You are on page 1of 11

Int. J. Pharm. Med. & Bio. Sc. 2014 K Naresh et al.

, 2014

ISSN 2278 – 5221 www.ijpmbs.com


Vol. 3, No. 3, July 2014
© 2014 IJPMBS. All Rights Reserved
Review Article

ULTRA PERFORMANCE LIQUID


CHROMATOGRAPHY
K Naresh1*, S Bhawani1 and T Maneesh Kumar1

*Corresponding Author: K Naresh  banu.sunkara@gmail.com

Ultra Performance Liquid Chromatography (UPLC) takes the advantage of technological strides
made in particle chemistry performance, system optimization, detector design and data
processing and control. Using sub 2 mm particles and mobile phases at higher linear velocities
and instrumentation that operates at higher pressures than those used in HPLC, dramatic
increases in resolution, sensitivity and speed of analysis can be obtained. This new category of
analytical separation science retains the practicality and principles of HPLC while creating a
step function improvement in chromatographic performance. This review introduces the theory
of UPLC and summarizes some of the most recent work in the field.

Keywords: UPLC, HPLC, Sensitivity, Resolution

INTRODUCTION capability to take full advantages of sub-2


µm particles (Michael, 2005).
UPLC refers to high Ultra Performance Liquid
Chromatography. It improves in three areas: UPLC can be regarded as new invention
chromatographic resolution, speed and sensitivity for liquid chromatography. UPLC brings dramatic
analysis. It uses fine particles and saves time improvements in sensitivity, resolution
and reduces solvent consumption. UPLC comes and speed of analysis can be calculated. It has
from High Performance Liquid Chromatography instrumentation that operates at
(HPLC). HPLC has been the evolution of the high pressure than that used in HPLC and in this
packing materials used to effect the separation. system uses fine particles (less than 2.5 µm) and
An underlying principle of HPLC dictates that as mobile phases at high linear velocities decreases
column packing particle size decreases, the length of column, reduces solvent
efficiency and thus resolution also increases. consumption and saves time. According to the
HPLC is approved technique as it has been used van Deemter equation, as the particle size
in laboratories worldwide over the past 30-plus decreases to less than 2.5 µm, there is a
years. HPLC technology simply doesn’t have the significant gain in efficiency, while the efficiency

1
Balaji Institute of Pharmaceutical Sciences, Laknepally, Narsampet, Warangal-506 331, Andhra Pradesh, India.

84
Int. J. Pharm. Med. & Bio. Sc. 2014 K Naresh et al., 2014

does not diminish at increased flow rates or linear 1


N
velocities (Swartz, 2005). w2
Therefore by using smaller particles, This illustrates that the narrower the peaks are,
speed and peak capacity (number of peaks the easier they are to separate from each other.
resolved per unit time in gradient separations) Also, peak height is inversely proportional to the
can be extended to new limits, termed UPLC. The peak width:
technology takes full advantage of
1
chromatographic principles to run separations H
w
Using columns packed with smaller particles (less
than 2.5 µm) and/or higher flow rates for So as the particle size decreases to increase
increased speed, this gives superior resolution N and subsequently Rs, an increase in sensitivity
is obtained, since narrower peaks are taller
and sensitivity .Now a days in industrial area
peaks. Narrower peaks also mean more peak
UPLC refers for some of the most recent work
capacity per unit time in gradient separations,
field.
desirable for many applications, e.g., peptide
CHEMISTRY OF SMALL maps. Still another equation comes into play when
migrating toward smaller particles:
PARTICLES
Smaller particles provide not only increased 1
Fxxx
efficiency, but also the ability to work at increased dpc
linear velocity without a loss of efficiency, providing
This relationship also is revealed from the van
both resolution and speed. Efficiency is the
Deemter plot. As particle size decreases, the
primary separation parameter behind UPLC,
optimum flow F opt to reach maximum N
since, it relies on the same selectivity and
increases. But since back pressure is
retentivity as HPLC. In the fundamental resolution
proportional to flow rate, smaller particle sizes
(Rs) equation: resolution is proportional to the
require much higher operating pressures, and a
square root of N.
system properly designed to capitalize on the
efficiency gains; A system that can both reliably
N   1   k 
Rs     deliver the requisite pressures and that can
4     k 1
System Efficiency Selectivity Re sentivity maintain the separation efficiency of the small
particles with tightly managed volumes. Efficiency
But since N is inversely proportional to particle is proportional to column length and inversely
size (dp): as the particle size is lowered by a proportional to the particle size.
factor of three, from, for example, 5 µm (HPLC
L
scale) to 1.7 µm (UPLC-scale), N is increased N
dp
by three and resolution by the square root of three
or 1.7. N is also inversely proportional to the Therefore, the column can be shortened by
square of the peak width: the same factor as the particle size without loss

85
Int. J. Pharm. Med. & Bio. Sc. 2014 K Naresh et al., 2014

of resolution. Using a flow rate three times higher v is the linear velocity, the carrier gas flow rate.
due to the smaller particles and shortening the
*The A term is independent of velocity and
column by one third (again due to the smaller
represents “eddy” mixing. It is smallest when the
particles), the separation is completed in 1/9 the
packed column particles are small and uniform.
time while maintaining resolution. Although high
efficiency, nonporous 1.5- particles are The B term represents axial diffusion or the
commercially available, they suffer from poor natural diffusion tendency of molecules. This
loading capacity and retention due to low surface effect is diminished at high flow rates and so this
area. Silica-based particles have good mechanical term is divided by v.
strength but can suffer from a number of *The C term is due to kinetic resistance to
disadvantages, which include a limited pH range equilibrium in the separation process. The kinetic
and tailing of basic analytes. Polymeric columns resistance is the time lag involved in moving from
can overcome pH limitations, but they have their the gas phase to the packing stationary phase
own issues including low efficiencies, limited and back again. The greater the flow of gas, the
loading capacities and poor mechanical strength. more a molecule on the packing tends to lag
Packing a 1.7 m particle in reproducible and behind molecules in the mobile phase. Thus term
rugged columns was also a challenge that needed is proportional to v. Therefore it is possible to
to be overcome, however. A smoother interior increase throughput, and thus the speed of
surface for the column hardware, and re- analysis without affecting the chromatographic
designing the end frits to retain the small particles performance. The advent of UPLC has demanded
and resist clogging were necessary. Packed bed the development of a new instrumental system
uniformity is also critical, especially if shorter for liquid chromatography, which can take
columns are to maintain resolution while advantage of the separation performance (by
accomplishing the goal of faster separations. reducing dead volumes) and consistent with the
pressures (about 8000 to 15,000 PSI, compared
PRINCIPLE with 2500 to 5000 PSI in HPLC). Efficiency is
The UPLC is based on the principal of use of proportional to column length and inversely
stationary phase consisting of particles less proportional to the particle size (Lars and
than 2.5 m (while HPLC columns are typically Honore, 2003). As shown in Figure 1,
filled with particles of 3 to 5 m). The underlying smaller particles provide increased efficiency as
principles of this evolution are governed by the well as the ability to work at increased linear
Van Deemter equation, which is an empirical velocity without a loss of efficiency, providing both
formula that describes the relationship between resolution and speed. Efficiency is the primary
linear velocity (flow rate) and plate height (HETP separation parameter behind UPLC since it relies
or column efficiency) (Van et al., 1956) . on the same selectivity and retentivity as HPLC.
H=A+B/v+Cv UPLC Stability Indicating Assay
where; UPLC Conditions
A, B and C are constants 1. Column: 2.1 x 30 mm 1.7 m ACQUITY BEH
C18

86
Int. J. Pharm. Med. & Bio. Sc. 2014 K Naresh et al., 2014

2. Temp: 30°C. A 45 s, 5-85% B linear gradient, • High-pressure fluidic modules


3. Flow rate: 0.8 mL/min was used. • Minimized system volume
4. Mobile phase: A was 10 mm ammonium • Negligible carryover
formate, pH 4.0, B was acetonitrile. UV
• Reduced cycle times
detection at 273 nm and 40 pts/s
• Last response detectors
5. Peaks in order: 5-nitroso-2, 4, 6-triaminopyrimidine,
4-amino-6-chloro-1, 3-benzenesulfanamide, • Integrated system software and diagnostics
hydrochlorthiazide, triamterine, and
The ACQUITY UPLC System’s high-pressure
methylbenzenesulfanamide; 5 µL injection, 0.1
fluidics optimizes flow rates to make the most of
mg/mL each.
small particle technology. The ACQUITY UPLC
Capitalizing on Smaller Particles System’s sample-handling design is designed to
Only small particles are not responsible for fast ensure exceptionally low carryover and reduced
resolution and sensitivity, some special cycle time. And when interfaced with the Sample
instrumentation system should be design. So Organizer, it increases unattended sample
some special kind of system capable of delivering capacity by ten times. High-speed detectors, both
the pressures required to realize the potential of optical and mass, contribute to increased
UPLC have been reported in the literature and sensitivity and help manage the heightened speed
elsewhere (Swartz, 2005). In early 2004, the first and resolution requirements of UPLC.
commercially available UPLC system that
• ACQUITY UPLC Systems are easily
embodied these requirements was described for
controlled, diagnosed, and monitored via a
the separation of various pharmaceutical related
small organic molecules, proteins, and peptides; graphical system console interface. The
it is called the ACQUITY UPLCTM System. The console offers:
ACQUITY UPLC System consists of a binary • Quick and easy access to critical instrument
solvent manager, sample manager (including the parameters.
column heater), detector, and optional sample
• Simple system start-up, elegant system status
organizer. The binary solvent manager uses two
monitoring and predictive performance
individual serial flow pumps to deliver a parallel
indicators to ensure maximum productivity.
binary gradient mixed under high pressure.
• Data management capabilities that are
Instrumentation
supported by both MassLynx™ and
The UPLC System has been holistically designed
Empower™ software.
to match the performance needs of innovative
column chemistries with robust hardware, easy- • The ACQUITY UPLC System is also supported
to-use software and specialized support services. by Intelligent Device Management technology
It consists of: with our Connections INSIGHT™ service,
• Small, pressure-tolerant particles providing instrument diagnostics.

87
Int. J. Pharm. Med. & Bio. Sc. 2014 K Naresh et al., 2014

1. Sample injection the device also needs to be minimal to reduce


potential band spreading. A fast injection cycle
2. UPLC columns
time is needed to fully capitalise on the speed
3. Column manager and heater or cooler afforded by UPLC, which in turn requires a high
4. Detectors sample capacity. Low volume injections with
minimal carryover are also required to increase
5. Softwares
sensitivity. There are also direct injection
6. Accessories approaches for biological samples.
7. Connection insight service (if provided by UPLC Columns
mfg.company.water provided it)
The design and development of sub-2 µm
Sample Inject particles is a significant challenge, and
Lee et al. described the design of injection valves researchers have been active in this area for
and separation reproducibility (Wu et al., 2001) some time, trying to capitalize on their advantages
and the use of a carbon dioxide enhanced slurry 2-4. Although high efficiency, nonporous 1.5 µm
packing method on the capillary scale for the particles are commercially available, they suffer
separation of some benzodiazepines, herbicides, from poor loading capacity and retention due to
and various pharmaceutical compounds (Wu et low surface area. Silica-based particles have
al., 1997). Jorgenson et al. modified a good mechanical strength but can suffer from a
commercially available HPLC system to operate number of disadvantages, which include a limited
at 17,500 psi and used 22 cm long capillaries pH range and tailing of basic analytes. Polymeric
packed with 1.5 mm C18-modified particles for columns can overcome pH limitations, but they
the analysis of proteins (Tolley et al., 2001). The have their own issues including low efficiencies,
calculated pressure drop at the optimum flow rate limited loading capacities, and poor mechanical
for maximum efficiency across a 15-cm long strength. In 2000, Waters introduced XTerra®, a
column packed with 1.7 µm particles is first generation hybrid chemistry that took
approximately 15,000 psi. Therefore, a pump advantage of the best of both the silica and
capable of delivering solvent smoothly and Polymeric column worlds. XTerra columns are
reproducibly at these pressures and that can mechanically strong, with high efficiency and
compensate for solvent compressibility, while operate over an extended pH range. They are
operating in both the gradient and isocratic produced using a classical sol-gel synthesis that
separation modes is required . With 1.7 µm incorporates carbon in the form of methyl groups.
particles, half-height peak In UPLC, sample But in order to provide the necessary mechanical
introduction is critical. Conventional injection stability for UPLC, a second generation Bridged
valves, either automated or manual, are not Ethyl Hybrid (BEH) technology was developed
designed and hardened to work at extreme called ACQUITY BEH, these 1.7 µm particles
pressure. To protect the column from extreme derive their enhanced mechanical stability by
pressure fluctuations, the injection process must bridging the methyl groups in the silica matrix.
be relatively pulse-free and the swept volume of Packing 1.7 µm particles into reproducible and

88
Int. J. Pharm. Med. & Bio. Sc. 2014 K Naresh et al., 2014

rugged columns was also a challenge that needed internal diameter (I.D.) and 150 mm in length, as
to be overcome. Requirements include a well as a bypass channel for flow injections. The
smoother interior surface of the column hardware ACQUITY UPLC System consists of a binary
and re-designing the end frits to retain the small solvent manager, sample manager including the
particles and resist clogging. Packed bed column heater, detector, and optional sample
uniformity is also critical, especially if shorter organizer. The binary solvent manager uses two
columns are to maintain resolution while individual serial flow pumps to deliver a parallel
accomplishing the goal of faster separations. All binary gradient. There are built-in solvent select
ACQUITY UPLC BEH columns also include valves to choose from up to four solvents. There
eCord™ microchip technology that captures the is a 15,000-psi pressure limit (about 1000 bar) to
manufacturing information for each column take full advantage of the sub-2 m particles. The
including the quality control tests and certificates sample manager also incorporates several
technology advancements. Using pressure
Figure 1: ACQUITY UPLC BEH
Column Chemistries assisted sample introduction, low dispersion is
maintained through the injection process, and a
series of pressures transducers facilitate self-
monitoring and diagnostics. It uses needle-in-
needle sampling for improved ruggedness and a
needle calibration sensor increases accuracy.
Injection cycle time is 25 s without a wash and
60 s with a dual wash used to further decrease
carry over. A variety of microtiter plate formats
(deep well, mid height, or vials) can also be
accommodated in a thermostatically controlled
of analysis. environment. Using the optional sample organizer,
the sample manager can inject from up to 22
Column Manager and Heater Cooler microtiter plates. The sample manager also
The ACQUITY UPLC Column Manager, with controls the column heater. Column
automatic column switching, is for high temperatures upto 65o can be obtained.
productivity UPLC sample processing, and its
Detectors
Column Heater/Cooler enables labs to use
temperature as a method parameter. The Half-height peak widths of less than one second
ACQUITY UPLC Column Manager allows users are obtained with 1.7 m particles, which gives
to take full advantage of the performance, range significant challenges for the detector. In order to
of stationary phases, and mechanical strength integrate an analyte peak accurately and
offered by ACQUITY UPLC BEH Columns. The reproducibly, the detector sampling rate must be
Column Manager provides temperature regulation high enough to capture enough data points across
from 10°C to 90°C, automated switching for up the peak. The detector cell must have minimal
to four columns with dimensions to 2.1 mm in dispersion (volume) to preserve separation

89
Int. J. Pharm. Med. & Bio. Sc. 2014 K Naresh et al., 2014

efficiency. Conceptually, the sensitivity increase Softwares


for UPLC detection should be 2-3 times higher ACQUITY UPLC Systems can be easily
than HPLC separations, depending on the controlled, diagnosed and monitored via a
detection technique. MS detection is significantly graphical system console interface with
enhanced by UPLC; an increased peak Empower™ and MassLynx™ software. Both
concentration with reduced chromatographic Empower and MassLynx provide the dynamic
dispersion at lower flow rates promotes increased data processing and information management
source ionization efficiencies. Waters detectors tools to convert the results generated by the
enhance ability to analyze a variety of compounds. ACQUITY UPLC System into valuable knowledge.
Its ACQUITY UPLC detectors.
Accesories
• The Photodiode Array (PDA),
Waters is continually extending the functionality
• Tunable UV (TUV), of the ACQUITY UPLC System: eCord™
Evaporative Light Scattering (ELS) are technology, available on all ACQUITY UPLC
optimized for UPLC system technology by virtue columns, records column history. Sample
of their low dispersion characteristics, high data Organizer that increases system capacity by
acquisition rates, and reliable performance with more than 10 times. FlexCart system platform
cost-effective maintenance support and parts. improves usability, accessibility and convenience

Tunable UV (TUV) Connection INSIGHTTM Service


The ACQUITY Tunable UV/Visible detector cell Connections INSIGHT™ uses Intelligent Device
consists of a light guided flow cell equivalent to Management technology to provide diagnostic
an optical fibre. Light is efficiently transferred down information for the ACQUITY UPLC Systems.
the flow cell in an internal reflectance mode that Connections INSIGHT creates a virtual technical
still maintains a 10 mm flow cell path length with support presence in your lab, enabling Waters to
a volume of only 500 mL. Tubing and connections provide you with proactive and timely service, with
in the system are efficiently routed to maintain the highest level of support and satisfaction.
low dispersion and to take advantage of leak
METHOD OPTIMIZATION
detectors that interact with the software to alert
the user to potential problems [26].
GUIDELINES AND
OBSERVATIONS
UPLC is an ideal inlet for the sensitivity and
During the course of optimizing the UPLC
specificity offered by mass spectrometry. The low
method, considerations to expedite future method
dispersion, high-speed detection performance of
transfers were developed, and the following
Waters MS Technologies, in combination with the
recommendations were made: Ø Increase elution
performance characteristics of UPLC, can
solvent strength to reduce run times taking
dramatically extend detection capabilities.
advantage of the high resolution potential of UPLC

90
Int. J. Pharm. Med. & Bio. Sc. 2014 K Naresh et al., 2014

columns. Increase mobile phase flow rate Partial loop-fill precision was good even at
secondarily to solvent strength in order to promote volumes up to 80% of the loop total volume .
longer column lifetimes. While high mobile phase Typical laboratory practice is to limit sample
linear velocities with good resolution are possible, volume injections to roughly 50% of the total loop
as with any column, routine operation at 80% volume. The UPLC injection system, which
maximum rated pressure led to shortened utilizes air-gap sandwiching of the sample, allows
lifetimes. UPLC operation around 8000 psi or less better utilization of the sample loop and higher
provided comparable or lower column cost per injection precision, reducing the need for use of
assay than HPLC. Maintaining low flows as much the full loop-fill mode. From a practical point of
as possible also reduces solvent and waste view, full loop fill requires substantially greater
disposal costs, although these are already an sample movement considering overfill functions.
order of magnitude less than HPLC. Reduce This likely increases subsequent needle washing,
column re-equilibration times by taking advantage this may impact sample throughput and increase
of the low system dwell volume. Programmed wear of the washing hardware. Larger sample
changes in the mobile phase take time to reach volume transfers also increases exposure to
the column. The small UPLC dwell volume sample particulates, lowering long-term
(measured as 110 L, 15% of that of the HPLC) instrument reliability. If full loop-fill mode is utilized,
allowed in part the abbreviation of the original perhaps for very high precision requirements
assay. Column re-equilibration accomplished ensure adequate loop overfilling. A significant
during next sample loading in the UPLC, further laminar flow velocity differential in the loading
increasing throughput. Reduce injection volumes sample between its wall interface and center is
appropriately for the column diameter to achieve created in the very narrow bore tubing of the UPLC
good peak shapes. Peak splitting occur when too injector. Overfilling the sample loop by at least
large of a strong sample solvent bolus four loop volumes was found necessary to fully
overwhelms the packing at the column head. displace wash solvent from the 5 L injector loop.
While this assay method tolerated 5 L injections, For this instrument, the manufacturer has
volumes of 1–3 L are more typical starting points. determined and set as the default the optimum
Smaller injection volumes may be compensated overfill volume with typical sample solvents for
by enhanced peak height from use of the high each sample loop size. Operators can specify
resolution columns and by the low carryover from other overfill volumes for unusual sample
the UPLC injector (measured as 10% of the compositions. Choose the proper composition
HPLC carryover for this analyte) to achieve an and volume of weak sample wash to obtain good
equivalent or even lower LOQ). An alternative to peak shape. A portion of the weak sample wash
smaller injection volumes might be to lower solvent will be co-injected with partial-loop filled
sample solvent strength to accomplish sample samples. The weak solvent wash should
focusing on the head of the column. Utilize partial therefore mimic the initial conditions mobile phase
loop-fill injections in preference to full loop-fill. in solvent strength. Utilizing the weak wash

91
Int. J. Pharm. Med. & Bio. Sc. 2014 K Naresh et al., 2014

Figure 2: Chromatograms (from top to bottom): Original HPLC, Initial Scaling to UPLC Showing
Peak Shape Improvement and Possibility for Further Method Optimization, and Final UPLC
Method. Order of Peak Elution: Internal Standard (IS) Then Cpd A

solvent as sample diluents in the sample loop 9. Reduces process cycle times, so that more
may enhance sample focusing onto the column. product can be produced with existing
The volume of the weak wash must be sufficient 10. Resources
to purge the former strong wash solvent from the
11. Increases sample throughput and enables
loop.
manufacturers to produce more material that
ADVANTAGES consistently meet or exceeds the product
1. Decreases run time and increases sensitivity specifications, potentially eliminating
variability, failed batches, or the need to re-
2. Provides the selectivity, sensitivity, and
work material.
dynamic range of LC analysis
12. Delivers real-time analysis in step with
3. Maintaining resolution performance.
manufacturing processes
4. Expands scope of Multiresidue Methods
13. Assures end-product quality, including final
5. UPLC’s fast resolving power quickly release testing
quantifies related and unrelated compounds
DISADVANTAGES
6. Faster analysis through the use of a novel
1. Due to increased pressure requires more
separation material of very fine particle size
maintenance and reduces the life of the
7. Operation cost is reduced
columns of this type. So far performance
8. Less solvent consumption similar or even higher has been demonstrated

92
Int. J. Pharm. Med. & Bio. Sc. 2014 K Naresh et al., 2014

by using stationary phases of size around 2 ii. UPLC can also be used to significantly improve
m without the adverse effects of high the success of the drug discovery process.
pressure. Drug discovery is heavily dependant upon the
early prediction of metabolic fate and
2. In addition, the phases of less than 2 m are
interactions of drug candidate molecules.
generally non-regenerable and thus have
limited use. iii. Sensitivity, selectivity, and analysis time
(sample throughput) are also some of the
APPLICATIONS challenges analysts face when analyzing
i. With UPLC increased resolution in shorter run environmental samples such as soil and water.
times can generate more information faster Explosives residues in soil or environmental
without sacrifices. Higher sample throughput waters are of both forensic and environmental
with more information per sample may interest. These types of assays prove
decrease the time to market, an important challenging because of the selectivity needed
driving force in today’s pharmaceutical to resolve positional isomers.
industry. The corresponding HPLC separation iv. In addition, for complex samples like natural
takes in excess of 12 min; UPLC accomplishes product extracts, added resolution can provide
the same separation in under 30 s.
more information in the form of additional

Comparision Between UPLC and HPLC


Characteristics HPLC Assay UPLC Assay

Column 150×3.2mm 150×2.1mm

Particle size 3 to 5µm Less than 2 µm

Flow rate 3.0 ml/min 0.6 ml/min

Needle wash methanol methanol

Injection volume 5µL 2 µL

Column temperature 30 0C 65 0C

Maximum back pressure 35-40 MPa 103.5 MPa

Total run time 10 min 1.5 min

Plate count 2000 7500

USP resolution 3.2 3.4

Delay volume 750 µL 110 µL

Note: Total solvent consumption (including 0.5 min of delay time in between injections). Acetonitrile: 10.5 ml, Water: 21.0 ml
Acetonitrile : 0.53 ml,Water: 0.66 ml.

93
Int. J. Pharm. Med. & Bio. Sc. 2014 K Naresh et al., 2014

peaks. HPLC versus UPLC separation of UPLC could be the higher back pressure than
comparison of a ginger root extract sample in conventional HPLC. This back pressure can
where both speed and resolution are improved. be reduced by increasing the column
temperature. But it seems that UPLC can offer
v. UPLC coupled with MS technology provided
parent and fragment mass information of lipids significant improvements in speed, sensitivity and

in one chromatographic run, thus, providing resolution compared with conventional HPLC.
an attractive alternative to current LC methods
REFERENCES
for targeted lipid analysis as well as lipidomic
1. Michael E Swartz (2005), “UPLCTM:
studies.
An Introduction and Review Waters
vi. Applications areas of UPLC specified in Corporation, Milford, Massachusetts, USA”,
Waters literature include high throughput Journal of Liquid Chromatography &
library screening, metabolite identification and Related Technologies, Vol. 28, pp. 1253–
bioanalysis, peptide mapping, stability 1263.
indicating analyses, and quantitative analysis. 2. Swartz M E (2005), “Ultra Performance
Liquid Chromatography (UPLC):
POTENTIAL AREAS OF USE An Introduction”, Separation Science Re-
i. Analysis of complex mixtures (e.g., impurity Defined, LCGC Supplement, May, pp. 8-13.
profiles, formulation inerts). 3. Van Deemter J J, Zuiderweg F J and
Klinkenberg A (1956), Chem. Eng. Sci., Vol.
ii. At-line analysis in manufacturing (analysis at
5, p. 271.
the vessel) .
4. Lars Y and Honore H S (2003), J.
iii. Analysis of large amounts of samples  For
Chromatogr., A, Vol. 1020, pp. 59–67.
LC/MS to get better spectra (improved signal
5. Swartz M E (2005), “Ultra Performance
to noise).
Liquid Chromatography (UPLC):
CONCLUSION An Introduction, and review”, Journal of
Liquid Chromatography & Related
UPLC gives increased resolution, speed and
Technologies, Vol. 28, pp. 1253–1263.
sensitivity for liquid chromatography therefore
6. Wu N, Lippert J A, Lee M L (2001), J.
due to UPLC new chemistry and instrumentation
Chromatogr. A , Vol. 911, p. 1.
technology can provide more information per unit
7. Lippert J A, Xin B, Wu N and Lee M L (1997),
of work. UPLC has main advantage over others
J. Microcol. Sepn., Vol. 1, p. 631.
is reduction of analysis time which also helps to
reduce solvent consumption. A negative aspect 8. Tolley L, Jorgenson J W and Mosely M A
(2001), Anal. Chem., Vol. 73, pp. 2985.

94

You might also like