Professional Documents
Culture Documents
Index terms: Acceptance testing . Images, processing . Images. quality . Radiography . Quality assurance Screens
#{149}
and films
I From the Department of Diagnostic Radiology, Mayo Clinic and Foundation, 200 First St, SW, Rochester, MN 55905.
From the AAPM/RSNA Physics Tutorial at the 1995 RSNA scientific assembly. Received June 26, 1996: revision requested
October 9 and received October 3 1 : accepted November 1 . Address reprint requests to the author.
. RSNA, 199’
177
of QC, it is essential that they be aware of the proximately 40,000 possible combinations of
effort required to produce quality radiographs. film, chemical solutions, processors, and inten-
This article introduces QC in general and QC sifying screens. Consequently, it is prudent to
for the screen-film imaging system and film pro- use the film, chemical solutions, processor, and
cessor. Guidelines for acceptance testing of all intensifying screens produced by one manufac-
components, along with acceptable variability, tuner or to use the combination recommended
are presented, and silver recovery and other by the film manufacturer. The manufacturer de-
ways of reducing chemical waste are discussed. signed these items as a system so that they are
Finally, artifact detection and diagnosis are illus- optimized as a system. Furthermore, if prob-
trated through a series of clinical examples. lems are encountered, it is almost impossible
for anyone to have a complete understanding of
U QUALITY ASSURANCE AND QUAHTY all the possible combinations and potential
CONTROL trouble spots.
It is important to differentiate between quality Film, chemical solutions, processors, and in-
assurance (QA) and QC. QA is an overall system tensifying screens should be thought of as an
for monitoring or assuring quality. QA includes imaging system. We would not think of buying
quality assessment, continuing education, QC, an x-ray tube from one company, an image in-
efficacy, utilization review, and outcomes as- tensifier from another, a video camera from a
sessment. QA is the broad concept or program third, and a video monitor from a fourth and
that focuses on an institution’s goals for quality trying to make these four items work together
medical care and quality outcomes. as a system. We rely on one manufacturer to
QC refers to the tests and controls devel- produce the imaging portion of our x-ray imag-
oped to monitor important variables in the pro- ing system. Likewise, we should rely on one
cess. SpecifIc technical measurements are made manufacturer to provide the radiographic film,
to monitor the output of a process, for ex- chemical solutions, processor, and intensifying
ample, the kilovoltage of an x-ray generator. If screens and we should use them as a system.
the control limits are reached or exceeded, QC
requires a change in the process. For example, . Control Charts
if the kilovoltage reaches or exceeds the eon- Control charts are the key to a QC program. A
trol limits, the generator must be recalibrated. control chart is a graphical presentation that
shows the value of interest (eg, the mid-density
. Goal of QC on a processor control chart) plotted as a fune-
The goal of QC is to provide consistent quality tion of time (Fig 1). (Mid-density is a film den-
output or a consistent quality product. For ex- sity of approximately 1 .20 above the base-plus-
ample, in photographic processing, the
goals fog level of the film.) A control chart allows for
are to provide consistent density or film speed, easy analysis of the data, as opposed to looking
consistent contrast, and a consistent base-plus- at numbers written in a log or in a table. More
fog level. One-hour photofinishers have had important, it allows for evaluation of trends,
stringent QC programs in place for many years. that is, slow changes in the operating level over
In fact, the 1-hour photofinishers do a better time (Fig 2). Even though these changes may
job of film processor QC than many radiology not result in the variable of interest reaching or
departments! The 1-hour photofinishers must exceeding the control limit, the trend indicates
provide optimal quality because if customers that, given sufficient time, this will occur. A
are not satisfied they will not return to that par- trend indicates that the development process is
ticular photofinisher. Unfortunately, the patient out of control. Consequently, whenever trends,
never knows the reason for the repeated radio- or drifts in data, are apparent, the cause should
graphs or the poor photographic processing be determined and corrected.
that results in a degraded x-ray image. Conse- All the values in film processing should be
quently, the radiology department is not under centered around the operating level (Fig 2).
the same pressure faced by the photofinisher to If the values are consistently above or below
produce a quality product. the operating level, the development process
should be adjusted to center the values around
. Optimization of Photographic Pro- the operating level; the operating level should
cessing not be adjusted so that it is centered in the
Photographic processing is a relatively simple data.
process that uses a developer solution, fixer so- A control chart also shows immediately if
lution, and wash. Unfortunately, there are ap- values reach or exceed the upper control limit
or the lower control limit. Any time a value
reaches or exceeds the control limit, it should
Drift
Zi5h1 flr.iL v,.
___
P,’(,,,’ “I rt, M,,h
-: UCL #{149}__..........7L
OL --;74:;/
LCL
I I I I I I
Date
Figure 2. Schematic control chart shows the up-
per control limit (UCL), operating level (OL), and
lower control limit (LCL). The data plotted here
clearly indicate a trend or drift. Even though all the
data points are within the control limits, with the
exception of the last point, the development pro-
cess is clearly out of control and needs to be modi-
fled to eliminate the drift. This drift should have
been apparent after the first five or six data points.
--]
The last data point exceeds the upper control limit,
since the process was allowed to continue even
S/i
though it was out of control. If corrective action had
been taken earlier when it was first apparent that
Figure 1. Typical photographic processor control
the process was out of control (ie, when the drift
chart shows the QC results for a mammography pro-
was first noted), the data points would not have cx-
cessor. In addition to the three sections on the chart
ceeded the control limits.
that are used to monitor the density difference, mid-
density, and base-plus-fog level of the film, there are
three other important sections. In the lower left cor-
ncr are the date and replenishment rate for the de- . Establishing Operating Levels and
‘eloper and fixer. Any time changes are made in the Control Limits
replenishment rates, the changes should be re- In many cases, it is easy to establish the operat-
corded in this area. The area immediately to the ing levels and control limits. For example, for x-
right of the replenishment rate section is used to
ray generator kilovoltage, the operating level is
record the date and the temperatures of the devel-
the level selected on the generator (eg, 80
oper and wash water. Any time changes are made in
kVp). The control limits are then selected on
either of these temperatures, the date and respective
temperatures should be recorded. In the lower right
the basis of appropriate criteria (eg, usually ±5
corner is a section labeled “ Remarks, “ which is used kVp).
to record other modifications made to the processor. For photographic processing, the selection
It is also important to record the date that new of operating levels is more difficult. The mid-
replenisher solutions were added to the replenish- density level must be established at a density of
ment tanks or when the processor was serviced. approximately 1 .20 or higher. One step on the
This information is often valuable when it is neces- sensitometric step tablet or step wedge that
san’ to troubleshoot problems with the photo- produces a density at the appropriate level
graliic processor.
should be selected. A small mark should be
placed on that step to ensure that the same step
is evaluated each time. The density difference
be plotted on the control chart and circled (Fig must be established as the difference between
1). Corrective action is then taken; that is, the the densities of two specific steps. The two
cause of the problem is determined and the steps that are usually selected result in densities
problem corrected. A new sensitometric con- of 0.45 or higher and approximately 2.50 or
trol strip is then processed, with the new data
being plotted on the same date.
* Applicable for the Kodak 480 BA processor (Eastman Kodak, Rochester, NY); special chemistry required.
It is often stated that the output of intensify- plenishment rates must be increased appropri-
ing screens changes with time. However, the ately. Furthermore, replenishment rates must
amount of change is relatively minor. The major be set properly and must be consistent. For ex-
variation in intensifying screen output is caused ample, more or less replenisher could be
by manufacturing batch-to-batch variability. pumped into the developer tank as the head
The best solution to the problems caused by pressure in the replenishment system increases
screen and cassette variability is to purchase or decreases (ie, as the amount of replenisher in
cassettes in large batches and ensure that the the replenisher tank increases or decreases).
screens and cassettes are part of the same man- The processor immersion time (Table) refers
ufacturing batch. The cassettes and screens to the time that a film is in a particular solution.
should be tested to determine their speed. The duration of the immersion is measured
Once this has been done, cassettes with differ- from the time that the leading edge of the film
ent speeds can be isolated to different areas. enters the solution of interest until that same
For example, all the slower cassettes could be leading edge enters the next solution. These im-
placed in one area of the radiology department, mersion times are very short, which empha-
and faster cassettes could be placed in another sizes the need to ensure proper chemical aetiv-
area. ity levels and proper processor operation.
One must monitor the processor sensitomet-
Film Emulsion Variability. -In addition to rically, that is, use freshly exposed sensitomet-
batch-to-batch variation in film emulsion, one rie control strips to determine the chemical ac-
must be concerned with aging and storage eon- tivity levels of the developer and fixer solu-
ditions. The characteristics of the film, includ- tions. All other methods of monitoring these
ing speed, contrast, and base plus fog, change levels do not provide useful information about
with time. No radiographic film should be used the way photographic chemical solutions inter-
after its expiration date. If the film is not stored act with a particular film. For example, there
under the proper conditions (eg, at a tempera- have been proposals to monitor the pH and
tune not higher than 70#{176}
F and a humidity not specific gravity of the developer solution. How-
higher than 50%), the aging effects may be ac- ever, both these measurements are meaningless
celerated. because there are so many components in the
Batch-to-batch variability among film may re- developer. The specific gravity will be the same
quire a change in all setup or calibration levels as long as the total amount of solids placed in
for phototimers and laser cameras. To minimize the solution is the same, regardless of what
this problem, film should be purchased in large these components are. Likewise, the developer
batches (eg, a 3-month supply at one time), and is a heavily buffered solution, so the addition of
all the film should have the same emulsion other components probably
will not affect pH.
batch number. There are two basic rules for minimizing pro-
cessor variability: (a) use the chemical solu-
Processor Variability. -There are many vari- tions, temperatures, and immersion times nec-
ables associated with the photographic proces- ommended by the film manufacturer, and
son. For example, the developer temperature (b) monitor the processor with all film types
must be consistent (to within ±0.5#{176}F). The used in that processor.
processing must speed
be consistent and as
specifIed the manufacturer.
by Depletion and Chemical Variability. -One final source of
oxidation of the chemical solutions must be variation is the photographic chemicals, either
compensated for by using the appropriate re- the processor or the replenisher solutions. In
plenishment rates. If for any reason the deple-
tion and oxidation rates are increased, the re-
U SUMMARY
A good photographic processing QC program is
essential for quality radiography. It is essential
to consider the screen, film, chemical solutions,
and processor as a system and to use the com-
ponents recommended by the film manufac-
tuner that are designed to provide optimal re-
51.llts. Part of a good QC program includes envi-
ronmental considerations such as minimizing
chemical wastes, recovery of silver, and recy- Figure 9. Uniformly exposed radiograph demon-
cling the polyester film base material. Artifact strates dank marks similar to ones that were show-
ing up on a substantial number of films coming from
one processor. Investigation proved that the marks
were caused by foam in the fixer. The marks occur
because the bubbles in the foam attach themselves
to the film and prevent the fixer from getting to the
film as quickly as necessary to stop the development
process. Consequently, the film develops further
wherever a bubble is located.
analysis is a skill required in QC to be able to Physics Monograph no. 20. New York, NY:
determine the processor problems that cause American Institute of Physics, 1994.
the artifacts. 4. Haus AG, ed. Film processing in medical imag-
ing. Madison, Wis: Medical Physics, 1993.
5. GrayjE, Winkler NT, StearsJ, Frank ED. Qual-
U REFERENCES
ity control in diagnostic imaging. Frederick, Md:
1 . Artifact identification program: a self teaching
Aspen, 1983.
guide, part no. 668066. Rochester, NY: East-
6. Quality assurance in diagnostic imaging. NCRP
man Kodak, 1977.
report no. 99. Bethesda, Md: National Council
2. Recommended specifications for new mam-
on Radiation Protection and Measurements,
mography equipment: screen-film x-ray sys-
1988.
tenTh, image receptors, and film processors.
Reston, Va: Anierican College of Radiology,
1995.
3. Seihert JA, Barnes (T, Gould RG. SpecitIca-
tion, tcceptince testing, and quality control of
diagm)stic x-ray imaging equipment . American
Association of Plwsicists in Medicine, Medical
This article meets the criteriafcr 1. 0 credit hour in category 1 of the AMA Physician ‘s Recognition
Au’ard. To obtain credit, see the questionnaire on pp 171-176