performed on both
radiographic equipment so basic radiography quality the acquisition and reporting* monitors
control testing is still needed. This would include:
under recommended working conditions. It
Calibration of the generator
Collimation
should be noted that cathode ray tube (CRT)
Beam Alignment
and flat panel displays (FPD) may have
Linearity of exposure different types of problems.
Spatial resolution (focal spot)
16 17 18
6
20/04/2021
19
19 20 21
7
20/04/2021
▪ There should be no blurring or bleeding of bright ▪ The gray scale is shown as a series of squares in
display areas into dark areas or aliasing of the the center of the image that range from black For Digital Systems:
spatial imaging patterns. (0%) to white (100%).
• Display SMPTE test pattern to evaluate initial quality of monitor
▪ All monitors used for interpretation should be ▪ It may be difficult to get the monitor to show ▪ Determine manufacturers recommended settings
tested at least monthly. both of these inner squares perfectly. ▪ Ensure monitor is set accordingly
▪ For the dynamic range test, both the 5% and 95% ▪ Most monitors do better showing the 95%
contrast areas should be visible and distinctly square than the 5% square. • If no guidance material available follow:
different than the adjacent 0% and 100% areas. – Video Monitor Manual Brightness & Contrast Setting Procedure
▪ However, you might try adjusting the ambient ▪ Diagnostic quality monitors typically include this
▪ Therefore: The brightness and contrast of the light to improve the visibility of the 5% square.
monitor is adequate if the 5% squares at both measurement in calibration check
ends of the gray scale are visible. ▪ Perform test following vendor recommendations
21
22 23 24
8
20/04/2021
Diagnostic video monitor luminance & Diagnostic video monitor luminance &
uniformity uniformity Monitor type Maximum luminance
Color CRT > 80 Cd/m2
Color LCD > 100Cd/m2
•Determine monitor type
• Use the TG-18 UNL -80 or SMPTE test Grey scale CRT secondary > 150 Cd/m2
•Calculate % difference of each of corners to centre
pattern ,or set an entirely white display
• Use a photo timer to measure Grey scale diagnostic >200 Cd/m2
luminance
Grey scale LCD 350 Cd/m2
-Measure maximum luminance in
the center and four corners of Minimum luminance
display CRT <1.0 Cd/m2
-Measure minimum luminance in LCD <1.5 Cd/m2
center of display % diff
All With in 15%
22 23 24
25 26 27
9
20/04/2021
Image quality uniformity & Image Quality – High Contrast Image Quality – Low Contrast
artifacts (Spatial) Resolution Resolution (detectability)
28 29 30
10
20/04/2021
28
31 32 33
11
20/04/2021
Weekly uniformity check Monthly AEC thickness check Monthly AEC thickness check
▪ If region of interest (ROI) facility is available
▪ Uniformity should be visually checked on a weekly basis ▪ Measure SNR and compare with baseline values
using the image of the test object or plain Perspex from the ▪ Set up test just as done for film.
daily test. ▪ Measure CNR and compare with baseline values
▪ Make 1st exposure on the thinner phantom.
▪ expose thicker phantom at the same setting
▪ By setting a narrow grey scale window areas of non- ▪ record the mAs ▪ Remedial levels
uniformity will be seen. ▪ Detector dose indicator baseline for that thickness ±
▪ If a printer is available , print both images and
measure optical densities. 10%
▪ Magnify or zoom the image electronically and inspect it in a
systematic fashion to look for artefacts such as faulty clusters ▪ If using the on-screen tool , select same size ▪ SNR baseline for that thickness ± 20%
of pixels or areas of unusually low noise (where the ▪ CNR baseline for that thickness ± 20%
region-of-interest (ROI) at same location on both
background mottle appears blurred or smoother than other ▪ No disturbing artefacts should be visible.
images and record average measurement.
areas of the image). ▪ If any of the levels are exceeded then take action to
correct the problem.
▪ This is particularly important for some DR systems.
34 35 36
12
20/04/2021
DETECTOR FLAT-FIELD CALIBRATION DETECTOR FLAT-FIELD CALIBRATION IMAGE PLATE (IP) CHECKS
(DR SYSTEMS ONLY) (DR SYSTEMS ONLY) (CR SYSTEMS ONLY)
▪ Some system require the user to carry out this flat-
▪ Some DR detectors may have a non-uniform Image plate erasure
response (due to variations in sensitivity, faulty fielding process periodically, and it is therefore
pixels etc.). Also, there are non-uniformities in the x-
ray beam due to the anode heel effect and x-ray included here although it is not strictly a QC test. ▪ It is recommends that any cassettes left in
beam divergence. ▪ On other systems this is carried out by the service the x-ray room, or left un-used for a few
▪ Some DR systems correct for these inherent non- engineer at routine service visits. days (e.g., over a weekend)
uniformities by a process of flat-fielding. Flat-field
correction maps are obtained using a standard beam
attenuator (usually a Perspex block) for one or more Method
▪ It should be erased using the image reader
exposure conditions.
▪ (e.g., different target/filter combinations and focal spot ▪ Carry out the flat field calibration according to ‘secondary erasure’ cycle before use, to
sizes) remove any ‘fog’ from the IP.
manufacturer’s protocols.
▪ Record and initial that the procedure has been
performed.
37 38 39
13
20/04/2021
40 41 42
14
20/04/2021
43 44 45
15
20/04/2021
Questions?
REFERENCES:
• Printer artefacts – check test pattern for printer
artefacts no disturbing artefacts should
be visible.
• If a densitometer is available, measure densities 1. Bushberg, J., Seibert, J.A., Leidholdt, Jr., E., Boone, J.
and compare with baseline values (2012). The Essential Physics of Medical Imaging. 2nd
. Edition. Philadelphia Lippincott Williams & Wilkins.
▪ Printer checks following software upgrade 2. Bushong, Stewart C. (2013). Radiologic Science for
▪ After software changes or an upgrade, it may be Technologist: Physics, Biology, and Protection. 10th
advisable to print both a test pattern and a clinical Edition. Mosby.
image to confirm that the hard copy remains
similar to the soft copy display 3. Thakur, Anju (2014). Department Of Radiodiagnosis
and Imaging Pgimer, Chandigarh
46 47 48
16