You are on page 1of 40

FONA ART Plus

Service Manual

English
FONA ART Plus Service Manual

ATTENTION X-ray equipment produces ionizing radiation that may be harmful if not
properly controlled. It is recommended that the equipment is operated by
trained personnel only, in accordance with existing law.
ATTENTION Even if compliant to specifications of electromagnetic compatibility, it is
recommended not to use the equipment in presence of external
electromagnetic fields, such as those generated by cellular phones.
ATTENTION Observe radiation protection guidelines as outlined in the Operating
Instructions.
X-rays are generated, when the exposure hand-switch is pressed.
Manufactured by FONA S.r.l. Via Galilei 11 - 20090 Assago (MI) Italy
Distributed by FONA Dental s.r.o.
Stefanikova 7 SK-811 06 Bratislava, Slovakia www.fonadental.com
NEW SINCE: 10.2015

List of Contents
1. Functional Description .....................................................................................................................3
1.1 Block Diagram ......................................................................................................................3
1.2 Mains Connection Module ......................................................................................................4
1.3 Power Supply Unit ................................................................................................................5
1.4 Column Motor Driver Board ...................................................................................................6
1.5 Power Switching Unit JED 28 .................................................................................................7
1.6 Control Unit ..........................................................................................................................8
1.7 Tube-Housing Assembly ........................................................................................................9
1.8 Panoramic Digital Sensor ..................................................................................................... 10
1.9 Module Interconnections ..................................................................................................... 10
2. Service Functions .......................................................................................................................... 11
2.1 PAN Alignment ................................................................................................................... 12
2.2 Languages.......................................................................................................................... 13
2.3 Exposure Setting ................................................................................................................ 14
2.4 Software Info ..................................................................................................................... 15
2.5 DAP Display ........................................................................................................................ 15
2.6 Cam Offset ......................................................................................................................... 16
2.7 Thermal Setting .................................................................................................................. 17
2.8 Keyboard Test .................................................................................................................... 18
2.9 LEDS Test .......................................................................................................................... 18
2.10 Automatic Test ................................................................................................................... 19
2.11 X-Ray Test ......................................................................................................................... 20
2.12 Encoders Test ..................................................................................................................... 21
2.13 Voltages Test ..................................................................................................................... 21
3. Service Procedures ........................................................................................................................ 22
3.1 Plastic and Metal Covers ...................................................................................................... 22
3.2 Replacement of Power Supply Unit ...................................................................................... 23
3.3 Replacement of Control Unit Board ...................................................................................... 24
3.4 Replacement of Power Switching and Regulation Unit JED 28 ................................................ 25
3.5 Replacement of Tube Housing Assembly .............................................................................. 26
3.6 Filament Calibration ............................................................................................................ 27
3.7 X-Ray Beam Alignment ........................................................................................................ 28
3.8 Replacement of Position Sensors ......................................................................................... 30
3.9 Sensor Calibration ............................................................................................................... 31
4. System Messages .......................................................................................................................... 32
4.1 Warning Messages .............................................................................................................. 32
4.2 Error Messages ................................................................................................................... 32
4.3 Other Alarms ...................................................................................................................... 34
5. Spare Parts ................................................................................................................................... 35
5.1 Positioning & Test Items ..................................................................................................... 35
5.2 Power Supply & Control Modules ......................................................................................... 36
5.3 Mechanical Items ................................................................................................................ 39

2/40 69 682 91510 - 151030


FONA ART Plus Service Manual

1. Functional Description
1.1 Block Diagram

DUAL POWER POWER TUBE-HEAD


SUPPLY SWITCHING X-RAY
115/230 V GENERATOR

MAINS X-RAY
POWER EMISSION

LASER REGULATOR KV & mA


LIGHTS FEEDBACK

MOTOR
FOR VERTICAL
TRANSLATION CONTROL
PANEL

X-RAY
FONA ART PLUS PUSHBUTTON
ROTATION SYNCRO
SYSTEM SIGNALS DIGITAL
CARRIAGE CONTROL IMAGING
TRANSLATION
UNIT SYSTEM

MOTION IMAGING
POSITION
SENSOR DATA

IMAGING
COMPUTER

 The mains line is connected to the terminal block on the right side of the
system.
 The mains power is fed to the Power Supply Unit located on top of the
unit.
 Supply conductors bring power from the Power Supply Unit to the Power
Switching Unit in the rotating arm.
 The power supply unit is providing necessary low voltages to feed the
circuits and the laser lights.
 A secondary winding on the low voltage transformer is devoted to the
motor for the vertical movement of the carriage.
 The power switching unit is connected to the control unit with a flat cable.
 The system control unit drives the control panel, the power switching unit
and the motors making use of the information from the position sensors.

69 682 91510 - 151030 3/40


FONA ART Plus Service Manual

1.2 Mains Connection Module

MAINS
FUSE
F 501

XC503

DIAGRAM OF X-RAY
PUSHBUTTON
CONTACTS
FOR DOOR
SWITCH

EXTERNAL LIGHT
X-RAY ON

EXTERNAL LIGHT
SYSTEM ON

MAINS INPUT

The mains connection module is placed on the right side of the unit,
behind the mirror cover.

On the left side of the board


 the supply line which goes up to the dual voltage supply board via fuse
F501, varistors, EMI filter and the low voltage transformer.

On the right side of the board


 the contacts for the external signal lights (system ON and X-ray ON),
 the contacts for the door switch, closed with a jumper if the door switch is
not connected to
 the connector XC503 for the X-ray pushbutton
(X-ray switch on wires 3 and 4).
 the block for low voltage distribution at top

4/40 69 682 91510 - 151030


FONA ART Plus Service Manual

1.3 Power Supply Unit

LOW VOLTAGE MAINS


TRAFO FILTER

MAINS
INPUT

FU 303

PLUG ON THE
RIGHT SIDE
FOR 230V
OUTPUT
LEFT SIDE
FOR 115V

FU 304 FU 305
CONNECTORS FOR
ADDITIONAL X-RAY
EXPOSURE LIGHTS

 The unit can be set to work at 115 or 230 V with configuration plug and
fuses (see table below).
 Mains power is fed via the EMI filter to the low voltage transformer and via
the relay to the line voltage rectifier. The stabilized voltages of +15V,
+24V and non-stabilized of +38V are generated.
 Specific low voltages are available for the motor for vertical displacement
and for the laser lights.
 During the preparation phase (filament heating) a signal from the control
unit causes the relay to close.
 During exposure the mains voltage rectified is connected to the Power
Switching Unit. The yellow light on the exposure hand-switch (push
button) and the buzzer are activated too.

System Fuses
Fuse Code Use 230 V Version 115 V Version
F 501 Mains Power Input 8 AT 16 AT
FU 303 Low Voltage Circuit 1.6 AT 3.15 AT
FU 304 Output Power to Switching 6.3 AF 6.3 AF
FU 305 Output Power to Switching 6.3 AF 6.3 AF
F MOT Column Motor Driver fuse 4 AT 4 AT

The equipment is protected by fuse F 501 (input power from mains), located on the top part of the mains
connection module, on the right side of the vertical carriage.
The circuit of the low voltage transformer is protected by fuse FU 303. The Output Power circuit (to the Power
Switching Unit) is connected by fuses FU 304 and FU 305

69 682 91510 - 151030 5/40


FONA ART Plus Service Manual

1.4 Column Motor Driver Board

LED 8: POWER ON

XD1
XD8: RESERVED
SUPPLY VOLTAGE
25.5 VAC
XD3: MOTOR ENABLE
F MOT FUSE
XD4: MOTOR ENCODER
LED SW-T
LED MOTOR ENABLE LED SW-B

XD5 & XD6


UP & DOWN XD7
PUSHBUTTONS TOP & BOTTOM
LEFT / RIGHT SIDE LIMIT
MICRO SWITCHES
UPPER LED
UP PUSHBUTTON UPPER LED
PRESSED ERR-ENC
ENCODER
LOWER LED ERROR
DOWN PUSHBUTTON
PRESSED
LOWER LED
ERR-T
TEMPERATURE
ERROR
XD2
MOTOR SUPPLY

The column motor driver board supplies the motor for vertical movement
of the carriage. The speed which is initially low is increased for long
travels. Limit switches at top and at bottom control the displacement.

CONNECTORS LIGHTS
Supply voltage 25.5 VAC.
XD1 LED 8 power supply ON
Yellow-green ground wire at top.
XD2 Motor supply LED SW-T ON if top micro-switch engaged
XD3 Motor enable LED SW-B ON if bottom micro-switch engaged
LED
Motor encoder (to be connected at system ON when the rotating arm is in
XD4 MOTOR
assembly) patient entry or patient exit position
ENABLE
XD5 Up and down pushbuttons LED UP
XD6 Up and down pushbuttons LED DOWN
Top and bottom limit micro-switches (to be LED Encoder error ON for 30 s if no
XD7
connected at system assembly) ERR-ENC movement is detected
ON if motor temperature
XD8 Reserved. LED ERR-T
exceeds 75°C.

6/40 69 682 91510 - 151030


FONA ART Plus Service Manual

1.5 Power Switching Unit JED 28

RP 103 GROUND
TRIMMER
mA
TEST
POINT 300 VDC
FDB INPUT
SA10
1
REGULATION
SECTION
GROUND RP 104
TRIMMER
CONTROL OF FDB
FILAMENT SUPPLY
CONTROL OF HIGH
VOLTAGE
GENERATION
OUTPUT
TO FILAMENT OUTPUT TO HV
TRANSFORMER TEST TRANSFORMER
POINT
KV & mA

300 VDC  During preparation and exposure the 300 VDC (rectified mains voltage) is
fed into the unit.
80 kHz pulses  Pulses at 80 kHz pass to the filament regulator driver, the output voltage
of which depends on the duty cycle of the driving pulses.
40 kHz pulses  Pulses at 40 kHz pass to the filament inverter driver, the output of which
goes to the filament transformer in the tube-head.
The inverter current is sensed by a dedicated resistor.
20 kHz pulses  During preparation, pulses at 20 kHz pass to the High Voltage (HV)
inverter driver. During exposure the pulses pass also to the High Voltage
regulator driver.
The output voltage of the regulator depends on the duty cycle of the
driving pulses.
 The output of the inverter is connected to the HV transformer in the tube-
head. The inverter current is sensed by a current transformer
 The regulation section controls the level of kV and mA by comparing the
feedback signals with the reference signals set by the control unit.
 The anode voltage (kV) is determined by the input voltage to the high
voltage transformer (tube-head).
 The anode current (mA) is determined by the filament supply.
 Calibration is required only for the filament supply.

69 682 91510 - 151030 7/40


FONA ART Plus Service Manual

1.6 Control Unit

The control unit handles the input from the operator when selecting
technique factors, type of projection, patient’s build, and output for of X-
ray emission and mechanical movements.
Error messages are generated for wrong commands or to signal non
correct or unsafe working conditions.

8/40 69 682 91510 - 151030


FONA ART Plus Service Manual

1.7 Tube-Housing Assembly

4
3
2 5

 Power for high voltage generation is supplied through points 4 and 5.


 High voltage is generated by means of a transformer plus a cascade
multiplier (by5).
 Filament supply is provided through points 2 and 3, via a dedicated
transformer.
 The feedback signals are provided for
 anode voltage (1V/10 kV)
 anode current current (1 V/mA)
 A sensor is available to generate a temperature feedback (7) and inhibit X-
ray generation when the temperature is out of limit.

69 682 91510 - 151030 9/40


FONA ART Plus Service Manual

1.8 Panoramic Digital Sensor

1.9 Module Interconnections


Control Unit ART Plus Pan
Syncro

XC7

Power
Exposure Supply Unit
XC101
Enable
Power
Switching & EXP
Regulation
Unit JED28

Interconnection
Board
XC204 12 V
DC

XC203

Power
supply

Tube
Housing
Assembly
Panoramic
Sensor Cat 5e PC WORKSTATION
SNAP150 Cable EN 60601-1/EN 60950

10/40 69 682 91510 - 151030


FONA ART Plus Service Manual

2. Service Functions
Service mode is activated by holding the BACK key down for 5 seconds. The desired functions are
selected with the INCREASE and DECREASE keys on the right of the display.

The functions of the other keys are:


(TEST) to exit from the Service menu.
(CONFIRM /BACK) to go back to the beginning of the Service menu.
(PLUS on the right of the display) to select the service function.
(MINUS on the right of the display) to select the service function.
(PROGRAM) to activate the program selected. Hold it down to store the current value (not
applicable during setting of the technical parameters in “Exposure Setting” where the LASER key

is used).
(RETURN) to activate some service functions.

Service Functions appear as listed below when scrolling downward with the DECREASE key.
Functions for Ceph mode appear only for systems configured as such .

Service Functions
1 PAN Calibration
2 Languages
3 Exposure Settings
4 Software Info
5 DAP Display
6 Cam Setting
7 Thermal Setting
8 Keyboard Test
9 LEDs Test
10 Automatic Test
11 X-ray Test
12 Encoders Test
13 Voltages Test

69 682 91510 - 151030 11/40


FONA ART Plus Service Manual

2.1 PAN Alignment


This function is intended to store reference positions of the rotating arm
after mechanical alignment.

 Rotation axis, to be done first, to set the angle offset of the rotating arm
NOTE
 X axis, to set the displacement offset of the rotating arm

By keeping pressed the BACK key for 5 s the system enters Service
Mode

Service +/- Service function can be selected using


PAN Alignment the INCREASE or DECREASE keys on the RIGHT side of the display.
Scroll the menu until PAN Alignment is displayed

PAN Align. +/- Press the PROGRAM key to enter PAN Alignment.
Rotation Axis
Possibility to operate on Rotation Axis is offered first

Offset Rot| <--> Press the PROGRAM key to enter setting Offset for Rotation Axis
M:0.2 deg >.....
The actual offset value stored in memory is shown

Offset Rot| <--> Press RETURN key to enter Rotation Axis edit mode
M:0.2 deg >.....
Press INCREASE or DECREASE keys on the RIGHT side of the display
to rotate the arm step by step till the desired angle.
The right position of the rotating arm is reached when it is parallel to the
NOTE
top part of the vertical carriage, with the generator aiming to the bite
block /chin rest.

Keep pressed the PROGRAM key to store new Offset Rot value

Press BACK key to exit PAN Alignment

PAN Align. +/- Press the PROGRAM key to enter PAN Alignment.
X Axis
Possibility to operate on Rotation Axis is offered first
Press INCREASE or DECREASE key to change to X Axis

Offset X | <--> Press the PROGRAM key to enter setting Offset for X Axis
M:27.4 mm >.....
The actual offset value stored in memory is shown

Offset X | <--> Press RETURN key to enter X Axis edit mode


M:27.4 mm >27.4
Press INCREASE or DECREASE keys on the RIGHT side of the display
to shift the arm backward or forward till the desired position.
 The right displacement of the rotating arm is reached when the panoramic
NOTE radiograph of the ball in the test phantom is round.
 Use INCREASE to move backward if the ball appears elongated
 Use DECREASE to move forward if the ball appears compressed

Keep pressed the PROGRAM key to store new Offset X value

Press BACK key to exit PAN Alignment

Press TEST Mode key to exit SERVICE MODE

12/40 69 682 91510 - 151030


FONA ART Plus Service Manual

2.2 Languages
This function allows for the selection of the message language
By keeping pressed the BACK key for 5 s
the system enters Service Mode
Service function can be selected using
Service +/-
Languages the INCREASE or DECREASE keys on the RIGHT side of the display.
Scroll the menu until Languages is displayed

Press the PROGRAM key to enter Languages.

The actual language currently selected is indicated


Sel Language
English

Press INCREASE or DECREASE key to change language:


Sel Language
German  English
 Italian
 French
 Spanish
 German
 Finnish
 AVAILABLE
 AVAILABLE

Keep pressed the PROGRAM key to enter new language

Sprache
Deutsch

Display Sprache:
Deutsch

Press BACK key to exit Languages

Press TEST Mode key to exit SERVICE MODE

69 682 91510 - 151030 13/40


FONA ART Plus Service Manual

2.3 Exposure Setting


This function is intended to assign kV and mA values to ADULT panoramic
programs (P1, P3, P4, P5, P6, P7) and CHILD panoramic program (P2) for
each one of the four types of patient build, for a total of 32 combinations.
By keeping pressed the BACK key for 5 s
the system enters Service Mode
Service function can be selected using
Service +/-
Expos. Settings the INCREASE or DECREASE keys on the RIGHT side of the display.
Scroll the menu until Exposure Settings is displayed

Press the PROGRAM key to enter Exposure Settings.

NOTE
Keep pressed the TEST key to reset all values to default (factory)
conditions
By pressing the PROGRAM key select the exposure program
for ADULT panorama (P1, P3, P4, P5, P6, P7) or
for CHILD panorama (P2)

By pressing the PATIENT key select the patient build,


Small, Medium, Large, Extra Large

Change the kV level to the desired value by pressing the


* SET kV / mA *
67 6.3 INCREASE or DECREASE keys at the LEFT side of the display

Change the mA level to the desired value by pressing the


INCREASE or DECREASE keys at the RIGHT side of the display

Keep pressed the LIGHT key to store the selected value

DFLT parameter! Repeat kV setting and mA setting


for the other types of patient build if requested
for the other CHILD or ADULT group of programs if requested
Press BACK key to exit EXPOSURE SETTINGS
ERROR Press TEST Mode key
#: 11 to exit SERVICE MODE
Do verify proper setting by changing PATIENT build for ADULT panorama
(programs P1, P3, P4, P5, P6, P7) and CHILD panorama (program P2)

14/40 69 682 91510 - 151030


FONA ART Plus Service Manual

2.4 Software Info


This function to provide the version of software of the control unit.
By keeping pressed the BACK key for 5 s the system enters Service Mode

Service function can be selected using


Service +/-
Software Info the INCREASE or DECREASE keys on the RIGHT side of the display.
Scroll the menu until Software Info is displayed

DRKOS V1.06r00
PantOs V4.00r05 First part of software version,

DRK Boot0 V1.3


CHIP: 2004.0003 Second part of software version

DRK Boot0 V1.3


DRKOS V1.06r00 Third part of software version

Msg. Table
V04.05 Fourth part of software version

Press BACK key to exit Software Info

Press TEST Mode key to exit SERVICE MODE

2.5 DAP Display


This function is intended to enable or disable the presentation of DAP
values on display.
By keeping pressed the BACK key for 5 s the system enters Service Mode

Service function can be selected using


Service +/- the INCREASE or DECREASE keys on the RIGHT side of the display.
DAP Display
Scroll the menu until DAP Display is displayed

DAP ON
Set OFF? The DAP presentation status is shown with request to change

Keep pressed the PROGRAM to change status


The confirmation message appears for few seconds,
DAP Display then the possibility to change is offered
Disabled

The DAP presentation status is shown with request to change


DAP OFF
Set ON?

Keep pressed the PROGRAM to change status


The confirmation message appears for few seconds, then the possibility to
DAP Display change is offered
Enabled

Press BACK key to exit Software Info

Press TEST Mode key to exit SERVICE MODE

69 682 91510 - 151030 15/40


FONA ART Plus Service Manual

2.6 Cam Offset


This function is intended to align the panoramic projection by checking
and setting the distance of the two pins of the test phantom.
The current value of Cam Setting appears on display when entering this
function.
Set up is fine when the pin distance is of 880 +/-20 pixels (860-900).

pixel 860-900

By keeping pressed the BACK key for 5 s


the system enters Service Mode

Service function can be selected using


Service +/-
Cam Offset the INCREASE or DECREASE keys on the RIGHT side of the display.
Scroll the menu until Cam Offset is displayed

Press the PROGRAM key to enter Cam Offset.

 DECREASE key on the right


CamStart Setting
Store :1.0 >1.0 to enlarge distance between pins.
 INCREASE key on the right
to diminish the distance between pins.

Keep pressed the PROGRAM key to store Cam Offset.

Press BACK key to exit Cam Offset

Press TEST Mode key to exit SERVICE MODE

16/40 69 682 91510 - 151030


FONA ART Plus Service Manual

2.7 Thermal Setting


This function is intended to enable or disable the use of the thermal
sensor and set the limit values for temperature error handling:
First part: Enable / Disable, Second part: Editor thermal values.
By keeping pressed the BACK key for 5 s the system enters Service Mode
Service function can be selected using
Service +/-
Thermal Setting the INCREASE or DECREASE keys on the RIGHT side of the display.
Scroll the menu until Thermal Setting is displayed

Press the PROGRAM key to enter Thermal Setting.

INCREASE or DECREASE keys on the RIGHT to change


Enable/Disable
Thermal Control  from Enable/Disable Thermal Control
 to Thermal Values Editor

Enable/Disable Thermal Control


Press the PROGRAM key to enter Enable / Disable Thermal Control.

Control ON
Set OFF?
Keep pressed the PROGRAM key enter selection.
Control Disabled
Control OFF
Control OFF
Set ON?
Keep pressed the PROGRAM key enter selection.
Control Enabled
Control ON
To exit see below

Thermal Values Editor


Press the PROGRAM key to enter Thermal Values Editor.

Service function can be selected using


Thermal Values
Editor the INCREASE or DECREASE keys on the RIGHT side of the display.
Scroll the menu until Thermal Setting is displayed

Press the PROGRAM key to enter Thermal Values Editor

Th.Set:1.2 >1.2 INCREASE or DECREASE keys on the RIGHT


Th.err:0.0 >4.7
to change first parameter Th.set

Th.Set:1.2 >1.2 INCREASE or DECREASE keys on the LEFT


Th.err:0.0 >4.7 to change second parameter Th.err

Keep pressed the PROGRAM key to store the values

NOTE Default values set in manufacturing:


Th.set: 1.2, TH.err: 4.7

Exit
Press BACK key to exit Thermal Setting

Press TEST Mode key to exit SERVICE MODE

69 682 91510 - 151030 17/40


FONA ART Plus Service Manual

2.8 Keyboard Test


This function is intended to display the code of the key buttons of the
control panel and of the Hand-Switch for X-ray exposure
Codes for the keys are reported here below.

X-Ray 2L+ 2L- 2R+ 2R- 4 13 5 6 7 8


By keeping pressed the BACK key for 5 s
the system enters Service Mode
Service function can be selected using
Service +/-
Keyboard Test the INCREASE or DECREASE keys on the RIGHT side of the display.
Scroll the menu until Keyboard Test is displayed

Press the PROGRAM key to enter Keyboard Test.

** Keyb. Test ** By keeping pressed the key button, the reference to the key is indicated
Key X-Ray as per above table (X-ray pushbutton pressed on the side example).

The EXIT functionality of the BACK and the TEST keys within the
NOTE
Keyboard Test Service Function are specially executed by keeping them
pressed.
Keep pressed BACK key to exit Keyboard Test

Keep pressed TEST Mode key to exit SERVICE MODE

2.9 LEDS Test


This function is intended to turns ON one by one all the lights on the
control panel as listed in the table below.

NOTE The yellow light on the X-Ray pushbutton is not included in this test

P1 P2 P3 P4 P5 P6 P7

S M L XL GREEN YELLOW RED TEST


By keeping pressed the BACK key for 5 s the system enters Service Mode

Service +/- Service function can be selected using


LEDS Test the INCREASE or DECREASE keys on the RIGHT side of the display.
Scroll the menu until LEDS Test is displayed

Press the PROGRAM key to enter LEDS Test.

** LED Test **

Press RETURN Arm Movement key to start LEDS Test.

Press BACK key to exit LEDS Test

Press TEST Mode key to exit SERVICE MODE

18/40 69 682 91510 - 151030


FONA ART Plus Service Manual

2.10 Automatic Test


This function is intended to perform a number of panoramic cycles for the
program and technique factors as set before entering this function.
By keeping pressed the BACK key for 5 s
the system enters Service Mode

Service +/- Service function can be selected using


Automatic Test the INCREASE or DECREASE keys on the RIGHT side of the display.
Scroll the menu until Automatic Test is displayed

Press the PROGRAM key to enter Automatic Test.

INCREASE or DECREASE keys on the RIGHT side of the display to select


Autom. Test +/-
100 PAN NO X Ray  100 PAN NO X Ray: to perform 100 panoramic movements
without X-ray emission
 10 PAN + X ray: to perform 10 panoramic movements
with X-ray emission

100 PAN NO X Ray


Press RETURN Arm Movement key to start 100 PAN NO X Ray.

 A Reset is done at the beginning


01
AUTO MOD: # 99  Program number selected is reported on the first line
 Number of cycles missing on second line

Press the EXPOSURE key to exit Automatic Test 100 PAN NO X Ray

NOTE EXIT from SERVICE MODE is done at the end of this function.

10 PAN + X ray
Press the EXPOSURE key to start 10 PAN + X ray

Possibility to change technique factors


*Set kv/mA-mAs*
61 6.3 INCREASE or DECREASE keys on the LEFT to change kV, 61 to 85 kV
INCREASE or DECREASE keys on the RIGHT to change mA, 4 to 10 mA
 A Reset is done at the beginning
02
WAIT...# 8  Program number selected is reported on the first line
Number of cycles missing on second line
Press BACK key to exit Automatic Test 10 PAN + X ray

NOTE EXIT from SERVICE MODE is done at the end of this function.

69 682 91510 - 151030 19/40


FONA ART Plus Service Manual

2.11 X-Ray Test


This function is intended to perform irradiation without motion.
By keeping pressed the BACK key for 5 s
the system enters Service Mode
Service function can be selected using
Service +/-
X-Ray Test the INCREASE or DECREASE keys on the RIGHT side of the display.
Scroll the menu until X-Ray Test is displayed

Press the PROGRAM key to enter X-Ray Test.

INCREASE or DECREASE keys on the RIGHT side of the display to select


X-Ray Test
Continued X-Ray  Continued X-Ray: set of kV and mA, time defined by the operator
 X-Ray Alignment: set of kV and mA
 Pulsed X-Ray: set of kV and mAs at 10 mA from 2 to 20
ATTENTION Wait for X-ray tube cool down after each exposure. Cool down diagrams
are reported on the Operating Instructions.
5 min waiting time is required after each 15 s panoramic of exposure at
85 kV, 10 mA (about 13000 J)
kV 61 64 67 70 73 76 79 82 85
mA 4 5 6.3 8 10
mAs 2 2.5 3.2 4 5 6.3 8 10 12.5 16 20
Continued X-Ray
Confirm Continued X-Ray with PROGRAM key

Continued X-Ray INCREASE or DECREASE keys on the LEFT to change kV, 61 to 85 kV


67 6.3
INCREASE or DECREASE keys on the RIGHT to change mA, 4 to 10 mA

Align X-Ray

NOTE Function not used.

Confirm Align X-Ray with PROGRAM key

X-Ray Alignment INCREASE or DECREASE keys on the LEFT to change kV, 61 to 85 kV


67 6.3
INCREASE or DECREASE keys on the RIGHT to change mA, 4 to 10 mA

Pulsed X-Ray
Confirm Pulsed X-Ray with PROGRAM key

INCREASE or DECREASE keys on the LEFT to change kV, 61 to 85 kV


Pulsed X-Ray
67 2.0 INCREASE or DECREASE keys on the RIGHT to change mAs from 2 to 20.
2.0 /2.5 /3.2 /4.0 /5.0 /6.3 /8.0 /10.0 /12.0 /16.0 /20.0
Exposures are done at 10 mA.

Exit
Press BACK key to exit X-Ray Test

Press TEST Mode key to exit SERVICE MODE

20/40 69 682 91510 - 151030


FONA ART Plus Service Manual

2.12 Encoders Test


This function is intended to perform test of position and motion sensors
By keeping pressed the BACK key for 5 s
the system enters Service Mode
Service function can be selected using
Service +/-
Encoders Test the INCREASE or DECREASE keys on the RIGHT side of the display.
Scroll the menu until Encoders Test is displayed

Press the PROGRAM key to enter Encoders Test.

Encoder Test +/- INCREASE or DECREASE keys on the RIGHT side of the display to select
Rotation Axis Rotation axis or X axis

Rotation Axis
Rot Axis <--> Use INCREASE or DECREASE keys on the RIGHT side of the display to
A:1 B:0 C:0 move the rotation arm and verify change of status of the position sensors
for Rotation Axis

X Axis
Use INCREASE or DECREASE keys on the RIGHT side of the display to
X Axis <-->
A:1 B:1 C:1 move the rotation arm and verify change of status of the position sensors
for X Axis

Exit
Press BACK key to exit Encoders Test

Press TEST Mode key to exit SERVICE MODE

2.13 Voltages Test


This function is intended to get indication of actual levels of low voltage
supply at 38, 24, 15, and 5V with indication on the display.
By keeping pressed the BACK key for 5 s
the system enters Service Mode

Service +/- Service function can be selected using


Voltages Test the INCREASE or DECREASE keys on the RIGHT side of the display.
Scroll the menu until Voltages Test is displayed

Press the PROGRAM key to enter Voltages Test.

The values of the low voltage supplies are indicated. In the example:
+38: 35 +24: 24
+15: 15 +5: 5.0  35 V out of 38
 24 v out of 24
 15 V out of 15
 5.0 out of 5 V
Press BACK key to exit Voltages Test

Press TEST Mode key to exit SERVICE MODE

69 682 91510 - 151030 21/40


FONA ART Plus Service Manual

3. Service Procedures
3.1 Plastic and Metal Covers

A
G
B

C
I

H E

A Vertical Upper Top cover Allen screw 3 mm, 2 at top and 2 on the back
B Vertical Upper Bottom cover Allen screw 3 mm, 2 on the back and 1 in front
C Mirror Cover Allen screw 3 mm, 4 on the back
D Display cover Phillips screw, 2 at top, 1 at bottom
E Inner Tube Head Cover Phillips screw, 2 on the back. Remove D first.
F Drive Lower Cover Phillips screw, 2 at bottom tube head side
G Drive Upper Metal Cover Allen screw 3 mm, 4 and 4 Phillips screws
H Inner Sensor Cover Phillips screw, 1 at bottom
I Outer Sensor Cover Phillips screw, 4 inside plus Allen screw 2 mm on knob.
J Shield on Power Supply Unit Phillips screw, 2 at top
K Shield on Power Switching Phillips screw, 4 at sides
L Back shield on Control Unit Phillips screw, 4

K
L

22/40 69 682 91510 - 151030


FONA ART Plus Service Manual

3.2 Replacement of Power Supply Unit

XC301
XT301 XT302

XC308
VT303
XT305

TEST
15V
38V
24V

EXP
ZH
SA301

XC303 XC304 XC302

 Remove Vertical Upper Top Cover (A) and Shield on Power Supply (J)
 Disconnect all cables, replace the board, and reconnect all cables
XC305:  Input of mains voltage from output of mains filter
XC303:  300 VDC output voltage to power switching unit
XC304:  Connection for additional buzzer
XC302:  Multipole cable
TEST 15, 38, 24:  Test points for low voltages 15, 24, and 38 VDC
ZH:  Filament supply signal
EXP:  Exposure signal
XC301:  Hand Switch connection with ground wire to XC302
XC308:  Laser supply at – 3.1 V
 VT303: jumper for 24 V supply to hand-switch
SA301:  ON for normal use,
 OFF to prevent exposures on demo units

NOTE  Verify Anode Current and in case perform filament calibration

69 682 91510 - 151030 23/40


FONA ART Plus Service Manual

3.3 Replacement of Control Unit Board

XC3: XC4:
CONNECTOR TO DISPLAY
CONTROL PANEL XC11 XC9

XC2: XC12 XC10


SERIAL
PORT XC14
XC15

JUMPER XC18
SIGNAL XC19 XC20
THERMAL
MOTOR SENSOR
X AXIS
INPUT
MOTOR SYNCRO 38VDC XC13
ROTATION SENSOR
XC17
XC5 XC6 XC7 TO
OUTPUT XC101
38VDC
 Remove Display Cover (D), Drive Upper Metal Cover (G) and Shield on
Power Switching unit (K), unlock the control unit module, bend it open and
remove the Back Shield (L).
 Disconnect all cables, mount the new control unit board, and connect back
all the cables as indicated.
 Configure the unit with original values:
 Upload firmware via serial port and null modem cable, if needed
Load Rotation and Translation Offset values. See chapter 0

 PAN Alignment, on page 12 (see label on top of power supply cover)


 Load CAM setting value, if needed (Prog 6)

24/40 69 682 91510 - 151030


FONA ART Plus Service Manual

3.4 Replacement of Power Switching and Regulation Unit JED 28


XC101 XC203
XC204 XC201

XT201
XT202

Remove Display Cover (D), Drive Upper Metal Cover (F), Shield on Power
switching Unit (K), unlock the control unit module and bend it open (see
chapter 3.1 Plastic and Metal Covers, on page 22)

Disconnect all cables, replace the board, and reconnect all cables

XC201: 300VDC power supply


XC202: Flat cable from control unit
XC203: Multiple cable
XC204: 38VDC supply to digital sensor
XT201: Supply to High Voltage transformer in the Tube Housing Assembly
XT202: Filament supply (ZH1, ZH2) and feedback signals (mA, kV)

NOTE Filament calibration is required.

69 682 91510 - 151030 25/40


FONA ART Plus Service Manual

3.5 Replacement of Tube Housing Assembly

POWER SWITCHING UNIT JED 28

FEEDBACK
FILAMENT SIGNALS
SUPPLY kV & mA

THERMAL
SENSOR HIGH VOLTAGE
SUPPLY
TER

 Remove Display Cover (D), Drive Upper Metal Cover (G), Shield on Power
Switching unit, unlock the control unit module, and swing it out open.
 Disconnect all wires of the Tube Housing Assembly (THA) from the Power
Switching Unit and ground wire from the frame.
 Dismount the tube housing assembly (THA) by removing the 4 mm Allen
screws in front (quantity 4) and mount the new one.
 Connect all wires previously disconnected
 Filament Supply (two wires: ZH1, ZH2)
 Ground connection (GND)
 mA feedback (mA)
 kV feedback (kV)
 TER Thermal Sensor
 High Voltage Supply (two wires plus ground : UN1, UN2, GND)
 Perform filament calibration
 Per form the X-Ray beam alignment

26/40 69 682 91510 - 151030


FONA ART Plus Service Manual

3.6 Filament Calibration

RP 103 GROUND
TRIMMER
mA
TEST
POINT
FDB
SA10
1

GROUND RP 104
TRIMMER
FDB
Vs and V-
mA TEST POINTS
TEST POINT

kV
TEST POINT

Filament calibration is required every time a Tube Housing Assembly or a


NOTE Power Switching Unit is replaced
ATTENTION
Wait for cooling time between exposures

NOTE Make sure the DC voltmeter to be used is calibrated


 Enter Service program X-RAY TEST /CONTINUED mode.
 Set the anodic voltage at 61 kV and anodic current at 10 mA.
 Switch to ON both switches SA101.
 Monitor the voltage between mA point and GROUND.
 Tune trimmer RP103 (mA) for 6V during exposure.
 Switch to OFF both switches SA101.
 If this procedure is done after replacement of a Power Switching Unit, the
NOTE
feedback signal (FDB) has to be calibrated too:
 Monitor the voltage between from FDB test point and GROUND.
 Set the anode voltage at 85 kV and anode current at 10 mA.
 Tune trimmer RP104 (FDB) for 10V during exposure.
 Exit Service Program X-RAY TEST /CONTINUED mode.

69 682 91510 - 151030 27/40


FONA ART Plus Service Manual

3.7 X-Ray Beam Alignment


 Switch the unit ON
 Keep pressed the BACK key for 5 s to enter Service Mode
 Select Service function with INCREASE or DECREASE keys on the
Service +/-
X-Ray Test RIGHT side of the display.
 Scroll the menu until X-Ray Test is displayed
 Press the PROGRAM key to enter X-Ray Test.

 INCREASE or DECREASE keys on the RIGHT side of the display to


Continued X-Ray
67 6.3 select Continued X-Ray

 Confirm Continued X-Ray with PROGRAM key

20 mm  Using masking tape, attach the 20 mm


Al filter Al filter over the X-Ray generator.
 Start the OrisWin Suite software and
click on Patients. Select the “demo”
patient from the list and start image
management.
 Click on Panoramic module. In the Ajat Panoramic Application, select
Tools > Maintenance > Mechanical Alignment.
 Set kV and mA as indicated on the screen.
Continued X-Ray
61 4.0 **  INCREASE /DECREASE keys on the LEFT to change kV
 INCREASE /DECREASE keys on the RIGHT to change mA
ATTENTION  Move to a safe position and expose
RADIATION
Picture Examination  While the panoramic unit is producing radiation, and assuming that
the beam falls on the surface of the sensor, you will see an
illuminated area representing the position of the actual beam
detected by the sensor.
 The numbers located on the top and bottom of the window indicate
the direction the sensor should be moved for proper alignment. (seen
from behind the sensor).
 If the beam is not detected by the software, or if it appears bent and
not vertical, laterally shifted, or vertically shifted, alignment has to be
performed.
 See next page for alignment instructions.

 If the green “Alignment ok” is displayed, the alignment is complete.

NOTE: Very minor deviations are tolerated.

NOTE: Be sure to check the alignment again after replacing any


covers.
NOTE: Perform the calibration procedure after the alignment has
been completed (refer to section 3.9 Sensor Calibration).

28/40 69 682 91510 - 151030


FONA ART Plus Service Manual

Beam Rotation If the beam is not vertical, correct the rotation by loosening the 4
screws (Allen 2.5 mm) of the holding plate of the collimator and by
rotating it as needed, within the mounting tolerances.

Beam Vertical Shitf If the beam is not vertical, correct the rotation by loosening the 4
screws (Allen 2.5 mm) of the holding plate of the collimator and by
rotating it as needed, within the mounting tolerances.

Beam Horizoltal Shitf Shift to the left or to the right the tube housing assembly, having
loosened the 4 screws (Allen 3 mm) holding it. A shift tuning screw
at side is available for fine adjustment (pulling action only).

THA HOLDING SCREWS


SHIFT TUNING
SCREW

KNOB
FRANKFURT
LIGHT COLLIMATOR
HOLDING
PLATE

69 682 91510 - 151030 29/40


FONA ART Plus Service Manual

3.8 Replacement of Position Sensors

 The position sensors are made of three electro-optical devices each


combining a light transmitter and a light receiver.
 The presence of light is detected, or the absence of the same due to a
non-transparent object cutting the beam of light.
 A metal disk and a linear plate with shaped profile and bearing a set of
holes are used to generate the signals for the rotation and translation
movement of the rotating arm (drive).
 Device A intercepts beginning and end of movement.
C B A  Device B intercepts the intermediate point of the movement
(mid-rotation or mid-translation).
 Device C is aligned with the holes to generate pulse signals

SENSORS FOR
ROTATION

SENSORS FOR
TRANSLATION

XC9 XC10

Sensors for Rotation  Switch the system OFF


 Remove Vertical Upper Top Cover (cover A).
 Disconnect connector XC9.
 Remove the two holding screws and replace the sensors module.
 Align device C on the holes of the encoder disk.
NOTE. Rotation Axis Offset.  Place back connector XC9 and switch the unit ON.
Set ROTATION OFFSET to have  Enter Service function Encoders Test and select Rotation Axis.
the rotating arm in the middle of  Verify the Rotation Axis OFFSET value and in case load new value.
rotation aiming normally to the
 Record date of change and new offset value.
mirror

Sensors for Translation  Switch the system OFF


 Remove Vertical Upper Top Cover (cover A).
 Disconnect connector XC10.
NOTE. X Axis Offset.  Remove the two holding screws and replace the sensor module.
Set TRANSLATION OFFSET to
 Align device C on the holes of the linear plate.
have the mid plane of the rotating
arm (vertical plane from  Place back connector XC10 and switch the unit ON.
collimator slit to imaging receptor)  Enter Service function Encoders Test and select X Axis.
when in PATIENT ENTRY  Verify the X Axis OFFSET value and in case load new value.
POSITION at 80 mm from the tip
 Record date of change and new offset value.
of the bite block

30/40 69 682 91510 - 151030


FONA ART Plus Service Manual

3.9 Sensor Calibration


The software calibration tool is used after any alignment or
hardware adjustments performed to the panoramic machine or
sensor.
Calibration is to be performed after the sensor has been
properly aligned, securely mounted and the entire panoramic
unit has been fully assembled with all covers.
Any alignment parts that may obstruct the XRAY beam must
be removed. This includes bite pieces, chin rests, and forehead
alignments.
Calibration allows the imaging software to get to know a
particular panoramic machine and makes software
adjustments for optimum performance.
For this process, you will need the 20 mm Al filter provided.
 To perform a calibration start the OrisWin Suite software and
click on Patients. Select any patient from the list and start
image management. Click on Panoramic module. In the Ajat
Panoramic Application, click Tools> Maintenance> Calibration.
 Attach the supplied 20 mm Al filter over the tube head
covering the focal spot.
 Click Calibrate when ready.
 The calibration process consists of several steps which include
making three exposures. Follow the on screen directions.
 Once the calibration is complete, the results will be displayed
on the right.
 If all the categories are green, the calibration was successful.
Yellow graphs indicate that maximum tolerances are close to
being exceeded and should be checked.
If the results are green or yellow, verify that the sensor
installation was successful by taking a test image.
Red graphs indicate that a maximum tolerance has been
exceeded.

69 682 91510 - 151030 31/40


FONA ART Plus Service Manual

4. System Messages
4.1 Warning Messages
X-ray generator hot The system is cooling down and the requested exposure would exceed the loading
capacity. Wait until ready.

CAM init Wait for system to complete initialization.

Adjust Arm Position  The rotation arm was moved out of position with unit switched OFF and cannot be
initialized.
 Turn the system OFF.
 Place manually the arm in “patient entrance” position by rotating and shifting the
carriage slowly.
 Turn system ON and perform initial reset.
Wait PC READY The communication line to the computer is missing or the image acquisition
program is not in input mode. Perform necessary correction.
This warning message is not active in test mode (service functions).

4.2 Error Messages


Error 1 kV reference signal out of range
 Measure the voltage reference for kV, 10 kV/V from 6.1 to 8.5 V on the Power
switching Unit during exposure in X-ray Test /Continued X-ray mode.
 Replace the Control Unit if value is not corresponding to the display setting.
Error 2 mA reference signal out of range
 Measure the voltage reference for mA, 1 V/mA from 4 to 10 V on the Power
switching Unit during exposure in X-ray Test /Continued X-ray mode.
 Replace the Control Unit if value is not corresponding to the display setting.
Error 3 Carriage rotation encoder failure
 Position sensors for rotation to be aligned or to be replaced.
Error 4 Carriage displacement encoder failure
 Position sensors for translation to be aligned or to be replaced.
Error 5 Carriage rotation sensor failure
 Position sensors for rotation to be aligned or to be replaced.
Error 6 Carriage displacement sensor failure
 Position sensors for translation to be aligned or to be replaced.
Error 9 Tube-head temperature exceeding limit
 Wait for tube-head to cool down.
Error 11 Exposure aborted during irradiation
 This is usually a consequence of the “dead man functionality” which terminates
irradiation upon release of the exposure hand-switch.
 If error appeared without request by the operator check the cable of the hand-
switch, its connector or the Control Unit.
Error 12 Exposure aborted before irradiation
 This is usually a consequence of the “dead man functionality” which terminates
irradiation upon release of the exposure hand-switch.
 Error generated by interruption during movement before irradiation. If error
appeared without request by the operator check the cable of the hand-switch, its
connector or the Control Unit.

32/40 69 682 91510 - 151030


FONA ART Plus Service Manual

Error 13 38 V DC supply out of range


 Check the value with a voltmeter on the test point of the Power Supply Board and
replace it if the value is not in the range 34 to 42 V.
 Conversely, if the electrical value is in the range, with the service function “Voltages
Test” verify the abnormal readout of the Control Unit: replace it in case.
 Check also the integrity of the connecting cables.
Error 14 24 V DC supply out of range
 Check the value with a voltmeter on the test point of the Power Supply Board and
replace it if the value is out of the range 23 to 25 V.
 Conversely, if the electrical value is in the range, with the service function “Voltages
Test” verify the abnormal readout of the Control Unit: replace it in case.
 Check also the integrity of the connecting cables.
Error 15 15 V DC supply out of range
 Check the value with a voltmeter on the test point of the Power Supply Board and
replace it if the value is out of the range 14.5 to 15.5 V.
 Conversely, if the electrical value is in the range, with the service function “Voltages
Test” verify the abnormal readout of the Control Unit: replace it in case.
 Check also the integrity of the connecting cables.
Error 16 5 V DC supply out of range
 Check the value with a voltmeter on the test point of the Power Switching Unit and
replace it if the value is out of the range 4.75 to 5.25 V.
 Conversely, if the electrical value is in the range, with the service function “Voltages
Test” verify the abnormal readout of the Control Unit: replace it in case.
 Check also the integrity of the connecting cables.
Error 17 High voltage failure
 The inverter current or tube voltage are increasing abnormally.
 Failure on the Power Switching Unit or in the Tube Head itself is possible.
 Replace the Power Switching Unit first and the tube-head in case.
Error 19 Exposure time exceeded
 Possible malfunction of the Control Unit. Replace it.
Error 20 Exposure aborted after irradiation
 This is usually a consequence of the “dead man functionality” which terminates
irradiation upon release of the exposure hand-switch.
 Error due to interruption during movement after irradiation. Re-take is not needed,
unless in the case of quitting in the middle of a TMJ procedure.
 If error appeared without request by the operator check the cable of the hand-
switch, its connector or the Control Unit.
Error 21 No tube current
 No or low anode current due to:
 Interruption of High Voltage power supply (Power Switching Unit)
 Interruption of Filament power supply (Power Switching Unit)
 Interruption of the regulation circuits (Power Switching Unit)
 Fault in the Tube-Head (input transformer or multiplying circuits).
 Check switch SA 301 (must be ON) and fuses FU 304 and FU 305 in the Power
Supply Unit
 If no defect is found, enter Service Function X-ray test /Continued, switch exposure
and check the HV inverter supplying voltage (between points VS and V- in the Power
Switching Unit) to be in range from 50 to 100 volts DC.
 If all is correct, the circuit between the secondary of High Voltage transformer and
the X-ray tube is broken.

69 682 91510 - 151030 33/40


FONA ART Plus Service Manual

 Depending on the location of the fault, replace the Power Switching Unit, the
connecting cables or the Tube-Head.
 This error can be also activated by sparking in the X-ray tube, when a very short
over-current signal switches off the High Voltage regulator, but is not trapped by the
Control Unit in order to activate error 17. In this case try to perform X-ray tube
seasoning (de-gassing with a number of exposures at low kV) or replace the Tube-
Head.
Error 23 Filament current out of range
 The filament current is exceeding the permitted range.
 Power Switching Unit to be replaced first and Control Unit in case.
Error 25 Thermal sensor failure
 The thermal sensor is broken or disconnected. Verify continuity and presence of
thermal sensor jumper on the control unit board.
 In case of failure of the thermal sensor inside the tube-head:
 Replace the Tube-Head.
 Alternatively through the Service Function “Thermal Settings” disable the
functionality for minimization of waiting time for cool down and control loading
manually using the cooling curves.
Error 32 Control unit fault
 Replace the Control Unit
Mains Fuse Blown Mains Fuse Blows at X-ray ON
 Check resistors R321, R322, R323, R324, R325 in the Power Supply Unit. The
problem can be generated in the Power Supply Unit itself, or by a short circuit in the
Power Switching Unit.
 Replace the Power Supply Unit or, in case of short circuit, the Power Switching Unit.

4.3 Other Alarms


Other Alarms  Take a note of the message generated and refer to the Technical Service.
 Internal Software Error, with error code xxh and
address yyyyyyyh. ?? ISE ??
Switch the system OFF and ON again. C: xxh A:yyyyyyyh
 Over-Current. This occurs for a short circuit on a
stepper motor: Over Current
A = Rotation Axis, B = X Axis Axe: A B
Switch the system OFF and remove the short circuit
on the motor,
then switch the system ON.
 TRIPOS, microprocessor’s operating system error.
 Application abort.

34/40 69 682 91510 - 151030


FONA ART Plus Service Manual

5. Spare Parts
5.1 Positioning & Test Items
A Bite Block
Code 76 680 10017

B Chin Rest + Bite block (kit)


Code 76 680 10027

C Nasal support for edentulous


Code 76 680 10018

D Temporal Resting Bars


Code 76 680 30056

E Line Laser Module


Code 76 680 20120

F Test Phantom
Code 76 680 10036

69 682 91510 - 151030 35/40


FONA ART Plus Service Manual

5.2 Power Supply & Control Modules


A Board Dual Voltage Supply 115-230V
Code 76 190 24760

B Mains Connection Module


Code 76 680 20095

C Power cable Inverter


Code 76 680 20200

D Signal Cable Inverter


Code 76 680 20300

E Low Voltage Transformer 2S


Code 76 190 20260

F Mains Switch
Code 76 190 26190

36/40 69 682 91510 - 151030


FONA ART Plus Service Manual

G EMI Filter
Code 76 190 24730
20A 120/250V 50-60Hz

H Set of Fuses 115 V 4 Types


Code 76 190 24475
Set of 10 for each type

I Sets of Fuses 230V 4 Types


Code 76 190 24470
Set of 10 for each type

J Board Control Unit


Code 76 680 20040

K Control Panel FONA ART Plus


76 680 20039

L Position Sensor
Code 76 680 20020

M Limit Switches Column


Code 76 680 20085

69 682 91510 - 151030 37/40


FONA ART Plus Service Manual

N Board Power Switching and Regulation JED28

Code 76 680 20370

O Tube Housing Assembly

Code 76 680 20160

P Hand Switch FONA with Cable

Code 76 680 20390

Q Kit Remote Hand-Switch

Code 93 190 00155

R Up/Down Pushbuttons FONA

Code 76 680 20079

38/40 69 682 91510 - 151030


FONA ART Plus Service Manual

5.3 Mechanical Items


A Side Nylon Bearing

Code 76 680 30075

B Eccentric Nylon Bearing

Code 76 680 30080

C Nylon Bearing

Code 76 680 30085

D Stepper Motor

Code 76 680 20060

E Raising Motor Kit


Code 76 680 30100

F Plastic cover for self standing base


Code 76 680 21600

69 682 91510 - 151030 39/40


We reserve the right to make any alterations which may be required due to clinical improvements

FONA ART PLUS Service Manual – English Edition 151030

*6968291510*
FONA Dental s.r.o.
Stefanikova 7 SK-811 06
Bratislava, Slovakia www.fonadental.com

You might also like