Professional Documents
Culture Documents
Triad TNT X-Ray Field Service/Calibration/Qa Kit: Model 10500amt
Triad TNT X-Ray Field Service/Calibration/Qa Kit: Model 10500amt
Service/Calibration/QA Kit
Model 10500AMT Save time in setup and procedure
• Automatic temperature and pressure
correction for faster operation in any
environment
• Time saving scroll functionality for
faster initial settings and recovery
• Recognizes and ignores spurious
background signals
Expanded range of functionality from
a single instrument
• Very Low Dose Rate (down to
20 nGy/sec at a resolution of 1 nGy/sec)
• Broader range of dental unit kV and
time measurement to include units with
Model 10500AT filament preheat
TRIAD TnT
• Pulse count mode for use with single
phase units with pulse count timers
• Easily measure small image intensifiers
(6 inch and smaller)
• AC adapter for use in lieu of batteries
Introduction Applications
The TRIAD™ TnT Kit is a full-function, x-ray dosimeter kit that performs fast, highly The TRIAD TnT Kit is ideal for
sensitive measurements in all modalities for x-ray field service, calibration, and quality government compliance testing,
assurance testing. The Model 10100AT TRIAD TnT Dosimeter Kit features the Model troubleshooting, repair of diagnostic x-ray
35050AT Dosimeter, a technologically advanced, microprocessor-controlled, x-ray equipment, installation and setup of new
radiation dosimeter. The kit also includes ion chambers and test stand, triaxial/coaxial equipment, preventative maintenance,
cable, AC adapter, HVL filter set, RS-232 interface cable with adapters, customization radiographic quality assurance
software, instruction manual CD and lightweight carrying case. measurements, and measurements
required for JCAHO accreditation. The
Features TRIAD TnT Kit performs measurements
• Bright display direct readout in user-selected units, including R, Sv, Gy, /sec, /min, for all modalities: radiographic,
/hr, /frame, or electrical units (C or A) fluoroscopic, mammographic (MQSA),
• Image intensifier measurements at 0.1 µR and 0.1 µR/sec resolution; cine in µR/frame CT, cine and dental.
• Expanded kVp and exposure time measurement capabilities
• Patented Linear Mammo Filter Pack kVp corrected based on Mammo K-Edge The TRIAD TnT incorporates the latest
Filter Pack design innovations to enhance ease of use,
• Simplified controls include auto-reset, auto-ranging, automatic offset & drift range of applications and save valuable
compensation, automatic power-down and automatic pressure and temperature time. TRIAD TnT continues the TRIAD
correction tradition as the most accurate, widely used
• Includes kVp and exposure time measurements when Model 35050AT is used with instrument of its kind available.
Model 35080B
• Optional TRIAD Toolkit for Excel for remote operation, waveform capture and
calibration, includes MQSA, Rad and Fluoro templates
• Self-checking features include Model 35050AT battery voltage, bias voltage, ion
chamber/cable leakage, and hardware/software checks
• Six AA alkaline batteries provide instrument power, as well as 300 V electronic ion
chamber bias. Auto power-down feature extends battery life
440.248.9300 www.flukebiomedical.com/rms
Specifications for TRIAD TnT Dosimeter Kit Measurement modes
kVp/Dose/Time Single shot, direct beam measurement of
(Model 10100AT)
exposure, kVp, and time. Auto-ranging across three decade ranges.
Exposure and exposure rate accuracy Basic accuracy of Model Auto reset between exposures. Display update after each exposure
35050AT is ± 1% of reading ± 2 range resolution steps over range of kVp/Rate Simultaneous measurement of kVp and exposure rate
18° to 28°C and ± 2% of reading ± 2 range resolution steps over the Full sensitivity dose Auto-ranging across five decades of
full operating temperature range of 0° to 50°C. 3% NIST-traceable sensitive ranges. Automatic drift & offset compensation, and
calibration provided with each system and includes effects of Models automatic post-exposure display hold
35050AT, 96035B and 96020C
Full sensitivity rate Measurement range covers a span from low-
Exposure time measurement Exposure time accuracy is ± 0.1% of level image intensifier measurements to unattenuated direct beams.
reading ± 0.2 msec. Maximum exposure time is 6.5 seconds and Automatic offset compensation and nonlinear filtering. Auto-
measurement resolution is 0.2 msec ranging provides five decades of sensitivity ranges. Display
Nominal measurement range updates once per second
Ion Effective Resolution Very Low Dose Rate (VLDR) This mode is only for making very
Units
chamber range*** step size low dose rate measurements. Nonlinear digital filtering and
15 cc R 100 µ to 20 1µ autoranging provide five decades of sensitivity ranges. Display
R/sec 100 µ to 20 1µ updates once per second. In this mode, automatic current offset
R/min 5 m to 1200 50 µ and drift compensation is disabled. As a result, the system can
R/hr 100 m to 72 k 1m
display very low dose rates
R/frame** 2 µ to 333 m 0.02 µ Power requirements Life ~30 hours with six AA alkaline batteries.
Automatic power-down after user-selected period of unattended
Gy 1 µ to 175 m 0.01 µ
operation (5-255 minutes). An AC adapter is supplied with each
Gy/sec 1 µ to 175 m 0.01 µ
35050AT. When the AC adapter is in use, the auto power down
Gy/min 50 µ to 10.5 0.5 µ feature is disabled, providing continuous operation. User selections
Gy/hr 1 m to 630 0.01 m for ion chamber, units, kV filter pack, temperature, pressure, and
Gy/frame** 0.02 µ to 2.9 m 0.2 n frame are stored in nonvolatile memory before automatic turnoff;
150 cc R 10 µ to 2 0.1 µ eliminates manual re-selection at power-up
R/sec 10 µ to 2 0.1 µ Bias voltage supply Fixed electronic bias (~300 V). Bias voltage is
R/min 0.5 m to 120 5µ removed from triaxial input connector at instrument turnoff
R/hr 10 m to 7.2 k 0.1 m Customization Allows user to modify contents of nonvolatile
R/frame** 0.2 µ to 33 m 0.002 µ memory including ion chamber and kV filter pack conversion factors,
Gy 0.1 µ to 17.5 m 0.001 µ temperature and pressure units, radiation units, and power down
Gy/sec 0.1 µ to 17.5 m 0.001 µ interval. A field customization software program is included for use
Gy/min 5 µ to 1050 m 0.05 µ with an IBM®-PC or compatible
Gy/hr 0.1 m to 63.0 0.001 m Display Two-line, 20-character alphanumeric PLED (polymer LED),
with 0.5 cm character height. Indicates all ion chamber/kV filter pack
Gy/frame** 0.002 µ to 290 µ 0.02 n
identification information, numerical measurement results, battery
150 cc VLDR R/sec 2 µ to 2* 0.1 µ
level, calibration date and other information
R/min 0.1 m to 120* 5µ
Weight 14 lb (6.4 kg)
R/hr 2 m to 7.2 k* 0.1 m
R/frame** 0.04 µ to 33 m* 0.002 µ
Gy/sec 0.02 µ to 17.5 m* 0.001 µ
Gy/min 1 µ to 1050 m* 0.05 µ
Gy/hr 0.02 m to 63.0* 0.001 m
Gy/frame** 0.4 n to 290 µ* 0.02 n
Electrical units C 1 p to 100 n 0.01 p
A 1 p to 100 n 0.01 p
Values for ion chambers are calculated using nominal sensitivities:
15 cc: 2.4 x 10 8 R/C
150 cc: 2.4 x 10 7 R/C
*Very Low Dose Rate effective range at 5% resolution steps
**At 60 frames/sec (1 to 120 frames/selectable)
***IEC 61674 effective range at 1% resolution steps
Connections Specifications for mA/mAs Meter (Model 35035)
Model 35080B interface Male, two lug BNC
The Model 35035
Computer interface RS-232, using RJ-45 connector. 9,600 baud Digital mA/mAs
8-bit, 1 stop, no parity, xon/xoff. Enables fully-programmable Meter is a versatile
operation and waveform display from a PC with optional Excel instrument that is
add-in. Powered when connected to computer used by x-ray
Ion chamber input Triax, BNC. Collector and guard positive- service engineers,
biased relative to ion chamber body and dosimeter chassis field service
Power 2.1 mm DC Power Jack, power input for an unregulated engineers and
9 V, 200 mA adapter with a center negative, 2.1 mm plug biomedical
Optional accessories engineers to
USB to RS-232 adapter (Model 38617) accurately measure mAs and fluoroscopic mA accurately for
Accessories supplied diagnostic, radiographic, and fluoroscopic imaging equipment. The
Test stand (Model 37581) Machined stainless steel upright rod Digital mA/mAs Meter operates with one, easily replaceable 9 volt
with base, ion chamber holder, and tray for HVL filters, which alkaline battery, facilitating convenience and portability.
includes one ion chamber stem Milliamp seconds (mAs)
Programming kit (Model 37594) Includes customization software Range Resolution Input impedance*
on CD, IBM-PC format and 6.5 ft (2 m) RS-232 interface cable
with adapters for PC-type and AT-type computers 200 mAs 0.1 mAs 10 ohms
Instruction manual (Model 35050ATCD) Model 10100AT 2000 mAs 1.0 mAs 1 ohm
* Does not include fuse resistance. Also, does not include effect
TRIAD Field Service Kit Operator’s Instruction Manual on CD
of bridge rectifier present when unit is set for AC
Cable (Model 38208) 20 ft (6 m) coax/triax cable
AC adapter 9 V, 200 mA, DC Current (mA)
HVL filter set (Model 37668) Set of 12 aluminum filters for half- Range Resolution Input impedance*
value layer measurements, which includes one 2 mm, two 1 mm, 20 mA 0.01 mA 100 ohms
two 0.5 mm, three 0.1 mm, one 0.2 mm and three 0.05 mm 200 mA 0.1 mA 10 ohms
Kit carrying case (Model 37500D) High density polyethylene 2000 mA 1 mA 1 ohm
(HDPE) plastic absorbs impact to protect contents. Custom-cut, * Does not include fuse resistance. Also, does not include effect of
high quality foam interiors surround and protect standard kit bridge rectifier present when unit is set for AC.
equipment and accessories. Dimensions 13 (w) x 18 (d) x 6 in (h) Signal input limits
(33 x 46 x 15 cm) Function Range Maximum input limit
Diagnostic ionization chambers (Models 96035B and 96020C)
Input shorted; 2.0 A maximum
Energy range mA OFF
(fuse protected)
96035B: 30 to 150 kVp; 20 to 50 kVp for mammographic 20 250 mA for 30 s*
96020C: 30 to 150 kVp 200 1.0 A for 30 s*
Nominal sensitivity 2000 2.0 A maximum (fuse protected)
96035B: 2.00 x 108 R/C (1.75 x 106 Gy/C) at 22°C and 1013 hPa mAs 200 1.0 A for 30 s*
2.21 x 108 R/C (1.94 x 106 Gy/C) at 22°C and 1013 hPa. Flat 2000 2.0 A maximum (fuse protected)
response suitable for conventional diagnostic radiography and *Limit set by power dissipation rating of shunt resistors
mammography
Controls 1) Power mA/mAs switch, 2) Reset switch, 3) Range
96020C: 2.08 x 107 R/C (1.82 x 105 Gy/C) at 22°C and
switch: 200 mA/mAs, 2000 mA/mAs and 20 mA range settings,
1013 hPa. Optimized for low-level image intensifier and cine
4) AC/DC switch
measurements
Accuracy 1% of reading ± two least significant digits for all ranges
Construction Environmental
96035B: Graphite-coated acrylic, parallel-plate, air-vented
Temperature range 5 to 35°C
96020C: Composite graphite-filled thermoplastic; parallel-
Relative humidity 0 to 80%
plate, air-vented
Storage temperature - 20 to 50°C
Volume Dimensions 2 (w) x 2 (d) x 3.50 in (h) (15 x 5.0 x 8.75 cm)
96035B: 15 cm3; 96020C: 150 cm3
Weight 0.78 lb (0.35 kg)
Display Liquid crystal display (LCD), 3.50 digit, 0.5 in (h) (13 mm)
Input Two banana jacks
Power requirements One 9 V alkaline battery with easy replacement
Accessories supplied
Test Lead Kit (Model CA-23)
440.248.9300 www.flukebiomedical.com/rms
Specifications for Non-Invasive kVp Divider Environmental
Temperature range 0 to 35°C
(Model 35080B) Relative humidity 20 to 80%
Storage temperature - 35º to 50°C
Orientation Long axis of the Model 35080B Non-invasive kVp
Divider oriented perpendicular to axis of x-ray tube to eliminate heel
effect
Power requirements One 9 V battery, 60 hours operation
“Battery-Check” function connects battery to output terminals for
voltage measurement
Dimensions 2.375 (w) x 3.5 (d) x 8.25 in (h) (6.0 x 9.0 x 21 cm)
Weight 1.50 lb (0.68 kg)
Accessories supplied
Wide-Range Filter Pack (Model 37617) For 50 to 150 kVp
The Model 35080B Non-Invasive kVp Divider quickly and accurately
Fitted carrying case
measures kV for all modalities. The unit checks both above and below
table tubes and displays the results on either the Model 35050AT Optional accessories
Dosimeter, a storage oscilloscope, or optional Excel Add-in software. CT Filter Pack (Model 33551) Provides precision measurements
The Model 35080B Non-Invasive kVp Divider is highly portable and in the 70 kVp to 140 kVp range with an accuracy of ± 2%. Active
eliminates the need for bulky and heavy high-voltage divider tanks. area is just 3.8 cm x 0.48 cm, to accommodate the narrow beam of
most scanners
A patented* wide range filter pack is included with the Model 35080B
Cadmium K-Edge and Linear Mammo Filter Pack Pair (Model
Non-Invasive kVp Divider and provides accurate readings for the
37351/37355) Cadmium K-Edge Mammo Pack uses the inherent
range of 50 to 150 kVp. Four optional filter packs are available for
stability of the cadmium k-edge as an absolute reference for
use with the Model 35080B Non-Invasive kVp Divider for CT,
precision measurements between 27.5 and 29.5 kVp with ± 0.5 kV
mammographic, and mobile applications.
accuracy. Use the Linear Mammo Pack with the Cadmium K-Edge
Features Mammo Pack for precision calibrations of ± 1.0 kV accuracy from
• Non-invasive technology eliminates the hazards of high-voltage 22 to 40 kVp
cables and the need for bulky divider tanks Mobile Filter Pack (Model 37946) Easily and accurately
• Optional filter packs enable testing in virtually all modalities measures kV on mid-frequency generators (> 1 kHz) with
• Rh/Rh measurement capability when Model 35080B Non-Invasive substantial kV ripple (> 8%). Provides precision measurements in
kVp Divider is used with Cadmium K-Edge and Linear Mammo the range of 50 to 135 kVp with accuracy of ± 2%
Filter Pack Pair Low Range Filter Pack (Model 38237) Provides precision
Range 50 to 150 kVp, using only the wide-range radiographic filter measurements in the 30 to 90 kVp range with an accuracy of
pack (Model Number 37617). Range and versatility are extended with ± 2%. For radiographic generators using tungsten target
the use of special optional filter packs
Accuracy Accuracy of ± 2% of reading in the range of 50 to 150 kVp,
exclusive of linearity, filtration, and gain effects. Linearity corrections
automatically applied when using Model 35080B Non-invasive kVp
Divider with Model 35050AT Dosimeter
Response time 150 µs (10% to 90%)
Calibration Internally generated signal provides a calibration check
Minimum time for valid reading 1 ms, 3-phase; one line cycle,
single-phase
Tube current Wide dynamic range from 4 to 3000 mA (3 phase), 2 to
1500 mA (single phase). Generator settings will vary in waveform and Tested. Meets applicable standards.
distance. Less than ± 1 kV effect for wide-range radiographic filter
pack covering 50 to 150 kVp. Specialty filter packs may have * Patent numbers 4,843,619 and 4,916,727.
different characteristics
Specifications for Diagnostic Ion Chambers Leakage current < 10 fA under normal bias conditions (300 V)
Collection efficiency
(Models 96020C and 96035B)
96020C 95% at 2,000 R/min
96035B 95% at 5,000 R/min
Wall material
96020C Composite graphite-filled thermoplastic
96035B Graphite-coated acrylic (methyl-methacrylate)
Window material
96020C 0.76 mm thick, graphite-coated polycarbonate
96035B Both entrance windows are made of 0.25 mm graphite-
coated polycarbonate
Window density
96020C 91 mg/cm2
96035B 32 mg/cm2
Active window area
96020C 100 cm2, centered within the chamber body
96035B Each side of the chamber has a circular active window
region centered 7.1 mm further from the BNC connector than the
center of the chamber body. Active window regions have an area
of 12.32 cm2
Collector plate
96020C 0.8 mm thick graphite-coated acrylic plate, 10.80 cm in
The Models 96020C and 96035B Diagnostic Ion Chambers are
diameter. A 2.16 cm x 2.85 cm guard region is electrically isolated
vented volume, parallel-plate air ionization chambers with side-
from the collector area
mounted BNC triaxial connectors. The Model 96020C Ion Chamber
has a nominal volume of 150 cm3; the Model 96035B has a nominal 96035B 0.25 mm thick, centrally mounted, graphite-coated,
volume of 15 cm3. Both ion chambers have a fully-guarded, centrally- polycarbonate plate, 3.18 cm, ± 0.01 cm in diameter. A 1.27 x 0.89 cm
located collector plate that provides superior collection efficiency. guard region is electrically isolated from the collector area
Connector Side-mounted, triaxial, two-lug BNC connector
The patented* Model 96035B has a dual energy range that enables
Calibration
both diagnostic and mammographic measurements. Diagnostic
measurements are taken using one side of the chamber as the entrance 96020C standard calibration The standard calibration is
window, and mammographic measurements are accomplished using performed at H60 (NIST defined as 60 kVp, first HVL of 6.0 mm Al,
the other side as the entrance window. homogeneity coefficient of 94)
96035B standard calibration The standard calibration is
Features
performed at one diagnostic and one mammographic beam quality.
• Very low leakage and low noise Calibration factors are normalized to 22°C and 760 mmHg
• Rugged mechanical construction
Diagnostic unattenuated beam Calibration on the diagnostic side
• These ionization chambers are supplied with triaxial BNC
of the chamber is performed at M80 (NIST defined as 80 kVp,
connectors
first HVL of 2.97 mm Al, homogeneity coefficient of 57)
Energy range
Mammographic beam Calibration on the mammographic side is
96020C 30 to 150 kVp performed at Mo/Mo28 (NIST defined as 28 kVp, first HVL of
96035B 30 to 150 kVp for diagnostic measurements. 20 to 50 kVp 0.332 mm Al, homogeneity coefficient of 74.3) or MV30 (PTB
for mammographic measurements defined as 30 kVp, first HVL of 0.337 mm Al)
Nominal volume
96020C 150 cm3; 11.30 cm diameter by 1.50 cm thick active volume
96035B 15 cm3; 3.96 cm diameter by 1.22 cm thick active volume
Nominal sensitivity
96020C H60: 2.08 x 107 R/C at 22°C and 760 mmHg. Optimized
for low-level image intensifier and cine measurements
96035B L100: 2.0 x 108 R/C at 22°C and 760 mmHg. MV30
(PTB Mammo Point): 2.21 x 108 R/C at 22°C and 760 mmHg. Flat
energy response suitable for conventional diagnostic radiography
and mammography * Patent number 5,508,526.
440.248.9300 www.flukebiomedical.com/rms
Available model(s) Optional accessories
Model 10100AT For all kits:
• Dosimeter (Model 35050AT) • CT Ion Chamber, 3.2 cm3 (Model 500-100)
• 15 cm3 Ion Chamber (Model 96035B) • Ultra-High Purity HVL Attenuators (Model 07-434), for mammo,
• 150 cm3 Ion Chamber (Model 96020C) set of 6
• Coax/Triaxial Cable, 20 ft (6 m) (Model 38208) • TRIAD Toolkit for Excel (Model 10500EXL)
• Programming Kit (Model 37594), includes customization software • USB to RS-232 adapter (Model 38617)
on CD, two-meter RS-232 interface cable and adapters
• Test Stand (Model 37581), Ion Chamber Stem, HVL Filter Tray For Models 10500AT & 10500AMT kits:
• HVL Filter Set (Model 37668) Mammographic Filter Pack Pair (Model 37355/37351) includes:
• AC Adapter, 9 V, 200 mA, DC • Cadmium K-Edge Mammo Filter Pack (27.5 to 29.5 kVp) ± 0.5 kV
• Instruction Manual (Model 35050ATCD) accuracy
• Kit Carrying Case (Model 37500D) • Linear Mammo Filter Pack (22 to 40 kVp) ± 1.0 kV accuracy
Note: Mammo Filter Packs are designed for molybdenum anode,
Model 10500AT beryllium window generators.
Components in Model 10100AT Kit, plus: • Mobile Filter Pack (Model 37946) (50 to 135 kVp) ± 2% accuracy
• Non-Invasive kVp Divider (Model 35080B) • CT Filter Pack (Model 33551) (70 to 140 kVp) ± 2% accuracy
• Wide-Range Filter Pack (Model 37617) (50 to 150 kVp)
Model 10500AMT
Components of Model 10500AT Kit, plus:
• mA/mAs Meter (Model 35035)
• Universal Test Lead Kit (Model CA-23)
Model 10500AMT
For more information, receive our full product catalog, or order online, contact
Radiation Management Services business of Fluke Biomedical: 440.248.9300 or
www.flukebiomedical.com/rms.
Specifications are subject to change without notice.
©2005 Fluke Biomedical. All rights reserved. TRIAD is a trademark of Fluke Corporation. IBM is a
trademark of International Business Machines. Printed in USA.
10500AMT-ds rev 5 03 jun 05
TRIAD™ Toolkit for Excel DI Diagnostic Imaging
Model 10500EXL
Introduction Features
The TRIAD™ Toolkit for Excel is a complete software package for the TRIAD that • Allows complete remote control of the
includes an Excel Add-In, called TRIAD Tool and Excel templates that may be used TRIAD
to evaluate the performance of radiographic, mammographic, and fluoroscopic x-ray • Automatically downloads Model
machines. The TRIAD Tool collects measured results from the Model 35050A 35050A Dosimeter configuration
Dosimeter and places the data in the active Excel worksheet. In addition, the TRIAD settings
Tool may be used to acquire and graph kV waveforms from the TRIAD as well as
remotely control the TRIAD Dosimeter. • Compatible with Windows® 95, 98,
ME®, NT® 4.0, 2000, and Microsoft®
Reproducibility Accuracy
Excel 97, 2000
Machine Settings
kVp mA FS SDD msec
80.0 200.0 small 26 in 100
kV Accuracy Timer Accuracy
kVp msec mR kV % Error mAs % Error
76.1 95.4 365.0 kVp Set kVp [<5.00%] Pass/Fail msec Set msec [<5.00%] Pass/Fail
75.4 94.9 354.0 50.00 53.06 6.12% Fail 200.0 193.7 -2.66% Pass
75.7 95.3 359.0 60.00 57.06 -4.90% Pass 200.0 144.7 -3.53% Pass
76.4 95.3 368.0 70.00 66.78 -4.60% Pass 200.0 120.1 0.08% Pass
75.3 95.1 364.0 80.00 76.80 -4.00% Pass 200.0 95.3 -4.70% Pass
90.00 85.80 -4.67% Pass 200.0 62.1 3.50% Pass
100.00 99.33 -0.67% Pass 200.0 45.5 -9.00% Fail
110.00 108.80 -1.09% Pass 200.0 28.8 -10.00% Fail
120.00 118.70 -1.08% Pass 200.0 20.4 2.00% Pass
0.1
0 0.5 1 1.5 2 2.5 3 3.5
mmAl
440.248.9300 www.flukebiomedical.com/rms
Specifications
Controls
The TRIAD Tools menu provides an interface for the user to remotely control the TRIAD, select filter packs, chambers, units,
retrieve kV waveforms, open templates, and change the options. A description of each control follows:
Connect/Disconnect Templates
Connects or disconnects the TRIAD to the communications port. Three templates are also provided with the TRIAD Toolkit for Excel:
Connect instructs the TRIAD Tool to read configuration information a radiographic template, a mammographic template and a fluoroscopic
such as the filter packs, chambers, and units template. Each template includes a help worksheet with detailed
instructions for its use.
Options The TRIAD Radiographic Template is used to perform the following
Several options can be specified including COM port, temperature, radiographic tests:
pressure, and frame rate • Reproducibility
• kVp accuracy
Measure • Timer accuracy
• Linearity
Sets the TRIAD up for a single exposure using the selected filter • Beam quality
pack, chamber, and unit
The TRIAD Mammographic Template may be used to perform
Auto Reset mammography tests required for ACR and MQSA. Measured data
from the Model 35050A Dosimeter may be automatically collected in
Informs the TRIAD Tool to automatically reset the TRIAD for another the following worksheets:
exposure, allowing a series of exposures to be made without used • kVp accuracy
intervention. At the time Auto Reset is clicked, the TRIAD will be set • kVp reproducibility
up for an exposure • Beam quality
Filter Packs • Breast entrance exposure
• AEC reproducibility
Selects the filter pack to use for kVp measurements • Average glandular dose
• Radiation output rate
Chambers
The TRIAD Fluoroscopic template is used to perform the following
Selects the ion chamber to use for exposure measurements fluoroscopic tests:
• kVp Accuracy
Units • Beam quality
• Fluoro Exposure Rate
Selects the unit to use for exposure measurements
These templates are easy to use and can be modified to fit the user’s
Download Waveform needs
The Waveform button will be enabled after an exposure is made and
before the TRIAD is setup for another exposure. When the Waveform System requirements
button is pressed, a dialog box opens and the user is prompted for the • Windows 95, 98, ME, NT 4.0, 2000
start and end times for the waveform chart, allowing the user to • Microsoft Excel 97, 2000
specify any portion of the kV waveform. The default start and end • One serial port (COM1 through COM4)
times are for the complete waveform. The waveform data will then be
placed in the active cell in the active workbook Available model(s)
Help 10500EXL TRIAD Toolkit for Excel
Opens the TRIAD Toolkit for Excel Instruction Manual
For more information, receive our full product catalog, or order online, contact
Radiation Management Services business of Fluke Biomedical: 440.248.9300 or
www.flukebiomedical.com/rms.
Specifications are subject to change without notice.
©2005 Fluke Biomedical. All rights reserved. TRIAD is a trademark of Fluke Corporation. Windows,
Windows ME, Windows NT, and Microsoft are trademark of Microsoft Corporation. Printed in USA.
10500EXL-ds rev 3 02 feb 05
NERO mAx ®
DI Diagnostic Imaging
X-ray Test Device
®
Victoreen Model 8000 • Non-invasive evaluator of radiation
outputs
• 100 kHz sampling speed captures
data from the most difficult machines
• 0.5 kV or 1% accuracy from 22 to
160 kV
• Measures kVp average, kV effective,
kV peak, time, exposure or rate, mA
or mAs, HVL, exposure/frame, and
mAs/frame
• Displays R or Gy
• Excel Add-in includes MQSA, Rad,
and Fluoro templates
• RS-232 computer interface
• Enhanced dental capabilities
Introduction
The Victoreen® NERO® mAx, Non-invasive Evaluator of Radiation Outputs, evaluates the widest range of x-ray machines on the
market today. Evaluation of pulsed fluoro, cine, computed tomography (CT), portable, mammographic, dental, radiographic,
fluoroscopic, low, medium, and high frequency machines is finally made possible with a single instrument. This fifth generation
instrument features 100 kHz sampling speed and direct mA/mAs measurements. The NERO mAx’s innovative Easy Flow Menu
(EFM) system and flexible softkeys provide an intuitive, user friendly operating environment for quick, accurate, and easy
measurements. All measurement modes and functions are displayed on the NERO mAx’s super bright LCD and are controlled by
the five softkeys directly below the display and three hard keys to the right.
Seven user selectable measurement modes and three system control modes are available and clearly displayed on the control
console screen for easy access and selection.
Applications
The NERO mAx consists of the control console, detector, detector cable, two filter cards, mAs leads, Excel Add-in, AC adapter,
HVL plates, instruction manual, and carrying case.
The compact control console houses the rechargeable battery, super bright easy to read backlit display, eight control buttons, and
the sophisticated electronics necessary for accurate, reproducible measurements. Connectors for power input, RS-232, printer,
scope output and the NERO mAx detector are located on the control console’s rear panel.
The NERO mAx detector contains sensors for simultaneously measuring kV, exposure or rate and mA or mAs. Solid state
detectors are used to measure kV. An ion chamber, located in the top of the detector, is used for exposure/rate measurements.
Connectors for external ion chambers and the NERO mAx detector interface are located on the rear panel of the detector. The
front panel has a keyed opening for the NERO mAx filter cards and a connector for mAs leads.
The filter cards contain the various filters needed to accurately measure kilovoltage. Each filter card is coded so that the NERO
mAx “knows” which filter is in use and its position. The NERO mAx also verifies that the filter card is valid for the selected
measurement mode. The two filter cards are keyed so that they may only be inserted properly. The W/Al filter card and the
Mammo filter card are clearly labeled as to the x-ray tube targets for which they are calibrated.
External chambers
External ion chambers for CT, mammographic, image intensifier tube, and special radiographic applications are available.
Chamber calibration factors can be stored in the NERO mAx for direct readout of measurements.
The Excel Add-in acquires measured data and waveforms directly into an Excel spread sheet to maximize flexibility for report
generation.
440.248.9300 www.flukebiomedical.com/rms
Real-time CT
Pulsed fluoro/AMSE
440.248.9300 www.flukebiomedical.com/rms
Specifications Physical (cont)
Size Volume is 3960 in3 (0.065 m3)
Kilovoltage Measured during the first 480 ms of exposure
Console 9.00 x 9.12 x 3.25 in (22.86 x 23.17 x 8.26 cm)
Measured quantity kVp Avg, kV Eff, kV Peak
Detector 6.56 x 3.70 x 2.58 in (16.66 x 9.4 x 6.55 cm)
Accuracy 0.5 kV or ± 1%
Filter cards 2.4 x 6.25 x 0.31 in (6.1 x 15.88 x 0.8 cm)
Reproducibility 0.5 kV or ± 1%
Weight
Range
Shipping 23 lb (10.43 kg)
Target/Filter Range Filtration Console 4.56 lb (2.067 kg)
W/Al 30 - 60 kV Detector (with card) 1.65 lb (0.747 kg)
50 - 100 kV Filter cards 2.9 and 3.2 oz (0.090 and 0.094 kg)
80 - 160 kV 1.2 mm of Al External chamber accessories
Mo/Mo 22 - 35 kV 30 µ of Mo Radiographic Ion Chamber, 30 cm3 (Victoreen Model 6000-528):
Mo/Rh 22 - 40 kV 25 µ of Rh energy response: within 7% from 30 to 150 kVp (15 to 65 keV);
Mo/Al 22 - 49 kV 1 mm of Al cable: 15 ft (4.5 m); chamber dimensions: 4 x 4 x 0.54 in thick
Rh/Rh 25 - 49 kV 25 µ of Rh (10.2 x 10.2 x 1.4 cm)
Rh/Al 25 - 49 kV 1 mm of Al Mammographic Ion Chamber, 3.3 cm3 (Victoreen Model 6000-529):
All calibrations performed with NIST energy response: within 5% from 0.2 to 5.0 mm Al HVL (16 to 90 kVp);
traceable calibration beam. cable: 15 ft (4.5 m); chamber dimensions: 4 cm Ø x 1.5 cm thick.
This chamber meets the needs of the MQSA for an external
Analyze/Display cycle time
transparent chamber. Option: Probe Holder (Model 6000-529-95)
Radio & mammo 3 seconds for 0.1 second exposure, 1 second for BRH2 test stand
for each 32 ms of exposure time
Image Intensifier Ion Chamber, 150 cm3 (Victoreen Model
Fluoro & AMSE 15 seconds for all exposures 6000-530B): energy response: ± 10% from 1.8 to 10 mm Al HVL;
Time cable: 10 ft (3.0 m); chamber dimensions: 6.26 x 8 x 0.63 in
Radio mode Measured during entire exposure at 90, 80, 75% (15.9 x 20.6 x 1.6 cm)
rise/fall of waveform, zero crossing, or pulse count Scatter Ion Chamber, 400 cm3 (Victoreen Model 6000-532B):
Mammo mode Measured during entire exposure at 90% rise/fall energy response: ± 5% from 32 to 662 keV; cable 10 ft (3.0 m);
of waveform chamber dimensions: 6.26 x 8 x 0.63 in (15.9 x 20.6 x 1.6 cm)
Accuracy 1 ms CT Ion Chamber, 3.2 cm3 (Victoreen Models 6000-100 and
Resolution 0.1 ms 500-100): energy response: ± 5% from 1 to 10 mm Al HVL;
Range All diagnostic exposures from 1 ms to 60 seconds cable: 3 ft (0.9 m); sensitive length: 4 in (10.0 cm); chamber
Exposure & rate Measured during entire exposure with automatic inside Ø: 0.25 in (6.4 mm)
energy, temperature, and pressure correction CT High Sensitivity Ion Chamber, 10 cm3, for multislice CT
Measured quantity Roentgens or Grays (Victoreen Models 6000-200 and 500-200): energy response: ± 5%
from 1 to 10 mm Al HVL; cable: 3 ft (0.9 m); sensitive
Accuracy ± 5%
length: 4 in (10.0 cm); chamber inside Ø: 0.45 in (11.44 mm)
Reproducibility Radio and mammo modes: ± 2% or 2 mR
Optional accessories
Resolution 0.1 mR
Ultra-High Purity HVL Attenuators (Model 07-434): for
Range All diagnostic exposure and rate measurements from mammo, set of 6
1 mR to 9999 R
Available model(s)
Fluoro rate 0.1 R/min to 999 R/min
8000 NERO mAx X-Ray Test Device, consists of a control console,
mAs and mA Measured invasively during entire exposure detector, detector cable, two filter cards, mAs leads, Excel Add-in,
Accuracy 2% HVL plates, instruction manual, and carrying case
Reproducibility ± 1% or 0.2 mAs Available AC adapters (specify with order)
Range 0.1 mAs to 9999 mAs, 0 to 1000 mA
Typical geo.
HVL Model Description
region
Accuracy ± 5%
14-328 110 VAC 12 VDC 1000 mA USA, Japan
Range 0.1 to 99.9 mmAl
14-401 230 VAC 12 VDC 1000 mA Europe
Physical
14-414 230 VAC 12 VDC 1000 mA UK
Display Super bright 240 x 60 pixel, super twist LCD with cold
cathode fluorescent backlight 14-414 and
230 VAC 12 VDC 1000 mA Australia
14-416 adapter
Detectors Ion chamber and solid state (kV detectors)
Ion chamber volume 45 cc nominal
Window area/density 38 mg/cm2 Polycarbonate Tested. Meets applicable standards.
HVL set 2.30 mm, 1.0 mm, 0.3 mm Al For more information, receive our full product catalog, or order online, contact
Radiation Management Services business of Fluke Biomedical: 440.248.9300 or
Power requirements 12 VDC 1 A external supply. Rechargeable www.flukebiomedical.com/rms.
internal batteries supply more than 4 hours of continuous service Specifications are subject to change without notice.
with overnight charge ©2005 Fluke Biomedical. All rights reserved. Victoreen and NERO are trademarks of Fluke Corporation.
Printed in USA.
8000-ds rev 4 12 may 05
NERO mAx Toolkit for Excel ®
DI Diagnostic Imaging
®
Victoreen Model 8000mAx
• Automatically collects measurement
results and places them in an Excel
worksheet
• Automatically captures kV
waveforms and charts them in an
Excel worksheet
• Templates are provided for QA tests
on radiographic, fluoroscopic, and
mammographic x-ray machines
• Templates may be modified to
perform user specific tests and
generate user customized reports
• Allows complete remote control of
the NERO mAx
• Complete on-line help speeds
learning
MEDICAL PHYSICIST'S MAMMOGRAPHY QC TEST SUMMARY • Compatible with Windows® 95, 98,
Preliminary Results
Site: ME®, NT® 4.0, 2000, and Microsoft®
Room ID: Excel 97, 2000
Survey Date:
X-Ray Unit Manufacturer:
Model:
This is only a preliminary list of findings. A full and final report will be mailed to you shortly.
Please call me if you have any questions about this summary.
440.248.9300 www.flukebiomedical.com/rms
Specifications
Controls
The NERO mAx menu provides an interface for the user to remotely control the NERO mAx and retrieve radiation and kV
waveforms. A description of each menu option follows:
System requirements
• Windows 95, 98, ME, NT 4.0, 2000
• Microsoft Excel 97, 2000
• One serial port (COM1 through COM4)
Available model(s)
8000mAx NERO mAx Toolkit for Excel
For more information, receive our full product catalog, or order online, contact
Radiation Management Services business of Fluke Biomedical: 440.248.9300 or
www.flukebiomedical.com/rms.
Specifications are subject to change without notice.
©2005 Fluke Biomedical. All rights reserved. Victoreen and NERO are trademarks of Fluke Corporation.
Windows, Windows ME, Windows NT, and Microsoft are trademarks of Microsoft Corporation. Printed
in USA.
8000max-ds rev 5 03 feb 05
X-Ray Test Device DI Diagnostic Imaging
®
Victoreen Model 4000M+
• Measures kVp maximum, kVp
average, kVp effective, dose and time
in one exposure
• Compact, lightweight design
• Displays R or Gy
• External ion chambers for Mammo,
CT, image intensifier and
phototiming measurements
• Automatic exposure reset for hands-
off operation
• Rechargeable Ni-Cd batteries
provide more than six hours of
continuous service
• RS-232 computer interface
• Storage scope output for real-time
waveform display
• Reversible display for fluoro
measurements
Introduction
The Victoreen® X-Ray Test Device, Model 4000M+ does it all. Simply place the instrument in the x-ray beam, make one exposure,
and it serially displays kVp Maximum, kVp Average, kVp Effective, dose, and time. The Model 4000M+ then automatically resets
for the next exposure. A CsI photodiode pair provide the kVp measurements through five user-selectable filter pairs. This ensures
optimum accuracy over the entire diagnostic range with minimum filtration dependence. Exposure measurements are made with a
parallel plate ionization chamber located above the filter wheel. Exposure time is measured with quartz crystal accuracy. Plus, a
variety of external ion chambers may be connected for even greater flexibility.
Specifications
Kilovoltage Exposure Detectors
Accuracy Measured during entire exposure; kV CsI/photodiode pair measures x-ray
1 kV Mo/Mo (22 to 35 kVp) kVp corrected transmission through differential attenuators
(Mammo generators w/30 µ Mo) Accuracy ± 5% Time Computed from kV waveform stored in
Range 10 mR to 10 R memory against quartz crystal time base
Range
Exposure Parallel plate ionization chamber
W/Al Tubes: 27 to 155 kVp
Volume 36 cm3
Mo/Mo Tubes: 21 to 50 kVp Fluoroscopic rate
Window 38 mg/cm2, 18.9 cm2 polycarbonate
Measured over one second intervals during
fluoro exposure Calibration Reference to a NIST traceable
Time voltage divider and a calibrated exposure
Accuracy ± 5%
Measured during entire exposure at 90% monitor during irradiation
rise/fall of waveform Range 0.5 to 200 R/min
Accuracy Within 2% or 2 ms, whichever is
greater
Range 1 ms to 10 seconds See next page for more specifications.
440.248.9300 www.flukebiomedical.com/rms
Specifications (cont) External chamber accessories
Physical Radiographic Ion Chamber, 30 cm3 (Victoreen Model 6000-528):
energy response: within 7% from 30 to 150 kVp (15 to 65 keV);
Display 16 character dot-matrix LCD cable: 15 ft (4.5 m); chamber dimensions: 4 x 4 x 0.54 in thick (10.2 x
Controls Model 4000+ Five rocker switches 10.2 x 1.4 cm)
On/Off Power switch Mammographic Ion Chamber, 3.3 cm3 (Victoreen Model 6000-529):
Radio/Fluoro Select radiographic or fluoro operation energy response: within 5% from 0.2 to 5.0 mm Al HVL (16 to
90 kVp); cable: 15 ft (4.5 m); chamber dimensions: 4 cm Ø x 1.5 cm
High/Low Select for sensitivity thick. This chamber meets the needs of the MQSA for an external
Roll Roll thru data transparent chamber. Option: Probe Holder (Model 6000-529-95) for
Exposure/All Select exposure only for external ion chamber BRH2 test stand
Mo/Mo or W/Al Select anode/filter of x-ray tube Image Intensifier Ion Chamber, 150 cm3 (Victoreen Model 6000-
530B): energy response: ± 10% from 1.8 to 10 mm Al HVL; cable:
Connectors 10 ft (3.0 m); chamber dimensions: 6.26 x 8 x 0.63 in (15.9 x 20.6 x
Power 9 VDC, 500 mA 1.6 cm)
Scope BNC for oscilloscope connection Scatter Ion Chamber, 400 cm3 (Victoreen Model 6000-532B):
RS-232 DB-9 connector configured as DCE. BNC and banana energy response: ± 5% from 32 to 662 keV; cable 10 ft (3.0 m);
plug for external Ion chamber chamber dimensions: 6.26 x 8 x 0.63 in (15.9 x 20.6 x 1.6 cm)
Power requirements 9 VDC 500 mA external supply. Rechargeable CT Ion Chamber, 3.2 cm3 (Victoreen Models 6000-100 and 500-100):
internal Ni-Cd batteries supply more than six hours of continuous energy response: ± 5% from 1 to 10 mm Al HVL; cable: 3 ft (0.9 m);
service with overnight charge sensitive length: 4 in (10.0 cm); chamber inside Ø: 0.25 in (6.4 mm)
Dimensions 8.5 (w) x 9 (d) x 3 in (h) (21.5 x 23 x 7.6 cm) CT High Sensitivity Ion Chamber, 10 cm3, for multislice CT
(Victoreen Models 6000-200 and 500-200): energy response: ± 5%
Weight Approximately 3.5 lb (1.59 kg) from 1 to 10 mm Al HVL; cable: 3 ft (0.9 m); sensitive length:
HVL set Aluminum filters: 2.3, 1.0, and 0.3 mm 4 in (10.0 cm); chamber inside Ø: 0.45 in (11.44 mm)
©2005 Fluke Biomedical. All rights reserved. Victoreen is a trademark of Fluke Corporation. Printed in USA.
4000-ds rev 3 16 may 05
4000 Toolkit for Excel ®
DI Diagnostic Imaging
®
Victoreen Model 4000EXL
• Automatically collects measurement
results and places them in an Excel
worksheet
• Captures radiation and kV
waveforms and charts them in an
Excel worksheet
• Templates are provided for QA tests
on radiographic, fluoroscopic, and
mammographic x-ray machines
• Templates may be modified to
perform user specific tests and
generate user customized reports
• Allows complete remote control of
the Model 4000M+ NERO
• Complete on-line help speeds
learning
• Compatible with Windows® 95, 98,
ME®, NT® 4.0, 2000, and Microsoft®
MEDICAL PHYSICIST'S MAMMOGRAPHY QC TEST SUMMARY
Preliminary Results Excel 95, 97, 2000
Site:
• Automatically detects the presence of
Room ID:
Survey Date: Model 4000M+ NERO
X-Ray Unit Manufacturer:
Model:
This is only a preliminary list of findings. A full and final report will be mailed to you shortly.
Please call me if you have any questions about this summary.
440.248.9300 www.flukebiomedical.com/rms
Specifications
Controls
The 4000 menu provides an interface for the user to remotely control the 4000M+ NERO and retrieve radiation and kV
waveforms. A description of each menu option follows:
System requirements
• Windows 95, 98, ME, NT 4.0, 2000
• Microsoft Excel 95, 97, 2000
• One serial port (COM1 through COM4)
Available model(s)
4000EXL 4000 Toolkit for Excel
For more information, receive our full product catalog, or order online, contact
Radiation Management Services business of Fluke Biomedical: 440.248.9300 or
www.flukebiomedical.com/rms.
Specifications are subject to change without notice.
©2005 Fluke Biomedical. All rights reserved. Victoreen and NERO are trademarks of Fluke Corporation.
Windows, Windows ME, Windows NT, and Microsoft are trademarks of Microsoft Corporation. Printed
in USA.
4000EXL-ds rev 4 03 feb 05
Crescent X-Ray Leakage DI Diagnostic Imaging
Detection System
Model 10970 • Ion Chamber array performs x-ray
leakage tests for diagnostic x-ray
Introduction tube manufacturers and tube
reloaders
The Crescent X-Ray Leakage
Detection System is custom- • Demonstrates product compliance to
designed for accurate low- the Radiation Control Act
level radiation measurements.
When used in a suitable • Standard 17-channel, 3-channel, or
configuration, the system custom designed systems are
demonstrates compliance to available
Title 21 CFR subchapter J,
Part 1020.30 (k) of the • 17 chambers mounted around a
Radiation Control Act. semicircle of 1 meter radius provides
The Model 10970 Crescent 180° coverage with overlap between
X-Ray Leakage Detection chambers
System consists of a
• Fast full spherical scan (2 minutes)
combination of Ion Chamber/
Electrometer Modules (Model • Direct readout in mR/hr
96010A/50300A), and a
Control Console (will include • Reliable modular construction
Model 70010A Dual Channel
Comparator Modules, 70020 • Expandable from one to eighteen
Reference Control Modules channels
and power supply). The air-
• Highest channel readout
filled ion chamber/electrometer module is the basic component of the system. Seventeen
of these modules are sufficient to provide a full spherical scan and are connected to the
control console through low-impedance cable, enabling remote monitoring. The Control
Console contains the high-voltage ionization potential, a precision comparator and trip
circuits, reference module, system control logic, and a spare HV power supply. The entire
system has a modular design for add-on capability,
interchangeability, and ease of maintenance. Leakage detection system configurations
Applications 17-channel 3-channel
system system
The only sure way to demonstrate compliance to leakage radiation
standards is with a full spherical scan of the x-ray emitting products, Module/Description No. req. No. req.
such as diagnostic x-ray tubes. X-ray leakage tests are made easy for 96010A Ion Chamber w/NIST traceable
17 3
x-ray tube manufacturers and tube reloaders. A rapid 100% calibration
50300A Electrometer* 17 3
production test can be less expensive than design analysis, analysis of 10970 Mainframe† 1 1
tolerances and tolerance buildup, and costly quality control 70010A Dual Channel Comparator** 9 2
procedures and inspection. With the X-Ray Leakage Detection 70020 Reference Control Module 1 1
System, you can perform a full scan of your product in only two 55 foot interface cable 9 2
minutes. High-voltage supply 1 1
Ion chamber mounting ring and hardware 1 --
System configurations High-voltage cable (17-channel) 1 --
Rack Cabinet Control Module 1 --
A basic three-channel leakage radiation system can be assembled High-voltage cable (3-channel) -- 1
from an Ion Chamber/Electrometer Module used in combination Calibration current source 1 1
with one Model 10970 Mainframe, two Model 70010A Dual * Requires one per channel
Channel Comparator Modules, one Model 70020 Reference Control ** Requires two per channel
† Specify 120 or 22 VAC
Module, and one High Voltage Power Supply. You can expand this
system up to 18 channels to provide a full spherical scan by adding Note: Typical systems are configured with either 17 channels or 3 channels. All
plug-in modules. Listed below are two standard system systems are custom designed. Spare components are recommended. On-site
installation is recommended. Calibration source and standard NEMA® rack
configurations: a 17-channel and a 3-channel system. Any number
cabinet are required.
of channels from 3 to 18 may be ordered. Details of system modules
are described in the table.
440.248.9300 www.flukebiomedical.com/rms
Mainframe (Model 10970) Dual Channel Comparator (Model 70010A)
• Pre-wired • Bright visual fault indication
mainframe • Altitude compensation
• Expandable system • Direct readout in mR/hr
• Line-operated • Front panel monitoring and test
• Self-checking • Readout of highest channel
Rack Mount Control Module
The Model 10970 The Model 70010A Dual Channel
Mainframe includes a 19 inch rack mount cage, 7 inches high Comparator compares outputs of the Model
which will accept up to nine Model 70010A plug-in Dual Channel 96010A/50300A Ion Chamber/Electrometer
Comparator Modules and one Model 70020 Reference Control and any desired reference radiation level. You
Module. The Mainframe incudes power supplies (± 15 volts and can select an input reference voltage of
6.3 VDC) for up to eighteen Model 50300A Electrometers, nine 500 mV (50 mR/hr) to demonstrate Dual Channel
Model 70010A Dual Channel Comparators, and one Model 70020 compliance to the Radiation Control Act. If Comparator
(Model 70010A)
Reference Control Module. the output of the ion chamber/electrometer
Model 50300A Electrometers are connected to the back panel of exceeds this value, a bright red light will go on and remain on
the Mainframe. Power required is 120 or 220 VAC (specify at until manually reset.
time of order). An output connector provides the contacts of the The Model 70010A Module has an adjustable gain and offset for
fault relay as well as the output of the highest channel in those the output of the Model 50300A Electrometer. Gain is adjusted to
systems using the Model 70010A. compensate for altitude, so readings are correct for installations as
high as 8000 feet above sea level. You can typically disregard
normal changes in barometric pressure as they will not exceed
± 3%. To correct reading for normal barometric pressure changes,
Ion Chamber/Electrometer Module use the correction table (included).
(Models 96010A/50300A)
The Model 70010A has two independent channels so you can
• Fully-guarded air- connect each channel to a Model 50300A Electrometer. Fifty-five
equivalent chamber feet of cable (included) connect the electrometers to the back of
• NIST-traceable the Model 10970 Mainframe. The Model 10970 supplies all
calibration required power.
• Electrometer directly The Model 70010A permits the highest output of all electrometers
connected to ion on the system to be read. You can use this output as the Y-axis on
an X-Y plot to demonstrate that no output exceeds the 100 mR/hr
chamber
legal level.
• Low-noise, high-
speed performance Ion Chamber/Electrometer Module
(Models 96010A/50300A)
Reference Control Module
The Model 96010A Ion (Model 70020)
Chamber is constructed of air-equivalent plastic and is vented. The Model 70020 Reference Control Module
The chamber has a window area of 100 cm2 (5 x 20 cm), and a sends a calibrated rejection level to the Model
volume of 500 cm3. 70010A Dual Channel Comparators. You can set
this level to between 0 and 100 mR/hr with a
When used to demonstrate compliance to the Radiation Control
direct-reading dial setting.
Act, seventeen chambers mounted around a semicircle of 1 meter
radius provide 180° coverage with overlap between chambers. The Model 70020 also provides a convenient
voltage source to check the actual trip levels of
The Model 50300A Electrometer may be combined with a Model each Dual Channel Comparator Module. All
96010A Ion Chamber to eliminate the problems associated with comparator modules are forced into the “fault”
high-impedance cable. The rise time is approximately 250 msec indication when the test switch is pressed. This
(10-90%) and the output noise is less than 10 mV, peak-to-peak quickly checks all channels. Reference
(1 mR/hr peak-to-peak). Based on the typical characteristics of the Control Module
There are two operational modes: in the Interrupt (Model 70020)
Model 96010A Ion Chamber, the electrometer module will be mode a fault relay closes whenever any channel
factory adjusted to provide a scale factor of 1 volt for 100 mR/hr goes above the trip level; in the Continuous mode the relay is not
NIST traceable ion chamber calibration. activated. The relay contacts can activate an alarm or stop the
drive motor when the system is scanning.
For more information, receive our full product catalog, or order online, contact
Radiation Management Services business of Fluke Biomedical: 440.248.9300 or
www.flukebiomedical.com/rms.
Specifications are subject to change without notice.
©2005 Fluke Biomedical. All rights reserved. NEMA is a registered trademark of the National Electrical
Manufacturers Association for its publication of voluntary standards and guidelines. NEMA is not a
certification mark. Printed in USA.
10970-ds rev 3 04 feb 05
CR RADCHEX DI Diagnostic Imaging
Model 07-CRX
• CR RADCHEX is an indispensable
tool for service engineers, physicists
and quality assurance personnel
Introduction Features
The CR RADCHEX System, consisting of a 24 x 30 cm electronic cassette, The electronic cassette once calibrated
electronic interface box, universal calibration tool, and Excel software, can assess produces an output signal directly
the performance and proper calibration of computed radiography (CR) automatic proportional to the light output from an
exposure control (AEC) systems. CR RADCHEX can be used to perform periodic exposed CR imaging plate when the plate
assessment. The system provides information to make recommendations for is stimulated by laser light (tracking ± 3%
corrective action and adjustments of clinical system speed and thus dose to patients. for energy and patient density). An
CR plate readers, AEC and automatic programmed radiography (APR) can all be interface box converts the output signal to
correlated to maximize the function of computed radiography systems in any clinical a digital number displayed on a LCD
setting. (CR-AEC number) that further
corresponds to a relative clinical system
speed number also displayed on the LCD.
Applications Excel software analyzes input (CR-AEC
CR RADCHEX is ideal for use by service engineers to initially calibrate and number) and compares it to the imaging
troubleshoot the CR plate reader, AEC and density selector settings. Physicists use plate signal output number (CR
CR RADCHEX to assess the performance of CR-AEC for compliance to clinical manufacturer: SAL# AGFA, EI# Kodak,
system speed objectives and patient dose. QA personnel can use CR RADCHEX to S# Fuji, Rex# Konica). Three templates
periodically document the performance of the CR system and to compare CR to are included for service engineers,
film/screen systems regarding desired ALARA objectives. medical physicists/radiation protection
inspectors and QC personnel.
CR RADCHEX is designed to work with all major brands of CR equipment.
440.248.9300 www.flukebiomedical.com/rms
Specifications Interface box
CR RADCHEX System The interface box is connected to the electronic cassette and converts
digital signals from the electronic cassette to numbers: an AEC
Accuracy Simulates output of Photostimulatable Phosphor Plate number and relative clinical system speed number, and displays them
(PSP) within ± 3% over a kVp range of 60 to 120 kVp and a patient on an LCD display.
equivalent thickness range of 5 to 25 cm (within specified operating
rates) Dimensions 4.92 x 4.13 x 1.96 in (12.5 x 10.5 x 5 cm)
CR AEC Calibration
Relative Clinical
Customer Desired Clinical System Speed CR AEC#
System Speed
Target Speed % Limit (+/-) Min: 4.77 440
400 10 Target: 5.25 400
Max: 5.78 364
Selector Setting: -1 -2 -3 -4 -5
CR Speed: 460 529 608 700 805
Target CR AEC#: 4.57 3.97 3.45 3.00 2.61
Selector Setting: 1 2 3 4 5
CR Speed: 348 302 263 229 199
Target CR AEC#: 6.04 6.94 7.98 9.18 10.56
440.248.9300 www.flukebiomedical.com/rms
Sample QC tracking graphs
40.00 400.00
P ercen tag e
30.00 300.00
20.00 200.00
10.00
0.00 100.00
0.00
4
4
00
00
/2
/2
4
18
28
00
00
8/
8/
/2
/2
18
28
8/
8/
Date Date
Chamber
40.00 14.00
CR AEC#
12.00
30.00 10.00
Center
20.00 Chamber
8.00
10.00 6.00
4.00 Right
0.00 2.00 Chamber
0.00
4
4
00
00
/2
/2
4
00
00
18
28
/2
/2
8/
8/
18
28
8/
8/
Date Date
15.00 1.40
CR AEC#
1.20
10.00 1.00
0.80
0.60
5.00 0.40
0.20
0.00 0.00
4
4
00
00
00
00
/2
/2
/2
/2
18
28
18
28
8/
8/
8/
8/
Date Date
Introduction Features
The AEC-6 System, consisting of an electronic screen-film cassette, an electronic The electronic cassettes, once calibrated,
interface box, and computer software, can assess the performance and calibration mimic the screen-film systems used in
of mammo/radiographic AEC systems and can provide information to estimate the a radiology department. Each electronic
relative speed and gradient of screen-film and processing systems. Once the AEC-6 cassette contains circuitry that analyzes
meter is calibrated to the screen-film characteristics and processing conditions, the the light output from the intensifying
meter provides the user with a readout in optical density for each exposure made. screen(s) and converts the light into a
digital electrical signal which is fed to an
onboard microprocessor for analysis. The
microprocessor sends information via the
Applications interface box to the computer for software
Service engineers use the AEC-6 meter to calibrate and make adjustments to analysis using mathematical algorithms
mammo/radiographic AEC systems instead of using film. Medical Physicists use that converts the microprocessor
the AEC-6 for performance assessment of mammo/radiographic AEC systems and information to optical density readouts.
based on the assessment, make recommendations for corrective action. Radiation
inspection officers use the AEC-6 to assess mammo/radiographic calibration and
suggest corrective action to x-ray equipment operators. Technologists use the
AEC-6 meter for teaching and routine quality control evaluation and balancing
optical density from room to room.
440.248.9300 www.flukebiomedical.com/rms
Specifications
AEC-6 System Electronic interface module
Accuracy Optical Density (OD) prediction ± 0.1 OD ranging from Interface box allows the electronic cassettes to be connected to the
0.5 to 3.0 computer.
Dynamic range 0.2 to 4.0 OD Dimensions 4.92 x 4.13 x 1.96 in (12.5 x 10.5 x 5 cm)
Setup and reset Manually using computer keys Weight 2.2 lb (1.0 kg)
Controls Manually using computer keys
Functions OD, Time, AEC# Computer Software for AEC-6 Systems
Power requirements 2 externally-accessible 9 V batteries Computer software CD-Rom containing compiled AEC-6 program
(DURACELL MN 1604 or equivalent) (Windows® 95 or higher)
Typical battery life > 40 hr Cassette interface cable 10 foot 25-pin straight thru serial cable F/M
Operating environment 59º to 95ºF (15º to 35ºC) (DB-25 male and DB-25 female)
Computer interface cable 15 foot serial mouse extension cable
(DB-9 male and DB-9 female)
Electronic cassettes
The electronic cassettes (mammography and radiography) simulate
the film-screen cassette used in the radiology department and provide Optional accessories
digital electrical signals to the computer. AEC-6 Interface (Model 07-AEC6I)
The mammographic cassette is specifically designed to mimic the Mammographic Cassette and Adapter (Model 07-AEC6MC)
attenuation of the film-screen cassette. An accessible compartment
Radiographic Cassette (Model 07-AEC6RC)
(drawer) allows various thicknesses of aluminum to be placed in the
cassette to match the attenuation of the film-screen cassette. Computer Software for AEC-6 Systems (Model 07-AEC6SW)
Mammographic cassette
Dimensions 7.5 x 14.37 x 0.5 in (19 x 36.5 x 1.3 cm) Available model(s)
Weight 4 lb (1.8 kg) 07-AEC6 AEC-6 System, consists of AEC-6 interface,
Radiographic cassette mammographic cassette and adapter, radiographic cassette, and
Dimensions 10.5 x 10.5 x 0.5 in (26.7 x 26.7 x 1.3 cm) computer software
Weight 1.7 lb (0.80 kg) 07-AEC6M AEC-6 Mammo System, consists of AEC-6 interface,
mammographic cassette and adapter, and computer software
07-AEC6R AEC-6 Rad System, consists of AEC-6 interface,
Mammographic attenuation filters radiographic cassette, and computer software
Six type 1100 aluminum filters
Five Filters 0.2 mm thickness by 10 x 8.1 cm
For more information or to receive our full product catalog, contact Fluke Biomedical at
One Filter 0.1 mm thickness by 10 x 8.1 cm 440.248.9300 or www.flukebiomedical.com/rms.
Specifications are subject to change without notice.
©2005, 2006 Fluke Biomedical. All rights reserved. Windows is a trademark of Microsoft Corporation.
Printed in USA.
07-AEC6-ds rev 3 24 feb 06
MAS-5 Meter DI Diagnostic Imaging
Model 07-MAS5
Introduction
For many years, service engineers have • An intelligent meter that measures
been asking for a meter that measures mAs, mAs, exposure time, and mA all at
exposure time, and mA at the same time. the same time
Diagnostic Imaging Specialists has
• Provides mA waveform information
produced this meter with extra features. Not
only will the MAS-5 provide mAs, • At a button press, the MAS-5 meter
exposure time, and mA, but the meter will ignore the first 10 milliseconds
provides mA waveform information. The of exposure
MAS-5 displays three 50 milliseconds mA
waveform samples so that radiographic and • AC and DC inputs
mammographic pre-heat circuits can be
analyzed and adjusted without using an
oscilloscope. Based on feedback from field
service engineers, the MAS-5 also has a
button that when pushed causes the meter
to ignore the first 10 milliseconds of the
mA waveform.
Applications Features
The MAS-5 intelligent meter uses a microcontroller to analyze the digital mA • A button that causes the meter to ignore
waveform and accurately displays the values essential for analyzing and calibrating the first 10 milliseconds of exposure
radiographic and mammographic equipment. • A diagnostic power-up sequence to
indicate operational status
The four line LCD displays the following:
• Auto LCD update
Line 1 mAs (average tube current (mA) times mA waveform exposure time)
• Optional manual reset
Line 2 Exposure Time (mA waveform exposure time in seconds)
• Automatic power-down when meter is
Line 3 mA (average tube current (mA) over the entire mA waveform) not used for more than five minutes
Line 4 Three sample mA waveform values:
• Displays when an exposure is detected
1st waveform value represents the average mA for the 1st 50 milliseconds of
exposure • Low battery indication
2nd waveform value represents the average mA for the 2nd 50 milliseconds
of exposure
3rd waveform value represents the average mA for the 3rd 50 milliseconds of
exposure
Specifications
Reset Auto LCD update; optional manual Operating temperature 59º to 95ºF Tested. Meets applicable standards.
reset (15º to 35ºC)
Power requirements One 9 V battery For more information, receive our full product catalog, or
Dynamic range 10 to 2000 mA; 0.1 to order online, contact Radiation Management Services
999.9 mAs; 1 ms to 6.535 sec Typical battery life > 40 hr business of Fluke Biomedical: 440.248.9300 or
www.flukebiomedical.com/rms.
Accuracy Size 4 x 6.5 x 0.33 in (10.16 x 16.51 x 0.84 cm) Specifications are subject to change without notice.
mAs ± 0.1 mAs or 1% (whichever is Weight 0.625 lb (0.28 kg) ©2005 Fluke Biomedical. All rights reserved. Printed in USA.
greater) 07-MAS5-ds rev 2 04 feb 05
Available model(s)
mA ± mA or 0.5% (whichever is greater)
07-MAS5 MAS-5 Meter
Time ± 1 ms or 1% (whichever is greater)
440.248.9300 www.flukebiomedical.com/rms
Diagnostic Imaging DI
Radchex/AEC
Model 07-RCX
Introduction
• An intelligent meter that monitors
daily quality control of the entire Radiographic machines
radiographic machine incorporating calibrated
automatic exposure control
• Convenient, fast, and accurate (AEC) systems are designed
assessment of the AEC Systems’ to maintain constant film
performance density regardless of the x-
ray energy (kVp) used or
• Separates x-ray machine from film
patient thickness/density.
processing quality control
The Radchex/AEC meter
• For AEC mammography provides a convenient, fast,
applications, see Mamchex/AEC and accurate assessment of
(Model 07-MCX) the performance of the AEC
system and provides daily
quality control of the entire
radiographic machine.
Applications
The Radchex meter consists
Features of a calibrated electronic radiographic cassette that responds to radiation the same
• Diagnostic power-up sequence to way a film/screen cassette responds to light from intensifying screens. No film is
indicate operational status needed to determine the performance of the AEC system or daily quality control of
the machine. If the AEC system is calibrated properly and the machine doesn’t drift,
• Manual reset using the Power/Reset then the digital number displayed by the Radchex meter should always be the same
button from day-to-day or for various x-ray energies or patient thickness.
• Automatic power-down when meter is To convert the Radchex number to an optical density number, software is provided
not used for more than one minute for the conversion; however, in most cases this conversion is unnecessary since the
• Displays when an exposure is being Radchex number should remain constant when kVp and thickness changes.
made When using the meter to assess AEC performance or daily machine quality control,
• Displays exposure time in no film/screen has to be used, therefore eliminating processing variations in the
milliseconds assessment.
• Low battery indication The Radchex is easy to operate:
1. Slip Radchex into Bucky, same as the film/screen cassette, until it locks into place.
2. Turn on.
3. Push reset button, wait until the LCD says “Ready for Exposure.”
4. Expose it.
5. Read the number and compare to a calibration number. The number tells whether
the operator has a “go” or “no go” situation.
Specifications
Reset Manual using Power/Reset button Short/Long term drift < 0.5% For more information, receive our full product catalog,
or order online, contact Radiation Management
Operating temperature 59º to 95ºF Typical battery life > 40 hr Services business of Fluke Biomedical: 440.248.9300
(15º to 35ºC) or www.flukebiomedical.com/rms.
Size 10.5 x 10.5 x 0.5 in (26.7 x 26.7 x 1.3 cm) Specifications are subject to change without notice.
Dynamic range 0 to 999.9 Weight 3.9 lb (1.8 kg) ©2005 Fluke Biomedical. All rights reserved. Printed in USA.
Power requirements Two 9 V batteries Available model(s)
07-RCX-ds rev 2 04 feb 05
Specifications
Range 3-decade auto-ranging and auto- Error prevention logic Low battery - auto AC recharge input 117 V ± 15% 50/60 Hz
indicating with proper decimal placement turnoff; OverRange indicator; Excessive power cord furnished 230 V ± 15% available
and measurement units. See selected ion intensity indicator; Probe disconnected on special order
chambers’ specifications for specific ranges indicator Dimensions 5.50 (w) x 8.50 (d) x 4.50 in (h)
Calibration Unnecessary. Response Display 3-digit LED legends; R/min, R/hr, (14 x 21.6 x 11.4 cm)
depends on frequency of internal crystal mR/hr, R, mR, mSv/min, mSV/h, mSv, µSv Weight 7 lb (3.2 kg) net
oscillator (± 0.01% from 10° to 40°C) Battery life 6 hours between charges at Accessories supplied Foam-lined carrying
Accuracy Depends only on crystal oscillator 20°C with normal usage. Automatic shutoff “brief” case, 2 and l0 ft ion chamber cables,
frequency in Exposure Rate mode. See when charge drops below usable level AC recharging cable, instruction and
selected probe’s specifications for Battery complement One 67.5 V collecting maintenance manual, probe (to fit in case)
measurement accuracy potential (shelf-life) and 4 rechargeable available separately
Controls Function switch: selects off, total 1.25 V “D” batteries for instrument power Available model(s)
exposure or exposure rate modes Battery charge time 10 hours to full 660 X-Ray & CT Exposure Measurement
Reset switch: selects run, stop, or reset (in recharge at 20°C (instrument OFF) Instrument, 110 V
Total Exposure Mode)
Intensity knob: controls readout brightness
See next page for available ion chambers.
440.248.9300 www.flukebiomedical.com/rms
Ion Chambers
660-3 660-4A 660-5
660-6
660-3DE 660-4DE 660-5DE
Application Diagnostic Beam Measurement Scatter CT
Probe volume 4 cm3 400 cm3 3.2 cm3
Measurement area 10 cm2 100 cm2 NA
Maximum rate 999 R/min 99.9 R/min 9.99 R/hr 999 R•cm/min
Maximum exposure 99.9 R 9.99 R 99.9 mR 99.9 R•cm
Resolution on Rate 10 mR/min 1 mR/min 0.1 mR/hr 0.01 R•cm/min
most sensitive Exposure 1 mR 0.1 mR 1 µR 0.001 R•cm
Intensity limit for 99%
40 R/sec 1.8 R/sec 10.8 R/hr 985 R/min
collection efficiency
Model 660-3 and 660-4A typical energy dependence Model 660-5 typical energy dependence
1.4
1.4
1.2 1.2
FRONT
Indicated / Actual
Indicated / Actual
1.0 1.0
SIDE
0.8 0.8
0.6 0.6
0.4 0.4
10 100 1000 10 100 1000
1.5 1.5
Indicated / Actual
Indicated / Actual
1.0 1.0
0.5 0.5
0.0 0.0
10 100 1000 10 100 1000
For more information, receive our full product catalog, or order online, contact
Radiation Management Services business of Fluke Biomedical: 440.248.9300 or
www.flukebiomedical.com/rms.
Specifications are subject to change without notice.
DI Diagnostic Imaging
Victoreen® Model 06-526
• Entrance skin exposure measurements
(ESE)
• Fluoroscopy exposure measurements
• Exposure checks; radiographic (mR/mAs)
• Beam quality; half value layer (HVL)
• mAs reciprocity; mA station
checks...plus many others, depending on
the remote external chambers used
• The American College of Radiology
(ACR) recommends this type of
product in their quality assurance
program
• Fast and easy-to-use. Battery
operation and built-in detector
“The original” x-ray exposure meter eliminate setup time
• Measures dose up to 2 R; dose rate up
Introduction to 20 R/min
Proven RAD-CHECK technology specifically designed to provide you with the
ultimate in versatility and cost-effective operation. • Energy response is ± 5% from 30 to
150 kVp for the RAD-CHECK PLUS
Accurate, lightweight, portable...this “industry standard” enables you to gain the internal chamber
critical edge in your QC program.
• Optional remote chambers for
Applications mammographic and cine imaging
Battery operation and built-in detector virtually eliminate setup time. Just place the systems
meter or external ion chamber on x-ray table; collimate, shoot, and read the result. • Extremely compact...6 x 6.25 x 2.75 inch;
Precision ion chamber and digital display ensure accuracy plus easy readability. weighs only 1.125 lb
Specifications Controls Auto or manual reset selector. CT Ion Chamber, 3.2 cm3 (Victoreen
Display zero reset button. Dose or dose rate Model 6000-100): energy response: ± 5%
Ranges 0.001 to 2 R, 0.01 to 20 R/min
output selector. Integral or remote ion from 1 to 10 mm Al HVL; cable: 3 ft (0.9 m);
Internal chamber 30 cc volume, energy chamber selector. On/off switch sensitive length: 4 in (10.0 cm); chamber
response ± 5% from 15 to 65 keV (30 to 150 kVp inside Ø: 0.25 in (6.4 mm)
Power 9 V alkaline battery, > 100 hours
filtered). 20.5 cm2 (5.1 cm Ø) effective
operation (50 hours in manual reset mode) Image Intensifier Ion Chamber, 150 cm 3
measurement area. Center of chamber 1.03 cm (Victoreen Model 6000-530B): energy response:
below top of chamber Dimensions 6 (w) x 6.25 (d) x 2.75 in (h)
(15.25 x 15.9 x 7 cm) ± 10% from 1.8 to 10 mm Al HVL; cable: 10 ft
Standard calibration At 75 kVp with 4 mm (3.0 m); chamber dimensions: 6.26 x 8 x 0.63 in
Al filtration at 22°C and one atmosphere Weight 1.125 lb (0.51 kg)
(15.9 x 20.6 x 1.6 cm)
Reproducibility Within 2% short-term over Optional accessories
Carrying Case (Model 89-525): holds
100 mR to 2 R range (1 to 20 mGy) Radiographic Ion Chamber, 30 cm3 RAD-CHECK PLUS and accessories
Electrometer drift 0.5 to 1 mR/min typical; (Victoreen Model 6000-528): energy Available model(s)
6 mR/min maximum (5 to 10 µGy; 60 µGy/min response: within 7% from 30 to 150 kVp
06-526 RAD-CHECK PLUS
maximum) (15 to 65 keV); cable: 15 ft (4.5 m);
chamber dimensions: 4 x 4 x 0.54 in thick 06-526-2200 RAD-CHECK PLUS, SI Units
Maximum exposure rate Minimum 90%
(10.2 x 10.2 x 1.4 cm)
collection efficiency at 20 R/sec
Automatic reset Resets display to zero; can
Mammographic Ion Chamber, 3.3 cm3 Tested. Meets applicable standards.
(Victoreen Model 6000-529): energy
also be reset manually
response: within 5% from 0.2 to 5.0 mm For more information, receive our full product catalog, or
Operating temperature 10 to 40°C Al HVL (16 to 90 kVp); cable: 15 ft (4.5 m); order online, contact Radiation Management Services
Relative humidity 0 to 90%, non-condensing business of Fluke Biomedical: 440.248.9300 or
chamber dimensions: 4 cm Ø x 1.5 cm www.flukebiomedical.com/rms.
Display 3.50 x 0.50 inch LCD, low battery thick. This chamber meets the needs of the Specifications are subject to change without notice.
indicator MQSA for an external transparent
©2005 Fluke Biomedical. All rights reserved. Victoreen and
chamber. Option: Probe Holder (Model RAD-CHECK are trademarks of Fluke Corporation. Printed in USA.
6000-529-95) for BRH2 test stand 06-526-ds rev 3 03 jun 05
440.248.9300 www.flukebiomedical.com/rms
Diagnostic Imaging DI
RAD-CHECK MICRO-R ®
Specifications
Ranges Operating temperature 50º to 104ºF Available model(s)
(10º to 40°C) 06-526-5240 RAD-CHECK MICRO-R
Low: 0.01 to 19.99 mR; 0.1 to 199.9 R/min
Relative humidity 0 to 90% 06-526-5242 RAD-CHECK MICRO-R,
High: 0.01 to 19.99 R; 0.1 to 1999 R/min
Display 3.50 x 0.50 inch LCD, low battery SI Units
Standard calibration At 75 kVp with 4 mm
indicator (Must have Model 06-524-3000 chamber
Al filtration at 22°C and one atmosphere
using Model 06-524-3000 chamber Control Reset button, dose or dose rate and must be calibrated at the same time)
output selector, high or low range selector,
Reproducibility Within 2% short-term over
on/off switch
100 mR to 2 R range (1 to 20 mGy)
Power requirements 9 V alkaline battery,
Maximum exposure rate Minimum 90% Tested. Meets applicable standards.
> 50 hour life
collection at 20 R/sec
Dimensions 6 (w) x 6.25 (d) x 2.75 in (h)
Electrometer drift For more information, receive our full product catalog, or
(15.25 x 6.25 x 7 cm) order online, contact Radiation Management Services
Low Range: 1 mR/min typical; 6 mR/min business of Fluke Biomedical: 440.248.9300 or
Weight 1.125 lb (0.51 kg) www.flukebiomedical.com/rms.
maximum
Optional accessories Specifications are subject to change without notice.
High Range: 10 µR/min typical; 60 µR/min
maximum Carrying Case (Model 89-525): holds ©2005 Fluke Biomedical. All rights reserved. Victoreen and
RAD-CHECK are trademarks of Fluke Corporation. Printed in USA.
RAD-CHECK MICRO-R and accessories 06-526-5240-ds rev 4 03 jun 05
Manual reset Resets display to zero
Wide-Range, Mammographic, DI Diagnostic Imaging
& Dental Digital kVp Meters
Models 07-494, 07-492 & 07-479
Introduction • Choose from three kVp meters
Whether you choose the
Wide-Range,
• Easy read digital display
Mammographic, or the
Dental Digital kVp
Meter, you will get • Painless setup
quick and accurate
measurements of your
diagnostic x-ray
generator tube potential.
These instruments need
no connection to the
x-ray generator.
Applications
These lightweight, rugged units are extremely easy to use: simply place on the x-ray Features
table, with the detector facing the x-ray source. With the beam’s central ray centered
on the detector, an exposure is made, and the reading appears immediately on the • Automatic display reset
large, easy-to-read liquid crystal display. • No remote control cables
Unique features are provided to ensure maximum efficiency and accuracy. Readings • Scope output for waveform analysis
remain on display until the next exposure is made, at which time the reading is
• Compact, lightweight and battery-
automatically updated. Automatic display indicators tell you when adjustment of
operated
exposure factors or battery replacement is necessary. Neither remote-control cables
nor time-consuming manual re-zeroing are needed.
A BNC connector is provided for radiation waveform display on a storage
oscilloscope.
Specifications
Ranges Controls Optional accessories
Wide-Range Wide-Range On/off, single/three-phase Carrying Case (Model 89-473)
Low: 50 to 90 kVp, 0.1 kVp resolution and range selection switch Available model(s)
High: 80 to 150 kVp, 0.1 kVp resolution Mammographic On/off and Moly/ 07-494 Wide-Range Digital kVp Meter
Mammographic 24 to 40 kVp, 0.1 kVp Tungsten selector switches 07-492 Mammographic Digital kVp Meter
resolution Dental On/off and single/three-phase 07-479 Dental Digital kVp Meter
Dental 45 to 90 kVp, 0.1 kVp resolution selector switches
Accuracy Operating temperature 50° to 104°F
Wide-Range ± 3% or 3 kVp (whichever is (10° to 40°C)
greater) Relative humidity 0 to 90%, non-
Mammographic ± 3% or 1.5 kVp condensing
(whichever is greater) Power requirements 9 V alkaline battery,
Dental ± 3% or 3 kVp (whichever is greater) 150 hours operation
mAs requirements Display 3.50 x 0.50 inch LCD. Automatic Tested. Meets applicable standards.
indication of (a) low battery condition,
Wide-Range (45.7 cm SDD) 18 mAs at
(b) need to adjust exposure factors For more information, receive our full product catalog, or
120 kVp; 50 mAs at 60 kVp, single phase.
Output signal BNC connector for order online, contact Radiation Management Services
Minimum exposure time 1/20 (0.05) sec business of Fluke Biomedical: 440.248.9300 or
waveform analysis
Mammographic (25 cm SDD) 100 mAs at www.flukebiomedical.com/rms.
24 kVp. Minimum exposure time 1/20 (0.05) sec Dimensions 8 (w) x 6 (d) x 2.50 in (h) Specifications are subject to change without notice.
(20 x 15 x 6 cm)
Dental 8.5 mAs at 45 kVp; 0.026 mAs at ©2005 Fluke Biomedical. All rights reserved. Printed in USA.
90 kVp Weight 2 lb (0.9 kg) 07-494-ds rev 3 03 jun 05
440.248.9300 www.flukebiomedical.com/rms
Diagnostic Imaging DI
Service & Quality Control Kits
Model 18-526 Series
• Each kit contains the essential Cardiac Cath/Special Procedures QC Kit
instruments that service personnel,
physicists, and QC technicians rely
(Model 18-526-3000)
on to check and calibrate today’s Contains the essential non-
most vital equipment invasive test tools for special
procedures:
• Every instrument selected for • Victoreen® RAD-CHECK®
MICRO-R (Model 06-526-5240)
inclusion in our Service/QC kits was
chosen for its reliability, accuracy, • Wide-Range Digital kVp Meter
and ease-of-use (Model 07-494)
Mammography Radiography/
QC Kit Fluoroscopy
(Model 18-526-1000) QC Kit
Everything you need to make (Model 18-526-2000)
compliance with ACR and All the test devices necessary to
MQSA regulations easy: perform QC in radiographic
• Victoreen® RAD-CHECK® and fluoroscopic suites:
PLUS (Model 06-526) • Victoreen® RAD-CHECK®
• Mammographic Digital kVp PLUS (Model 06-526)
Meter (Model 07-492) • Wide-Range Digital kVp
• Digital X-Ray Pulse-Counter/ Meter (Model 07-494)
Timer (Model 07-453)
• Digital X-Ray Pulse-Counter/Timer (Model 07-453)
• Mammographic Ion Chamber, 3.3 cm3 (Model 6000-529)
• Radiographic Ion Chamber, 30 cm3 (Model 6000-528)
• Carrying Case (Model 89-426)
• Carrying Case (Model 89-426)
Dental QC Kit
(Model 18-526-4000)
Service and QC will be easier and more cost-
effective than ever with this comprehensive kit:
• Victoreen® RAD-CHECK® PLUS (Model 06-526)
• Dental Digital kVp Meter (Model 07-479)
For more information, receive our full product catalog,
• Digital X-Ray Pulse-Counter/Timer (Model 07-453) or order online, contact Radiation Management
• Remote Sensor (Model 07-453-2000): for use with Services business of Fluke Biomedical: 440.248.9300
or www.flukebiomedical.com/rms.
Model 07-453
Specifications are subject to change without notice.
• Radiographic Ion Chamber, 30 cm3 (Model 6000-528)
©2005 Fluke Biomedical. All rights reserved. Victoreen and
• Carrying Case (Model 89-426) RAD-CHECK are trademarks of Fluke Corporation. Printed in USA.
18-526-ds rev 4 15 jun 05
Digital X-Ray Pulse DI Diagnostic Imaging
Counter/Timer
Nuclear Associates Model 07-453
• Easy-to-read digital display
Introduction • Automatic reset; holds a reading until
Poor or inconsistent quality of x-ray the next exposure
images is often caused by inaccuracy
or inconsistency of the generator’s • Battery-operated, light weight; fits
timer. This results in repeat easily into tool box or pocket
examinations, which cost time and
money. A poorly maintained system is • Output connector (included) allows a
also a hazard to the patient; when a radiation output waveform to be
malfunction in the timer occurs, the viewed on an oscilloscope
patient may receive unnecessary
radiation doses. It is for important • For added operator convenience, the
reasons such as these, that regular remote sensor is available as an
monitoring of x-ray systems and option
timers is an essential part of a good
quality assurance program.
Applications Features
The Digital X-Ray Pulse Counter/Timer is a non-invasive, solid-state instrument that • Measures duration of radiation output
can be used to measure the exposure time of either AC or DC x-rays. It can also produced by x-ray generators
measure the duration of radiation output produced by a wide variety of medical and • Measures AC or DC x-rays
dental x-ray systems. A sensitive x-ray detector in the instrument allows direct
• Gives direct readings (time or pulses)
measurement of exposure from the x-ray head. Pulses produced by half-wave and
full-wave x-rays are measured as 60 or 120 pulses per second. For DC, capacitor • Can be used for medical or dental x-ray
discharge and three-phase x-rays, the Digital X-Ray Pulse Counter/Timer measures systems
the exposure time in milliseconds. When testing x-ray timers and controls, the time of • Designed specifically to allow service
relay contact closure can be measured using the AC input feature. personnel to accurately and easily
An output connector on the side of the Digital X-Ray Pulse Counter/Timer allows the assess the performance of x-ray
user to view a radiation output waveform on an oscilloscope. Using this feature, generators, timers, and controls
technicians can diagnose and troubleshoot problems with x-ray generators. • Reduces repeat examinations; saves
time and money
Specifications
Accuracy Controls/Indicators Dimensions 3.15 (w) x 5.8 (d) x 1.6 in (h)
AC input ± 1 count • Three-position switch: Pulse, Off, (8 x 14.7 x 4 cm)
DC input 2%, ± 1 count Milliseconds Weight 0.5 lb (0.21 kg)
X-ray detection ± 1 count • Four-digit LCD (0.4 inch character) Optional accessories
Sensitivity • Low battery indicator Remote Sensor with 10 ft cable
AC input 65 VAC minimum • “Low Batt” appears in display when (Model 07-453-2000)
X-ray input 50 kVp, 5 mA at 5 cm from battery voltage reaches 4.8 V ± 0.5 V Oscilloscope Leads (Model 88-453)
top surface of case, pointed to target on case • Power-on: LED (green); oscilloscope Carrying Case (Model 89-453)
Range 9999 pulses; 9999 milliseconds output Available model(s)
Display 0.4 inch liquid crystal Remote sensor The optional Remote Sensor 07-453 Digital X-Ray Pulse Counter/Timer
Power requirements 9 V battery, alkaline or can be used when the user would like to have
equivalent, 24 hours continuous; typically the unit in their hand, so that readings can be
6 months of normal use seen without having to walk back-and-forth Tested. Meets applicable standards.
AC input jacks 130 volts AC maximum; from the x-ray table to the control room after
65 V AC minimum; input circuit not affected each exposure. The Remote Sensor can also For more information, receive our full product catalog, or
be used when placement of the Digital X-Ray order online, contact Radiation Management Services
by reversed polarity business of Fluke Biomedical: 440.248.9300 or
Pulse Counter/Timer in the beam is www.flukebiomedical.com/rms.
Output signal BNC connector for waveform
questionable, such as in a Panorex dental
analysis Specifications are subject to change without notice.
x-ray unit
Connections None required for direct ©2005 Fluke Biomedical. All rights reserved. Printed in USA.
exposure measurement 07-453-ds rev 3 03 jun 05
440.248.9300 www.flukebiomedical.com/rms
Diagnostic Imaging DI
Dual-Range Digital mAs Meter
Nuclear Associates Model 07-487
Applications
The greatest use for the Nuclear
Associates’ Model 07-487 mAs meter is
in calibrating the high-current, short-time
station where a conventional mAs meter Introduction
is precluded by tube ratings. The The Nuclear Associates’ Dual-Range Digital mAs Meter, Model 07-487, allows
instrument can be used (after verifying service personnel to check and adjust the mA settings of x-ray generators. This easy-
the generator accuracy) to set all mA to-use instrument is calibrated directly in mAs, thus eliminating the need for the
stations and check that phototiming error calculations typically required with more complicated and expensive equipment.
does not exceed the limits of good
practice. To use, simply connect the The digital mAs meter is very sensitive. It can measure increments of 0.1 mAs. It has
cable to the x-ray generator and make the a low range of 0 to 199.9 mAs; push a button and the range expands to 0 to
required exposure. The mAs reading 1999 mAs.
appears instantaneously on the four-digit
LCD. A display indicator warns of the
need for battery replacement.
Specifications
Range 0 to 199.9 mAs (“+” overrange Power requirements Single 9 V alkaline Available model(s)
indicator above 160 mAs). Also 0 to battery; typical life 80 hours 07-487 Dual-Range Digital mAs Meter
1999 mAs (“+” overrange indicator above Input jack Uses 2 banana jacks
1600 mAs)
Dimensions 3.50 (w) x 6.63 (d) x 1.38 in (h) Tested. Meets applicable standards.
Accuracy ± 2% of reading (8.9 x 16.8 x 3.5 cm)
Input 25 to 1000 mA Weight 0.44 lb (0.2 kg) For more information, receive our full product catalog, or
order online, contact Radiation Management Services
Drift Zero Accessories supplied 24 inch cable with business of Fluke Biomedical: 440.248.9300 or
www.flukebiomedical.com/rms.
Operating temp. 50º to 100ºF (15º to 30ºC) banana plugs and insulated alligator clips on
Specifications are subject to change without notice.
Controls POWER (on/off), RANGE (high- opposite ends
©2005 Fluke Biomedical. All rights reserved. Printed in USA.
low), and RESET 07-487-ds rev 2 13 jun 05
The Wave DI Diagnostic Imaging
Precision High-Voltage Divider
Nuclear Associates Model 07-469
440.248.9300 www.flukebiomedical.com/rms
Diagnostic Imaging DI
VeriLUM ®
Specifications Introduction
®
Minimum computer requirements Intel VeriLUM is an innovative tool for ensuring consistent video display performance. It
PC, CD-Rom, Microsoft Windows 95, and a provides an easy and efficient way to judge whether a video display system is
video board capable of 1024 x 768 pixels continuing to function normally or needs adjustment or replacement.
with a minimum of 256 colors or a grayscale
video board Applications
Calibration Traceable to a NIST source
VeriLUM can be used for acceptance testing of a video display system.
Luminance accuracy ± 2%
It also provides a quick visual check for the user. A SMPTE test pattern is
Luminance repeatability ± 1% displayed on each video monitor. If the gray scale range and stability is adequate and
Luminance range 0.05 to 1000 cd/sqm if all of the monitors have essentially the same look-and-feel, then the video system is
Weight 1 lb (0.45 kg) ready for use.
Available model(s) VeriLUM makes a rapid set of measurements of the display luminance for
tracking consistent performance over time. These measurements take less than 30
18-116 VeriLUM Color Dual Mode Pod
(please specify Serial or USB connector)
seconds per monitor and the history chart can be printed when hard copy
documentation is needed.
VeriLUM can be used to perform gamma correction in conformance with the
DICOM Part 14 Grayscale Standard Display Function or any other user-defined
luminance response model. VeriLUM provides on-board gamma correction using
BARCO/Metheus, DOME, Image Systems, Matrox, and RealVision gray scale video
boards. If the operating system is Microsoft Windows 98, 2000, XP®, ME®, or NT®,
For more information, receive our full product catalog, or and if the color video card supports downloadable gamma ramps, then VeriLUM will
order online, contact Radiation Management Services
business of Fluke Biomedical: 440.248.9300 or
use that capability.
www.flukebiomedical.com/rms. VeriLUM software can be installed on as many workstations as desired; no additional
Specifications are subject to change without notice.
licenses are required. This allows the VeriLUM luminance pod to be taken from
©2005 Fluke Biomedical. All rights reserved. VeriLUM is a
trademark of IMAGE Smiths, Inc. DICOM is a trademark of Digital
workstation to workstation to perform measurements. The VeriLUM luminance pod
Imaging and Communications in Medicine. Microsoft, Windows, and supports all CRTs and all LCD panels. It uses a serial communications port or a USB
Windows XP, ME, and NT are trademarks of Microsoft Corporation.
Intel is a trademark of Intel Corporation. Printed in USA. port on the PC, thus a simple extension cable (provided) allows for simple
18-116-ds rev 2 07 feb 05
connection.
Nuclear Associates DI Diagnostic Imaging
Precision Photometer
Model 07-621
• NIST-traceable
Introduction Features
Photometers are required for medical, scientific, and laboratory applications. This • Easy to read digital display
Precision Photometer utilizes a filtered sensor with spectral response tightly • Small...convenient to carry and
calibrated to the CIE photopic response. The illuminance receptors closely follow supplied with its own carrying case
the Cosine Law relative sensitivity versus angle of illuminance.
• Rugged construction
The Precision Photometer is a highly accurate instrument designed to measure both • Measure button: press it to get
illuminance (the amount of light falling on a surface) in lux (lumens per m2) and continuously updated readings.
luminance (the amount of light emitted from a surface) in nit (candela per m2). Releasing the measure button freezes
the last reading for convenient
Applications reference
The Precision Photometer easily and quickly verifies that an x-ray collimator light • Range button: adjusts the measurement
and/or CT system light source used for patient alignment and localization, is in display for the resolution desired
accordance with regulations and guidelines. It also measures the brightness and
uniformity of an x-ray view box, quickly detecting non-uniformity (which may • LED display: visible in very dim light,
appear as artifacts, causing misdiagnosis). This battery-operated photometer has a as well as direct sunlight. LED displays
bright LED display and only two operating controls: “Measure” for taking a reading, are inherently robust in comparison to
and “Range” to adjust the meter display to the light level being measured. liquid crystal displays (LCDs)
• The battery-powered photometer
provides tens of thousands of readings
440.248.9300 www.flukebiomedical.com/rms
Specifications
Capabilities Illuminance measured in lux (foot-candles); or Optional accessories
2
luminance measured in candela/m (nit) Flexible 12 inch Fiber-optic Probe
Readout display Three-digit LED, 0.25 inch high • Designed for measuring relative densities
Range 0.1 to 999,000 lux or nits (candela/m2), equivalent to 0.01 to of areas on radiographs, or luminance of
99,000 foot candles video screens and view boxes
Accuracy Within 7% of full scale range, for light sources between
2500° and 5400° Kelvin
Electrical accuracy 1% plus 2 digits Rigid 1 inch
Sensor Silicon photodiode with photometric filter Fiber-optic Probe
Response curve shape Close match to CIE photopic response curve • Designed for
Power requirements Type A-76 alkaline button cells or silver oxide readings taken
equivalents against a monitor
Dimensions 2.8 (w) x 1.2 (d) x 4 in (h) (7 x 3 x 10 cm) or view box
Weight 0.25 lb (0.11 kg)
Optional accessories
Fiber-optic Probe, Flexible 12 inch (Model 07-634)
(must be calibrated with meter)
Fiber-optic Probe, Rigid 1 inch (Model 07-634-1000)
(must be calibrated with meter)
Rotating Illuminance Receptor (Model 07-634-1100)
(must be calibrated with meter)
Carrying Case (Model 89-621)
Available model(s) Rotating Illuminance Receptor
07-621 Precision Photometer • Designed for reading background light
07-621-6341 Precision Photometer with rotating illuminance and field light
receptor
07-621-6342 Precision Photometer with rigid 1 inch fiber-optic probe
07-621-6343 Precision Photometer with flexible 12 inch fiber-optic probe
Factory recalibration available
Model 03-310
• Static control spray used on viewing screens and other static generating
surfaces
• Water based, nonflammable, no solvents
For more information, receive our full product catalog, or order online, contact Radiation Management Services
business of Fluke Biomedical: 440.248.9300 or www.flukebiomedical.com/rms.
Specifications are subject to change without notice.
©2005 Fluke Biomedical. All rights reserved. Radi-Stat is a trademark of Fluke Corporation. Printed in USA.
07-621-ds rev 3 15 feb 05
Collimator/Beam DI Diagnostic Imaging
Alignment Test Tool
Model 07-661-7662
Collimator Verification of the proper end of the 15 cm high clear plastic cylinder.
alignment of the collimator light field with the When the balls are positioned over one
x-ray field is essential in radiographic quality another and at a right angle to the film, their
control. This device readily indicates a 1% or images will appear as one if the central ray is
2% misalignment at a 40 inch focal-film truly perpendicular to the film. The
distance (FFD), but it may be used at any FFD. approximate degree of improper angulation
It consists of a flat 8 x 10 inch plate with a can also be determined.
14 x 18 cm pattern etched on its surface. It Specifications
can also be used to check fluoroscopy
alignment and collimation. Dimensions beam alignment
5.9 (h) x 2.5 in OD (15 x 006.3 cm)
Beam alignment Improper central ray
Weight beam alignment 0.54 lb (0.24 kg)
alignment will distort a radiographic image.
This device provides a simple means of Weight collimator 0.41 lb (0.19 kg)
determining if the x-ray beam is Available model(s)
perpendicular to the image receptor and 07-661-7662 Collimator/Beam Alignment
centered with respect to the light field. A steel Test Tool
ball is mounted in the center of a disc at each
For more information, receive our full product catalog, or order online, contact Radiation Management Services business of Fluke Biomedical: 440.248.9300 or
www.flukebiomedical.com/rms.
Specifications are subject to change without notice.
440.248.9300 www.flukebiomedical.com/rms
Diagnostic Imaging DI
Star Patterns
Models 07-503 to 07-551
Star X-Ray Test Patterns for Measuring Focal Spot Size
Focal spot size can be determined by observing the regions of blurring which occur when the pattern is
radiographed by an x-ray source of finite dimensions. Radiation from different areas of the focal spot will
cause a periodic blurring of the pattern due to penumbra effects. Knowledge of the geometric factors, and
the distance from the center of the pattern to the region where blurring occurs, will permit the calculation of
the focal spot size with the same accuracy as measurements made with a pinhole camera.
Shown in Model 07-509-2000
Available model(s)
07-503-2000 High-Precision Star X-Ray Test Pattern, 55 mm Ø. 07-543-1000 Ultra-High Precision Star X-Ray Patterns.
For measuring focal spots from 0.1 to 0.3 mm. Has four 15° (Same specifications as Model 07-543-2000)
patterned sectors with a 0.5° angle of a single line within a sector. 07-550 Ultra-High Precision Star X-Ray Patterns. (Same as Model
Lead-foil thickness 0.03 mm 07-503-2000 except it has four 45° patterned sectors, for easier
07-503-1000 Ultra-High Precision Star X-Ray Test Pattern. interpretation). Lead-foil thickness 0.03 mm
(Same specifications as Model 07-503-2000) 07-551 Ultra-High Precision Star
07-509-2000 High-Precision Star X-Ray Test Pattern, 55 mm Ø. X-Ray Pattern. (Same as Model
For measuring focal spots from 1 mm and up. Has four 45° sectors 07-503-2000 except it has four 15°
with a 2° angle of a single line within a sector. Lead-foil thickness patterned sectors with a 0.25° angle).
0.05 mm Lead-foil thickness 0.03 mm
07-509-1000 Ultra-High Precision Star X-Ray Test Pattern. 07-510-2000 High-Precision Star
(Same specifications as Model 07-509-2000) X-Ray Test Pattern, 55 mm Ø. For
07-542-2000 Precision Star X-Ray Test Pattern, 55 mm Ø. For measuring focal spots from 1 mm and
measuring focal spots from 0.3 to 0.6 mm. Has four 28 patterned up. Has one 360° pattern sector with
sectors with a 1° angle of a single line within a sector. Lead-foil a 2° angle of a single line within a
thickness 0.03 mm sector. Lead-foil thickness 0.05 mm
07-542-1000 Ultra-High Precision Star X-Ray Test Pattern. 07-510-1000 Ultra-High Precision
(Same specifications as Model 07-542-2000) Star X-Ray Test Pattern. (Same Shown in Model 07-510
07-543-2000 High-Precision Star X-Ray Test Pattern, 55 mm Ø. specifications as Model 07-510-2000)
For measuring focal spots from 0.8 to 1.5 mm. Has four 35°
patterned sections with a 1.5° angle of a single line within a sector.
Lead-foil thickness 0.03 mm
For more information, receive our full product catalog, or order online, contact Radiation Management Services business of Fluke Biomedical: 440.248.9300 or
www.flukebiomedical.com/rms.
Specifications are subject to change without notice.
©2005 Fluke Biomedical. All rights reserved. Printed in USA. 07-501-ds rev 2 22 feb 05
440.248.9300 www.flukebiomedical.com/rms
Diagnostic Imaging DI
X-Ray Output Detector
Model 07-451
• Timer calibration (single-phase, at standard diagnostic conditions (80 kVp,
three-phase or CP units) 100 mA). This extremely high output permits
• Loading characteristics the simple interpretation of oscilloscope
• Rectifier malfunctions displays. Since the detector rise time is better
• Contactor problems than 1 microsecond, no alteration of the true
• Cable or connector arcing x-ray output pulse shape is introduced.
• Shutter calibration, etc. Specifications
This low-cost X-Ray Output Detector offers Shock-resistant, solid-state diode detector.
a dynamic means of demonstrating x-ray Power source: none required. Rise Time:
generator performance. It is used with a Less than 1 µsec.
storage or camera oscilloscope to display the Dimensions 1.25 x 1.25 x 0.50 in
intensity-time relationship of an x-ray beam. (3.175 x 3.175 x 1.27 cm)
Scope tracing examples of typical To use, the detector is placed in the x-ray Weight 0.58 oz (16.6 g)
x-ray generator diagnostics
beam, and the output cable is connected to Optional accessories
the oscilloscope input. The resulting Cable, 20 ft (6 m), BNC to BNC
waveshape patterns are used to calibrate and/ (Model 88-222)
or diagnose malfunctions in the x-ray Available model(s)
generator.
Timer Timer Loading 07-451 X-Ray Output Detector, includes
The detector supplies a crisp 200 mV signal BNC Output Connector
Introduction Features
The PACS Test Pattern provides a simple, objective approach to system evaluation Evaluates important benchmarks of
and quality control, allowing for quick, objective testing of image quality by a single radiographic image quality,
observer. Simply digitize the PACS Test Pattern by loading the test image from the including:
disk and display it at full resolution on your display monitor. Then, using your
systems controls, magnify and view different areas of the test pattern. • High contrast resolution
• Low contrast discrimination
Applications • Linearity of gray scale response
A regular quality control procedure is mandatory to ensure a diagnostic level of • Geometric distortion
image quality with PACS and teleradiology. A quick, objective and reproducible QC
test is needed to ensure optimization of the digitizer and display system. • Reproduction of continuous fine lines
• Digitizer noise
As an integral part of your regular QC for a laser scanner or CCD digitizer, the digital
format PACS Test Pattern will prove itself as an invaluable QC tool for testing image
display systems and film printers.
“A teleradiology/PACS quality control test pattern may help identify problems that should be addressed to optimize printer
quality. Regular interval testing of paper printers with a dedicated test pattern should be recommended as a part of the
quality control program for all teleradiology/PACS systems.”*
* A.D. Maidment, Ph.D; M, Albert, Ph.D; and E.J. Halpern, M.D., “A Quality Control Program for Paper Printers Used with Teleradiology/PACS,”
Radiology, 205 (P) (November 1997), 307.
440.248.9300 www.flukebiomedical.com/rms
PACS/Teleradiology Test Pattern is superior Specifications
to the SMPTE Test Pattern for PACS/ Horizontal, vertical, and diagonal high contrast line Pairs
Teleradiology quality control Tests resolution up to 6 lp/mm when printed on 14 x 17 inch film
Gray scale Tests full range of your printer or display
The PACS/Teleradiology Test Pattern tests:
Low contrast pattern Digital values of low contrast squares differ
• The ability to discriminate between 16 different gray scale from the background density by 9%
levels from its lowest to its highest optical density, which is a
Weight < 1 lb
greater number of gray scale levels than is tested by the
SMPTE Test Pattern. Available model(s)
• For low contrast discrimination with a rose-hole-type pattern 07-450-1024 PACS Digital Test Pattern (1024 x 768)
which is not available on the SMPTE Pattern. 07-450-4000 PACS Digital Test Pattern (4 x 4 K)
Note: Digital PACS Test Pattern supplied on a single, 3.5 inch floppy
• The ability to discriminate between different gray levels on disk
both sides of the pattern, as well as in the center of the
pattern. These features, designed specifically for QC in
teleradiology systems, provide a more superior test of gray
For more information, receive our full product catalog, or order online, contact
scale reproduction than the SMPTE Pattern. Radiation Management Services business of Fluke Biomedical: 440.248.9300 or
www.flukebiomedical.com/rms.
• High contrast resolution in horizontal, vertical and diagonal
Specifications are subject to change without notice.
axes, while the SMPTE Pattern test only horizontal and
©2005 Fluke Biomedical. All rights reserved. Printed in USA.
vertical axes. 07-450-ds rev 2 01 jul 05
Introduction Features
A microprocessor and programmable logic-based design provide significant • Pushbutton operation including last
performance improvement over the common memory-based monitor testers. Up to setting memory
one hundred scan formats are available, including all current Video Electronics • Four video output connections for
Standards Association (VESA) standards and most fixed frequency monitor scan direct connection to PC, SUN, MAC
formats, and the Model 07-894 is field-upgradeable to custom frequencies. Four and BNC monitors
video patterns (full raster, color bars, gray scale and dotted cross hatch) with
16 colors and intensity control provide all the patterns and colors normally required • Features convenient VGA, MAC and
to test and align any monitor. SUN groupings of commonly used
scan formats
• Selectable sync outputs for composite,
sync-on green, and horizontal and
vertical
• Selectable polarity settings for
horizontal and vertical sync outputs
• Tests for color • Tests for • Tests for • Tests for color • Optional 10 minute timeout on video
to prevent CRT burn-in
balance intensification convergence, focus purity
and geometry • Four video patterns with 16 colors and
• Shows 16 split • Displays 16 • Shows 11 x 11 lines • Displays full intensity control provided
bars with half bars with 10 x 10 dots color window
intensity
440.248.9300 www.flukebiomedical.com/rms
Applications Specifications
The newly designed Model 07-894 is an easy-to-use, portable Size 8 x 6.6 x 2.7 in
instrument for testing, repairing and aligning analog computer Patterns The four video patterns with 16 colors and intensity control
monitors, LCD displays and video projection systems. Use the provide all the patterns and colors normally required to test and align
Model 07-894 to cut repair time and reduce equipment any monitor.
downtime. Battery power enables you to repair monitors at the Power supply 9 VDC 500 mA wall transformer or six “C” cell
job site, eliminating the need to disconnect the monitor and alkaline batteries
transport it to another location for testing. Projectors can be Compatibility Over 100 frequencies cover a wide range of monitors,
aligned onsite, without the host workstation being present. including all industry standard PC, MAC and SUN frequencies
Connectors
• VGA (15 pin HD female)
• MAC II (15 pin D female)
• SUN (13W3 D female)
• RGHHV for BNC (5)
Available model(s)
07-894 Computer Monitor Tester
For more information, receive our full product catalog, or order online, contact
Radiation Management Services business of Fluke Biomedical: 440.248.9300 or
www.flukebiomedical.com/rms.
Specifications are subject to change without notice.
* Designed by Joel E. Gray, Ph.D., Professor Emeritus, Department of Radiology, Mayo Clinic®, Rochester,
MN 55905. Manufactured under licensing agreement with Mayo Foundation for Medical Education and
Research.
Radiopaque Rulers
Model 07-533
For more information, receive our full product catalog, Available model(s)
or order online, contact Radiation Management 07-533 Radiopaque Ruler, 30 cm long,
Services business of Fluke Biomedical: 440.248.9300
07-533-1100 Radiopaque Ruler, 110 cm long,
or www.flukebiomedical.com/rms. 2 mm divisions. Weight: 0.05 lb (0.04 kg) 2 mm divisions. Weight: 0.5 lb (0.24 kg)
Specifications are subject to change without notice. 07-533-1000 Radiopaque Ruler, 100 cm long, 07-533-3600 Radiopaque Ruler, 36 cm long,
©2005 Fluke Biomedical. All rights reserved. NEMA is a registered
2 mm divisions. Weight: 0.5 lb (0.24 kg) 2 mm divisions. Weight: 0.05 lb (0.04 kg)
trademark of the National Electrical Manufacturers Association for
its publication of voluntary standards and guidelines. NEMA is not a
certification mark. Mayo Clinic is a registered trademark of Mayo
Foundation. Printed in USA.
07-622-ds rev 2 18 feb 05
X-Ray Pinhole Assemblies DI Diagnostic Imaging
Models 07-611 to 07-633
• For measuring the dimensions of focal spots of x-ray tubes
• Custom Pinholes Available on Request
One of the specifications and chief features of an x-ray tube is its focal spot size, a
very important factor in the resolution to be achieved during a radiologic
examination. A small focal spot size will provide the maximum resolution. However,
there are other factors in the construction of an x-ray tube, such as the heat
dissipation within the target, which limits the minimum size of the target.
In order to determine the focal spot size of an x-ray tube, a small and precise pinhole
is often used. Its size must be very small compared to the focal spot to be measured.
Any of four precision pinholes can be used with our Model 07-623 and 07-622 test
stands. The test stand height is adjustable in order to maintain the minimum
magnification factor of two. A fluoroscopic screen is part of the test stand and is
used for centering the focal spot before exposing the film.
The 0.010 mm pinhole diameter is for focal spot sizes from 0.5 to 0.10 mm; the
0.030 mm pinhole diameter is for focal spot sizes below 1.0 mm; the 0.075 mm
diameter is for focal spots from 1.0 to 2.5 mm; the 0.1000 mm diameter is for sizes
above 2.5 mm. Pinhole assemblies
The pinhole diaphragm is made from a 90:10 gold-platinum alloy in accordance with
the table and figure.
Pinhole diaphragm dimensions Nominal
Model (mm) pinhole diameter
D L (mm)
07-633 0.010 ± 0.005 0.020 ± 0.010 0.010
07-613 0.030 ± 0.005 0.075 ± 0.010 0.030
07-617 0.075 ± 0.005 0.350 ± 0.010 0.075 Cross-sectional view of pinhole
07-611 0.100 ± 0.005 0.500 ± 0.010 0.100
High-Quality Comparators
Models 07-620 & 07-635
• For magnification and measuring focal spot sizes
Magnification 25X
Supplied with a 0-5 mm reticle. Features
a knurled ring which adjusts the focal
point to personal preference. Transparent
body allows illumination to fall on Magnification 7X
magnified area. Supplied with protective
Supplied with its own leather case. Fits in
storage case. An excellent, high-quality
the palm of your hand. Transparent body
precision magnifier.
allows illumination to fall on magnified
Weight 0.20 lb (0.10 kg) area. Accurate, easy to use, versatile, and
Available model(s) truly portable.
07-635 Comparator, 25X Magnification
Weight 0.05 lb (0.04 kg)
For more information, receive our full product catalog, or order online, contact Radiation Management Services Available model(s)
business of Fluke Biomedical: 440.248.9300 or www.flukebiomedical.com/rms.
Specifications are subject to change without notice. 07-620 Comparator, 7X Magnification
©2005 Fluke Biomedical. All rights reserved. Printed in USA. 07-611-ds rev 2 18 feb 05
440.248.9300 www.flukebiomedical.com/rms
Diagnostic Imaging DI
TransMat ®
Specifications Introduction
Dimensions 26 x 72 in (66 x 183 cm) This lightweight, durable mat is designed to facilitate movement of patients to and
Weight 6.8 lb (3.1 kg) from gurneys, stretchers, and examination tables. Except when going to and from the
Available model(s) hospital bed, the patient never has to leave the mat...it travels with the patient. The
mat’s 26 x 72 inch size makes it compatible with all stretchers, tables, and beds. Its
17-355 TransMat Patient Transfer
fully-washable surface permits easy removal of stains, including spilled contrast
Mattress
media.
For more information, receive our full product catalog, or order online, contact Radiation Management Services
business of Fluke Biomedical: 440.248.9300 or www.flukebiomedical.com/rms.
Specifications are subject to change without notice.
©2005 Fluke Biomedical. All rights reserved. TransMat is a trademark of Fluke Corporation. Printed in USA.
17-355-ds rev 2 18 feb 05
Anthropomorphic Phantoms DI Diagnostic Imaging
®
PIXY , Sectional and Mammo II
• Teaching/Training in Diagnostic
Radiology
Phantom?
440.248.9300 www.flukebiomedical.com/rms
PIXY The Anthropomorphic Training/ The matching of skeletons and soft-tissues permits external and
bony landmarks to coincide. The position of bones within the
Teaching Phantom soft tissues is anatomically correct.
PIXY was designed specifically The detail cast into Fluke Biomedical phantom skeletons is
for training radiologic considered a triumph of the sculptural moldmaker’s craft. The
technologists. PIXY is 156 cm skull, for example, has frontal and sphenoidal sinuses,
(5 ft 1 in) tall and weighs 48 kg ethmoidal and mastoid air cells and the auditory ossicles. Bone
(105 lb). She is repeatable, sutures are radiographically visible.
virtually indestructible and a Opaque PIXY
Soft tissues
convenient substitute for patients.
PIXY is available in opaque or transparent tissue-equivalent
Pixy is made of tissue-equivalent materials and has life-like materials. The transparent PIXY has visible organs and
articulations. Students have no difficulty in maneuvering PIXY skeleton except at the hips, knees, and elbows, which are
into most desired positions. opaque because, as on the opaque PIXY, latex coverings are
• An anatomically and radiologically correct female needed to retain tissue-equivalent gels for soft-tissue continuity
• Small size and low weight simplify positioning at these articulations. Two-ply coverings protect against gel
• Can be positioned for most views leakage.
• Permits evaluation of student performance Lungs
• Organs accept contrast media Standard PIXY lungs are molded of tissue-equivalent foam
• Opaque or transparent with the mass density of inflated human lungs (0.30 g/cc). They
PIXY is used to demonstrate anatomy and evaluate positioning are connected to the oro-nasal cavity by the stem bronchi and
and imaging techniques, including kVp, mAs, contrast, optical trachea. The oro-nasal pharynx is filled with a nearly air-
density, OFD and TFD. Radiographs of PIXY are optically equivalent foam.
equivalent in density and contrast to human patients. Optional animal lungs, which duplicate the intricate detail of
PIXY permits unlimited exposures and tolerates trainee errors. the vascular trees, are available. These lungs are fixed in the
inflated state and molded to conform to the pleural cavities of
PIXY anatomy the phantom. The pulmonary arteries are injected with a blood-
PIXY shoulders have ball and socket joints; elbows and knees equivalent plastic. In addition, low, medium or high contrast
flex 90° to 100°. Hips flex 130° with 30° hyperextension. material may be selected by the user.
A “frog position” is made possible by lateral flexion at the hips. Refurbishment
The right hand is molded with fingers positioned for an AP
Fluke Biomedical phantoms offers a PIXY refurbishment
view, while the left hand is in an oblique-lateral position. The
service which, after wear and tear from usage over an extended
left foot is in full plantarflexion; the right foot is in a neutral
period of time, restores PIXY to its original condition. This
position.
service includes repair of minor bone fractures of hands and
C1, C2, C6 and C7 were converted to mechanical nylon joints feet. Quotes are furnished upon request for more extensive
because educators in the field prefer full positioning damage.
capabilities for the head. The design permits the remaining
neck vertebrae to be fixed in a normal position, while assuring Available
Model
model(s)
Description
a full range of head motion. Opaque PIXY Phantom with stomach, gall bladder,
PIXY contains abdominal and pelvic organs: stomach, gall RS-102 urinary bladder, kidneys, rectum and sigmoid flexure.
bladder, urinary bladder, kidneys, rectum and sigmoid flexure. Permanent shipping/storage case
RS-102T Same as RS-102 but transparent
These are air-filled, but accept water or contrast media and can
RS-157* Animal lungs
be easily flushed after use. Custom fractures and custom
pathologies are optional. Custom fractures and pathologies. Depressed skull
fracture at any desired location. Hairline fracture of the
PIXY materials scaphoid bone of the wrist (with no fragment separation).
Skeletons RS-102SP* Fracture of the superior pubic ramus. Rib fracture at the
mid-axillary line. Unseparated fracture of the patella.
Skeletons are in continuous production, Stress fracture of the 5th tarsal bone of the extended foot.
so prompt shipments are routine. Pathologies based on user's requirements
Nevertheless, human skeletons are
RS-102R Standard PIXY refurbishment
available for users who desire them. Transparent PIXY *Must be ordered with phantom (cannot be retrofitted).
There is a surcharge to cover the high *Highly detailed polymer skeletons which reproduce the shape, mass density and attenuation coefficients of
cost of scarce natural skeletons and for added labor needed to the cortical bone and spongiosa, allow continuous production of phantoms instead of sporadic production
due to limited availability, variable size and uncertain chemical composition of human skeletons.
rework them to fit PIXY molds.*
Sectional Phantoms
Sectional Phantoms, with the anatomic and radiofidelity of
PIXY, are used widely in teaching/training, with many other
applications. They represent an average male 175 cm (5 ft 9 in)
tall and weighs 74 kg (162 lb). They are rugged, easily
transported and shatter-proof.
RS-116T RS-117
Available model(s)
Model Description
RS-108 or RS-108T Head with cervical spine (C1-C7)
RS-109 or RS-109T Head without cervical spine
RS-111 or RS-111T Thorax
RS-115 RS-118T RS-119
RS-113 or RS-113T Pelvis
RS-114 or RS114T Hand/wrist (natural position), right or left
RS-115 or RS115T Hand/wrist (oblique position), left only
RS-116 or RS-116T Foot/ankle (natural position), right or left
440.248.9300 www.flukebiomedical.com/rms
MAMMO II Phantom The breast is mounted on a post with adjustments for height
and the angles needed for various views.
MAMMO II Phantom is a patient substitute that allows
instructors to teach mammography positioning. Mammograms,
taken within the range of standard technical factors, provide
realistic images with high contrast.
MAMMO II simplifies classroom procedures, allowing
trainees to take as many exposures as are needed to develop
expertise.
• Teaches positioning for craniocaudal and mediolateral oblique Phantom mammogram
views
• Teaches spot-compression procedures
• Sensitivity training for patient comfort
• Helpful in training male students (no more loaded vests to simulate Positioning for mediolateral
breasts) oblique view
A molded gel provides realistic compressibility. Mammograms
are taken at a breast thickness of 5 cm, using normal technical
factors. A small indicator box is placed next to the breast to signal
when the 5 cm breast thickness is reached. An “ouch” is
represented by a red warning light which informs the trainee
that a patient’s pain level can be reached at about this
compression. The light is actuated 5 mms before the dead stop
produced by the box. This procedure emphasizes the care
needed for patients when the pain zone is reached.
Indicator box
Available model(s)
RS-750 MAMMO II Phantom complete with stand and
carrying/storage case
For more information, receive our full product catalog, or order online, contact
Radiation Management Services business of Fluke Biomedical: 440.248.9300 or
www.flukebiomedical.com/rms.
©2005 Fluke Biomedical. All rights reserved. PIXY® is a reqistered trademark of Radiology Support
Devices, Inc. Printed in USA.
AnthroPhantoms-ds rev 1 13 jun 05
Lung/Chest Phantoms DI Diagnostic Imaging
Model RS-310 to RS-350
440.248.9300 www.flukebiomedical.com/rms
Custom pathologies Available model(s)
The Lung/Chest Phantom is available either with the
Model Description
diaphragm permanently sealed to the interior of the phantom,
fixing the lungs in place, or with a removable diaphragm, Lung/Chest Phantom with permanently sealed
RS-310
which permits the interchange of lungs, and provides an diaphragm
unlimited number of configurations and disease simulations. Lung/Chest Phantom with permanently sealed
RS-315
diaphragm and left coronary artery
Custom pathologies available as wrap-around sheets on either Lung/Chest Phantom with removable diaphragm,
RS-320
or both lungs. lung pair insert without coronary artery
Lung/Chest Phantom with removable diaphragm,
All models are available with blood-equivalent pulmonary RS-330
lung pair insert coronary artery
arteries, or with low or high contrast media added. RS-340 Lung pair insert without coronary artery
RS-345 Lung pair insert with coronary artery
RS-350 Wrap-around pathology sheets
For more information, receive our full product catalog, or order online, contact
Radiation Management Services business of Fluke Biomedical: 440.248.9300 or
www.flukebiomedical.com/rms.
Specifications are subject to change without notice.
• Full or half-head
• Opaque or transparent
Introduction Features
Fluke Biomedical Angiographic Head Phantoms bridge the gap between physical and • The phantom is virtually shatterproof
anatomical information requirements. Molded in tissue-equivalent material, an
accurate male skull contains a 3-dimensional, high-contrast vascular simulation to • The phantom can be used
facilitate correlation of the radiologist’s subjective evaluation of angiographic image omnidirectionally
quality with actual measurements of resolution and contrast under the same exposure
conditions.
Each Angiographic Head Phantom contains a 5-step wedge. A 2-10 line pair/mm test
pattern is optional. Other patterns or test objects and custom pathologies are
available on special order.
The Angiographic Head Phantom provides a “dry run” to completely check out
angiographic equipment and ensure that it is operating satisfactorily in all significant
details before subjecting patients to radiological procedures. If the imaging system
malfunctions, the phantom also plays a critically important role in isolating and
verifying correction of the problems. The Angiographic Head Phantom confines the
variables to components and technique and is an invaluable service and teaching tool.
440.248.9300 www.flukebiomedical.com/rms
Applications Available model(s)
Unique RSD soft-tissue equivalent material is used for the skin Model Description
tissue and neck and is adjusted to brain density in the cranium.
This material closely duplicates brain tissues in radio- RS-230 Opaque Full Angiographic Head with stepwedge
absorptive and scatter properties. The phantom is virtually Transparent Full Angiographic Head with
RS-230T
shatterproof. stepwedge
Opaque Full Angiographic Head with resolution test
RS-235
A high-contrast 3-dimensional vascular pattern is placed in the pattern (2-10 lp/mm) and stepwedge
median plane. The pattern closely resembles an internal carotid
Transparent Full Angiographic Head with resolution
arteriogram in its early phase. The radio-density of the RS-235T
test pattern (2-10 lp/mm) and stepwedge
angiographic pattern has been adjusted to represent a 75%
contrast-media. The phantom can be used omnidirectionally. RS-240 Opaque Half Angiographic Head with stepwedge
Transparent Half Angiographic Head with
The Angiographic Head RS-240T
stepwedge
Phantom is available as Opaque Half Angiographic Head with resolution
RS-245
either a full or half head. test pattern (2-10 lp/mm) and stepwedge
The full head has the
Transparent Half Angiographic Head with
advantage of representing RS-245T
resolution test pattern (2-10 lp/mm) and stepwedge
the full scatter properties
and subject contrast of a
human head. The half-head
model is well suited to non-
grid techniques. It has the For more information, receive our full product catalog, or order online, contact
Radiation Management Services business of Fluke Biomedical: 440.248.9300 or
advantages of lighter weight, www.flukebiomedical.com/rms.
lower cost and smaller size, Specifications are subject to change without notice.
easily fitting into a briefcase. ©2005 Fluke Biomedical. All rights reserved. Printed in USA.
RS-230-ds rev 1 09 mar 05
• Anthropomorphic
Section A: Aluminum plates, 0.4 mm thick, set at a 45° angle, Section E: ¼ inch diameter
measure beam width and slice thickness. aluminum rod on the
phantom’s longitudinal axis,
checks general alignment,
and “Dry Water” gives a
realistic “noise check.
Section B: The dosimetry section provides patient-exposure
controls, and checks on abnormal internal doses before actual
patient scans. Customization to accommodate TLD with rods or
chips is available.
Available model(s)
RS-250 Computed Tomography (CT) Head
Phantom
Section C: Cylindrical tumors of graded sizes and radiodensities
establish high and low-contrast resolution and demonstrate partial
beam-averaging effects.
440.248.9300 www.flukebiomedical.com/rms
Kohrman Injection Phantom
Diagnostic Imaging DI
(KIP)
Model RS-1300
Features Introduction
• KIP has the capability for injections not KIP has been designed to provide realistic functions, while avoiding complications
only in the spine, but also at one not essential to its use. Proper fluoroscopic needle placement techniques can be
shoulder, one hip, and at the pubis taught or practiced without fear of biological contamination hazards associated with
symphysis fresh or frozen cadavers.
• KIP has skeletal structures covered Fluoroscopically, the look and feel of the anatomical landmarks are important and
with a soft gel and a latex skin, can be demonstrated with the use of KIP. Needle placement for caudals, epidurals,
supported by a hard, synthetic material selective nerve and root blocks, medial branch blocks, facet injections and
• The materials and design processes in sympathetic blocks can all be demonstrated. Practice RF needle-placement along
KIP are made not only to simplify the with disc needle-placement plus injection techniques for shoulder, hip and symphysis
overall process, but also to keep it as pubis, all in the convenience and safety of your own particular laboratory or teaching
realistic as possible facility. There is now no need to deal with local, state, and federal regulations
regarding biohazards associated with cadavers.
You can keep KIP in the closet and take it out whenever needed.
Applications
KIP has been developed to KIP has been tested to
help one learn and to hone determine the probable
skills for proper needle effective life of injection
placement for a variety of sites; a gel/skin section
interventional techniques. It was used for a trial and
was not practicable to six hundred injections
design the phantom with a were made in a dime-
capability of demonstrating sized area. There was no
realistic flow patterns perceptible degradation
associated with diagnostic of the latex skin or of
dyes, so it is not designed the gel itself. Considering the far greater area available in KIP,
for actual injections. it is unlikely that repairs or refurbishments will be needed for a
However, the all important very long time.
fluoroscopic guidance and
needle-placement are The gel-filled injectable
realistic in KIP. parts of KIP are
completely encased in a
Further simplifications are latex skin. Many latices
the elimination of skeletal are known allergens, but
articulations that permit the latex skin of KIP is
joint or spinal motions and made of the same
the elimination of rib material that is used in
numbers 2 to 9; the rigid the Fluke Biomedical
flesh makes these phantom. Hundreds of these phantoms are in use, but no user
superfluous. Positioning of has ever complained of allergic reactions. The Fluke
KIP as a single unit Biomedical phantom technicians, who have worked for years
provides complete and with this latex, have also never had such reactions. Allergic
realistic positioning reactions are not expected to be a concern for the vast majority
capabilities. of users. The materials and design processes in KIP are made
not only to simplify the overall process, but also to keep it as
realistic as possible.
Available model(s)
RS-1300 Kohrman Injection Phantom (KIP)
For more information, receive our full product catalog, or order online, contact
Radiation Management Services business of Fluke Biomedical: 440.248.9300 or
www.flukebiomedical.com/rms.
Photographs courtesy of Michael J. Ayers - www.theayers.com
Fluoroscopic Images courtesy of Michael j. Kohran, M.D.
Lima Pain and Rehab - Lima, Ohio
440.248.9300 www.flukebiomedical.com/rms
Diagnostic Imaging DI
CLEAR-Pb ®
Compensation Filters*
Nuclear Associates Model 57-4 Series
• Improve image quality
• Filters are 30% lead by weight
• Reduce the need for multiple
exposures
• Ensure a more uniform image
density
• Reduce patient exposure by
selectively attenuating the x-ray
beam
• Lightweight, easy to use
• Mounts to any collimator
• Collimator light field is never
blocked Installation of filter holder into
Lightweight CLEAR-Pb lead-plastic filter is fully transparent
collimator accessory tray
CLEAR-Pb Compensation
Filters eliminate the problems
inherent in imaging a wide
range of densities on one
radiograph. Because they are
far superior, they replace the
bulky, heavy aluminum filters
that block the collimator light
field. In addition, CLEAR-Pb
filters are only one-fifth as
heavy as aluminum filters.
And, they eliminate the use of
gradient-speed intensifying
screens.
CLEAR-Pb filters are made of
lightweight plastic that is 30%
lead by weight. A unique
“quick-stik” magnetic
mounting system plus a filter
Single-exposure spinal radiographs: the Single-exposure AP view of the foot: radiograph at left was
holder that slides into the taken without a CLEAR-Pb filter. Notice “burnout” at the toes
radiograph at left was obtained without a
CLEAR-Pb filter. Notice “burnout” in collimator tray ensure that the due to increased technique needed to properly expose the dense
cervical and thoracic areas. At right is tarsal bones. The radiograph on the right was taken with a
filter is held firmly in place. It CLEAR-Pb filter. The image density is uniform from instep to
radiograph obtained with a CLEAR-Pb
AP/PA filter. Note uniform density also permits instant toes
throughout the spinal column repositioning as the area and/
or degree of filter coverage
changes.
* Patents Nos. 4.129.524 and 4.182/821.
Full-Spine Scoliosis Filter Lead-Plastic Filters
Tested and proven in FDA and specialized radiography Optional accessories recommended for use with Full-Spine
studies…enables scoliosis radiography with more diagnostic Scoliosis Filters:
detail and less radiation exposure.
Breast Shields (Model 57-409) Provides protection to the
radiosensitive breast and lung parenchyma adjacent to the
spine. Fully adjustable; may be used on the Filter Holder
Chest X-Ray Compensation Filter Assembly (57-426) with or without a compensation filter.
For greater diagnostic detail over entire lung with fewer Consists of two 3 x 3 inch steel/lead shields with magnetic
repeats. tape on one side.
Reveals details that conventional x-rays usually miss. Adult Gonad Shield (Model 57-408) Shamrock-shaped insert
has three overlapping lead circles (each 0.50 inch Ø) cemented
The CLEAR-Pb Chest X-Ray Filter eliminates a problem to clear plastic.
commonly encountered in chest x-rays: a portion of the lung
field is usually severely underexposed because it is hidden Pediatric Gonad Shield (Model 57-444) For imaging
behind the hilum. With the CLEAR-Pb filter, you can increase children and the sacroiliac joints of adults. Overlapping lead
the beam intensity sufficiently so that all details of the lung circles are 0.25 inch Ø.
and the posterior mediastinal field are clearly revealed. The
Thin Buildup Filter (Model 57-430) Used with AP/PA filters
CLEAR-Pb filter attenuates the x-ray beam while protecting
for patients with measurements from 14 to 25 cm. Provides
the rest of the field from overexposure. You get clear
additional filtration in the cervical area to compensate for the
diagnostic detail of the lungs, heart and spine.
added exposure that may be needed in the thoracic/lumbar
area.
Reference
1. Gray J.E., Hoffman, A.D., Peterson H.A., Mayo Clinic: “Dose Reduction in Radiography for Scoliosis.” J. Bone Joint Surg., (January, 1983), 5-12.
2. “Patient Exposure Reduction During Scoliosis Radiography,” FDA Publication: 85-8251, August, 1985. Request Reprint No. 361B.
3. American Academy of Orthopedic Surgeons Bulletin 32:1, January, 1984.
4. Downey, E.F., Jr., Butler, P. “Less Radiation and Better Images: a New Scoliosis Radiography System,” Milit. Med., 149 (September, 1984), 526-528.
5. Butler, P.F., Thomas, A.W., Thompson, W.E., Wollerton, M.A., Rachlin, J.A., “Simple Methods to Reduce Patient Exposure During Scoliosis Radiography.”
Radiologic Technology, 57:5 (1986).
6. Petersen, T.D., Rohr, W. “Improved Assessment of Lower Extremity Alignment Utilizing New Radiographic Techniques,’’ Clinical Orthopedics, (June, 1987).
7. Feczko PJ, et al. “Compensation Filtration for Decubitus Radiography During Double-Contrast Barium Enema Examinations,” Radiology, 149:3 (December,
1983), 848-850.
440.248.9300 www.flukebiomedical.com/rms
CLEAR-PB Compensation Filters
Filter holder
Model Filter Weight Length Width Application Configuration
required
57-429 Chest 16 oz 5.125 in 5.125 in No PA View (72 inch FFD)
3 inch Wedge
Yes
57-433 Lateral Decubitus at 5.5 oz 6.5 in 3 in • Suggested for children
(see 57-426)
40 inch FFD
• AP foot
57-440 Foot & Ankle Filter 3.9 oz 6.5 in 2 in No
• Podiatric equipment
AP/PA Yes
57-405* 5 oz 6.5 in 2.5 in • Lateral chest
(72 inch FFD) (see 57-426)
AP/PA Yes
57-415 9 oz 6.5 in 4.125 in • Sectional, AP cervical thoracic
(40 inch FFD) (see 57-426)
Yes
57-430 Thin Buildup 2.5 oz 6.5 in 2.5 in
(see 57-426)
Yes
57-434 Thick Buildup 3.5 oz 6.5 in 2.5 in
(see 57-426)
Thin Wedge; for
Yes • AP scanograms, for determinationi of
57-437** conventional 8 oz 6.5 in 4.5 in
(see 57-426) long-leg length discrepancies
machines
• Full-leg radiography
Thick Wedge; for
Yes (under bodyweight load)
57-438** conventional 13 oz 6.5 in 4.5 in
(see 57-426) • Orthopedic angiographic
machines • Cross table lateral hips
Yes
57-408 Adult Gonad Shield 0.50 oz 6.5 in 1 in
(see 57-426)
Pediatric Gonad
Yes
57-444 Shield 0.25 oz 6.5 in 1 in
(see 57-426)
(72 inch FFD)
Basic Full-Spine Filter Set. Includes CLEAR-Pb AP/PA Filter (57-405), CLEAR-Pb Lateral Cervical Filter (57-406), Lateral Thoracic
57-402
Filter (57-407), Adult Gonad Shield (57-408), Breast Shield Set (57-409), and Filter Holder (57-426). Weight: 3 lb (1.4 kg)
Deluxe Full-Spine Filter Set. Includes CLEAR-Pb AP/PA Filter (57-405), CLEAR-Pb Lateral Cervical Filter (57-406), Lateral Thoracic
57-445 Filter (57-407), Thin Buildup Filter (57-430), Thick Buildup Filter (57-434), Adult Gonad Shield (57-408), Pediatric Gonad Shield
(57-444), Breast Shield Set (57-409), and Filter Holder (57-426). Weight: 3 lb (1.4 kg)
X-Ray Protection System. Includes Adult Gonad Shield (57-408), Breast Shield Set (57-409), Filter Holder (57-426), and Pediatric Gonad
57-404
Shield (57-444).
®
* These models manufactured under licensing agreement with the Mayo Clinic Mayo Foundation.
** These models manufactured under licensing agreement with Alvarado Orthopedic Research Company.
Lead-Steel Sandwich Plastic Magnetic Tape Lead-Plastic
For more information, receive our full product catalog, or order online, contact Radiation Management Services business of Fluke
Biomedical: 440.248.9300 or www.flukebiomedical.com/rms.
Specifications are subject to change without notice.
©2005 Fluke Biomedical. All rights reserved. CLEAR-Pb is a trademark of Fluke Corporation. Mayo Clinic is a registered trademark of the Mayo Foundation. Printed in USA.
57-4-ds rev 2 18 feb 05